QFJ · Intravascular Bleed Monitor

Cardiovascular · 21 CFR 870.1345 · Class 2

Overview

Product CodeQFJ
Device NameIntravascular Bleed Monitor
Regulation21 CFR 870.1345
Device ClassClass 2
Review PanelCardiovascular

Identification

An intravascular bleed monitor is a probe, catheter, or catheter introducer that measures changes in bioimpedance and uses an algorithm to detect or monitor progression of potential internal bleeding complications.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the intravascular bleed monitor is subject to the following special controls: - (1) In vivo animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use and evaluate the following: - (i) Device performance characteristics; - Adverse effects, including gross necropsy and histopathology; and (ii) - (iii) Device usability, including device preparation, device handling, and user interface. - (2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: - (i) Tensile testing of joints and materials; - Mechanical integrity testing: (ii) - (iii) Friction testing; - (iv) Flush testing: - (v) Air leakage and liquid leakage testing; - (vi) Latching and unlatching testing; - (vii) Kink and bend testing; - (viii) Insertion force testing; - Torque testing; (ix) - Corrosion testing; and (x) - (xi) Dimensional tolerance testing. - (3) Performance data must support the sterility and pyrogenicity of the device components intended to be provided sterile. - (4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. - (5) The patient contacting components of the device must be demonstrated to be biocompatible. - (6) Software verification, validation, and hazard analysis must be performed. - (7) Performance data must demonstrate electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety. - (8) Human factors performance evaluation must demonstrate that the user can correctly use the device, based solely on reading the directions for use. - (9) Labeling must include: - Instructions for use: (i) - (ii) A shelf life and storage conditions: - (iii) Compatible procedures: - (iv) A sizing table; and - Quantification of blood detected. (v)

*Classification.* Class II (special controls). The special controls for this device are:(1) In vivo animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use and evaluate the following: (i) Device performance characteristics; (ii) Adverse effects, including gross necropsy and histopathology; and (iii) Device usability, including device preparation, device handling, and user interface. (2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Tensile testing of joints and materials; (ii) Mechanical integrity testing; (iii) Friction testing; (iv) Flush testing; (v) Air leakage and liquid leakage testing; (vi) Latching and unlatching testing; (vii) Kink and bend testing; (viii) Insertion force testing; (ix) Torque testing; (x) Corrosion testing; and (xi) Dimensional tolerance testing. (3) Performance data must support the sterility and pyrogenicity of the device components intended to be provided sterile. (4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (5) The patient contacting components of the device must be demonstrated to be biocompatible. (6) Software verification, validation, and hazard analysis must be performed. (7) Performance data must demonstrate electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety. (8) Human factors performance evaluation must demonstrate that the user can correctly use the device, based solely on reading the directions for use. (9) Labeling must include: (i) Instructions for use; (ii) A shelf life and storage conditions; (iii) Compatible procedures; (iv) A sizing table; and (v) Quantification of blood detected.

Recent Cleared Devices (2 of 2)

RecordDevice NameApplicantDecision DateDecision
K230273Saranas Early Bird Bleed Monitoring SystemSaranas, Inc.May 25, 2023SESE
DEN180021Early Bird Bleed Monitoring SystemSaranas, Inc.Mar 1, 2019DENG

Top Applicants

Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...