K210548 · Centurion Medical Products · DQX · May 17, 2022 · Cardiovascular
Device Facts
Record ID
K210548
Device Name
Medline Integrated Arterial Catheter
Applicant
Centurion Medical Products
Product Code
DQX · Cardiovascular
Decision Date
May 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Intended Use
The Medline Integrated Arterial Catheter permits access to the peripheral arterial circulation system for short-term access (less than 30 days).
Device Story
Single-use, sterile, peripherally inserted arterial catheter; 20-Gauge catheter, 22-Gauge echogenic introducer needle, 0.018" stainless steel guidewire. Used in ICU, OR, Anesthesiology, and Cath Lab by physicians and nurses. Principle: Percutaneous arterial puncture; visual verification of blood flashback in needle hub and shaft notch; guidewire advancement via lever; catheter advancement over guidewire; guidewire removal; connection to hemodynamic monitoring tubing. Provides real-time blood pressure monitoring and blood access. Benefits: Enables continuous monitoring in shock, hypertensive emergency, stroke, or complex surgery; facilitates frequent blood draws.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by non-clinical bench testing, including ISO 80369-7 (leakage, separation, overriding), ISO 11070 (corrosion, tensile force, fracture, flex, insertion force), and ISO 10555 (leakage, surface inspection) standards, alongside biocompatibility testing per ISO 10993-1.
Indicated for adult patients requiring short-term (less than 30 days) access to the peripheral arterial circulation system for blood pressure monitoring, frequent blood sampling, or cardiac function monitoring.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K160018 — Arrow Quickflash Arterial Catheterization Device · Arrow International, Inc. (Subsidiary of Teleflex, Inc.) · Jun 13, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
May 17, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Centurion Medical Products Sheri Deisler Regulatory Affairs Manager 100 Centurion Way Williamston, Michigan 48895
#### Re: K210548
Trade/Device Name: Medline Integrated Arterial Catheter Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: April 18, 2022 Received: April 18, 2022
#### Dear Sheri Deisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K210548/A001
Device Name Medline Integrated Arterial Catheter
Indications for Use (Describe)
The Medline Integrated Arterial Catheter permits access to the peripheral arterial circulation system for short-term access (less than 30 days).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Below the word "CENTURION" is a gray line, and below the line, the words "MEDICAL PRODUCTS" are written in smaller, gray letters.
# SECTION 5 510(k) Summary [AS REQUIRED BY 21CFR807.92(c)]
#### Submitter / 510(k) Sponsor
Centurion Medical Products 100 Centurion Way Williamston, MI 48895 Registration Number: 3008403546
#### Contact Person(s)
Sheri Deisler Manager, Regulatory Affairs Phone: 517-545-1136 Email: SDeisler@medline.com
Amanda Baughman Regulatory Affairs Specialist Phone: 517-545-1119 Email: ABaughman@medline.com
Summary Preparation Date April 11, 2022
Type of 510(k) Submission Traditional 510(k)
#### Device Name / Classification
Proprietary Name: Medline Integrated Arterial Catheter Classification Name: Wire, Guide, Catheter Product Code: DQX Classification Panel: Cardiovascular Regulatory Class: II Regulation Number: 870.1330
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Below the word "CENTURION" is a thin gray line, and below that line are the words "MEDICAL PRODUCTS" in smaller, gray letters.
Primary Predicate Device
Health Line ARTLINE K160448
### Secondary Predicate Device
Arrow Radial® Artery Catherization Set K810675
### Device Description
The Medline Integrated Arterial Catheter is a single-use device supplied sterile and nonpyrogenic; it will be packaged inside convenience kits alongside various components outside the scope of this 510(k) submission. The proposed device is a single lumen peripherally inserted catheter device designed to permit short-term access to the peripheral arterial circulatory for less than 30 days. The device assembly comprises a 20-Gauge catheter, 22-Gauge introducer needle, and a 0.018" guidewire with slide advancer. The catheter is made of radiopaque polyurethane tubing, echogenic needle, and removable stainless steel guidewire with slide advancer. The subject device has a kink-resistant catheter design. The device features dual flash with two distinct points of visualization, one with the clear introducer needle proximal hub and the second point at the notch cutout in the needle shaft. The final finished device is packaged with a suture wing (ALCS004). The suture wing is a Class I, 510(k) exempt device.
