BOSTON SCIENTIFIC ULTRA-SOFT SV BALLOON DILATATION CATHETER
K021735 · Boston Scientific Scimed, Inc. · DQY · Aug 8, 2002 · Cardiovascular
Device Facts
Record ID
K021735
Device Name
BOSTON SCIENTIFIC ULTRA-SOFT SV BALLOON DILATATION CATHETER
Applicant
Boston Scientific Scimed, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The proposed Ultra-soft SV Balloon Dilatation Catheter is indicated for percutaneous transluminal angioplasty (PTA) of the iliac, femoral, iliofemoral, popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Device Story
Ultra-soft SV is a Monorail balloon dilatation catheter; dual-lumen coaxial distal segment; outer lumen for balloon inflation; wire lumen accommodates guidewires (≤0.018 in). Used by physicians in clinical settings for percutaneous transluminal angioplasty (PTA) to dilate stenotic arteries or obstructive lesions in dialysis fistulae. Device is advanced over a guidewire to the target lesion; balloon is inflated to a known diameter and length at recommended pressures to restore vessel patency. Benefits include minimally invasive treatment of obstructive vascular disease.
Clinical Evidence
Bench testing and biocompatibility testing only. No clinical data provided.
Technological Characteristics
Monorail balloon dilatation catheter; dual-lumen coaxial distal segment; compatible with guidewires ≤0.018 in; available in various balloon diameters (4.0mm–7.0mm) and lengths (15mm–20mm).
Indications for Use
Indicated for patients requiring percutaneous transluminal angioplasty (PTA) of iliac, femoral, iliofemoral, popliteal, or renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Boston Scientific Gazelle™ Balloon Dilatation Catheter (K001134)
Related Devices
K050389 — ULTRA-SOFT SV BALLOON DILATATION CATHETER · Boston Scientific Corp · May 25, 2005
K051519 — FOX SV PTA CATHETER (82XXX SERIES). FOX PLUS PTA CATHETER (12XXX SERIES) · Abbott Laboratories (Perclose, Inc.) · Jun 21, 2005
K121265 — AMPHIRION PLUS PTA CATHETER · Medtronic, Inc. · May 22, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
#### Summary of Safety and Effectiveness Section 4
#### (Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990)
#### General Provisions
| Submitter's Name<br>and Address | Boston Scientific, Scimed, Inc.<br>One Scimed Place<br>Maple Grove, Minnesota 55311 |
|---------------------------------|-----------------------------------------------------------------------------------------------|
| Contact Person | Candice Burns<br>(763) 494-2845 |
| Classification Name | Percutaneous Transluminal Angioplasty<br>Balloon Dilatation Catheter<br>(21CFR Part 870.1250) |
| Common or Usual Name | Balloon Dilatation Catheter |
| Proprietary Name | Boston Scientific Ultra-soft™ SV Balloon<br>Dilatation Catheter |
#### Classification
According to Section 13 of the Federal Food, Drug and Cosmetic Act, the device classification is Class II, Performance Standards
#### Predicate Devices
Boston Scientific Gazelle™ Balloon Dilatation Catheter (K001134; cleared April 7, 2000)
### Device Description
The Ultra-soft SV Percutaneous Transluminal Angioplasty (PTA) catheter is a Monorail™ catheter with a balloon near the distal tip. The balloon is designed to provide an inflatable segment of known diameter and length at recommended pressures. The distal segment of the balloon catheter is dual lumen and coaxial. The outer lumen is used for inflation of the balloon. The wire lumen permits the use of guidewires (≤0.018 in / 0,46 mm) to facilitate advancement of the catheter to and through the stenosis to be dilated. The Ultra-soft SV will be available in the following sizes.
{1}------------------------------------------------
## Section 4
# Summary of Safety and Effectiveness
| Length<br>Balloon/Catheter | Balloon Diameter | | | | | | |
|----------------------------|------------------|-------|-------|-------|-------|-------|-------|
| | 4.0mm | 4.5mm | 5.0mm | 5.5mm | 6.0mm | 6.5mm | 7.0mm |
| 15 mm/90 cm | × | × | × | × | × | × | × |
| 15 mm/150 cm | × | × | × | × | × | × | × |
| 20 mm/90 cm | × | × | × | × | × | × | × |
| 20 mm/ 150 cm | × | × | × | × | × | × | × |
#### Device Description (continued)
#### Intended Use
The proposed Ultra-soft SV Balloon Dilatation Catheter is indicated for percutaneous transluminal angioplasty (PTA) of the iliac, femoral, iliofemoral, popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
#### Technological Characteristics
The Ultra-soft SV catheter will incorporate a substantially equivalent design, packaging, fundamental technology, manufacturing, sterilization and intended use as those featured in predicate Boston Scientific legally marketed balloon catheters.
### Safety and Performance
Bench testing and biocompatibility testing were performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing regimen.
### Conclusion
Based on the Indications for Use, technological characteristics, and safety and performance testing, the Ultra-soft SV Balloon Dilatation Catheter has been shown to be adequate for its intended use and is considered to be substantially equivalent to the Gazelle PTA Catheter (K001134; cleared April 7, 2000).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus-like symbol consists of three parallel lines that curve and undulate, suggesting movement and dynamism.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 0 8 2002
Ms. Candice Burns Boston Scientific Scimed, Inc. One Scimed Place Maple Grove, MN 55311-1566
K021735 Re:
> Ultra-soft™ SV Balloon Dilatation Catheter Regulation Number: CFR: 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, LIT Dated: May 24, 2002 Received: May 28, 2002
Dear Ms. Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);
{3}------------------------------------------------
Page 2 - Ms. Candice Burns
labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Robert R. Tiller, Ph.D.
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Notification 510(K) Notification
Boston Scientific Ultra-soft™ SV Balloon Dilatation PTA Catheter
# Section 2
Indications for Use
510(k) Number (if known) _K021733 ك
Device Name: Ultra-soft™ SV Balloon Dilatation Catheter
Indications for Use:
The Ultra-soft SV Balloon Dilatation Catheters are recommended for percutaneous transluminal angioplasty of the iliac, femoral, ilio-femoral, popliteal, renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Division of Cardiovascular & Respiratory Devices | |
|--------------------------------------------------|---------|
| 510(k) Number | K021735 |
| Prescription Use<br>(Per 21 CFR 801.109) | |
|------------------------------------------|--|
|------------------------------------------|--|
OR
| Over The Counter Use | |
|----------------------|--|
|----------------------|--|
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.