K092105 · Radi Medical Systems AB · DQK · Oct 9, 2009 · Cardiovascular
Device Facts
Record ID
K092105
Device Name
RADIANALYZER XPRESS
Applicant
Radi Medical Systems AB
Product Code
DQK · Cardiovascular
Decision Date
Oct 9, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
RadiAnalyzer® Xpress is intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological and blood flow parameters based on the output from one or more electrodes, transducers or measuring devices. RadiAnalyzer® Xpress is indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters with Pressure Wire®.
Device Story
RadiAnalyzer® Xpress is a diagnostic computer used in catheterization labs by clinicians to process signals from Pressure Wire® and external transducers. It accepts physiological inputs (pressure, temperature) and computes parameters including systolic/diastolic/mean blood pressure, heart rate, Fractional Flow Reserve (FFR), Coronary Flow Reserve (CFR), and first time derivative of pressure (dP/dt). The device displays numerical values and waveforms on an integrated screen; data can be exported to cardiac monitors, printers, or PCs running RadiView® or PhysioMon™ software. It assists clinicians in diagnosing and treating coronary or peripheral artery disease by providing real-time hemodynamic data. The system includes a remote control and necessary cabling for integration into the clinical workflow.
Clinical Evidence
Bench testing only. Software verification and validation were conducted to ensure the device and its accessories (PhysioMon™ and RadiView®) satisfy safety and performance requirements. No clinical data was required or provided.
Technological Characteristics
Programmable diagnostic computer; Class II; Product codes DOK, DSK. Features integrated display, remote control, and connectivity for external transducers/electrodes. Software-based processing for hemodynamic parameters. Includes algorithms for dP/dt and IMR (in RadiView®).
Indications for Use
Indicated for patients undergoing measurement of physiological parameters with Pressure Wire® in catheterization and cardiovascular specialty laboratories to provide hemodynamic information for diagnosis and treatment.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
# OCT - 9 2009
#### 5 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements in the Safe Mcdical Device Act 1990 and 21 CFR §807.92.
### Submitter's Information:
| Name: | Radi Medical Systems AB |
|-------------------------|--------------------------------------------|
| Address: | Palmbladsgatan 10, |
| | SE-754 50 Uppsala, Sweden |
| Phone/Fax: | +46-18-16 10 00 / +46-18-16 10 99 |
| Contact Person: | Mats Granlund |
| Date of Preparation: | July 8, 2009 |
| Device Name: | |
| Trade Name: | RadiAnalyzer® Xpress |
| Common Name: | Diagnostic computer |
| Classification Name: | Programmable Diagnostic Computer §870.1425 |
| | Blood pressure computer § 870.1110 |
| Predicate Device Names: | RadiAnalyzer® Xpress (K042628) |
| | RadiView® (K013943) |
#### Device Description:
RadiAnalyzer® Xpress is a diagnostic computer designed to compute, record and display information from Pressure Wire® and other external transducers. The information is displayed as graphs as well as numerical values on the integrated screen and may also be transferred to a cardiac monitor, RadiAnalyzer® Printer and/or PC with external viewing software installed, such as RadiView® or PhysioMon™. Data includes: systolic, diastolic and mean blood pressure, heart rate, and Fractional Flow Reserve (FFR). RadiAnalyzer® Xpress can be upgraded with additional software, such as Thermo Option software to enable assessments of invasive temperature, Coronary Flow Reserve (CFR) or other physiological parameters such as first time derivative of pressure (dP/dt).
RadiAnalyzer® Xpress comes with one remote control, two monitor cables, one adapter cable, and one mains cable (standard configuration).
#### Intended Use:
RadiAnalyzer® Xpress is intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological and blood flow parameters based on the output from one or more electrodes, transducers or measuring devices.
{1}------------------------------------------------
The indication has been modified to address that the device is not only suitable for usage for diagnosis and treatment of coronary or peripheral artery disease but for patients that undergo measurement of physiological parameters with PressureWire®.
## Indication for Use:
RadiAnalyzer® Xpress is indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters with Pressure Wire®.
## Technical Characteristic:
The mechanical, clectrical and signal properties of RadiAnalyzer® Xpress are identical to the predicate device with the addition of the algorithm for calculation of the first time derivative of pressure and with the addition of the IMR (Index of Microcirculatory Resistance) calculation in the accessory RadiView®.
## Functional/Safety testing:
The software verification/validation conducted indicate that RadiAnalyzer® Xpress and its accessories PhysioMon™ and RadiView® satisfy safety and performance requirements of the device specifications and do not raise additional safety issues.
### Conclusion:
On the basis of the testing conducted, it may be concluded that RadiAnalyzer® Xpress and its accessories PhysioMon™ and RadiView® satify specificd safety and performance
requirements and are substantially equivalent to predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Radi Medical Systems AB Mr. Mats Granlund Director, Quality & Regulatory Affairs Palmbladsgatan 10, SE-754 50 Uppsala SWEDEN
OCT ~ 9 2009
Re: K092105
> Trade/Device Name: RadiAnalyzer® Xpress Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Codes: DOK, DSK Dated: July 8, 2009 Received: July 14, 2009
Dear Mr. Granlund:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. Mats Granlund
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
M. A. Hillebrenner
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number:
Kogzios
Device Name:
RadiAnalyzer® Xpress
Indications for Use:
RadiAnalyzer® Xpress is indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters with Pressure Wire®.
RadiAnalyzer® Xpress is intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological and blood flow parameters based on the output from one or more electrodes, transducers or measuring devices.
Prescription Use X (Per 21 CFR 801.Subpart D)
'
.. .
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. A. Hillhem
Page 1 of _
ivision Sian-Off) Division of Cardiovascular Devices
09 2105 510(k) Number