THE EMBOSHIELD EMBOLIC PROTECTION SYSTEM

K052454 · Abbott Vascular Devices · NTE · Sep 14, 2005 · Cardiovascular

Device Facts

Record IDK052454
Device NameTHE EMBOSHIELD EMBOLIC PROTECTION SYSTEM
ApplicantAbbott Vascular Devices
Product CodeNTE · Cardiovascular
Decision DateSep 14, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Emboshield Embolic Protection System is indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of the Filtration Element placement should be between 2.8 mm and 6.2 mm.

Device Story

Temporary percutaneous transluminal filtration system; captures embolic material (thrombus/debris) during carotid artery angioplasty and stenting. System components: RX BareWire (PTFE-coated stainless steel guidewire with platinum radiopaque tip); RX Delivery Catheter (deploys filtration element); Filtration Element (polyurethane membrane with Nitinol arms, 140-micron perfusion pores); RX Retrieval Catheter (retractable centering catheter with distal expansile tip). Used in clinical settings by physicians; device tracks over guidewire to site; filtration element deployed to capture debris; retrieved post-procedure. Benefits: reduces risk of distal embolization during carotid interventions.

Clinical Evidence

Clinical evidence provided via the SECuRITY Registry Trial, a prospective study enrolling 399 patients across 30 sites. Results support the safety and performance of the device and its components for the intended indication.

Technological Characteristics

Components: PTFE-coated stainless steel guidewire, polyurethane membrane, Nitinol arms. Filtration element sizes: 3.0 mm to 6.0 mm. Perfusion pores: 140 microns. Radiopaque marker bands for positioning. Mechanical filtration principle. No software or energy source.

Indications for Use

Indicated for patients undergoing carotid artery angioplasty and stenting procedures requiring embolic protection; suitable for vessel diameters between 2.8 mm and 6.2 mm.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of a human figure with three heads, which are represented by curved lines. The figure is positioned to the right of the logo. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the figure. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Abbott Vascular Devices c/o Mr. Ric Ruedy Director, Regulatory Affairs 400 Saginaw Drive Redwood City, CA 94063 Re: K052454 Emboshield® Embolic Protection System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NTE Dated: March 23, 2005 Received: September 6, 2005 Dear Mr. Ruedy: This letter corrects our substantially equivalent letter of September 14, 2005. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. NOV 2 8 2005 {1}------------------------------------------------ #### Page 2 - Mr. Ric Ruedy Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Dinh R. Vahner Image /page/1/Picture/7 description: The image contains a signature on the left side and the words "Branch Director" on the right side. The signature is illegible, but it appears to be written in cursive. The words "Branch Director" are written in a simple, sans-serif font. The image is likely a document that has been signed by a branch director. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use | 510(k) Number (if known): | K052454 | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | The Emboshield <sup>®</sup> Embolic Protection System | | Indications For Use: | The Emboshield Embolic Protection System is indicated for use as a guidewire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of the Filtration Element placement should be between 2.8 mm and 6.2 mm. | Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Khemmer - (Diston Sign-Off) Division of Cardlovascular Devices 510(k) Number Kos 2454 {3}------------------------------------------------ SEP 1 4 2005 APPENDIX A. 510(k) SUMMARY Abbott Vascular Devices Emboshield® Embolic Protection System KOS2454 510(k) Premarket Notification CONFIDENTIAL | Sponsor/Submitter: | Abbott Vascular Devices (on behalf of MedNova, Ltd) | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 400 Saginaw Drive | | | Redwood City, CA 94063 | | Contact Person: | Debbie Cogan | | | Regulatory Affairs Manager | | | Phone:(650) 474-3263 | | | Fax:(650) 474-3041 | | Date of Submission: | March 22, 2005 | | Device Trade Name: | Emboshield® Embolic Protection System | | Device Common Name: | Embolic Protection System | | Device Classification: | Class II | | Regulation Number: | 21 CFR 870.1250 | | Classification Name: | Device, coronary saphenous vein bypass graft, temporary, for<br>embolization protection | | Product Code: | NFA | | Predicate Device: | Guidant RX Accunet Embolic Protection System (K042908) | | Intended Use: | The Emboshield® Embolic Protection System is indicated for<br>use as a guidewire and embolic protection system to contain<br>and remove embolic material (thrombus/debris) while<br>performing angioplasty and stenting procedures in carotid<br>arteries. The diameter of the artery at the site of the Filtration<br>Element placement should be between 2.8 and 6.2 mm. | | Device Description: | The Emboshield® Embolic Protection System is a temporary<br>percutaneous transluminal filtration system designed to capture<br>embolic material released during angioplasty and stent<br>procedures within carotid arteries. The system consists of the<br>following components:<br>RX BareWire™™ | The RX BareWire™ is a 0.014" PTFE coated stainless steel guidewire with a 3.0cm (0.014") platinum radiopaque distal tip The RX BareWire™ is available in two lengths, section. 315cm and 190cm and a variety of stiffness. {4}------------------------------------------------ ## RX Delivery Catheter The RX Delivery Catheter usable length is 135 cm. The crossing profile is between 0.048" and 0.051", depending on Filtration Element size. A pull handle is used to deploy the loaded Filtration Element from the pod. Two pairs of indicator bands are provided along the catheter shaft. A proximal pair (90 cm and 100 cm from the catheter tip) to indicate the catheter tip position during advancement through the guide catheter, and a distal pair to indicate the proximity of the RX exit port during catheter retraction. A radiopaque marker band is positioned proximal to the pod. ## Filtration Element The Filtration Element consists of a polyurethane membrane with internal Nitinol arms. There are two proximal triangular entry ports and multiple distal perfusion pores of 140microns. There is a proximal and a distal marker band. The Filtration Element is available in 3.0 mm, 4.0 mm, 5.0 mm and 6.0 mm diameters. ## RX Retrieval Catheter The RX Retrieval Catheter has a 139 cm usable length, a retractable Centering Catheter, and a distal expansile tip. The maximum outer diameter is 0.072''. A handle with a flushing port is situated at the proximal end. Two pairs of indicator bands are provided along the catheter shaft. Another proximal pair of indicator bands (90 cm and 100 cm from the catheter tip) allows indication of the catheter tip position during advancement through the guide catheter and a distal pair to indicate the proximity of the RX Retrieval Catheter exit port during catheter retraction. A radiopaque marker band is positioned approximately 3.5cm from the distal tip on the outer sheath. Abbott Vascular Devices has submitted information on Summary of Substantial Equivalence: indication for use, design and principle of operation, biocompatibility and performance characteristics to establish that Emboshield Embolic Protection System is substantially equivalent to the currently marketed predicate device. > The Emboshield Embolic Protection System has the same intended use as the predicate device. Results of scientific testing have ensured that all materials are biocompatible, no new adverse effects were introduced and physical properties are appropriate for the intended use. In vitro and in vivo testing was conducted. The SECuRITY Registry Trial was conducted {5}------------------------------------------------ enrolling a total of three hundred ninety nine (399) patients at 30 sites. In summary, the results of the testing and clinical study support the safety and performance of the Emboshield device and its components for the intended indication when used in accordance with the Instructions for Use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...