← Product Code [NLH](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH) · K153153

# Reprocessed Electrophysiology Catheter (K153153)

_Innovative Health, LLC · NLH · Mar 14, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH/K153153

## Device Facts

- **Applicant:** Innovative Health, LLC
- **Product Code:** [NLH](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH.md)
- **Decision Date:** Mar 14, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Catheters are indicated for multiple electrode electrophysiological mapping of the heart, i.e. recording or stimulation only. The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Catheters are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed LASSO NAV and Reprocessed LASSO 2515 NAV eco Variable Catheters provide location information when used with compatible CARTO 3 EP Navigation Systems. (These catheters are not compatible with CARTO 3 EP Navigation Systems prior to Version 2.3.)

## Device Story

Electrophysiology catheters used for cardiac mapping; input consists of electrical signals (electrograms) from atrial tissue; platinum/iridium electrodes capture signals; device provides location data when integrated with CARTO 3 EP Navigation System. Used in clinical electrophysiology labs by physicians; catheters inserted via 8F guiding sheath into atria. Nitinol loop design allows variable sizing (15-25mm) for custom fit in cardiac veins. Output used by physicians to visualize cardiac electrical activity and guide diagnostic procedures. Reprocessed single-use device; limited to one reprocessing cycle; includes cleaning, decontamination, and functional testing to ensure performance equivalent to original device.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, cleaning validation, sterilization validation, functional testing (visual inspection, dimensional verification, electrical performance, simulated use, mechanical characteristics, electrical safety, dielectric and current leakage), and packaging validation.

## Technological Characteristics

Catheters feature platinum/iridium electrodes on a circular spine (Nitinol loop for variable models). Sizes: 4.5F or 4F distal tips; 7F French size; 115cm length. Compatible with CARTO 3 EP Navigation Systems (v2.3+). Reprocessed via cleaning, decontamination, and functional testing. Limited to one reprocessing cycle.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- LASSO 2515 NAV eco Variable Catheter and LASSO NAV eco Catheter ([K113213](/device/K113213.md))
- LASSO NAV Catheter ([K093376](/device/K093376.md))
- Variable LASSO NAV Catheter, Models: D-1290-01, D-1290-2 ([K081258](/device/K081258.md))

## Reference Devices

- Reprocessed 2515 NAV Variable Electrophysiology Catheter ([K112292](/device/K112292.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes an abstract symbol of a human figure. The symbol is composed of three stylized profiles facing to the right, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the seal.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 14, 2016

Innovative Health, LLC Rafal Chudzik Director of Engineering 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257

Re: K153153

Trade/Device Name: Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: February 5, 2016 Received: February 8, 2016

Dear Rafal Chudzik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Description                                                    | Item<br>Number | Number of<br>Electrodes | Electrode<br>Spacing (mm) | Loop<br>Diameter<br>or Curve<br>(mm) | French<br>Size | Length<br>(cm) |
|----------------------------------------------------------------|----------------|-------------------------|---------------------------|--------------------------------------|----------------|----------------|
| Lasso Nav<br>eco Fixed<br>Diagnostic<br>EP Catheter            | D134901        | 10                      | 4.5                       | 15                                   | 7              | 115            |
|                                                                | D134902        | 20                      | 4.5 pairs                 | 15                                   | 7              | 115            |
|                                                                | D134903        | 10                      | 6.0                       | 20                                   | 7              | 115            |
|                                                                | D134904        | 20                      | 6.0 pairs                 | 20                                   | 7              | 115            |
|                                                                | D134905        | 10                      | 8.0                       | 25                                   | 7              | 115            |
|                                                                | D134906        | 20                      | 8.0 pairs                 | 25                                   | 7              | 115            |
| Lasso 2515<br>Nav eco<br>Variable<br>Diagnostic<br>EP Catheter | D134301        | 20                      | 2-6-2                     | 25-15                                | 7              | 115            |
|                                                                | D134302        | 10                      | 8.0                       | 25-15                                | 7              | 115            |

## Reprocessed Single-Use Device Models Included in Clearance:

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K153153

Device Name

Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheter

Indications for Use (Describe)

The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Catheters are indicated for multiple electrode electrophysiological mapping of the heart, i.e. recording or stimulation only. The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Catheters are designed to obtain electrograms in the atrial regions of the heart.

