Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL Catheter
K240225 · Shockwave Medical, Inc. · PPN · Mar 22, 2024 · Cardiovascular
Device Facts
Record ID
K240225
Device Name
Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL Catheter
Applicant
Shockwave Medical, Inc.
Product Code
PPN · Cardiovascular
Decision Date
Mar 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Shockwave Medical IVL System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, iliofemoral, popliteal, infra-popliteal, and renal arteries. Not for use in the coronary, carotid or cerebral vasculature.
Device Story
System uses catheter-based lithotripsy to treat calcified peripheral artery stenoses. Device consists of IVL catheter with integrated balloon and lithotripsy emitters, connected to an external IVL generator via a cable. Physician delivers catheter to lesion site; generator sends energy to emitters, creating acoustic pressure pulses that disrupt calcium. Disruption allows balloon dilatation at low pressure. Used in clinical settings by physicians. System components exchange data during treatment. Benefits include improved vessel compliance and successful dilatation of difficult-to-treat calcified lesions.
Clinical Evidence
Bench testing only. No clinical data required. Verification and validation testing included guidewire/sheath compatibility, balloon diameter/length/inflation/deflation, crossing profile, tensile/torsional strength, emitter/marker integrity, system leakage, RBP, fatigue, sonic output, pulse count/rate, temperature rise, particle count, and simulated use.
Technological Characteristics
Percutaneous catheter system; integrated balloon with lithotripsy emitters; inflation/guidewire lumens; radiopaque marker bands. Energy source: electrical (generator-driven acoustic pressure pulses). Connectivity: data exchange between generator, cable, and catheter. Sterilization: standard medical device methods. Software: embedded system for generator-catheter data exchange.
Indications for Use
Indicated for patients requiring lithotripsy-enhanced balloon dilatation of calcified stenotic lesions in peripheral arteries (iliac, femoral, iliofemoral, popliteal, infra-popliteal, renal). Contraindicated for use in coronary, carotid, or cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Shockwave Medical Intravascular Lithotripsy System (K191840)
Reference Devices
Shockwave Medical Intravascular Lithotripsy System (K203365)
Related Devices
K240954 — Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL Catheter · Shockwave Medical, Inc. · May 7, 2024
K203365 — Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave M5+ Peripheral IVL Catheter · Shockwave Medical, Inc. · Apr 22, 2021
K221852 — Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL Catheter · Shockwave Medical, Inc. · Aug 16, 2022
K191840 — Peripheral Intravascular Lithotripsy (IVL) Catheters, IVL Generator and Connector Cable Kit, IVL Connector Cable (Replacement) · Shockwave Medical, Inc. · Aug 7, 2019
K242213 — Shockwave Intravascular Lithotripsy (IVL) System with the Javelin Peripheral Intravascular Lithotripsy (IVL) Catheter · Shockwave Medical, Inc. · Sep 27, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 22, 2024
Shockwave Medical, Inc. Anna Bushart Senior Regulatory Affairs Specialist 5403 Betsy Ross Dr. Santa Clara, California 95054
Re: K240225
Trade/Device Name: Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PPN Dated: January 26, 2024 Received: January 26, 2024
Dear Anna Bushart:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Digitally signed by Ariel Ariel G. Ash-G. Ash-shakoor -S Date: 2024.03.22 shakoor -S 15:15:08 -04'00'
For
Gregory O'Connell Assistant Director
{2}------------------------------------------------
DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240225
Device Name
Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL Catheter
Indications for Use (Describe)
The Shockwave Medical IVL System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, iliofemoral, popliteal, infra-popliteal, and renal arteries. Not for use in the coronary, carotid or cerebral vasculature.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Shockwave Medical. The word "SHOCKWAVE" is written in a bold, sans-serif font, with the letters in a dark gray color. To the left of the word "SHOCKWAVE" are three concentric circles in a light blue color. Below the word "SHOCKWAVE" is the word "MEDICAL" in a smaller, lighter gray font.
Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL Catheter K240225 510(k) Summary Page 1 of 3
# K240225 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92.
# Name, Address, and Phone Number of Applicant
Shockwave Medical, Inc. 5403 Betsy Ross Drive Santa Clara, CA 95054 Phone: (510) 279-4262
# Contact Person
Anna Bushart
# Date Prepared
January 26, 2024
#### Device Name and Classification
| Trade Name: | Shockwave Intravascular Lithotripsy (IVL) System with the<br>Shockwave E8 Peripheral IVL Catheter |
|----------------------|---------------------------------------------------------------------------------------------------|
| Common Name: | Catheter, lithotripsy, peripheral, transluminal |
| CFR Classification: | 21 CFR 870.1250 |
| Classification Name: | Percutaneous catheter |
| Product Code: | PPN |
#### Predicate Device
The predicate device is the Shockwave Medical Intravascular Lithotripsy System, K191840, cleared by FDA on August 7, 2019.
