NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter
K223026 · Northeast Scientific, Inc. · OWQ · Feb 3, 2023 · Cardiovascular
Device Facts
Record ID
K223026
Device Name
NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter
Applicant
Northeast Scientific, Inc.
Product Code
OWQ · Cardiovascular
Decision Date
Feb 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheters are designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in cerebral vessels. The NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.
Device Story
Reprocessed diagnostic intravascular ultrasound (IVUS) catheter; utilizes cylindrical ultrasound transducer array to radiate acoustic energy and detect echoes; generates real-time cross-sectional images of coronary and peripheral vessel lumen and wall structures. Used in clinical settings as adjunct to angiography; operated by physicians. Catheter tracks over 0.014" guidewire; features radiopaque markers for fluoroscopic visualization; includes hydrophilic coating. Requires connection to Volcano s5 or CORE Series imaging systems for operation. Reprocessing includes cleaning, decontamination, inspection, and functional testing; limited to one reprocessing cycle by manufacturer.
Clinical Evidence
Bench testing only. Validation included cleaning, sterilization, visual inspection, dimensional verification, simulated use, mechanical characteristics, hydrophilic coating performance, system compatibility, drying, packaging, and biocompatibility.
Technological Characteristics
Cylindrical ultrasound transducer array; 3.5F size; 150 cm length; 0.014" guidewire compatible; hydrophilic coating; radiopaque markers. Reprocessed device; restricted to one cycle. Compatible with Volcano s5/CORE Series systems.
Indications for Use
Indicated for evaluation of vascular morphology in coronary and peripheral vasculature; adjunct to conventional angiography for imaging vessel lumen and wall structures. Not indicated for cerebral vessels.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter (K210655)
NES Reprocessed Visions PV .014p RX Digital IVUS Catheter (K200195)
Related Devices
K200195 — NES Reprocessed Visions PV .014P RX Digital IVUS Catheter · Northeast Scientific, Inc. · Jun 22, 2020
K250592 — NES Reprocessed Visions PV .018 Digital IVUS Catheter (R-86700) · Northeast Scientific, Inc. · May 12, 2025
K210655 — Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter · Innovative Health, LLC · May 25, 2022
K230584 — Vein360 Reprocessed Visions PV.035 Digital IVUS Catheter · Vein 360, LLC · Jun 6, 2023
K181126 — Reprocessed Visions PV .035 Digital IVUS Catheter · Innovative Health, LLC · Jan 18, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
February 3, 2023
Northeast Scientific, Inc. Matt Farley Director of Product Development 2142 Thomaston Ave. Waterbury, Connecticut 06704
Re: K223026
Trade/Device Name: NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWO Dated: November 30, 2022 Received: November 30, 2022
Dear Matt Farley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
The item numbers included in the scope of this submission are as follows:
| Description | Item Number | French Size | Guide Wire | Minimum<br>Sheath | Length (cm) |
|------------------------------------------------|-------------|-------------|------------|-------------------|-------------|
| Eagle Eye Platinum RX<br>Digital IVUS Catheter | 85900P | 3.5F | 0.014" | 5F | 150 cm |
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K223026
Device Name
NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter
Indications for Use (Describe)
The NES Reprocessed Eagle Eve Platinum RX Digital IVUS Catheters are designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in cerebral vessels.
The NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
### SECTION 5: 510(k) SUMMARY
As required by 21 CFR 807.92(c)
#### Submitter's Name and Address:
Northeast Scientific, Inc. (NES) 2142 Thomaston Ave. Waterbury, CT 06704
### Contact Name and Information:
Matt Farley Director of Product Development Northeast Scientific, Inc. 203-756-2111 (office) 203-757-5532 (fax) matt@smarthealth-care.com
### Date Prepared:
November 30th, 2022
### Device Information:
| Trade/Proprietary Name: | NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter |
|-------------------------|-------------------------------------------------------------|
| | |
| Common or Usual Name: | Ultrasonic Imaging Catheter |
| Classification Name: | Diagnostic Intravascular Catheter |
| Review Panel: | Cardiovascular |
| Classification Number: | Class II, 21 CFR 870.1200 |
| Product Code: | OWO. OBJ |
### Predicate Device:
| 510(k) Number | 510(k) Title | Manufacturer | Applicant |
|---------------|--------------------------------------------------------------|----------------------------|-------------------------------|
| K210655 | Reprocessed Eagle Eye Platinum RX<br>Digital IVUS Catheter | Northeast Scientific, Inc. | Innovative Health, LLC. |
| K200195 | NES Reprocessed Visions PV .014p RX<br>Digital IVUS Catheter | Northeast Scientific, Inc. | Northeast<br>Scientific, Inc. |
Table 5.1 – Predicate Device
#### Device Description:
The NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter incorporates a cylindrical ultrasound transducer array. The array radiates acoustic energy into the surrounding tissue and detects the subsequent echoes. The information from the echoes is used to generate real-time images of the coronary and peripheral vessels.
The NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter utilizes an internal lumen that allows the catheter to track over the 0.014" (0.36 mm) guide wire exits from the guide wire lumen approximately 24 cm proximal to the catheter is introduced percutaneously or via surgical cutdown into the vascular system.
{5}------------------------------------------------
Three 1 mm-long radiopaque markers are incorporated on the internal lumen positioned 10 mm apart from distal edge to distal edge, starting 10 mm from the proximal edge of the scanner marker tube normally visible under fluoroscopy.
The Eagle Eye catheter may only be used with the Volcano s5 Series or CORE Series of Systems Operator's Manual. This catheter will not operate if connected to any other imaging system.
A hydrophilic coating is applied externally to a distal portion of the catheter.
| The item numbers in scope of this submission are as follows: | |
|--------------------------------------------------------------|--|
|--------------------------------------------------------------|--|
| Description | Item Number | French Size | Guide Wire | Minimum<br>Sheath | Length (cm) |
|------------------------------------------------|-------------|-------------|------------|-------------------|-------------|
| Eagle Eye Platinum RX<br>Digital IVUS Catheter | 85900P | 3.5F | 0.014" | 5F | 150 cm |
Table 5.2 – Device Scope
## Indications for Use:
The NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a crosssectional image of such vessels. This device is not currently indicated for use in cerebral vessels.
The NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.
## Technological Characteristics:
The purpose, design, materials, function, and intended use of the Reprocessed Catheter are identical to the predicate devices. There are no changes to the claims, clinical applications, patient populations, performance specifications, or method of operation. In addition, Northeast Scientific's reprocessing of the Catheter includes removal of visible soil and decontamination. Each device is inspected and functionally tested prior to packaging and labeling.
## Functional and Safety Testing:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter. This included the following:
- . Cleaning Validation
- . Sterilization Validation
- Functional testing
- Visual Inspection O
- Dimensional Verification O
- o Simulated Use
- Mechanical Characteristics O
- o Hydrophilic Coating
- O System Compatibility
- Drying Validation ●
- Packaging Validation
- Biocompatibility
K223026 Traditional 510(k)
{6}------------------------------------------------
The catheter is reprocessed no more than one (1) time. Each device is marked during reprocessing. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Northeast Scientific. Northeast Scientific restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
## Conclusion:
Northeast Scientific, Inc. concludes that the NES Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter is as safe and effective as the predicate devices described herein.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.