CARTO XP EP NAVIGATION SYSTEM, SOFTWARE VERISON 8

K042999 · Biosense Webster, Inc. · DQK · Mar 21, 2005 · Cardiovascular

Device Facts

Record IDK042999
Device NameCARTO XP EP NAVIGATION SYSTEM, SOFTWARE VERISON 8
ApplicantBiosense Webster, Inc.
Product CodeDQK · Cardiovascular
Decision DateMar 21, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2

Indications for Use

The CARTO™ XP EP Navigation System is intended to acquire real time catheter based cardiac electrophysiogical maps in patients who are eligible for a conventional electrophysiological study. The CARTO™ XP EP Navigation System is restricted for use by licensed medical practitioner who participated in a CARTO™ training course. There are no special contraindications when using the CARTO™ XP EP Navigation System. The intended use of the CARTO™ XP mapping system is catheter-based atrial and ventricular mapping. The CARTO™ XP mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as electrical activation maps, electrical propagation maps, electrical potential maps, and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed in real time on the display screen. The intended use of the modified device is the same as for Predicate Device A cleared under 510 (k) Number K020863. The indications for use have been expanded to include the CARTOMERGE™ capability to import, register and merge CT or MRI structural images with CARTO maps physiological information and real time catheter navigation.

Device Story

System acquires, analyzes, and displays electro-anatomical maps of human heart; inputs include intracardiac electrograms and body surface ECG. CARTOMERGE module imports, registers, and merges pre-acquired CT or MRI images (DICOM) with real-time CARTO maps. System displays electrical activation, propagation, potential, and chamber geometry maps in real-time. Used in clinical settings by trained licensed medical practitioners. Output provides visual cardiac mapping to assist physicians in electrophysiological procedures; facilitates navigation and clinical decision-making during cardiac interventions.

Clinical Evidence

No clinical data provided. Safety and effectiveness supported by non-clinical bench and animal testing.

Technological Characteristics

Cardiac mapping system; integrates intracardiac electrograms and body surface ECG. Supports DICOM-based CT/MRI image import and registration. Complies with UL 2601-1:97, CSA C22.2 NO.601.1, IEC 60601-2-25:93, and IEC 60601-2-27:94 safety standards.

