K030026 · Alliance Medical Corp. · NLH · May 2, 2003 · Cardiovascular
Device Facts
Record ID
K030026
Device Name
REPROCESSED ELECTROPHYSIOLOGY CATHETER
Applicant
Alliance Medical Corp.
Product Code
NLH · Cardiovascular
Decision Date
May 2, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
Reprocessed Electrophysiology Catheters are intended for temporary intracardiac sensing, recording, stimulation, and electrophysiological mapping of cardiac structures.
Device Story
Reprocessed diagnostic EP catheters; transmit electrical impulses; used for endocardial recording or stimulation. Device components: handpiece, flexible shaft, distal tip with diagnostic electrodes. Deflectable models allow curve manipulation via handpiece; fixed curve models have established shapes. Used in clinical electrophysiology studies by physicians. Reprocessing involves removal of adherent visible soil, decontamination, and functional testing of components prior to packaging. Output consists of electrical signals for cardiac mapping/stimulation; aids clinical decision-making regarding cardiac rhythm management.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility testing, validation of the reprocessing procedure, sterilization validation, and functional testing of device components to ensure performance remains consistent with original specifications.
Technological Characteristics
Diagnostic electrode catheter; handpiece-actuated distal tip deflection or fixed-curve geometry. Materials and design identical to OEM predicates. Reprocessing includes decontamination and functional verification. Sterilization validation performed. No software or electronic processing components.
Indications for Use
Indicated for temporary intracardiac sensing, recording, stimulation, and electrophysiological mapping of cardiac structures in patients undergoing electrophysiology studies. Catheter cables are indicated for use with appropriate electrode catheters during these studies.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 0 6 2004
Alliance Medical Corporation c/o Ms. Moira Barton Senior Regulatory Affairs Specialist 10232 South 51st Street Phoenix, Arizona 85044
Re: K030026
Trade Name: Reprocessed Electrophysiology (EP) Catheters & Cables Regulation Number: 21 CFR 870.1220 & 870.2900 Regulation Name: Electrode recording catheter and Patient transducer and electrode cable (including connector) Regulatory Class: Class II (two) Product Code: NLH & DSA Dated: April 7, 2003
Dear Ms. Barton:
Received: April 7, 2003
This letter corrects our substantially equivalent letter of May 02, 2003 regarding one of the product codes for your devices. The code was erroneously written as NHL when it should have been NLH.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Moira Barton
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Blyumman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### II. Indications for Use Statement
K030026 510(k) Number (if known):
Device Name: Alliance Medical Corporation Reprocessed Electrophysiology (EP) Catheters
Indications for Use: Reprocessed Electrophysiology (EP) Catheters are indicated for temporary intracardiac sensing, recording, stimulation and electrophysiological mapping of cardiac structures.
Indications for Use: Reprocessed electrophysiology catheter cables are indicated for use with the appropriate electrode catheter during electrophysiology studies.
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K050026 |
|---------------|---------|
|---------------|---------|
| Prescription Use (per 21 CFR 801.109) | <img alt="Prescription Use" src="prescription_use.png"/> | or | Over-the-Counter Use | __________ |
|---------------------------------------|----------------------------------------------------------|----|----------------------|------------|
|---------------------------------------|----------------------------------------------------------|----|----------------------|------------|
CONFIDENTIAL
Alliance Medical Corporation
Reprocessed Electrophysiology Catheters
Traditional 510(k)
13
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol featuring three horizontal lines that curve and converge, resembling a human figure in profile.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
1 2004 NOV
Alliance Medical Corporation c/o Ms. Moira Barton Regulatory Affairs Manager 10232 South 51st Street Phoenix, AZ 85044
Re: K030026 - Supplemental Validation Submission Trade Name: See Enclosed List Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter and Patient Transducer and Electrode Cable (including connector) Regulatory Class: Class II (two) Product Code: NLH & DSA Dated: January 2, 2003 Received: January 3, 2003
Dear Ms. Barton:
The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on May 2, 2003. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
{4}------------------------------------------------
#### Page 2 - K030026
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Bhimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
## Page 3 – K030026
.
## Enclosure – List of Devices
and the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of th
| K030026 |
|-----------------|
| Models OEM (n)) |
| (total=37) |
| Daig (37) |
| 401206 |
| 401210 |
| 401211 |
| 401222 |
| 401223 |
| 401225 |
| 401226 |
| 401227 |
| 401228 |
| 401260 |
| 401261 |
| 401262 |
| 401275 |
| 401276 |
| 401281 |
| 401287 |
| 401305 |
| 401306 |
| 401309 |
| 401311 |
| 401328 |
| 401353 |
| 401356 |
| 401357 |
| 401362 |
| 401379 |
| 401380 |
| 401381 |
| 401382 |
| 401383 |
| 401386 |
| 401389 |
| 401392 |
| 401393 |
| 401399 |
| 401400 |
.
