PrecisionTAVI (v1.1)

K223809 · Dasi Simulations · QQI · May 30, 2023 · Cardiovascular

Device Facts

Record IDK223809
Device NamePrecisionTAVI (v1.1)
ApplicantDasi Simulations
Product CodeQQI · Cardiovascular
Decision DateMay 30, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1405
Device ClassClass 2
AttributesSoftware as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K223809 · May 30, 2023PrecisionTAVI (v1.1)Dasi SimulationsRetrospective clinical patient CT imaging (pre-TAVR and post-TAVR)The software's predictive accuracy for THV frame deformation was validated by comparing simulated deployment results against actual post-procedural clinical CT data from 89 patients.Retrospective clinical data; Post-TAVR CT imaging; Validation study

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
THV deployment validation; Retrospective validation study comparing simulated outcomes to clinical post-operative imaging89 patients with tricuspid aortic valve morphology who received a SAPIEN S3/Ultra THV; Sample Size: 89Not applicable for this studyTHV frame deformation (inflow, waist, and outflow diameter; aspect ratio)

Indications for Use

PrecisionTAVI is an optional, non-invasive, post processing software solution that is indicated for patient-specific simulations of Transcatheter Aortic Valve Replacement (TAVR) during procedural planning. The software performs computer simulation to predict post TAVR in vivo valve frame deformation of clinician selected Transcatheter Heart Valve (THV) device types and sizes. The information provided by PrecisionTAVI is intended for use by cardiologists, and clinical specialists, and is not intended in any way to eliminate, replace, or substitute for, in part, the healthcare provider's judgment and analysis of the patient's condition receiving the images retains the responsibility for interpreting and validating all information and making all patient treatment decisions. PrecisionTAVI is not intended to replace the simulated device's instructions for use for final TAVR device selection and placement.

Device Story

PrecisionTAVI is a non-invasive software solution for TAVR procedural planning. Input consists of 3D anatomical models generated from patient multi-slice cardiac CT images. DASI Simulations operators use a proprietary computational modeling workflow to simulate the mechanical interaction between a predefined Transcatheter Heart Valve (THV) model and the patient-specific aortic root anatomy. The output is a report with 3D visualization of the predicted post-deployment THV frame deformation. The report is accessed via a web-based portal. Clinicians use this information as an adjunct to their own judgment for TAVR planning and device selection. The device does not replace the manufacturer's instructions for use for the specific THV device. Benefits include improved visualization of potential device-anatomy interaction prior to the procedure.

Clinical Evidence

Bench testing using retrospective clinical data from 89 patients with tricuspid aortic valve morphology who received SAPIEN S3/Ultra THV. Validation compared simulated THV deformation against post-TAVR CT-derived geometries. Results: 100%, 97%, and 98% of cases showed <±2mm difference in inflow, waist, and outflow diameters respectively; 97%, 98%, and 98% showed <±0.1 difference in aspect ratio (meeting ≥95% success criteria). Qualitative assessment by 5 clinicians and 3 engineers showed 96% agreement in case evaluations and >80% agreement in eccentricity/apposition metrics.

Technological Characteristics

Software-based computational simulation. Inputs: DICOM-based 3D anatomical models from cardiac CT. Outputs: 3D visualization of THV deformation. Connectivity: Web-based portal for report access. Architecture: Proprietary computational modeling technology for mechanical interaction simulation. No physical materials or energy sources.

Indications for Use

Indicated for patient-specific simulations of Transcatheter Aortic Valve Replacement (TAVR) during procedural planning for patients undergoing TAVR. Intended for use by cardiologists and clinical specialists.

Regulatory Classification

Identification

An interventional cardiovascular implant simulation software device is a prescription device that provides a computer simulation of an interventional cardiovascular implant device inside a patient's cardiovascular anatomy. It performs computational modeling to predict the interaction of the interventional cardiovascular implant device with the patient-specific anatomical environment.