Indications for Use

Indicated for patients eligible for conventional electrophysiological study requiring catheter-based atrial and ventricular mapping. No specific age or gender restrictions. No special contraindications.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 2 1 2005 CARTO XP Version 8 EP Navigation System Special 510(k) ## C. 510(K) Summary The 510(k) summary for the CARTO XP V8 Mapping System is provided below. ## 510(k) Summary for the CARTO™ XP V8 EP Navigation System Biosense Webster, Inc. 510(k) Notification Submitted by: 3333 Diamond Canyon Rd. Diamond Bar, CA 91765 USA Phone: +1-800-729-9010 Fax: +1-909-839-8804 Diana Thorson Contact Person: Project Manager, Regulatory Affairs CARTO™ XP EP Navigation System Proprietary Device Name: Programmable diagnostic computer Classification Name: (per 21 CFR 870.1425) Cardiac mapping system Common Devicc Name: A. CARTOTM XP QwikMap EP Navigation Predicate Devices: System B. Siemens Medical System singo Multimodality Workstation Biosense Webster (Israel) Ltd. Manufacturer: POB 2009 Tirat HaCarmel, 39120 Israel {1}------------------------------------------------ K042999/52 P2/2 CARTO XP Version 8 EP Navigation System Special 510(k) ### Indications For Use The CARTO™ XP EP Navigation System is intended to acquire real time catheter based cardiac electrophysiogical maps in patients who are eligible for a conventional electrophysiological study. The CARTO™ XP EP Navigation System is restricted for use by licensed medical practitioner who participated in a CARTO™ training course. There are no special contraindications when using the CARTO™ XP EP Navigation System. #### General Device Description The CARTO™ XP EP Navigation system is designed to acquire, analyze, and display electro-anatomical maps of the human heart. The maps are reconstructed using the combination of information gathered from the integration of intracardiac electrograms with their respective endocardial locations. Maps may be displayed as electrical activation maps, electrical propagation maps, electrical potential maps, and chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed in real time on the display screen. The CARTOMERGE module provides for the import, processing, visualization and analysis of pre-acquired cardiac images, superimposed to the CARTO XP EP maps. The CARTOMERGE supports import of Computed Tomography (CT) and Magnetic Resonance (MRI) images in DICOM format. The CARTO™ XP EP Navigation System complies with the following safety standards: UL 2601-1:97/CSA C22.2 NO.601.1 IEC 60601-2-25:93 and Al(99) IEC 60601-2-27:94 The non-clinical bench and animal testing show that the device is as safe and as effective as the previously marketed devices to which it is being compared and does not raise any new questions of safety or effectiveness. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a circular seal with text around the perimeter and a symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The symbol in the center is a stylized eagle with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 1 2005 Biosense Webster, Inc. c/o Ms. Diana Thorson Project Manager, Regulatory Affairs 3333 Diamond Canyon Road Diamond Bar, CA 91765 Re: K042999 Trade Name: CARTO™ EP Navigation System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: II (two) Product Code: DQK Dated: January 31, 2005 Received: February 03, 2005 Dear Ms. Thorson: We have reviewed your Section 510(k) premarket notification of intent to market the indication we have reviewed your Section 910(x) promotion is substantially equivalent (for the indications for use stated in the enclosure) the acted predicate devices marketed in interstate for use stated in the encrosule) to regard manative posice Amendments, or to commerce provided to May 20, 1978, the exactions of the Federal Food, Drug, devices that have occh recaisoned in assess approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approval of a premarket approval ap and Cosment ACT (Act) that do not require apper o the general controls provisions of the Act. The You may, therefore, market the devices, books of the more of the manual registration, listing of general controls provisions of the rise, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) me etails in the regulations affecting your device can be may be subject to such additional econtroller interest and to 898. In addition, FDA may oc found in the Code of Pederal Regarations, your device in the Federal Register. {3}------------------------------------------------ Page 2 – Ms. Diana Thorson Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease oc devisou and I Drimination that your device complies with other requirements of the Act that I Drimas Intacted and regulations administered by other Federal agencies. You must or any I edetar statutes and registements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF IT rate 677, idoomig (21 (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quinn control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product indiation control to begin marketing your device as described in your Section 510(k) I mis letter will and will yourse finding of substantial equivalence of your device to a legally prematics notication "The classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you t the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Whoordinating of responsibilities under the Act may be obtained from the Oiner general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, B. Zimmerman for R. Z. Zimmerman, M.D. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # 1042999 CARTO XP Version 8 EP Navigation System Special 510(k) ## D. Indications For Use Statement 510(k) No (if known): k042999 Device Name: CARTO™ EP Navigation System #### Indication for Use: The intended use of the CARTO™ XP mapping system is catheter-based atrial and ventricular mapping. The CARTO™ XP mapping system allows real-time display of cardias maps in a uumber The CARTO'" XF inapping System anows of cardiac electrical activation maps, cardiac of different formals. Maps may of expression maps, and cardiac chamber electrical propagation naps, carenas t signals, including body surface ECG and geomeny maps. The acquired panelle organia, and time on the display screen. The intended use of the modified device is the same as for Predicate Device A cleared I he intended use of the modified dovied to also for use have been expanded to include under 510 (k) Number K020863. The indications for use have been expanded to include the CARTOMERGE™ capability to import, register and merge CT or MRI structural the CARTOMERGE™ Capability to import, rogister and the catheter navigation. images with CARTO maps physiological information and real time catheter navigation. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Blumamoto (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K043999 Page 1 of 1 Brosense Webster, Inc Page 15 of 111
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...