#### Principle of Operation
After puncturing the skin and advancement into the artery, arterial cannulation is verified visually by observing blood flashback into the catheter shaft and the clear hub of the introducer needle. The introducer needle is then advanced a maximum of 1-2 mm further into the vessel. The user stabilizes the introducer needle and then advances the guidewire using the actuating lever until it coincides with the clear tube's black indicator line. After that, the guidewire is deployed into the artery. The catheter is then advanced over the guidewire into the artery using a slight rotating motion. The catheter is held in place, and the assembly is then pulled back, removing the guidewire. After pulsatile blood is confirmed, a sterile primed hemodynamic monitoring tube with a stopcock is secured to the catheter tube. The catheter position is secured to the patient with a removable suture wing, provided with the proposed device and suture or a suture-free securement device per hospital protocol.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Below the word "CENTURION" is a thin gray line, and below that is the phrase "MEDICAL PRODUCTS" in smaller, gray letters.
#### Proposed Conditions of Use
An arterial catheter is intended for use in the general adult patient population. It is primarily used for real-time, dynamic monitoring of blood pressure in patients, including those experiencing shock, hypertensive emergency, stroke, patients on titratable vasoactive medications, and those undergoing complex surgical procedures. The device provides access for frequent blood draws, including ABG's in ventilated patients, and for monitoring cardiac function in conjunction with separate technologies (i.e., pulse pressure variation). The proposed device will be used in several hospital environments, including the ICU, OR, Anesthesiology, and Cath Lab. These environments will typically be well lit and climatecontrolled. The Medline Integrated Arterial Catheter is a single-use device in which no reuse will occur, and sterile technique should be used during placement. The catheter may remain in place for up to 30 days and be utilized for continuous monitoring while therapy is being given or accessed intermittently for blood sampling.
#### Anatomical Location of Use and Description of Users
Arterial Catheters are inserted percutaneous into an artery (blood vessel); they are typically inserted into the radial artery but can be placed in the elbow (brachial), groin (femoral), or foot (Dorsalis pedis). Typical users of the proposed device include trained ICU physicians, intensivists, anesthesiologists, surgeons, and ICU nurses. Great care needs to be taken to avoid vessel wall puncture and inadvertent subarterial placement. Therefore, the intended users' will be nurses and physicians (e.g., anesthesiology, surgery, critical care) with education ranging from undergraduate to graduate level. Intended users should have basic knowledge and training of catheter placement techniques, contraindications, and necessary precautions. Intended users should have experience with the modified Seldinger technique, and only minimal device-specific training is needed (e.g., reading and understanding the IFU).
The Medline Integrated Arterial Catheter will be available as outlined below:
| Centurion Medical<br>Products Item Number | Description | Gauge | Length |
|-------------------------------------------|-----------------------------------------|-------|-------------|
| RA20G175* | Medline Integrated<br>Arterial Catheter | 20Ga | 1.75 inches |
The final finished device, RA20G175, is packaged with a Suture Wing (ALCS004).
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Below that, separated by a thin gray line, the words "MEDICAL PRODUCTS" are written in smaller, gray letters.
#### Indications for Use
The Medline Integrated Arterial Catheter permits access to the peripheral arterial circulation system for short-term access (less than 30 days).
## Figure 1: Medline Integrated Arterial Catheter with protective tube (Photograph 1 of 3)
Image /page/6/Picture/5 description: The image shows a long, thin, clear plastic tube with a white tip on one end and a black piece attached to the other end. The tube is lying on a dark gray surface. The white tip appears to be a swab, and the black piece may be a handle or a marker.
Figure 2: Medline Integrated Arterial Catheter (Photograph 2 of 3)
Image /page/6/Picture/7 description: The image shows a medical catheter laying on a dark surface. The catheter is long and thin, with a needle at one end and a white cap at the other. There is a black piece attached to the catheter near the white cap. The catheter appears to be unused and sterile.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Below the word "CENTURION" is a thin, gray line. Below the line, the words "MEDICAL PRODUCTS" are written in smaller, gray letters.
