Toro 88 Superbore Catheter

K252297 · Toro Neurovascular, Inc. · QJP · Jan 16, 2026 · Cardiovascular

Device Facts

Record IDK252297
Device NameToro 88 Superbore Catheter
ApplicantToro Neurovascular, Inc.
Product CodeQJP · Cardiovascular
Decision DateJan 16, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Toro 88 Superbore Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

Device Story

Toro 88 Superbore Catheter is a variable stiffness, single-lumen catheter for introducing interventional devices into vasculature. Device features stainless steel/nitinol wire reinforcement, internal lubricious liner, and distal hydrophilic coating. Single radiopaque tip marker allows fluoroscopic visualization. Proximal end includes Luer lock hub. Used in clinical settings by physicians to navigate vasculature over steerable guidewires or microcatheters. Device facilitates delivery of interventional tools; benefits patient by enabling minimally invasive access to peripheral, coronary, and neurovascular sites.

Clinical Evidence

Bench testing only. No clinical data. Testing included dimensional verification, simulated use, tensile strength, coating lubricity, particulate evaluation, kink resistance, liquid/air leakage, flowrate, distal tip stiffness, radiopacity, connector compatibility, and static burst pressure. Biocompatibility testing per ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, complement activation, thrombogenicity, PTT, platelet/leukocyte counts) confirmed safety.

Technological Characteristics

Variable stiffness single-lumen catheter; stainless steel and nitinol wire reinforcement; polymers of varying durometers; internal lubricious liner; distal hydrophilic coating. Dimensions: 118 cm length, 0.088" ID. Sterilized via Ethylene Oxide (SAL 10^-6). Complies with ISO 10555-1:2023, ISO 80369-7:2021, ISO 80369-20:2024, and IEC 62366-1:2015/Amd 1:2020.

