Sim&Size enables visualization of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery. Sim&Size also allows for the ability to computationally model the placement and deployment of neurointerventional devices. General functionalities are provided such as: - Segmentation of neurovascular structures - Automatic centerline detection - Visualization of X-ray based images for 2D review and 3D reconstruction - Placing and sizing tools - Reporting tools Information provided by the software is not intended in any way to eliminate, replace or in part, the healthcare provider's judgment and analysis of the patient's condition.
Device Story
Sim&Size is radiological image processing software for preoperational planning of intracranial aneurysm treatments. Input: angiographic DICOM data (e.g., 3D rotational angiography). Operation: software segments neurovascular structures, performs automatic centerline detection, and computationally models the placement, deployment, and apposition of neurointerventional devices (e.g., flow diverters, intrasaccular devices). Output: 2D/3D visualizations of vessels and simulated device deployment, plus reporting tools. Usage: installed on Windows or Mac OS personal computers; operated by clinicians. The software provides patient-specific visualization to assist in sizing and planning; it does not replace clinical judgment. Benefits: improved pre-procedural assessment and device selection for neurovascular interventions.
Clinical Evidence
Bench testing only. Testing included validation of DICOM importation, patient management, image display/processing, anatomic reconstruction, computational modeling of neurovascular devices, and physical device deployment testing to verify the accuracy of the computational model. All tests met required specifications.
Technological Characteristics
Software-based radiological image processing; runs on Windows or Mac OS. Inputs: DICOM datasets (CT, enhanced CT, 3D rotational angiography). Features: segmentation, automatic centerline detection, 3D reconstruction, and device modeling. Connectivity: local file import via physical media (USB). Rendering: local OpenGL. No patient contact.
Indications for Use
Indicated for preoperational planning and sizing for neurovascular interventions and surgery, including computational modeling of neurointerventional device placement and deployment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 17, 2019
Sim&Cure % Scott Blood Regulatory Consultant MEDIcept, LLC 22 Nichols Street #2 Salem, Massachusetts 01970
Re: K190049
Trade/Device Name: Sim&Size Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: PZO Dated: August 2, 2019 Received: August 6, 2019
Dear Scott Blood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brian Pullin, M.S Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190049
Device Name Sim&Size
Indications for Use (Describe)
Sim&Size enables visualization of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery. Sim&Size also allows for the ability to computationally model the placement and deployment of neurointerventional devices.
General functionalities are provided such as:
- · Segmentation of neurovascular structures
- · Automatic centerline detection
- · Visualization of X-ray based images for 2D review and 3D reconstruction
- · Placing and sizing tools
- · Reporting tools
Information provided by the software is not intended in any way to eliminate, replace or in part, the healthcare provider's judgment and analysis of the patient's condition.
Type of Use (Select one or both, as applicable)
|× Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K190049
Image /page/3/Picture/1 description: The image shows the logo for Sim&Cure. The word "Sim" is in black, followed by a red ampersand and the word "Cure" in black. Below the logo is the text "SECURE YOUR TREATMENT" in a smaller, sans-serif font.
## 510(k) Summary
#### 1. Sponsor Name
Submitter's Name: Sim&Cure Address: 1682 Rue de la Valsière 34790 Grabels FRANCE Phone: +33 769 51 37 98
Contact Person: Scott Blood Regulatory Consultant MEDIcept 978.729.5978 scottqara@gmail.com
Date of Preparation: September 17, 2019
#### 2. Device Information
| Trade Name: | Sim&Size |
|--------------------|---------------------------------------------|
| Common Name: | Radiological Image Processing Software |
| Regulation Number: | 892.2050 |
| Regulation Name: | Picture Archiving and Communications System |
| Class: | II |
| Product Code: | PZO |
#### 3. Predicate Device
K171534 - SurgicalPreview, EndoVantage LLC
#### 4. Device Description
Sim&Size is software that allows for the preoperational planning of medical device sizes for the treatment of intracranial aneurysms. The computational modeling of neurointerventional devices, such as flow diverters and intrasaccular devices, are supported by the software to provide a patient-specific visualization of the deployment of the device from angiographic DICOM data, Information provided by the software is not intended in any way to eliminate, replace or substitute for, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition.
The Sim&Size software is intended to be loaded on any Windows- or Mac-OS personal computer. A user license(s) must be purchased from the company in order to use the software
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Sim&Cure. The word "Sim" is in black, followed by a red ampersand. The word "Cure" is in black. Below the logo is the text "SECURE YOUR TREATMENT" in a smaller font size.
after installation. The software also has a training module that offers learning assessments that are relevant to the software.
The software interacts with a patient's DICOM images, the user, and the device. To do so, the graphical interface is organized into three graphical pages (or screens):
- a. Patient selection page
- b. Module selection page
- Simulation page C.
