Reprocessed Carto Vizigo Bi-Directional Guiding Sheath

K212165 · Innovative Health, LLC · PNE · Mar 10, 2022 · Cardiovascular

Device Facts

Record IDK212165
Device NameReprocessed Carto Vizigo Bi-Directional Guiding Sheath
ApplicantInnovative Health, LLC
Product CodePNE · Cardiovascular
Decision DateMar 10, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The Reprocessed Carto Vizigo Bi-Directional Guiding Sheath is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. The sheath curve can be visualized when used with compatible Carto 3 EP Navigation Systems.

Device Story

Reprocessed steerable guiding sheath designed for cardiac catheter introduction/maneuverability. Features hemostasis valve, sideport with three-way stopcock, and handle with rotating collar for bi-directional tip deflection (≤ 180°). Includes distal vent holes and radiopaque tip marker for fluoroscopic visualization. Compatible with Carto 3 EP Navigation Systems for sheath curve visualization. Used by physicians in clinical settings to facilitate catheter access to cardiac anatomy. Reprocessed by Innovative Health via cleaning, decontamination, inspection, and functional testing. Limited to one reprocessing cycle; serialized and tracked.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility, cleaning/sterilization validation, visual inspection, dimensional verification, dynamic continuity, simulated use, leak testing, mechanical characteristics, and electrical safety (dielectric and current leakage).

Technological Characteristics

Steerable guiding sheath; 8.5F inner diameter, 11.5F outer diameter, 71 cm length. Features hemostasis valve, sideport, and bi-directional deflection handle. Radiopaque tip marker. Compatible with Carto 3 EP Navigation Systems. Reprocessed via cleaning, decontamination, and functional testing. Single-use reprocessing limit.

Indications for Use

Indicated for introducing cardiovascular catheters into the heart, including the left side via the interatrial septum. For use by clinicians in a prescription-only setting.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of a human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 10, 2022 Innovative Health, LLC. Amanda Babcock Regulatory Affairs Manager 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257 Re: K212165 Trade/Device Name: Reprocessed Carto Vizigo 8.5F Bi-Directional Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: PNE Dated: March 7, 2022 Received: March 8, 2022 Dear Amanda Babcock: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # K212165 - Amanda Babcock The item number included in the scope of this submission is as follows: | Product Description | Item Number | Curve size | Outer Diameter | Usable length | Inner Diameter | Number of Electrode | |--------------------------------------------------------------------|----------------|---------------|----------------|---------------|----------------|---------------------| | Reprocessed<br>Carto Vizigo<br>Bi-Directional<br>Guiding<br>Sheath | D138501 | Small (17 mm) | 11.5F | 71 cm | 8.5F | 4 | | D138502 | Medium (22 mm) | 11.5F | 71 cm | 8.5F | 4 | | | D138503 | Large (50 mm) | 11.5F | 71 cm | 8.5F | 4 | | {4}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212165 Device Name Reprocessed Carto Vizigo 8.5F Bi-Directional Guiding Sheath Indications for Use (Describe) The Reprocessed Carto Vizigo Bi-Directional Guiding Sheath is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. The sheath curve can be visualized when used with compatible Carto 3 EP Navigation Systems. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|----------------------------------------------------------------------| | <span style="text-decoration: underline;"></span> | | | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ## SECTION 5: 510(k) SUMMARY As required by 21 CFR 807.92(c) #### Submitter's Name and Address: Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257 ### Contact Name and Information: Amanda Babcock Requlatory Affairs Manager Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7705 (fax) ababcock@innovative-health.com #### Date prepared: July 09, 2021 #### Device Information: Trade/Proprietary Name: Reprocessed Carto Vizigo 8.5F Bi-Directional Guiding Sheath Common or Usual Name: Guiding Sheath Classification Name: Reprocessed Catheter Introducer Classification Number: Class II, 21 CFR 870.1340 Product Code: PNE ### Predicate Device: | 510(k) Number | Device | Manufacturer | |---------------|--------------------------------------------|------------------| | K170997 | Carto Vizigo Bi-Directional Guiding Sheath | Biosense Webster | #### Reference Device: | 510(k) Number | Device | Manufacturer | |---------------|---------------------------------------------|-------------------------| | K170311 | Reprocessed Agilis NxT Steerable Introducer | Innovative Health, LLC. | ### Device Description: The Reprocessed Carto Vizigo Bi-Directional Guiding Sheath is designed to provide accessibility and maneuverability in the cardiac anatomy. The steerable sheath is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with three-way stopcock is provided for air or blood aspiration, and fluid infusion. A handle equipped with a rotating collar to deflect the tip clockwise ≤ 180° and counterclockwise ≤ 180°. The steerable sheath features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to allow fluoroscopic visualization. Note: Only the steerable sheath and dilator are subject of this submission. The quidewire is purchased off-the shelf (K935170) and packaged with the reprocessed devices. {6}------------------------------------------------ The item numbers in scope of this submission are as follows: | Product<br>Description | Item<br>Number | Curve<br>size | Outer<br>Diameter | Usable<br>length | Inner<br>Diameter | Number of<br>Electrode | |--------------------------------------------------------------------|----------------|-------------------|-------------------|------------------|-------------------|------------------------| | Reprocessed<br>Carto Vizigo<br>Bi-Directional<br>Guiding<br>Sheath | D138501 | Small<br>(17 mm) | 11.5F | 71 cm | 8.5F | 4 | | | D138502 | Medium<br>(22 mm) | 11.5F | 71 cm | 8.5F | 4 | | | D138503 | Large<br>(50 mm) | 11.5F | 71 cm | 8.5F | 4 | Table 5.1: Device Scope ## Indications for Use: The Reprocessed Carto Vizigo Bi-Directional Guiding Sheath is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. The sheath curve can be visualized when used with compatible Carto 3 EP Navigation Systems. ## Technological Characteristics: The purpose, design, materials, function, and intended use of the Reprocessed device is identical to the predicate device. There are no changes to the claims, clinical applications, patient populations, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of the reprocessed device includes removal of visible soil and decontamination. Each device is inspected, and function tested prior to packaging and labeling. ## Functional and Safety Testing: Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Carto Vizigo Bi-Directional Guiding Sheath. This included the following: - Biocompatibility ● - Cleaning Validation . - Sterilization Validation - Functional testing ● - · Visual Inspection - = Dimensional Verification - . Dynamic Continuity - . Simulated Use - . Leak - Mechanical Characteristics ■ - . Electrical Safety Testing - . Dielectric and Current Leakage - o Packaging Validation The device is reprocessed no more than one (1) time. Each device is marked, serialized and tracked. After the device has reached the maximum number of reprocessing cvcles. the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors. ## Conclusion: Innovative Health concludes that the Reprocessed Carto Vizigo Bi-Directional Guiding Sheath is substantially equivalent to the predicate devices described herein.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...