Special Controls

In combination with the general controls of the FD&C Act, the interventional cardiovascular implant simulation software device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Software verification, validation, and hazard analysis, with identification of appropriate mitigations, must be performed, including a full verification and validation of the software according to the predefined software specifications. (2) Computational modeling verification and validation activities must be performed to establish the predictive capability of the device for its indications for use. (3) Performance validation testing must be provided to demonstrate the accuracy and clinical relevance of the modeling methods for the intended implantation simulations, including the following: (i) Computational modeling results must be compared to clinical data supporting the indications for use to demonstrate accuracy and clinical meaningfulness of the simulations; (ii) Agreement between computational modeling results and clinical data must be assessed and demonstrated across the full intended operating range ( *e.g.,* full range of patient population, implant device sizes and patient anatomic morphologies). Any selection criteria or limitations of the samples must be described and justified;(iii) Endpoints ( *e.g.,* performance goals) and sample sizes established must be justified as to how they were determined and why they are clinically meaningful; and(iv) Validation must be performed and controls implemented to characterize and ensure consistency ( *i.e.,* repeatability and reproducibility) of modeling outputs:(A) Testing must be performed using multiple qualified operators and using the procedure that will be implemented under anticipated conditions of use; and (B) The factors ( *e.g.,* medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data.(4) Human factors evaluation must be performed to evaluate the ability of the user interface and labeling to allow for intended users to correctly use the device and interpret the provided information. (5) Device labeling must be provided that describes the following: (i) Warnings that identify anatomy and image acquisition factors that may impact simulation results and provide cautionary guidance for interpretation of the provided simulation results; (ii) Device simulation inputs and outputs, and key assumptions made in the simulation and determination of simulated outputs; and (iii) The computational modeling performance of the device for presented simulation outputs, and the supporting evidence for this performance.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration". May 30, 2023 DASISimulations Lakshmi Dasi Chief Technology Officer 5115 Parkcenter Ave Suite 205 Dublin, Ohio 43017 Re: K223809 Trade/Device Name: PrecisionTAVI (v1.1) Regulation Number: 21 CFR 870.1405 Regulation Name: Interventional cardiovascular implant simulation software device Regulatory Class: Class II Product Code: QQI Dated: December 19, 2022 Received: April 26, 2023 Dear Dr. Dasi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jaime Raben -S Jaime Raben, PhD Acting Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K223809 Device Name PrecisionTAVI v1.1 #### Indications for Use (Describe) PrecisionTAVI is an optional, non-invasive, post processing software solution that is indicated for patient-specific simulations of Transcatheter Aortic Valve Replacement (TAVR) during procedural planning. The software performs computer simulation to predict post TAVR in vivo valve frame deformation of clinician selected Transcatheter Heart Valve (THV) device types and sizes. The information provided by PrecisionTAVI is intended for use by cardiologists, and clinical specialists, and is not intended in any way to eliminate, replace, or substitute for, in part, the healthcare provider's judgment and analysis of the patient's condition receiving the images retains the responsibility for interpreting and validating all information and making all patient treatment decisions. PrecisionTAVI is not intended to replace the simulated device's instructions for use for final TAVR device selection and placement. Type of Use (Select one or both, as applicable) | <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K223809 510(K) SUMMARY ## SUBMITTER | Company Name: | DASI Simulations | |---------------|-------------------------------| | Address: | 5115 Parkcenter Ave Suite | | | Dublin OH 43017 United States | | Contact: | Lakshmi Prasad Dasi | | Phone number: | 614-586-6760 | | Email: | lakshmi.dasi@dasisim.com | | Date prepared | May 26, 2023 | ## DEVICE | Device Name: | PrecisionTAVI | |----------------------|------------------------------------------------------------------| | Classification Name: | Interventional Cardiovascular Implant Simulation Software Device | | Regulation Number: | 870.1405 | | Product Code: | QQI | ## PREDICATE DEVICE | Predicate # | K214066 | |---------------|------------------| | Device Name: | FEops HEARTguide | | Product Code: | QQI | {4}------------------------------------------------ ## DEVICE DESCRIPTION DASI Simulations PrecisionTAVI is a computer simulation device that predicts implant frame deformation after implantation of a Transcatheter Heart Valve (THV) device. The simulation combines a predefined THV device model and size with a patient-specific model of the patient's anatomy thereby predicting the post deployment deformation of the THV and the anatomy. The simulation results are intended to be used by qualified clinicians as additional information for planning transcatheter aortic valve replacement (TAVR). PrecisionTAVI conducts TAVR device deployment simulation using proprietary computational modeling technology. The input for the simulation is a 3D model of the patient anatomy. The 3D model is generated from 2D medical images of the patient anatomy (multi-slice Cardiac Computed Tomography). The simulation output is a report with 3D visualization capability to depict the predicted deformed THV in the deformed patient-specific anatomy of the aortic valve and root. The 3D model generation and the report generation from the simulation is performed by trained operators at DASI Simulations using an established workflow. The report is accessible to the end user as a download from the DASI Simulations portal with a standard web browser. ### INDICATIONS FOR USE PrecisionTAVI is an optional, non-invasive, post processing software solution that is indicated for patient-specific simulations of Transcatheter Aortic Valve Replacement (TAVR) during procedural