← Product Code [NLH](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH) · K200205

# Reprocessed Advisor FL Circular Mapping Catheter, Sensor Enabled (K200205)

_Stryker Sustainability Solutions · NLH · May 13, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH/K200205

## Device Facts

- **Applicant:** Stryker Sustainability Solutions
- **Product Code:** [NLH](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH.md)
- **Decision Date:** May 13, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ is a sensorenabled steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The catheter can be used to map the atrial regions of the heart. The catheter is used with the EnSite Precision™ System to combine and display magnetic processed patient positioning and navigation mapping information.

## Device Story

Steerable, flexible electrophysiology catheter; used for recording intracardiac signals and cardiac stimulation; maps atrial regions of heart. Input: intracardiac electrical signals and magnetic positioning data. Operation: physician manipulates handle-based control mechanism to deflect distal loop; catheter integrates with EnSite Precision System (impedance and magnetic field technology) to display 3D cardiac mapping and patient positioning. Output: electrical signals for diagnostic analysis and navigation data for 3D mapping. Used in clinical electrophysiology labs by physicians. Benefits: enables precise cardiac mapping and signal acquisition for diagnostic studies. Reprocessed device limited to one cycle; includes cleaning, decontamination, and functional testing.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, reprocessing validation, sterilization validation, and functional performance testing (EEPROM verification, location sensor testing, visual inspection, dimensional testing, electrode continuity/isolation, deflection testing, rigidity, leakage, tensile testing, locking mechanism, and electrical safety).

## Technological Characteristics

Thermoplastic elastomer construction; noble metal electrodes. Bi-directional steerable shaft with handle-based deflection. Sensor-enabled for magnetic navigation. Compatible with EnSite Precision System. Reprocessed via cleaning and decontamination; sterilized. Single-use limit (max 1 cycle).

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Advisor FL Circular Mapping Catheter, Sensor Enabled ([K160335](/device/K160335.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

May 13, 2020

Stryker Sustainability Solutions Mia McCorkel Senior Regulatory Affairs Specialist 18101 W Drake Drive Tempe, Arizona 85283

Re: K200205

Trade/Device Name: Reprocessed Advisor FL Circular Mapping Catheter, Sensor Enabled Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: April 14, 2020 Received: April 15, 2020

Dear Mia McCorkel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Model numbers intended for reprocessing:

| Model Number    | Description                                    | Curve | Number of<br>Loop<br>Electrodes | Electrode<br>Spacing | Loop<br>Diameter |
|-----------------|------------------------------------------------|-------|---------------------------------|----------------------|------------------|
| D-AVSE-DF10-F15 | Bi-Directional<br>Circular Mapping<br>Catheter | DF    | 10                              | 3-3-3                | 15               |
| D-AVSE-DF10-F20 | Bi-Directional<br>Circular Mapping<br>Catheter | DF    | 10                              | 5-5-5                | 20               |

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K200205

Device Name

Reprocessed Advisor FL Circular Mapping Catheter, Sensor Enabled

Indications for Use (Describe)

The Reprocessed Advisor FL Circular Mapping Catheter, Sensor Enabled is a sensor-enabled steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The catheter can be used to map the atrial regions of the heart.

The catheter is used with the EnSite Precision System to combine and display magnetic processed patient positioning and navigation mapping information.

| Type of Use (Select one or both, as applicable)                                  |                                                                      |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Stryker Sustainability Solutions. The word "stryker" is in bold, black, sans-serif font on the top line. Below that, the words "Sustainability Solutions" are also in bold, black, sans-serif font, but in a smaller size.