# Figure 3: Suture Wing (Photograph 3
Image /page/7/Picture/3 description: The image shows a small, light blue plastic clip with a unique design. It features two circular pads with holes, connected by a U-shaped bridge. The clip is placed on a textured, dark gray surface, possibly a foam or fabric material. The lighting highlights the clip's shape and the surface's texture, creating a clear contrast.
The final finished device, RA20G175, is packaged with a Suture Wing (ALCS004), pictured above.
#### Summary of Technological Characteristics
The Medline Integrated Arterial Catheter is similar in design and technological characteristics to the predicate devices. Minor differences from the predicate devices design are minor and do not raise any additional questions on safety and effectiveness.
Table 5-1 provides a side-by-side comparison of the key attributes with the proposed device, the Medline Integrated Arterial Catheter, and the predicate devices, the Health Line ARTLINE (K160448), and the Arrow Radial® Artery Catherization Set (K810675).
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Below that is a thin gray line, and below that is the phrase "MEDICAL PRODUCTS" in smaller, gray letters.
# Table 5-1: Proposed and Predicate Device(s) Comparison
| Device<br>Characteristics | Proposed Device<br>Medline Integrated Arterial<br>Catheter | Primary Predicate<br>Device<br>Health Line ARTLINE | Secondary Predicate<br>Device<br>Arrow Radial ® Artery<br>Catheterization Set | Comparative Analysis |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory History | New Device | K160448 | K810675 | Same clearance pathway |
| Classification Panel | Cardiovascular | Cardiovascular | Cardiovascular | Same |
| Classification Name | Guide, Wire, Catheter | Guide, Wire, Catheter<br>Catheter, Percutaneous | Guide, Wire, Catheter | Same |
| Device Class | Class II | Class II | Class II | Same |
| Device Code | DQX | DQX/DQY | DQX, | The proposed device and<br>secondary predicate have the<br>same device code. The primary<br>predicate device has two product<br>codes DQX/DQY. |
| Intended Use | The catheter is intended to<br>permit access to peripheral<br>vessels | The catheter is intended to<br>permit access to peripheral<br>vessels | The catheter is intended to<br>permit access to peripheral<br>vessels | Same |
| Indication for Use | The device permits access<br>to the peripheral arterial<br>circulation system for<br>short-term access (less than<br>30 days). | The ARTLINE device permits<br>access to the peripheral<br>arterial circulation system for<br>short-term access (less than 30<br>days). | The Arrow Arterial<br>Catheterization device<br>permits access to the<br>peripheral arterial<br>circulation or to other<br>small vessels. | The indications for use are the<br>same for the proposed device and<br>primary predicate. The secondary<br>predicate device varies slightly<br>but still fall under the same<br>intended use. |
| Device<br>Characteristics | Proposed Device<br>Medline Integrated Arterial<br>Catheter | Primary Predicate<br>Device<br>Health Line ARTLINE | Secondary Predicate<br>Device<br>Arrow Radial ® Artery<br>Catheterization Set | Comparative Analysis |
| Single Use Vs.<br>Disposable | Single-Use | Single-Use | Single-Use | Same |
| Sterile Vs. Non-Sterile | Sterile | Sterile | Sterile | Same |
| Non-Pyrogenic | Yes | Yes | Yes | Same |
| Accessories | Suture Wing | Suture Wing | Suture Wing | Same |
| Available in Sterile Kit<br>Configurations | Yes | Yes | Yes | Same |
| Sterilization Method | Ethylene Oxide Gas | Ethylene Oxide Gas | Ethylene Oxide Gas | Same |
| Shelf Life | 1 Year | 5 Year | Not indicated | N/A |
| Catheter body OD | 20 Ga | 20 Ga | 18, 20 Ga | All three devices feature 20 Ga<br>catheter body. |
| Catheter length | 1 ¾ inches (4.45 cm) | 1 ¾ inches (4.45 cm) | 1 ¾ inches (4.45 cm) | Same |
| Catheter body material | Polyurethane | Polyurethane | Polyurethane | Same |
| Dual blood flashback | Yes – clear introducer<br>needle proximal hub and<br>notch cutout in the needle<br>shaft | Yes – clear introducer needle<br>proximal hub and notch cutout<br>in the needle shaft | No - clear introducer<br>needle proximal hub | The proposed device and primary<br>predicate device feature dual<br>flash. The secondary predicate<br>device features one blood<br>flashback. |
| Echogenic needle | Yes | Yes | Not indicated | The proposed device and primary<br>predicate device feature an<br>echogenic needle. The secondary |