Indications for Use

Indicated for introduction of interventional devices into peripheral, coronary, and neuro vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION January 16, 2026 Toro Neurovascular, Inc. Ryan Walker Co-Founder and Chief Operating Officer 470 Wald Irvine, California 92618 Re: K252297 Trade/Device Name: Toro 88 Superbore Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: December 18, 2025 Received: December 19, 2025 Dear Ryan Walker: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252297 - Ryan Walker Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252297 - Ryan Walker Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, NAIRA MURADYAN -S Naira Muradyan, PhD Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252297 | | | Device Name Toro 88 Superbore Catheter | | | Indications for Use (Describe) The Toro 88 Superbore Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} toro neurovascular K252297 # 510(k) Summary | Date Prepared: | January 13, 2026 | | --- | --- | | Primary Device Classification Name: | Catheter, Percutaneous, Neurovasculature | | Secondary Device Classification Name: | Catheter, Percutaneous | | Device Name: | Toro 88 Superbore Catheter | | Applicant: | Toro Neurovascular, Inc. 470 Wald Irvine, CA 92618 | | Applicant Contact: | Ryan Walker Chief Operating Officer 657-229-1108 | | Regulation Number: | 21 CFR 870.1250 | | Primary Classification Product Code: | QJP | | Secondary Classification Product Code: | DQY | | Regulation Medical Specialty: | Cardiovascular | | 510(k) Review Panel: | Neurology | | Predicate Device: | TracStar LDP Large Distal Platform; Zoom 88 Large Distal Platform; Zoom 88 Large Distal Platform Support (K240948) | ## Device Description: The Toro 88 Superbore Catheter is a variable stiffness, single lumen catheter designed to introduce interventional devices into the vasculature. The catheter shaft design includes stainless steel and nitinol wires and polymers of varying durometers. The catheter incorporates an internal lubricious liner to facilitate its advancement over a steerable guidewire or microcatheter. To reduce friction during manipulation, a hydrophilic coating is applied to the distal exterior section of the catheter shaft. A single radiopaque tip marker provides visualization under fluoroscopy. The proximal end of the catheter includes a hub with Luer lock. The catheter is supplied sterile, for single use only. ## Indications for Use Statement: The Toro 88 Superbore Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. Page 1 of 7 {5} C toro neurovascular K252297 Comparison with the Predicate Device: | Device Attribute | TracStar Large Distal Platform Zoom 88 Large Distal Platform Zoom 88 Large Distal Platform Support (K240948) | Toro 88 Superbore Catheter (K252297) | | --- | --- | --- | | Indications for Use Statement | The TracStar LDP Large Distal Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The Zoom 88 Large Distal Platform and Zoom 88 Large Distal Platform Support are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | The Toro 88 Superbore Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | | Primary Classification Product Code | QJP | Same as K240948 | | Secondary Classification Product Code | DQY | Same as K240948 | | Regulation Number | 21 CFR 870.1250 | Same as K240948 | | Regulation Medical Specialty | Cardiovascular | Same as K240948 | | Review Panel | Neurology | Same as K240948 | | Design Feature(s) | | | | Tip | Beveled Distal Edge, Soft, Flexible, Atraumatic | Straight Distal Edge, Soft, Flexible, Atraumatic | | Dimensional Specifications | | | | Proximal Inner Diameter | 0.088" | Same as K240948 | | Proximal Outer Diameter | 0.110" | 0.108" | | Distal Inner Diameter | 0.088" | Same as K240948 | | Distal Outer Diameter | 0.106" | Same as K240948 | | Effective Length | TracStar LDP: 80 – 105 cm Zoom 88 LDP: 110 cm Zoom 88 LDP Support: 100 cm | 118 cm | | Materials | | | | Distal Catheter Shaft | Reinforced with metals and polymers | Same as K240948 | | Proximal Catheter Shaft | Reinforced with metals and polymers | Same as K240948 | | Coating | Hydrophilic coating | Same as K240948 | {6} C toro neurovascular K252297 | Device Attribute | TracStar Large Distal Platform Zoom 88 Large Distal Platform Zoom 88 Large Distal Platform Support (K240948) | Toro 88 Superbore Catheter (K252297) | | --- | --- | --- | | Materials | Commonly used medical grade plastics and metals with hydrophilic coating | Same as K240948 | | Accessories and Packaging | | | | Accessories | Rotating Hemostasis Valve | Split Introducer Sheath | | Packaging Configuration | The catheters are placed in a protective polyethylene tube, mounted with accessory RHV onto a polyethylene packaging card, placed into a pouch, sealed and labeled. The sealed pouch and IFU are placed in a labeled shelf carton box. | The catheter is placed in a protective polyethylene tube, mounted with split introducer sheath onto a high-density polyethylene backer card, placed into a pouch, sealed, and labeled. The sealed pouch and Instructions for Use are placed inside a labeled shelf carton. | | Shelf Life | | | | Shelf Life | 36 Months | 12 Months | | Sterility | | | | Sterilization Method | Ethylene Oxide | Same as K240948 | | Sterility Assurance Level (SAL) | 10^{-6} | Same as K240948 | Page 3 of 7 {7} C toro neurovascular K252297 Biocompatibility: | Test | Test Method | Acceptance Criteria | Results | | --- | --- | --- | --- | | L929 MEM Elution Test – ISO | ISO 10993-5:2009 | To be non-cytotoxic, the device must achieve a Grade of 2 or lower. | Pass: Non-cytotoxic. | | Kligman Maximization Test – ISO | ISO 10993-10:2021 | Using the scoring system of Magnusson and Kligman, the device must achieve a Grade < 1. | Pass: Non-sensitizing. | | Intracutaneous Injection Test – ISO | ISO 10993-23:2021 | To be a non-irritant, the mean score difference between the test device and vehicle control must be ≤ 1.0. | Pass: Non-irritant. | | Systemic Injection Test – ISO | ISO 10993-11:2017 | The Toro 88 Superbore Catheter passes if none of the animals injected with device extracts show a significantly greater biological response than the control animals. | Pass: Non-toxic. | | Rabbit Pyrogen Test (Material-Mediated) – ISO | ISO 10993-11:2017 USP <151> | To meet pyrogen absence requirements, no rabbit should show a temperature increase of 0.5°C or more above baseline. | Pass: Non-pyrogenic. | | Rabbit Blood Hemolysis Test (Direct Method) – ASTM | ASTM F756-17 ISO 10993-4:2017 | To be non-hemolytic, the hemolytic index must be < 2%. | Pass: Non-hemolytic. | | Rabbit Blood Hemolysis Test (Indirect Method) – ASTM | ASTM F756-17 ISO 10993-4:2017 | To be non-hemolytic, the hemolytic index must be < 2%. | Pass: Non-hemolytic. | | SC5b-9 Complement Activation Test (Direct Contact) – ISO | ISO 10993-4:2017 | To be a non-activator of the complement system, there should be no statistically significant increase between SC5b-9 concentrations in the plasma exposed to the device and that of the plasma exposed to the negative control article, the untreated control, and/or a commercially available device. | Pass: Non-activator of the complement system. | | Thrombogenicity Study in Dogs – ISO | ISO 10993-4:2017 | The Toro 88 Superbore Catheter is non-thrombogenic if thrombus formation is comparable to or lesser than a commercially available device. | Pass: Comparable. | | Partial Thromboplastin Time Test (Direct Contact) – ASTM | ASTM F2382-18 ISO 10993-4:2017 | The Toro 88 Superbore Catheter is non-reactive if the PTT% is not significantly different from the untreated control, negative control, and/or a commercially available device. | Pass: No effect on coagulation of human plasma. | | Platelet and Leukocyte Count Test (Direct Contact) – ASTM | ASTM F2888-19 ISO 10993-4:2017 | The Toro 88 Superbore Catheter will be considered comparable to the reference article if the platelet count of the test article is within 80-120% of the reference article. | Pass: No effect on platelet and leukocyte concentrations. | Page 4 of 7 {8} toro neurovascular Packaging: | Test | Test Method Summary | Result | | --- | --- | --- | | Visual Inspection | The Toro 88 Superbore Catheter was evaluated per ASTM F1886/F1886M-16. | Pass | | Seal Width | The Toro 88 Superbore Catheter was evaluated by measuring pouch seals to verify sufficient seal width to ensure sterility after distribution. | Pass | | Bubble Leak | The Toro 88 Superbore Catheter was evaluated per ASTM F2096-11 (2019). | Pass | | Seal Strength | The Toro 88 Superbore Catheter was evaluated per ASTM F88/F88M-23. | Pass | Sterilization: | Test | Test Method Summary | Result | | --- | --- | --- | | Ethylene Oxide Residual | The Toro 88 Superbore Catheter was evaluated per ISO 10993-7:2008/Amd 1:2019. | Pass | | Ethylene Chlorohydrin Residual | The Toro 88 Superbore Catheter was evaluated per ISO 10993-7:2008/Amd 1:2019. | Pass | | Bacterial Endotoxin Testing | The Toro 88 Superbore Catheter was evaluated per AAMI/ANSI ST72:2019 and USP-<85>. | Pass | Performance Data – Bench: | Test | Test Method Summary | Result | | --- | --- | --- | | Design Verification | | | | Dimensional Verification | The Toro 88 Superbore Catheter was evaluated to ensure the dimensional specifications were met for inner diameter, outer diameter, coating length, and working length. | Pass | | Visual Inspection | The Toro 88 Superbore Catheter was evaluated to ensure the device met the surface requirements per ISO 10555-1:2023, §4.7. | Pass | | Simulated Use | The Toro 88 Superbore Catheter was evaluated to verify the performance of the device as it relates to device preparation, maneuverability, durability, compatibility, burst pressure, and torque strength. | Pass | | Tensile Strength | The Toro 88 Superbore Catheter was tested in accordance with ISO 10555-1:2023, Annex B to ensure the device has sufficient strength to withstand tensile forces throughout a procedure. | Pass | | Coating Lubricity | The Toro 88 Superbore Catheter was evaluated to ensure the frictional forces are comparable to or less than the predicate device. | Pass | Page 5 of 7 {9} Page 6 of 7 | Test | Test Method Summary | Result | | --- | --- | --- | | Particulate Evaluation | The Toro 88 Superbore Catheter was evaluated to ensure the particulate count and size distribution (≥10μm, ≥25μm, and ≥50μm) are comparable to or less than the predicate device. | Pass | | Coating Integrity | The Toro 88 Superbore Catheter was evaluated post-simulated use to ensure the coating was free of defects that contribute to unacceptable particulate generation. | Pass | | Kink Resistance | The Toro 88 Superbore Catheter was evaluated to ensure the catheter shaft does not kink when navigating anatomically relevant bends with varying radii. | Pass | | Liquid Leakage | The Toro 88 Superbore Catheter was evaluated to ensure the device remains leak free when tested in accordance with ISO 10555-1:2023, Annex C. | Pass | | Flowrate | The Toro 88 Superbore Catheter was evaluated to ensure the flowrate of the device was comparable to the predicate device when tested in accordance with ISO 10555-1:2023, Annex E. | Pass | | Distal Tip Stiffness | The Toro 88 Superbore Catheter was evaluated to ensure the distal tip stiffness of the device was comparable to or lower than the predicate device. | Pass | | Radiopacity | The Toro 88 Superbore Catheter was evaluated to ensure the marker band is radiopaque. | Pass | | Small-Bore Connectors | The Toro 88 Superbore Catheter was evaluated to ensure the dimensional and performance requirements were met per ISO 80369-7:2021 and ISO 80369-20:2024. | Pass | | Compatibility | The Toro 88 Superbore Catheter was evaluated to ensure it is compatible with syringes and contrast media. | Pass | | Air Leakage | The Toro 88 Superbore Catheter was evaluated to ensure the device remains leak free when tested in accordance with ISO 10555-1:2023, Annex D. | Pass | | Static Burst Pressure | The Toro 88 Superbore Catheter was tested in accordance with ISO 10555-1:2023, Annex F to ensure the device does not burst under pressure that could be seen when performing contrast injections. | Pass | {10} toro neurovascular | Test | Test Method Summary | Result | | --- | --- | --- | | Torque Strength | The Toro 88 Superbore Catheter was evaluated to ensure the torque strength of the device was comparable to commercially available devices. | Pass | | Design Validation and Usability Engineering | | | | Design Validation | The Toro 88 Superbore Catheter was evaluated to validate the performance of the device, including its packaging and labeling, with intended users. | Pass | | Usability Engineering | The Toro 88 Superbore Catheter was evaluated for usability per IEC 62366-1:2015/Amd 1:2020. | Pass | Performance Data – Animal: A determination of substantial equivalence is based upon successful completion of bench performance testing. Performance Data – Clinical: A determination of substantial equivalence is based upon successful completion of bench performance testing. Conclusion: The Toro 88 Superbore Catheter is substantially equivalent to the TracStar Large Distal Platform, Zoom 88 Large Distal Platform, and Zoom 88 Large Distal Platform Support based on the same intended use and similar technological characteristics. The differences do not raise new or different questions of safety and effectiveness. The successful completion of non-clinical testing supports that the device meets the design specifications and performs as intended. Page 7 of 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...