K032050 · Boston Scientific Corp · DQX · Aug 8, 2003 · Cardiovascular
Device Facts
Record ID
K032050
Device Name
PERIVAC KIT
Applicant
Boston Scientific Corp
Product Code
DQX · Cardiovascular
Decision Date
Aug 8, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The PeriVac Kit is intended for use in pericardial aspiration and drainage in the presence of pericardial effusion or tamponade.
Device Story
PeriVac Kit is a procedural tray for pericardial aspiration and drainage; contains components for site preparation, anesthesia, puncture, drainage, collection, and dressing. Used by clinicians in clinical settings for patients with pericardial effusion or tamponade. Device facilitates fluid removal to relieve pressure; single-use kit.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Procedural kit containing various medical components for pericardial aspiration. EO sterilized. Single-use. No electronic or software components.
Indications for Use
Indicated for patients requiring pericardial aspiration and drainage due to pericardial effusion or tamponade.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K964519 — CURITY THORACENTESIS TRAY · Kendall Healthcare Products Co. Div.Of Tyco Health · Dec 20, 1996
K960574 — ARGYLE TURKEL NEONATAL/PEDIATRIC THORACIC CATHETER INSERTION TRAY AND THORACIC CATHETER SYSTEM · Sherwood Medical Co. · Aug 5, 1996
K163321 — Rocket Indwelling Pleural Catheter (IPC) Insertion Kit · Rocket Medical Plc · Aug 17, 2017
K051084 — DENVER PLEURX DRAINAGE KIT AND DENVER PLEURX VACUUM BOTTLE WITH DRAINAGE LINE · Denver Biomedicals, Inc. · Jun 13, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K032050 pg 1 of 2
# AUG - 8 2003
510(k) Notification PeriVac Kit
# 510(K) SUMMARY
| Category: | Comments |
|--------------------------------------|------------------------------------------------------------------------------------------------------|
| Sponsor: | Boston Scientific Corporation/EP Technologies, Inc.<br>2710 Orchard Parkway<br>San Jose, CA 95134 |
| Correspondent: | Ronald C. Allen, Ph.,D.<br>Manager, Regulatory Affairs<br>2710 Orchard Parkway<br>San Jose, CA 95134 |
| Contact Numbers: | Phone: 408.895.3670<br>Fax: 408.895.2202<br>Email: allenr@bsci.com |
| Device Common Name | Pericardiocentesis Kit |
| Device Proprietary Name | PeriVac Kit |
| Device Classification | Class II |
| Predicate Device | Manfield PeriVac Kit |
| Predicate Device Manufacturer(s) | Mansfield, Boston Scientific Corporation |
| Predicate Device Proprietary Name(s) | Pericardiocentesis Kit |
| Predicate Device Classification(s) | Class II |
# Date Summary Was Prepared:
May 3, 2003.
# Description of the Device:
The PeriVac Kit is a complete procedure tray for the purpose or pericardial aspiration and/or drainage. It contains the necessary components for site preparation, anesthesia, puncture, drainage, collection, and dressing.
### Intended Use:
The PeriVac Kit is intended for use in pericardial aspiration and drainage in the presence of pericardial effusion or tamponade.
# Comparison to Predicate Devices:
| | Predicate Device | Modified Device |
|--------------------|-------------------------|-----------------|
| 510(k) Reference | Pre-Amendment- Class II | TBD |
| Intended Use | Pericardial aspiration | Same |
| Device Description | Procedural Kit | Same |
| Single Use? | Yes | Same |
CONFIDENTIAL
{1}------------------------------------------------
#### KQ32Q50 pg 20f 2
-
510(k) Notification
PeriVac Kit
:
| | Predicate Device | Modified Device |
|--------------------------|----------------------|-----------------|
| EO Sterilized? | Yes | Same |
| Manufacturer | BSC/ EP Technologies | Same |
| Device<br>Classification | Class I | Same |
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, possibly representing health and human services.
AUG - 8 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Boston Scientific Corp. c/o Dr. Ronald Allen EP Technologies, Inc. 2710 Orchard Parkway San Jose, CA 95134
Re: K032050
PeriVac Kit, models 4304, 4305, 4314, and 4315 Regulation Number: 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: June 30, 2003 Received: July 2, 2003
Dear Dr. Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{3}------------------------------------------------
#### Page 2 – Dr. Ronald Allen
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FIDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains 5 ml of 1% HCL Lidocaine which are subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing this drugs, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled. "Misbranding by reference to premarket notification'(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers,
{4}------------------------------------------------
Page 3 – Dr. Ronald Allen
International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure ·
{5}------------------------------------------------
KO32050 510(k) Noufication PeriVac Kit
#### INTENDED USE STATEMENT
510(k) Number:
PeriVac Kit Device Name:
Indication for Use:
The intended use of the subject kit remains the same as the pre-Amendment kit, and reads as follows:
The PeriVac Kit is intended for use in pericardial aspiration and drainage in the presence of pericardial effusion or tamponade.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED
Qehta
**510(k) Number** K032050
Prescription Use Only
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.