{6}------------------------------------------------
## Page 4 - K030026
1
Image /page/6/Figure/1 description: The image shows a label with the text "K030026" at the top, followed by "M.4.1.1.5 [OEM](n)1" in the middle, and "L-370" below that. At the bottom of the label, there is a box with the number "401401" inside. The text is in black and the background is a light gray color.
100 Career States
: 上一篇:
1. Participant
: 2017-02-20 1
{7}------------------------------------------------
AY - 2 2003
Image /page/7/Picture/1 description: The image shows the text "K030026" on the top line and "Q1/2" on the bottom line. The text is written in a handwritten style, with thick, dark strokes. The numbers and letters are clearly legible, although the handwriting is not perfectly neat.
Image /page/7/Picture/2 description: The image shows the word "ALLIANCE" in a serif font, with a stylized graphic above it. The graphic consists of three curved lines that are thick and black. The lines are arranged in a parallel fashion, with the top line being the shortest and the bottom line being the longest, creating a sense of depth and movement.
RATION
# PART B: 510(k) SUMMARY
| TEL | 480.763.5300 |
|-----------|--------------------------|
| FAX | 480.763.5310 |
| Toll Free | 888.888.3433 |
| | www.alliance-medical.com |
| Submitter: | Alliance Medical Corporation<br>10232 South 51st Street<br>Phoenix, Arizona 85044 | TEL 480.763.53<br>FAX 480.763.53<br>Toll Free 888.888.34<br>www.alliance-medical. |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Contact: | Moira Barton<br>Senior Regulatory Affairs Specialist<br>(480) 763-5300 (o)<br>(480) 763-5310 (f) | |
| Date of preparation: | 12/30/2002 | |
| Name of device: | <i>Trade/Proprietary Name:</i> Reprocessed Electrophysiology<br>Catheter<br><i>Common or Usual Name:</i> Electrophysiology Catheter or<br>Electrode Recording Catheter<br><i>Classification Name:</i> Electrode Recording Catheter | |
| Predicate device(s): | Daig Response™ legally marketed Electrophysiology Catheter<br>devices under various 510(k) premarket notifications. | |
| | K894500 Daig Electrophysiology Catheter<br>K913940 Response™ Steerable Catheter<br>K942379 Daig Diagnostic Electrophysiology Catheter<br>K002976 Daig Electrophysiology Catheter - Response™ and Supreme™<br>K012708 Reprocessed Electrophysiology Catheters | |
| Device description: | Diagnostic electrophysiology (EP) catheters are specially<br>designed electrode catheters that transmit electrical impulses<br>and can be positioned for endocardial recording or stimulation.<br>Diagnostic EP catheters incorporate a handpiece, a flexible<br>shaft and a distal tip section containing diagnostic electrodes.<br>The distal tips of deflectable catheters can be deflected into a<br>curve by manipulating the handpiece; fixed curve catheters have<br>an established distal tip shape. | |
| Intended use: | Reprocessed Electrophysiology Catheters are intended for<br>temporary intracardiac sensing, recording, stimulation, and<br>electrophysiological mapping of cardiac structures. | |
| Indications<br>statement: | Reprocessed diagnostic EP catheters are indicated for<br>temporary intracardiac sensing, recording, stimulation and<br>electrophysiological mapping of cardiac structures.<br><br>Reprocessed electrophysiology catheter cables are indicated for<br>use with the appropriate electrode catheter during<br>electrophysiology studies. | |
{8}------------------------------------------------
K030026
p.2/2
| Technological characteristics: | The design, materials, and intended use of Reprocessed Electrophysiology (EP) Catheters and Electrophysiology Catheter Cables are identical to the predicate devices. The mechanism of action is identical to the predicate devices in that the same standard mechanical design, materials and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Alliance Medical Corporation's reprocessing of electrophysiology catheters includes removal of adherent visible soil and decontamination. Each individual electrophysiology catheter is tested for appropriate function of its components prior to packaging and labeling operations. |
| Performance data: | Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Electrophysiology (EP) Catheters. |
| | <ul><li>Biocompatibility</li><li>Validation of reprocessing</li><li>Sterilization Validation</li><li>Function test(s)</li></ul> |
| | Performance testing demonstrates that Reprocessed Electrophysiology (EP) Catheters perform as originally intended. |
| Conclusion: | Alliance Medical Corporation concludes that the modified device (the Reprocessed Electrophysiology (EP) Catheter) is safe, effective and substantially equivalent to the predicate devices as described herein. |
.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.