XC11 ICE System, USA

K250955 · Yorlabs, Inc. · OBJ · Oct 22, 2025 · Cardiovascular

Device Facts

Record IDK250955
Device NameXC11 ICE System, USA
ApplicantYorlabs, Inc.
Product CodeOBJ · Cardiovascular
Decision DateOct 22, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1200
Device ClassClass 2
AttributesPediatric

Indications for Use

The XC11 ICE Box Ultrasound System is intended for intra-cardiac and intra-luminal ultrasound imaging when operated with the XC11 ICE Catheter and XC11 ICE System Handle. The XC11 ICE Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. The XC11 ICE Catheter is for use with YorLabs XC11 ICE System Handle only.

Device Story

System provides real-time 2D intracardiac/intraluminal ultrasound imaging for interventional cardiology guidance. Comprised of 9F single-use sterile catheter with 64-element phased array transducer and XC11 ICE Box processing unit with touchscreen tablet. Catheter features four-way articulation for steering and on-handle imaging controls. System uses wireless connectivity between components. Operated by clinicians in interventional settings to visualize anatomy and devices during procedures. Output displayed on tablet for real-time guidance; assists in procedural navigation and device placement. Benefits include precise visualization of cardiac structures without diagnostic ultrasound requirements.

Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, thermal, electrical, electromagnetic, and mechanical safety testing, and a GLP animal study.

Technological Characteristics

9F single-use catheter with 64-element phased array transducer; four-way articulation steering. Wireless-enabled connectivity. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, IEC 62359, and ANSI C63.27. System includes image acquisition modules and touchscreen tablet interface.

Indications for Use

Indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy, physiology, and other devices in adult and pediatric patients during cardiac interventional percutaneous procedures. For imaging guidance only; not for treatment delivery.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 22, 2025 Yorlabs, Inc. Wanda Carpinella Regulatory Consultant 2210 Faraday Avenue, Suite 100 Carlsbad, California 92008 Re: K250955 Trade/Device Name: XC11 ICE System, USA Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ, IYN, IYO, ITX Dated: September 22, 2025 Received: September 22, 2025 Dear Wanda Carpinella: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K250955 - Wanda Carpinella Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250955 - Wanda Carpinella Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, **Hetal B. Odobasic -S** for Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} XC11 ICE System | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250955 | ? | | Please provide the device trade name(s). | | ? | | XC11 ICE System, USA | | | | Please provide your Indications for Use below. | | ? | | The XC11 ICE Box Ultrasound System is intended for intra-cardiac and intra-luminal ultrasound imaging when operated with the XC11 ICE Catheter and XC11 ICE System Handle. The XC11 ICE Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. The XC11 ICE Catheter is for use with YorLabs XC11 ICE System Handle only. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K250955 Page 1 of 3 Summary Traditional Premarket Notification 510(k) Summary YorLabs XC11 ICE System, USA 510(k) Number K250955 Date Prepared: October 21, 2025 I. SUBMITTER YorLabs, Inc. 2210 Faraday Avenue, Suite 100 Carlsbad, CA 92008 Contact Person Glen McLaughlin, PhD YorLabs, Inc. 2210 Faraday Avenue, Suite 100 Carlsbad, CA 92008 II. DEVICE Name of Device: XC11 ICE System, USA Common or Usual Name: Intracardiac imaging system Classification Regulation: 21 CFR § 897.1200 Regulation Name: Diagnostic Intravascular Catheter Product Codes: OBJ (Class 2) – Catheter, Ultrasound, Intravascular IYN, IYO, ITX III. PREDICATE Primary Predicate: Acuson P500 Ultrasound Imaging System (K163396) Secondary Predicate: AcuNav Diagnostic Ultrasound Catheter 8F, 10F (K071234) IV. DEVICE DESCRIPTION The XC11 ICE System, USA is a fully integrated imaging platform designed for wireless communications and a compact footprint. It provides precise, real-time visualization of intracardiac anatomy and devices positioned within the heart. The System is comprised of two main components: XC11 ICE Catheter (2D), USA A 9F, single-use, sterile device capable of real-time 2D side-view imaging for intracardiac and intraluminal ultrasound in adult and pediatric patients. The catheter features easy steering maneuverability via deflection and rotation, along with on-handle imaging controls for enhanced usability. XC11 ICE Box Ultrasound System, USA This unit houses the image data acquisition and computer modules for image display and includes a connector for interfacing with the XC11 ICE System Handle. The XC11 ICE Tablet with a touchscreen interface, enables technicians to control the system software efficiently. {5} K250955 Page 2 of 3 # V. INDICATIONS FOR USE The XC11 ICE Box Ultrasound System is intended for intra-cardiac and intra-luminal ultrasound imaging when operated with the XC11 ICE Catheter and XC11 ICE System Handle. The XC11 ICE Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. The XC11 ICE Catheter is for use with YorLabs XC11 ICE System Handle only. # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATES ## Intended Use The subject device and predicate are ultrasound system intended for intracardiac echocardiography (ICE) using a catheter-based ultrasound probe. The XC11 ICE System is designed primarily for interventional cardiology procedures that utilize catheter-based ICE imaging. It is not intended for diagnostic ultrasound imaging. As such, reference to diagnostic ultrasound procedures have been removed from the indications for use for the subject device. ## Technological Features The XC11 ICE Box Ultrasound System shares similar technological features to the ACUSON P500 Ultrasound System, and the XC11 ICE Catheter is technologically consistent to the AcuNav Ultrasound Catheter 8F and 10F used for intracardiac echocardiography. The design of the XC11 ICE Box Ultrasound System is simplified compared to the ACUSON P500 Ultrasound System with only the features necessary for interventional catheterization procedures that use ICE imaging for guidance only. The XC11 ICE Box Ultrasound System, like the ACUSON P500 Ultrasound System, support wireless-enabled connectivity to its components. The XC11 ICE Catheter, like the AcuNav Ultrasound Catheter 8F and 10F, incorporates a 64-element phased array transducer embedded at catheter's distal tip for 2D imaging, and it has steering mechanism controls through four-way articulation to manipulate the imaging orientation to facilitate intracardiac imaging. # VII. Summary of Non-Clinical Tests The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The XC11 ICE System complies with FDA-recognized standards: - IEC 60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety, 2005/A2:2021 - IEC 60601-1-2 Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbance - Requirements and Tests, Edition 4.1, 2020 - IEC 60601-2-37, Medical Electrical Equipment – Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, Edition 2.1, 2015 - ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within Risk Management Process, Fifth edition, 2018 - ISO 14971, Application of risk management to medical devices, 2019 {6} - IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related medical diagnostic ultrasonic fields, Edition 2.1, 2017 - ANSI C63.27 /D1.0: American National Standards Institute - Standard for Evaluation of Wireless The following quality assurance measures are applied to the development of the system: - Risk Analysis - Requirements Reviews - Design Reviews - Testing on unit level (Module verification) - Integration testing (System verification) - Performance testing (Verification & Validation) - GLP Animal Study ## VIII. Summary of Clinical Tests The subject of this premarket submission, XC11 ICE System, did not require clinical studies to support substantial equivalence. ## IX. CONCLUSION In conclusion, the XC11 ICE System is as safe and effective as the commercially available predicate for its intended use as demonstrated by the performance data for the subject device and is deemed substantially equivalent. K250955 Page 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...