K113454 · Micro Therapeutics, Inc. · DQX · Feb 24, 2012 · Cardiovascular
Device Facts
Record ID
K113454
Device Name
14 WIRE STANDARD, 14 WIRE LS
Applicant
Micro Therapeutics, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Feb 24, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Intended Use
The 14 Wire Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective the optime of catheters in the peripheral and cerebral vasculature during diagnostic and/or therapeutic procedures. The device is not intended for use in the coronary arteries. The Guidewire Torque Device is intended to facilitate guidewire manipulation during interventional procedures. The Guidewire Introducer is intended to facilitate guidewire insertion into a catheter hub or through a hemostasis valve.
Device Story
The 14 Wire is a stainless steel guidewire with a radiopaque distal segment and a hydrophilic coating. It is used by physicians in clinical settings to assist in the selective placement of catheters within the peripheral and cerebral vasculature. The device includes a torque device for manual manipulation and an introducer to aid insertion into catheter hubs or hemostasis valves. By providing a trackable, lubricious path, the guidewire facilitates the navigation of diagnostic or therapeutic catheters to target sites, potentially improving procedural success and reducing vessel trauma. It is a manual, non-powered device.
Clinical Evidence
Bench testing only. Performance data included visual and dimensional inspection, tip flexibility, buckling force, torque response, shapeability/retention, flexing, coating adherence/lubricity, navigation, kink resistance, tensile/fracture strength, and corrosion resistance. Biocompatibility, shelf-life, bioburden, pyrogen, EtO residuals, and sterility testing were also performed.
Technological Characteristics
Stainless steel guidewire with radiopaque distal segment and hydrophilic coating. Includes torque device and introducer. Mechanical device; no energy source. Sterilized via EtO. Performance specifications include tip flexibility, torque response, and tensile strength.
Indications for Use
Indicated for patients requiring selective catheter placement in peripheral and cerebral vasculature during diagnostic or therapeutic procedures. Not for use in coronary arteries.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
K113454
FEB 2 4 2012
#### 3. 510(k) Summary
# 510(K) Summary
:
| 510(k) Owner: | Micro Therapeutics d/b/a ev3 Neurovascular<br>9775 Toledo Way<br>Irvine, CA 92618<br>Establishment Registration No. 2029214 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Gregory J. Geissinger<br>Manager, Regulatory Affairs<br>Telephone: (949) 297-9781<br>E-mail: ggeissinger@ev3.net |
| Date Summary<br>Prepared: | 10 February 2012 |
| Trade Name of<br>Device: | 14 Wire Standard Hydrophilic Guidewire &<br>14 Wire LS Hydrophilic Guidewire |
| Common Name of<br>Device: | Catheter Guidewire |
| Classification of<br>Device: | DQX, Catheter Guidewire (21 CFR 870.1330), Class II |
| Predicate Devices: | 0.014" Transend EX Platinum Guidewire (K971254)<br>0.014" Transend EX Guidewire (K964611) |
| Device<br>Description: | The 14 Wire is a stainless steel guidewire with a radiopaque<br>distal segment. The distal portion of the guidewire is<br>hydrophilically coated. Included within the sterile pouch is a<br>torque device to assist in guidewire manipulation and a guidewire<br>introducer to ease the introduction of the guidewire into the<br>catheter hub and/or hemostasis valve. |
| Intended Use: | The 14 Wire is indicated for general intravascular use to aid in<br>the selective placement of catheters in the peripheral and<br>cerebral vasculature during diagnostic and/or therapeutic<br>procedures. The device is not intended for use in the coronary<br>arteries. |
| | The Guidewire Torque Device is intended to facilitate guidewire<br>manipulation during interventional procedures. |
| | The Guidewire Introducer is intended to facilitate guidewire<br>insertion into a catheter hub or through a hemostasis valve. |
| Non-Clinical<br>Performance Data: | Biocompatibility testing, extensive bench testing, and an in vitro<br>design validation study were performed as well as shelf-life<br>testing and an assessment of bioburden, pyrogen, EtO residuals,<br>and sterility. |
{1}------------------------------------------------
## Non-Clinical Tests
- Visual Inspection
- Dimensional Inspection .
- Tip Flexibility .
- Tip Buckling Force .
- Torque Response .
- . Tip Shapeability
- Tip Shape Retention .
- . Flexing Test
- . Coating Adherence
- Navigation .
- Kink Resistance ●
- Torque Strength .
- Tensile Strength ●
- . Coating Lubricity
- . Fracture Test
- Corrosion Resistance .
### Conclusion:
The 14 Wire is substantially equivalent to the 0.014" Transend EX and EX Platinoon Guidewires based on the successful completion of non-clinical testing, identical principles of operation and nearly identical indications for use, as well as similarities in the design, materials, dimensions, accessories, packaging, and performance specifications.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Micro Therapeutics, Inc. c/o Mr. Gregory J. Geissinger Manager, Regulatory Affairs 9775 Toledo Way Irvine, CA 92618
FEB 2 4 2012
Re: K113454
Trade/Device Name: 14 Wire Standard Hydrophilic Guidewire & 14 Wire LS Hydrophilic Guidewire
Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guidewire Regulatory Class: Class II Product Code: DOX Dated: February 10, 2012 Received: February 13, 2012
Dear Mr. Geissinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{3}------------------------------------------------
Page 2 – Mr. Gregory J. Geissinger
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
M.S. Hillebenin
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use Statement 2.
! !.
#### Indications for Use
510(k) Number (if known):
Device Name: 14 Wire Standard & 14 Wire LS
Indications for Use: The 14 Wire Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective the optime of catheters in the peripheral and cerebral vasculature during diagnostic and/or therapeutic procedures. The device is not intended for use in the coronary arteries.
The Guidewire Torque Device is intended to facilitate guidewire manipulation during interventional procedures.
The Guidewire Introducer is intended to facilitate guidewire insertion into a catheter hub or through a hemostasis valve.
AND/OR
Prescription Use __ X
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use
..
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.L. Kelleher
Page 1 of 1
(Division Sign-Off) Division of Cardiovascular Devices
K-113454 510(k) Number_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.