| Device<br>Characteristics | Proposed Device<br>Medline Integrated Arterial<br>Catheter | MEDICAL PRODUCTS<br>Primary Predicate<br>Device<br>Health Line ARTLINE | Secondary Predicate<br>Device<br>Arrow Radial ® Artery<br>Catheterization Set | Comparative Analysis |
| | | | | predicate device does not have<br>this feature. |
| Radiopaque single<br>lumen catheter | Yes | Yes | Yes | Same |
| Integrated Guidewire | Yes | Yes | Yes | Same |
| Removable Guidewire | Yes | Yes | No | The proposed device and primary<br>predicate device feature a<br>removable guidewire. The<br>secondary predicate device does<br>not have this feature. |
| Kink Resistant | Yes | Yes | Yes | Same |
| Guidewire material/size | Stainless Steel 0.018",<br>Straight, soft tip | Stainless Steel 0.018",<br>Straight, soft tip | Stainless Steel 0.018",<br>Straight, soft tip | Same |
{9}------------------------------------------------
# CENTURION
#### MEDICAL PRODUCTS
{10}------------------------------------------------
# CENTURION
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Below the word "CENTURION" is a thin gray line, and below that line, the words "MEDICAL PRODUCTS" are written in smaller, gray letters.
#### Shelf Life and Sterilization
The Medline Integrated Arterial Catheter is sterilized by Ethylene Oxide (EO). The method used to validate the sterilization cycle for this device was conducted in accordance with ISO 11135:2014, Sterilization of Health Care Products - Ethylene Oxide - Requirements for Development, Validation, and Routine Control of a Sterilization Process for Medical Devices, to ensure that a Sterility Assurance Level (SAL) of 1 x 106 is achieved. The proposed device has also been evaluated for EO/ECH residuals in accordance with ISO-10993-7:2008, Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals.
In accordance with ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, aging studies have been conducted to verify a oneyear shelf-life of the subject device and to ensure functionality and sterility are successfully maintained throughout the duration of this shelf life.
#### Summary of Testing
To evaluate performance and functionality, non-clinical verification of the Medline Integrated Arterial Catheter has been conducted. The results of these test have demonstrated the proposed devices' substantial equivalence in accordance with relevant test methods, and ultimately support a substantial equivalence determination.
#### Functional Performance Testing
Stress Cracking in accordance with ISO 80369-7 Positive Pressure Liquid Leakage in accordance with ISO 80369-7 Sub Atmospheric Pressure Air Leakage in accordance with ISO 80369-7 Resistance to Separation from Axial Load in accordance with ISO 80369-7 Resistance to Separation from Unscrewing in accordance with ISO 80369-7 Resistance to Overriding with ISO 80369-7 Kink Resistance Bend Stiffness Corrosion Resistance in accordance with ISO 11070 Guidewire Rebound Peak Tensile Force in accordance with ISO 11070 and ISO 10555 Fracture in accordance with ISO 11070
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for Centurion Medical Products. The word "CENTURION" is written in large, bold, blue letters. Underneath the word "CENTURION" is a thin gray line, and below that, the words "MEDICAL PRODUCTS" are written in smaller, gray letters.
Flex in accordance with ISO 11070 Insertion Force Loosening Force Ouick Flash Strength of Union in accordance with ISO 11070 Visual Inspection Simulated Use Testing Print Durability Freedom from Leakage in accordance with ISO 10555 Surface Inspection (Catheter Tube) in accordance with ISO 10555 Needle Bevel Interface Inspection in accordance with ISO 11070 Dimensional Verification Echogenicity
#### Biocompatibility Testing
The biological evaluation of the subject device was conducted in accordance with ISO 10993-1:2018 Biological Evaluation of the Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process as recognized by the FDA. The proposed device is classified an externally communicating device with prolonged duration of use (>24 hours, to ≤ 30 days) contact with circulating blood.
#### Summary of Clinical Testing
Not applicable. No clinical testing was conducted on the proposed device.
#### Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification. Centurion Medical Products concludes that Medline Integrated Arterial Catheter is substantially equivalent to the predicate device, Health Line ARTLINE (K160448).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.