This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Device Story
OptiCross™ HD is a sterile, short-rail intravascular ultrasound (IVUS) imaging catheter. It consists of an imaging core with a 60 MHz radial ultrasonic transducer at the distal tip and a catheter body with a distal imaging window. The device connects to a Motordrive Unit (MDU5 PLUS™) via an electro-mechanical interface. During use, the catheter is flushed with heparinized saline to provide acoustic coupling. It is operated by a physician in a clinical setting (e.g., cath lab) during coronary interventions. The transducer rotates to capture cross-sectional ultrasound images of coronary vessels. The output is displayed on an IVUS system, allowing the physician to visualize intravascular pathology, assess lesion length, and guide interventional procedures. The device benefits patients by providing high-resolution imaging to inform clinical decision-making during coronary procedures.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance data, including bench-top testing (deliverability, crossability, image resolution, measurement accuracy, pullback reliability), biocompatibility (ISO 10993-1), acoustic output testing (FDA Track 1 limits), electromagnetic compatibility (IEC 60601-1-2), and a GLP pre-clinical porcine study to validate in-vivo image quality.
Technological Characteristics
60 MHz radial ultrasonic transducer; short-rail catheter design; electro-mechanical connector interface; heparinized saline acoustic coupling; radiopaque markers; telescoping shaft for 15 cm linear movement; compatible with minimum 5F guide catheter and 0.014" guidewire. Complies with IEC 60601-1-2 (EMC) and ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for patients who are candidates for transluminal coronary interventional procedures requiring ultrasound examination of coronary intravascular pathology.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
November 16, 2017
Boston Scientific Corporation Mugdha Dongre Regulatory Affairs Specialist II 47215 Lakeview Boulevard Fremont, California 94538
Re: K173284
Trade/Device Name: OptiCross™ HD 60 MHz Coronary Imaging Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ, ITX Dated: October 13, 2017 Received: October 16, 2017
Dear Mugdha Dongre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Mugdha Dongre
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Wilhelm
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173284
Device Name
OptiCross™ HD 60 MHz Coronary Imaging Catheter
Indications for Use (Describe)
This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary per 21 CFR 807.92
| Submitter's Name and | Boston Scientific Corporation | | |
|----------------------|-----------------------------------------------------------------|---------|------------------|
| Address | 47215 Lakeview Boulevard | | |
| | Fremont, CA 94538 | | |
| | USA | | |
| Contact Name and | Mugdha Dongre | | |
| Information | Regulatory Affairs Specialist II | | |
| | Tel: 510.624.1314 | | |
| | Fax: 510.440.7698 | | |
| | Email: Mugdha.Dongre@bsci.com | | |
| Date Prepared | October 13, 2017 | | |
| Trade Name | OptiCross™ HD 60MHz Coronary Imaging Catheter | | |
| Common Name | Diagnostic Intravascular Catheter, Ultrasound Transducer | | |
| Classification Name | Catheter, Ultrasound, Intravascular (OBJ) has been | | |
| | classified as Class II per 21 CFR 870.1200 | | |
| | Transducer Ultrasonic (ITX) has been classified as Class II | | |
| Predicate Name | per 21 CFR 892.1570 | | |
| | OptiCross™ 40 MHz<br>Coronary Imaging Catheter | K123621 | 15-Apr-2013 |
| Reference Devices | OptiCross™ 6, 40 MHz | K153617 | 25-February-2016 |
| | Coronary Imaging Catheter | | |
| | | | |
| | OptiCross™ 18, 30 MHz | K160514 | 22-June-2017 |
| | Peripheral Imaging Catheter | | |
| | | | |
| Device Description | OptiCross™ HD (60 MHz Coronary Imaging Catheters) is a | | |
| | sterile, short rail imaging catheter. | | |
| | It consists of two main assemblies: | | |
| | 1. Imaging Core | | |
| | 2. Catheter Body | | |
| | The imaging core is composed of a hi-torque, flexible, rotating | | |
| | drive cable with a radial looking 60 MHz ultrasonic transducer | | |
| | at the distal tip. An electro-mechanical connector interface at | | |
| | the proximal end of the catheter makes the connection to the | | |
| | Motordrive Unit (MDU5 PLUS™). The MDU5 PLUS-catheter | | |
| | interface consists of an integrated mechanical drive socket and | | |
| | electrical connection. | | |
| | The catheter body is comprised of three sections: | | |
| | 1. Distal Imaging Window Lumen | | |
| | 2. Proximal Shaft Lumen | | |
| | 3. Telescoping Section | | |
| | The distal imaging window lumen and proximal shaft lumen | | |
| | sections comprise the "working length" of the catheter, and the | | |
{4}------------------------------------------------
| | telescoping section remains outside of the guiding catheter. |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The catheter body has a distal imaging window lumen with |
| | proximal exit 1.6 cm from the distal end. A radiopaque (RO) |
| | marker is embedded in the catheter body at 0.5 cm from the |
| | distal tip. In addition, two insertion depth markers are located |
| | on the proximal shaft lumen at 90 cm and 100 cm from the |
| | distal tip to aid in estimating catheter position relative to the |
| | distal guide catheter tip. The proximal shaft lumen is attached |
| | to the telescoping section via a strain relief connection. |
| | The telescoping shaft (section) allows the imaging core to be |
| | advanced and retracted for 15 cm of linear movement. |
| | The corresponding movement of the transducer occurs from |
| | the proximal end of the guidewire exit port to the proximal end |
| | of the distal imaging window lumen. The telescope section has |