planning. The software performs computer simulation to predict post TAVR in vivo valve frame deformation of clinician selected Transcatheter Heart Valve (THV) device types and sizes. The information provided by PrecisionTAVI is intended for use by cardiologists, and clinical specialists, and is not intended in any way to eliminate, replace, or substitute for, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition. The clinician receiving the images retains the responsibility for interpreting and validating all information and making all patient treatment decisions. PrecisionTAVI is not intended to replace the simulated device's instructions for use for final TAVR device selection and placement. #### INDICATIONS FOR USE COMPARISON PrecisionTAVI has similar indications and the same intended use in comparison to the predicate device. The difference in indications for use is attributed to the different disease/anatomy (aortic valve vs. left atrial appendage). #### TECHNOLOGICAL COMPARISON - 1. Technical Method: Both the subject and predicate devices perform simulations to predict the {5}------------------------------------------------ post-implantation device deformation by combining a predefined device model with a patient specific model of the patient anatomy. - 2. Target Area: The predicate device's target area is left atrial appendage, while the subject device's target area is the aortic valve/aortic root. - 3. Functionality: Both the subject and predicate device enable visualization of CT scan images (2D and 3D), generate patient specific anatomical models of aortic root (subject device) and left atrial appendage (predicate device); enable visualization of post-deployment device deformation: Transcatheter Heart Valve device (subject device) and Transcatheter Atrial Appendage Occlusion device (predicate device). The predicate device has measurement tools in addition to the functionality described above. - 4. Interpretation of Images: Both the subject and predicate devices outputs are interpreted by health care professionals. - 5. Modeling Strategy: Both the subject and predicate devices provide simulation outputs of the mechanical interaction between the device specific computational model and patient specific anatomical model. - Software Architecture: Both subject and predicate devices have DICOM data provided by 6. clinical users as input, while the simulations results are prepared by qualified engineers/ analysts and made available with the use of web-based viewers. - Simulated Devices: Subject device provides simulation for the Balloon-expandable Edwards 7. SAPIEN 3 or Ultra devices, while the predicate device provides simulation results for Boston Scientific WATCHMAN, Boston Scientific WATCHMAN FLX, and Abbott Amplatzer Amulet devices. ### TESTING SUMMARY PrecisionTAVI performance has been validated against post-TAVR CT images of 89 patients and with a registered error of < ±2mm for mean inflow, waist, and outflow diameter of the nominally deployed THV in ≥95% of cases, and ≤ ±0.1 for THV aspect ratio for the out-of-round deformation of the nominally deployed THV in ≥95% of cases. More than 80% of cases have been found in agreement in both clinician and engineer qualitative assessments. THV deployment validations were performed using data derived from clinical deployments in anatomy of 89 patients that had a tricuspid aortic valve morphology, had received a SAPIEN S3/Ultra THV (Edwards Lifesciences, Irvine, CA, USA), and had pre-TAVR as well as post-TAVR CT imaging available. THV deformation prediction validation was performed by comparing simulated and physical/clinical device deployments. A computational model of the pre-operative aortic root geometry was reconstructed from the pre-TAVR CT image data. Then, the 3D THV model was virtually deployed into the pre-TAVR anatomy using the PrecisionTAVI software. Also, the THV geometries were reconstructed from post-operative CT image data. Qualitative and quantitative validations were then performed by comparing the simulated THV to post-TAVR CT derived deformed THV. {6}------------------------------------------------ The mean THV diameter at the inflow, waist and outflow regions was used to compare the agreement in predicting the THV deformation and the aspect ratio was used to evaluate agreement in predicting the out-of-round deformation of the THV. Precision TAVI showed excellent accuracy in predicting the deployment as seen on post-TAVR CTA with 100%, 97% and 98% of all cases showing less than ± 2mm difference in THV diameter measured at the inflow, waist, and outflow regions respectively and 97%. 98% and 98% of all cases showing less than ±0.1 difference in THV aspect ratio for the out-of-round deformation between Precision TAVI simulated output and post-TAVR CTA, thus satisfying the acceptance criteria of ± 2mm for mean inflow, waist, and outflow diameter of the nominally deployed THV in ≥95% of cases, and ±0.1 for THV aspect ratio for the out-of-round deformation in ≥95% of cases. Qualitative performance of Precision TAVI was evaluated by five (5) experienced independent clinicians in the TAVR space and three (3) trained DASI Simulations engineers. This was comprised of a side-by-side comparison of the TAVR post-procedural images (clinical outputs) with the Precision TAVI images (simulated outputs) for the (1) inflow, (2) waist, (3) outflow regions and (4) vertical cross section for each case. Clinicians were presented with comparison image pairs spanning a range of all device sizes, age, sex, and calcium distributions and asked to respond whether they found the simulated outputs to be in close agreement with the clinical outputs. Overall, 96% of all case evaluations (48/50) were found to be in agreement with 90.5% (181/200) cut planes testing successfully, demonstrating excellent qualitative performance. Engineers were presented with comparison image pairs spanning a range of all device sizes, age, sex, and calcium distributions and asked to respond whether they found the simulated outputs to be in close agreement with the clinical outputs. Overall, 97% and 99% of cases were found to be in agreement in eccentricity and apposition of the THV stent, thus satisfying the acceptance criteria of > 80%. ## CONCLUSIONS The characteristics that determine the functionality and performance of DASI Simulations PrecisionTAVI, the subject device, are substantially equivalent to the predicate device. The testing indicates that the subject device is as safe, as effective, and performs as well as the predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...