The Reprocessed LASSO NAV and Reprocessed LASSO 2515 NAV eco Variable Catheters provide location information when used with compatible CARTO 3 EP Navigation Systems. (These catheters are not compatible with CARTO 3 EP Navigation Systems prior to Version 2.3.)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## SECTION 5: 510(k) SUMMARY

As required by 21 CFR 807.92(c)

#### Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

#### Contact Name and Information:

Rafal Chudzik Director of Engineering (602) 326-7716 (cell) (480) 525-6006 (office) (844) 965-9359 (fax) rchudzik@innovative-health.com

#### Date prepared:

October 30, 2015

## Device Information:

| Trade/Proprietary Name: | Reprocessed LASSO® NAV eco and LASSO 2515 NAV<br>eco Variable Electrophysiology (EP) Catheter |
|-------------------------|-----------------------------------------------------------------------------------------------|
| Common Name:            | Electrophysiological Mapping Catheter                                                         |
| Classification Name:    | Catheter, Recording, Electrode, Reprocessed                                                   |
| Classification Number:  | Class II, 21 CFR 870.1220                                                                     |
| Product Code:           | NLH                                                                                           |

## Predicate Device:

| 510(k) Number | 510(k) Title                                                         | Manufacturer     |
|---------------|----------------------------------------------------------------------|------------------|
| K113213       | LASSO® 2515 NAV eco Variable<br>Catheter and LASSO® NAV eco Catheter | Biosense Webster |
| K093376       | LASSO® NAV Catheter                                                  | Biosense Webster |
| K081258       | Variable LASSO® NAV Catheter,<br>Models: D-1290-01, D-1290-2         | Biosense Webster |

## Reference Device:

| 510(k) Number | 510(k) Title                                                | Reprocessor                         |
|---------------|-------------------------------------------------------------|-------------------------------------|
| K112292       | Reprocessed 2515 NAV Variable<br>Electrophysiology Catheter | Stryker Sustainability<br>Solutions |

## Device Description:

#### Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter

{5}------------------------------------------------

The Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter have been designed to facilitate electrophysiological mapping of the atria of the heart with the CARTO 3 EP Navigation System and a reference device. Both catheters are deployed in the right or left atrium through an 8F quiding sheath. Both deflectable catheters consists of a 4.5 F (Lasso NAV eco) or 4 F (LASSO NAV eco Variable) circular spine on the distal tip, with platinum/iridium electrodes that can be used for stimulation and recording.

The Reprocessed LASSO 2515 NAV eco Variable Catheter features a Nitinol loop design that allows the expansion and contraction of the loop to custom-fit veins with different sizes, ranging from 25mm to 15mm diameter (± 15%).

The Reprocessed LASSO NAV eco Catheter is a fixed catheter with three fixed catheter sizes, namely 15, 20, 25 mm loop sizes to accommodate different vein sizes, Each loop size will be available with either 10 or 20 electrodes.

## Indications for Use:

#### Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter

The Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter are indicated for multiple electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. These catheters are designed to obtain electrograms in the atrial regions of the heart.

The Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter provide location information when used with compatible CARTO EP Navigation Systems. (These catheters are not compatible with CARTO 3 EP Navigation Systems prior to Version 2.3).

Note: Only the catheter is the subject of this submission. Any other related equipment is not included in the scope of this submission.

#### Technoloqical Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

## Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter. This included the following:

{6}------------------------------------------------

- Biocompatibility ●
- Cleaning Validation .
- Sterilization Validation .
- Functional Testing ●
	- Visual Inspection ●
	- Dimensional Verification
	- Electrical Performance ●
	- Simulated Use .
	- Mechanical Characteristics ●
	- Electrical Safety Testing
	- Dielectric and Current Leakage .
- . Packaging Validation

The Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter are reprocessed no more than one (1) time. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

## Conclusion:

●

Innovative Health concludes that the Reprocessed LASSO 2515 NAV eco Variable Catheter and Reprocessed LASSO NAV eco Catheter are as safe and effective as the predicate devices described herein.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH/K153153](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLH/K153153)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