| | proximal markers for lesion length assessment, consisting of a |
| | series of marks spaced 1 cm apart on the telescope body. |
| | A flush port with a one-way check valve is used to flush the |
| | interior of the catheter body and maintain a flushed condition. |
| | The catheter must be flushed with heparinized saline prior to |
| | use, as this provides the acoustic coupling media required for |
| | ultrasonic imaging. The one-way check valve helps retain |
| | saline in the catheter during use. |
| Intended Use/Indications for<br>Use | This catheter is intended for ultrasound examination of<br>coronary intravascular pathology only. Intravascular<br>ultrasound imaging is indicated in patients who are candidates<br>for transluminal coronary interventional procedures. |
| Device Technology<br>Characteristics and<br>Comparison to Predicate<br>Device | The OptiCross™ HD catheter is a 60 MHz ultrasound coronary<br>imaging catheter intended to operate with a BSC IVUS system<br>for diagnostic imaging. It will be used with a minimum 5F<br>(>0.058" ID) guide catheter / introducer sheath and a 0.014"<br>guidewire. |
| | The OptiCross™ HD Coronary Imaging Catheter incorporates<br>a transducer with new design features intended to utilize a<br>60MHz frequency to enhance image resolution while<br>maintaining similar depth of penetration as well as similar<br>performance characteristics as that of its predicate OptiCross™<br>40Mhz Coronary Imaging Catheter. |
| | Modifications with respect to the predicate device include: |
| | • A new transducer (new design with new materials) |
| | • Female telescope. (OCHD 5F will leverage female<br>telescope from OptiCross™ 6 [K153617] instead of<br>OptiCross) |
| | • PCA board populated with two zero Ohm resistors to<br>generate DC coupling. |
| | |
| | In support of a substantial equivalence determination, BSC has<br>compared and evaluated the material and design differences<br>between the subject and predicate device. |
| | Non-clinical performance evaluations, as described below,<br>indicate that the subject device is substantially equivalent to,<br>and at least as safe and effective as the predicate device<br>OptiCross™ 40 MHz Coronary Imaging Catheter (K123621). |
| Non-Clinical Performance<br>Data | Determination of substantial equivalence is based on an<br>assessment of non-clinical performance data. |
| | Non-clinical data includes bench-top performance testing,<br>packaging validation, biological safety, electromagnetic<br>compatibility, acoustic output testing, and a GLP pre-clinical<br>animal study to validate imaging performance in the porcine<br>model. |
| | Bench Testing:<br>Bench testing was performed to evaluate physical integrity,<br>functionality, and overall performance of the catheter.<br>Performance criteria includes deliverability, crossability,<br>guide catheter compatibility, lubricity, retraction capability,<br>image resolution, image penetration, non-uniform rotational<br>distortion (NURD), measurement accuracy, pullback<br>reliability, dimensional requirements, visibility under<br>fluoroscopy, interface with ancillary devices, environmental<br>requirements, user interface requirements, catheter robustness<br>and simulated use structural integrity. |
| | Pre-clinical Evaluation:<br>The purpose of this study was to support design validation and<br>business need evaluations claims associated with in-vivo image<br>quality. |
| | Biological Safety Testing:<br>OptiCross™ HD catheter utilizes existing materials from the<br>commercially available OptiCross™ family of catheters which<br>have similar configurations and are manufactured under the<br>same processing conditions. Hence the biocompatibility test<br>data has been leveraged from the commercially available<br>OptiCross™ Imaging Catheter to meet the requirements of ISO<br>10993-1. |
| | |
| | Electrical and Mechanical Safety:<br>Acoustic Output was evaluated in accordance with FDA<br>Guidance, Information for Manufacturers Seeking Marketing<br>Clearance of Diagnostic Ultrasound Systems and<br>Transducers (September 9, 2008). Acoustic Output test results<br>for OptiCross™ HD are below the FDA Track 1 limits.<br>Electromagnetic compatibility testing was also conducted;<br>demonstrating compliance to IEC 60601-1-2 (3rd Edition).<br><br>Packaging Validation:<br>Packaging of OptiCross™ HD catheters is equivalent to the<br>packaging of other commercially available BSC IVUS<br>catheters in the OptiCross product family. Thus, the OptiCross<br>HD packaging will leverage the packaging DV results of the<br>currently available OptiCross catheters to demonstrate it meets<br>the required product specifications.<br><br>Conclusion:<br>Non-clinical performance evaluations, as described above,<br>indicate that the subject device is substantially equivalent to,<br>and at least as safe and effective as the predicate device,<br>OptiCross™ 40 MHz Coronary Imaging Catheter (K123621). |
| Clinical Performance Data | Not applicable; determination of substantial equivalence is<br>based on an assessment of non-clinical performance data. |
| Conclusion | With respect to the predicate, design modifications<br>incorporated by OptiCross HD do not alter the fundamental<br>scientific technology or the indications for use. Based on<br>Failure Mode Effects Analysis (FMEA), comprehensive<br>verification and validation activities were successfully<br>completed; raising no new issues of safety or effectiveness.<br>Non-clinical performance data supports a determination that<br>the subject device, OptiCross HD, is substantially equivalent to<br>the predicate device, OptiCross 40 MHz Coronary Imaging<br>Catheter (K123621); and that it is at least as safe and effective<br>for its intended use |
{5}------------------------------------------------
{6}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.