#### Reference Device
The reference device is the Shockwave Medical Intravascular Lithotripsy System, K203365, cleared by FDA on April 22, 2021.
#### Indications for Use / Intended Use
The Shockwave Medical IVL System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, iliofemoral, popliteal, infra-popliteal, and renal arteries. Not for use in the coronary, carotid or cerebral vasculature.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Shockwave Medical. The word "SHOCKWAVE" is in large, bold, gray letters. The word "MEDICAL" is in smaller, gray letters below the word "SHOCKWAVE". There are blue circles around the letter "O" in "SHOCKWAVE".
# Device Description
The IVL Catheter is a proprietary lithotripsy device delivered through the peripheral arterial system of the lower extremities to the site of an otherwise difficult to treat calcified stenosis. Energizing the lithotripsy device will generate acoustic pressure pulses within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. The IVL Catheter is comprised of an integrated balloon with an array of integrated lithotripsy emitters for the localized delivery of acoustic pressure pulses. The system consists of an IVL Catheter, an IVL Connector Cable and an IVL Generator.
The Shockwave E Peripheral IVL Catheter shaft contains an inflation lumen, a guidewire lumen, and the lithotripsy emitters. The emitters are positioned along the length of the balloon working length for delivery of acoustic pressure pulses. The balloon is located near the distal tip of the catheter. Two radiopaque marker bands within the balloon denote the length of the balloon to aid in positioning of the balloon during treatment. The balloon is designed to provide an expandable segment of known length and diameter at a specific pressure.
The IVL Generator and Connector Cable are used with a Shockwave Medical IVL Catheter to deliver localized, lithotripsy-enhanced, balloon dilatation of calcified, stenotic arteries. The IVL Generator, IVL Connector Cable and IVL Catheters are designed to exchange data during patient treatment.
# Technological Comparison
This Traditional 510(k) Premarket Notification describes dimensional modifications to the predicate Shockwave S4 Peripheral IVL Catheter and reference Shockwave M** Peripheral IVL Catheter including increased balloon length, increased catheter length and additional emitter channel.
The catheter labeling was updated to reference the name of the modified device, Shockwave E Peripheral IVL Catheter. Additionally, the Shockwave E6 labeling includes a sterile cable sleeve in the packaging to form a convenience kit.
The IVL System has the same intended use, similar principles of operation and has substantially equivalent technological characteristics including same fundamental scientific technology, design, energy source, shelf life, and sterilization as the 510(k) cleared IVL System.
#### Summary of Performance Data
Objective evidence demonstrating that the IVL System design output meets the product design input requirements as well as that device performance characteristics conform to user needs and intended uses as defined in the product specification was provided. Testing was conducted in accordance with Shockwave Medical's Risk Analysis procedures, applicable FDA guidance documents and relevant international standards. Testing included:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Shockwave Medical. The word "SHOCKWAVE" is in large, bold, gray letters. The "O" in "SHOCKWAVE" has a design of concentric circles emanating from it. Below "SHOCKWAVE" is the word "MEDICAL" in smaller, gray letters.
- IVL Catheter design verification and validation testing: ●
- Guidewire compatibility O
- Introducer sheath compatibility O
- Nominal balloon diameter O
- Balloon diameter at Rated Burst Pressure (RBP) O
- Balloon length O
- Balloon inflation time O
- Balloon deflation time O
- Useable catheter length O
- Balloon crossing profile O
- Catheter distal tip O
- Distal tip durability O
- Catheter bonds tensile strength O
- Catheter torsional strength O
- O Emitter and marker band integrity
- System leakage O
- Minimum balloon RBP O
- Balloon fatigue (multiple inflations) O
- O Sonic output
- Catheter pulse count and pulse rate O
- Temperature rise O
- Catheter particle count O
- Catheter connector length O
- Catheter connection O
- Catheter identification O
- Catheter sterility (visual inspection) O
- Cable sleeve packaging O
- Catheter compatibility with materials and accessories commonly used in Over-the-O Wire (OTW) peripheral balloon angioplasty procedures
- Simulated use testing o
Results demonstrated that the performance of the IVL System meets its design specifications and demonstrates substantial equivalence for its intended use; therefore, additional clinical data were not required.
# Basis for Substantial Equivalence
The IVL Catheter with dimensional modifications shares the same intended use, similar principles of operation, overall technical and functional capabilities, and similar design and materials as the identified predicate device. Any differences between the IVL Systems were evaluated through design verification and validation testing which demonstrated device performance and confirmed that there are no new questions of safety or effectiveness. The modified IVL Catheter is therefore substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.