# SECTION 5: 510(k) SUMMARY

## Submitter:

Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, Arizona 85283

## Contact:

Mia McCorkel Senior Regulatory Affairs Specialist 480-343-1855 (c) 480-763-2965 (f) mia.mccorkel@stryker.com

#### Date of Preparation: January 27, 2020

### Name of Device:

|  | Trade/Proprietary Name: | Reprocessed Advisor FL™ Circular Mapping Catheter,<br>Sensor Enabled™                               |
|--|-------------------------|-----------------------------------------------------------------------------------------------------|
|  | Common Name:            | Diagnostic Electrophysiology Catheter, Electrode Recording<br>Catheter or Electrode Recording Probe |
|  |                         | Classification Information: Cardiovascular (21 CFR § 870.1220, NLH, Class II)                       |
|  | Model Numbers:          | D-AVSE-DF10-F15, D-AVSE-DF10-F20                                                                    |

### Predicate Device:

| Model Numbers                      | 510(k) Number | 510(k) Title                                            | Original Manufacturer |
|------------------------------------|---------------|---------------------------------------------------------|-----------------------|
| D-AVSE-DF10-F15<br>D-AVSE-DF10-F20 | K160335       | Advisor FL Circular Mapping<br>Catheter, Sensor Enabled | St. Jude Medical      |

### Device Description:

The Reprocessed Advisor™ FL Circular Mapping Catheters, Sensor Enabled™ are steerable, flexible, insulated electrophysiology catheters constructed of thermoplastic elastomer material and noble metal electrodes. The shaft curvature is manipulated by the control mechanism located on the handle at the catheter's proximal end. To adjust the curve on the bi-directional catheter, use the actuator to deflect the catheter in either direction. The distal loop is oriented counter-clockwise as viewed from the handle.

K200205

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Stryker Sustainability Solutions. The word "stryker" is written in a bold, sans-serif font in black. Below the word "stryker" is the phrase "Sustainability Solutions," also written in a bold, sans-serif font in black. The logo is simple and modern, and it conveys a sense of strength and innovation.

This device is compatible with the EnSite Precision™ System which is a three dimensional (3-D) cardiac mapping system that combines impedance and magnetic field technology.

| Model Numbers   | Description                                                         | Curve | Loop Diameter |
|-----------------|---------------------------------------------------------------------|-------|---------------|
| D-AVSE-DF10-F15 | Reprocessed Advisor FL Circular<br>Mapping Catheter, Sensor Enabled | DF    | 15mm          |
| D-AVSE-DF10-F20 | Reprocessed Advisor FL Circular<br>Mapping Catheter, Sensor Enabled | DF    | 20mm          |

The model numbers included in the scope of this submission are as follows:

# Intended Use:

The Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ is a sensorenabled steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The catheter can be used to map the atrial regions of the heart.

The catheter is used with the EnSite Precision™ System to combine and display magnetic processed patient positioning and navigation mapping information.

# Summary of Technological Characteristics:

The design, materials, and intended use of Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ are equivalent to the predicate device. The mechanism of action of the reprocessed device is identical to the predicate device in that the same standard mechanical design, materials, and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Stryker Sustainability Solutions' reprocessing of the device includes removal of adherent visible soil and decontamination. Each individual device is tested for appropriate function of its components prior to packaging and labeling operations.

# Performance Data:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™. This included the following tests:

- Biocompatibility ●
- Validation of Reprocessing ●
- Sterilization Validation
- Functional Performance Testing ●
	- о EEPROM verification
	- Location sensor testing o
	- Visual inspection O
	- Dimensional testing O
	- Electrode continuity and isolation testing O
	- Curve and in-plane deflection testing O
	- Rigidity testing o
	- Freedom from leakage O

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Stryker Sustainability Solutions. The word "Stryker" is in a bold, sans-serif font, with the letters slightly overlapping. Below "Stryker" is the phrase "Sustainability Solutions" in a smaller, bold, sans-serif font. The logo is black on a white background.

- o Tensile testina
- Locking mechanism testing
- . Electrical Safety Testing

The performance testing demonstrates that reprocessed devices are as safe and effective as the predicate and operate as originally intended.

The Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ is reprocessed no more than one (1) time. Each reprocessed device is tracked with a serial number label which is affixed to the catheter shaft. Once the device reaches the maximum number or reprocessing cycles, it is rejected and taken out of service. Reprocessing is conducted only by Stryker Sustainability Solutions. Stryker Sustainability Solutions restricts its reprocessing to exclude devises previously reprocessed by other reprocessors.

# Conclusion:

The results of bench and laboratory testing demonstrate that the Reprocessed Advisor™ FL Circular Mapping Catheters, Sensor Enabled™ are at least as safe and effective and perform as well as the identified legally marketed predicate device as described herein.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH/K200205](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH/K200205)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