Sim&Size was designed to enable the visualization of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery. Sim&Size also allows for the ability to computationally model the placement, deployment and apposition of neurointerventional devices.
#### 5. Intended Use
Sim&Size enables visualization of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery. Sim&Size also allows for the ability to computationally model the placement and deployment of neurointerventional devices. General functionalities are provided such as:
- Segmentation of neurovascular structures ●
- Automatic centerline detection
- Visualization of X-ray based images for 2D review and 3D reconstruction
- Placing and sizing tools
- Reporting tools
Information provided by the software is not intended in any way to eliminate, replace or substitute for, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition.
#### 6. Comparison of Technological Characteristics
The predicate device for the Sim&Size software is SurgicalPreview™, cleared per premarket notification K171534. Both Sim&Size and SurgicalPreview have the same indications for use for preoperational planning of neurovascular procedures using existing image data. A summary comparison of technological characteristics in provided below:
| Characteristic | Predicate device<br>SurgicalPreview<br>K171534 | Subject device<br>Sim&Size<br>This submission |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | SurgicalPreview enables visualization<br>and measurement of cerebral blood<br>vessels for preoperational planning and | Sim&Size enables visualization of<br>cerebral blood vessels for<br>preoperational planning and sizing for |
| Characteristic | Predicate device<br>SurgicalPreview<br>K171534 | Subject device<br>Sim&Size<br>This submission |
| | sizing for neurovascular interventions<br>and surgery. SurgicalPreview also<br>allows for the ability to computationally<br>model the placement and deployment of<br>neurointerventional devices.<br>General functionalities are provided<br>such as:<br>• Segmentation of neurovascular<br>structures<br>• Automatic centerline detection<br>• Visualization of CT scan images for<br>2D review and 3D reconstruction<br>• Measurement and annotation tools<br>• Reporting tools<br>Information provided by the software is<br>not intended in any way to eliminate,<br>replace or substitute for, in whole or in<br>part, the healthcare provider's judgment<br>and analysis of the patient's condition | neurovascular interventions and<br>surgery. Sim&Size also allows for the<br>ability to computationally model the<br>placement and deployment of<br>neurointerventional devices.<br>General functionalities are provided<br>such as:<br>• Segmentation of neurovascular<br>structures<br>• Automatic centerline detection<br>• Visualization of X-ray based images<br>for 2D review and 3D<br>reconstruction<br>• Placing and sizing tools<br>• Reporting tools<br>Information provided by the software<br>is not intended in any way to<br>eliminate, replace or substitute for, in<br>whole or in part, the healthcare<br>provider's judgment and analysis of<br>the patient's condition |
| Interface to Image<br>Sources | DICOM Image Data | Same |
| Import of Patient Data | Manual through keyboard/mouse,<br>automatic import with image file, study<br>creation list | Same |
| List Image<br>Functionality | Deleting, anonymizing, search | Same |
| Image Processing | Segmentation by user with clinician<br>review and comment | Same |
| 3D Assessment | Linear (length and diameter)<br>measurements, volume measurements | 3D assessment based on 3D model of<br>the simulated device inside the vessels |
| Image and 3D Display | Orthogonal, color volume rendering, 2D<br>slide review, active presets, 3D view of<br>assemblies of devices | Same |
| DICOM Support | Compatible with all scanner vendor<br>DICOM datasets, storage SCP, import<br>DICOM files, DICOM compliance for<br>CT and enhanced CT, import from<br>DICOMDIR, storage SCU,<br>query/retrieve SCU | DICOM compliance for CT and<br>enhanced CT, Read DICOM images<br>from 3D rotational angiography<br>stations |
| Computer OS<br>Compatibility | MS Windows and Mac OS | Same |
| Data<br>Interchange/Transfer | Secure Internet File Server | Transfer by physical media; i.e. USB<br>memory stick |
| Characteristic | Predicate device<br>SurgicalPreview<br>K171534 | Subject device<br>Sim&Size<br>This submission |
| Output File Format | Web browser via WebGL | Local openGL rendering |
| Preoperational<br>Planning | Yes | Yes |
| Patient Contact | No | No |
| Human Intervention<br>for Interpretation of<br>Images | Yes | Yes |
{5}------------------------------------------------
# Sim&Cure Secure Your Treatment
{6}------------------------------------------------
#### 7. Performance Testing
Performance tests on the bench were performed on the Sim&Size. The following were performed:
- Tests of importation of DICOM images ●
- Patient manager tests
- Tests of image display and processing
- Functioning tests for visualization of anatomic reconstruction
- . Functioning tests for computational modeling of currently-approved neurovascular devices
- Physical device deployment testing to verify the computational model .
- . Report creation and visualization tests
All bench testing has been performed and the software has met the required specifications for the completed tests.
#### 9. Summary
Sim&Cure has demonstrated that the Sim&Size is substantially equivalent to its listed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.