SENSI

K110704 · Diacoustic Medical Devices (Pty) , Ltd. · DQD · Jun 24, 2011 · Cardiovascular

Device Facts

Record IDK110704
Device NameSENSI
ApplicantDiacoustic Medical Devices (Pty) , Ltd.
Product CodeDQD · Cardiovascular
Decision DateJun 24, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1875
Device ClassClass 2
AttributesPediatric

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Heart murmur identificationSensitivity of 91%, Specificity of 94%Design verification of a CAA algorithm
Heart murmur identificationSensi accuracy 70.8%, Cardioscan accuracy 67.9%Comparative Study between the FDA approved Zargis system and the Sensi

Indications for Use

The Sensi device consisting of the Sensi Diagnostic Heart Murmur Software, the WelchAllyn Master Elite Stethoscope and WelchAllyn USB Analyzer is a decision support device intended to be used on a single patient to assist the medical examiner in analyzing cardiac sounds for the identification and classification of suspected murmurs. It is used to distinguish between normal/physiological and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyze these signals. The acoustic heart signal is analyzed to identify specific heart sounds that may be present. Identified sounds include S1, S2, and suspected systolic murmurs. Sensi indicates whether or not a recorded heart sound contains a suspected heart murmur and the confidence with which the analysis was made The device must be used in a clinical setting by trained personnel with the prescribed accessories and all relevant patient information must be taken into consideration before a diagnosis is made. The interpretations of heart sounds offered by the Sensi device are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data. Sensi is not intended to be a diagnostic device. It does not supersede the judgment of the qualified medical personnel. The device is intended to aid the physician in the evaluation of heart sounds. The physicians responsible for reviewing and interpreting the results, along with the auscultatory findings and medical history, in making a referral decision.

Device Story

Sensi is a decision support system for heart murmur analysis. Inputs: acoustic heart sounds (four auscultation positions) and 3-lead ECG signals captured via WelchAllyn Master Elite Stethoscope and USB Analyzer. Operation: software runs on Windows PC; uses ECG QRS peaks to segment heart beats; analyzes acoustic signals to identify S1, S2, and suspected systolic murmurs. Output: classification of heart sounds (normal vs. pathological) and confidence score. Used in clinical settings by trained personnel. Benefits: aids physician evaluation of heart sounds to inform referral decisions. Not a standalone diagnostic; requires physician over-read and integration with patient history.

Clinical Evidence

Clinical validation included a feasibility/usability study in clinical settings (users scored 4/5); comparative study vs. Zargis Cardioscan (Sensi 70.8% accuracy vs. 67.9% for Cardioscan); design verification of CAA algorithm (94% specificity, 91% sensitivity); and validation of murmur classification in pediatric population (1568 heart sounds analyzed).

Technological Characteristics

System includes WelchAllyn Master Elite Stethoscope, WelchAllyn USB Analyzer, and Sensi Diagnostic Heart Murmur Software. Operates on Windows-based PC. Signal processing uses ECG-synchronized acoustic analysis. Software algorithm performs heart sound segmentation and murmur classification.

Indications for Use

Indicated for use by trained medical personnel in clinical settings to assist in identifying and classifying suspected heart murmurs in patients. Aids in distinguishing between normal/physiological and pathological murmurs by analyzing acoustic heart signals and ECG. Not for diagnostic use; requires physician over-read.

