The Reprocessed AcuNav Diagnostic Ultrasound Catheter is intended for intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients. The reprocessed device is not indicated for use with pediatric patients.
Device Story
Reprocessed AcuNav Diagnostic Ultrasound Catheter; distal end features ultrasound transducer for 2-D imaging; four-way articulation steering mechanism for image plane orientation. Used for intracardiac and intraluminal visualization of cardiac/great vessel anatomy and physiology and other devices in heart. Operated by clinicians in clinical settings. Device is reprocessed single-use catheter; Innovative Health performs cleaning, decontamination, inspection, and functional testing. Reprocessed maximum one time; tracked via marking. Benefits include clinical visualization of cardiac structures and devices during procedures.
Clinical Evidence
Bench testing only. Testing included biocompatibility, cleaning validation, sterilization validation, functional testing (visual inspection, dimensional verification, ultrasound transducer testing, simulated use, mechanical characteristics), electrical safety testing (dielectric and current leakage), and packaging validation.
Technological Characteristics
Intravascular ultrasound catheter with 2-D transducer and four-way articulation steering. Reprocessed device; materials and design identical to predicate. Includes cleaning, decontamination, and functional testing. Single-use, reprocessed max one time. Electrical safety per dielectric and current leakage standards.
Indications for Use
Indicated for intracardiac and intraluminal visualization of cardiac and great vessel anatomy, physiology, and other devices in adult patients. Not indicated for pediatric patients.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
AcuNav Diagnostic Ultrasound Catheter 8F and 10F (K071234)
K170474 — Reprocessed SoundStar 3D Diagnostic Ultrasound Catheter · Innovative Health, LLC · Jun 16, 2017
K033436 — REPROCESSED DIAGNOSTIC ULTRASOUND CATHETER · Alliance Medical Corp. · Mar 16, 2005
K250545 — Medline ReNewal Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter (8F, 8FG, 10F, 10FG) · Surgical Instrument Service and Savings, Inc. · Jun 5, 2025
K243101 — Medline ReNewal Reprocessed Siemens ACUSON AcuNav 10F Ultrasound Catheter (for Siemens Systems) (08255790 ); Medline ReNewal Reprocessed Siemens ACUSON AcuNav 10F Ultrasound Catheter (for GE Systems) (10043342) · Surgical Instrument Service and Savings, Inc. · Mar 10, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black and are connected by flowing lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2016
Innovative Health, LLC Rafal Chudzik Director of Engineering 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257
Re: K153090
Trade/Device Name: Reprocessed AcuNav Diagnostic Ultrasound Catheters Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWQ Dated: February 5, 2016 Received: February 8. 2016
Dear Rafal Chudzik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Item<br>Number | Description | Length<br>(cm) | French Size |
|----------------|------------------------------------------|----------------|-------------|
| 08255790 | AcuNav Diagnostic<br>Ultrasound Catheter | 90 | 10 |
| 10135936 | AcuNav Diagnostic<br>Ultrasound Catheter | 90 | 8 |
| 10135910 | AcuNav Diagnostic<br>Ultrasound Catheter | 90 | 8 |
| 10043342 | AcuNav Diagnostic<br>Ultrasound Catheter | 90 | 10 |
Reprocessed Single-Use Device Models Included in Clearance:
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153090
#### Device Name
Reprocessed AcuNav Diagnostic Ultrasound Catheter
Indications for Use (Describe)
The Reprocessed Acunav Diagnostic Ultrasound Catheter is intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients,
The reprocessed device is not indicated for use with pediatric patients.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# SECTION 5: 510(k) SUMMARY
As required by 21 CFR 807.92(c)
# Submitter's Name and Address:
Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257
# Contact Name and Information:
Rafal Chudzik Director of Engineering (602) 326-7716 (cell) (480) 525-6006 (office) (844) 965-9359 (fax) rchudzik@innovative-health.com
# Date prepared:
October 23, 2015
### Device Information:
| Trade/Proprietary Name: | Reprocessed AcuNav™ Diagnostic Ultrasound Catheters |
|-------------------------|-----------------------------------------------------|
| Classification Name: | Reprocessed Intravascular Ultrasound Catheter |
| Common Name: | Diagnostic Ultrasound Catheter |
| Classification Number: | Class II, 21 CFR 870.1200 |
| Product Code: | OWQ |
### Predicate Device:
| 510(k) Number | 510(k) Title | Manufacturer |
|---------------|-----------------------------------------------------|------------------------------|
| K071234 | AcuNav Diagnostic Ultrasound Catheter<br>8F and 10F | Siemens Medical<br>Solutions |
### Reference Device:
| 510(k) Number | 510(k) Title | Reprocessor |
|---------------|-----------------------------------------------|--------------------------------|
| K063076 | Reprocessed Diagnostic Ultrasound<br>Catheter | Ascent Healthcare<br>Solutions |
### Device Description:
The Reprocessed AcuNav Diagnostic Ultrasound Catheters distal end has an ultrasound transducer providing 2-D imaging. A steering mechanism controls the image plane orientation through four-way articulation of the tip.
{5}------------------------------------------------
# Indications for Use:
The Reprocessed AcuNav Diagnostic Ultrasound Catheter is intended for intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients.
The reprocessed device is not indicated for use with pediatric patients.
# Technoloqical Characteristics:
The purpose, design, materials, function, and intended use of the Reprocessed Diagnostic Ultrasound Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of the Diagnostic Ultrasound Catheter includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.
# Functional and Safety Testing:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Diagnostic Ultrasound Catheters. This included the following:
- · Biocompatibility
- · Cleaning Validation
- · Sterilization Validation
- · Functional Testing
- · Visual Inspection
- · Dimensional Verification
- · Ultrasound Transducer Testing
- Simulated Use
- · Mechanical Characteristics
- · Electrical Safety Testing
- · Dielectric and Current Leakage
- · Packaging Validation
The Reprocessed Diagnostic Ultrasound Catheters are reprocessed no more than one (1) time. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
# Conclusion:
Innovative Health concludes that the Reprocessed Diagnostic Ultrasound Catheter is as safe and effective as the predicate devices described herein.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.