Regulatory Classification

Identification

Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110704 page 1 of 3 ## 510(k) Summary 807.92(c) JUN 2 4 2011 ### SPONSOR Company Name: 807.92(a)(1) Diacoustic Medical Devices (Pty) Ltd Company Address 19 Quantum Street Techno Park Stellenbosch, South Africa 7600 Telephone: Fax: Contact Person: 27 21 880 2033 27 86 557 4391 Matthys L. Cronje Summary Preparation Date: March 2, 2011 ### DEVICE NAME Trade Name: Common/Usual Name: Classification Name: Regulation Number: Product Code: Device Class: ## 807.92(a)(2) Sensi with Diagnostic Heart Murmur Software Electronic Stethoscope/Heart Sounds Analyzer Electronic Stethoscope; Phonocardiograph 21 CFR 870.1875. 870.2390 DQD, DQC Class II ### PREDICATE DEVICE 807.92(a)(3) | Legally Marketed Equivalent Device | | | |------------------------------------|-----------------------------------|----------| | Company | Product | 510(k) # | | Zargis Medical Corp | Zargis Acoustic Cardioscan | K083309 | | Meditron AS | Meditron II Thestethoscope System | K013725 | ### DEVICE DESCRIPTION ### 807.92(a)(4) The Sensi with Diagnostic Heart Murmur Software is a decision support device intended to acquire, record, and analyze heart sounds. It is used to distinguish between normal/physiological and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyzing these signal. The complete system is a CD comprising Sensi Diagnostic Heart Murmur Software that runs on a computer with Windows operating system, instructions for use, the WelchAllyn Master Elite Stethoscope and the WelchAllyn Meditron Analyzer that captures the acoustic heart and ECG signals. The Sensi Diagnostic Heart Murmur Software distinguishes between normal/ physiological and pathological heart murmurs by analyzing the acoustic heart and ECG signals captured with the WelchAllyn Meditron Analyzer. The device will record the acoustic sound of the heart at the four main auscultation positions and a three lead ECG signal. The ECG signal is not intended for diagnostic use, only to synchronize with the {1}------------------------------------------------ beginning of each heart cycle. The acoustic heart signal is analyzed to identify heart sounds that may be present, identified sounds include S1, S2 and suspected systolic murmurs. ### DEVICE INDICATIONS FOR USE 807.92(a)(5) The Sensi device consisting of the Sensi Diagnostic Heart Murmur Software, the WelchAllyn Master Elite Stethoscope and WelchAllyn USB Analyzer is a decision support device intended to be used on a single patient to assist the medical examiner in analyzing cardiac sounds for the identification and classification of suspected murmurs. It is used to distinguish between normal/physiological and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyze these signals. The acoustic heart signal is analyzed to identify specific heart sounds that may be present. Identified sounds include S1, S2, and suspected systolic murmurs. Sensi indicates whether or not a recorded heart sound contains a suspected heart murmur and the confidence with which the analysis was made The device must be used in a clinical setting by trained personnel with the prescribed accessories and all relevant patient information must be taken into consideration before a diagnosis is made. The interpretations of heart sounds offered by the Sensi device are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data. Sensi is not intended to be a diagnostic device. It does not supersede the judgment of the qualified medical personnel. The device is intended to aid the physician in the evaluation of heart sounds. The physicians responsible for reviewing and interpreting the results, along with the auscultatory findings and medical history, in making a referral decision. ### COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a)(6) The Sensi software device features were directly compared with the FDA cleared Zargis Acoustic Cardioscan and Meditron Stethoscope Analyzer. Synopsis of the comparison analysis: - All three software systems uses equivalent computer platforms. . - Both Sensi and Meditron uses equivalent accessories (electronic stethoscope and ● USB Analyzer - ECG). The acoustic performance between the Littmann and WelchAllyn electronic stethoscopes are equivalent. - Both Sensi and Cardioscan uses equivalent signal processing algorithms. Sensi . uses the ORS peak information from the ECG to segment the heart beats. No additional issues are introduced using this standard processing method and makes the process only more safe and effective. - Patient information and signal display handling are in all three cases equivalent. . - Sensi and Cardioscan's clinical performances are equivalent. ● {2}------------------------------------------------ K110704 page 3 of 3 ### SAFETY and EFFECTIVENESS 807.92(b) A comprehensive list of verification and validation testing was performed in accordance to Diacoustic's Design Control procedures. Software validation was performed for all aspects of the Sensi System and Software. The graphical user interface and usability were compared to the predicate devices. Validation of the Sensi was performed to ensure that the Sensi system consistently fulfills its intended use and the needs of the user. A clinical software validation was performed to insure the performance of the software algorithm | Study Type | Results | |------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feasibility & Usability Study | Usability validation was performed within real life clinical<br>settings by intended users.<br>On average all users scored the usability of the Sensi<br>Software more than 4 out of 5. | | Comparative Study between the FDA<br>approved Zargis system and the Sensi | Sensi achieves overall accuracy of 70.8%<br>Cardioscan achieves an accuracy of 67.9% | | Design verification of a CAA<br>algorithm | Specificity of 94% and sensitivity of 91% | | Comparison of the Sensi software<br>program's graphical user interface and<br>usability to that of three other<br>comparable software programs | The graphical user Sensi compares favorably to the three<br>auscultation analysis packages with similarities between all<br>the major display and interface functions | | Validate algorithms used to<br>distinguish between functional and<br>pathological heart murmurs in the<br>pediatric population. | 1568 heart sounds were accepted to meet the criteria of<br>good quality and match the recorded pathological condition | ### CONCLUSION ### 807.92(b)(3) Based upon the indications for use, technological characteristics and safety and performance testing, it is the conclusion of Diacoustic Medical that the Sensi device consisting of the Sensi Diagnostic Heart Murmur Software, the WelchAllyn Master Elite Stethoscope and WelchAllyn USB Analyzer is as safe and effective as the predicate devices and raises no new issues of safety and effectiveness. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol in blue, which is a symbol associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged around the symbol in a circular fashion. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUL 22 2011 Diacoustic Medical Devices (Pty) Ltd. c/o Mr. E.J. Smith Smith Associates 1468 Harwell Ave. Crofton, MD 21114 Re: K110704 Trade Name: Sensi Electronic Stethoscope Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II (two) Product Code: DQD Dated: June 6, 2011 Received: June 6, 2011 Dear Mr. Smith: This letter corrects our substantially equivalent letter of June 24. 2011. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. E.J. Smith Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bran D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K110704 # Indications for Use 510(k) Number (if known): Device Name: Sensi Indications for Use: The Sensi device consisting of the Sensi Diagnostic Heart Murmur Software, the WelchAllyn Master Elite Stethoscope and WelchAllyn USB Analyzer is a decision support device intended to be used on a single patient to assist the medical examiner in analyzing cardiac sounds for the identification and classification of suspected murmurs. It is used to distinguish between normal/physiological and pathological heart murmurs by recording the acoustic signal of the heart and the ECG signal simultaneously and analyze these signals. The acoustic heart signal is analyzed to identify specific heart sounds that may be present. Identified sounds include S1, S2, and suspected systolic murmurs. Sensi indicates whether or not a recorded heart sound contains a suspected heart murmur and the confidence with which the analysis was made The device must be used in a clinical setting by trained personnel with the prescribed accessories and all relevant patient information must be taken into consideration before a diagnosis is made. The interpretations of heart sounds offered by the Sensi device are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data. Sensi is not intended to be a diagnostic device. It does not supersede the judgment of the qualified medical personnel. The device is intended to aid the physician in the evaluation of heart sounds. The physicians responsible for reviewing and interpreting the results, along with the auscultatory findings and medical history, in making a referral decision. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |------------------------------------|---------| | Division of Cardiovascular Devices | | | 510(k) Number | K110704 | | Page | 1 of 1 |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...