← Product Code [NLG](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLG) · K190785

# Reprocessed PentaRay Nav eco High-Density Mapping Catheter (K190785)

_Innovative Health, LLC · NLG · Jun 25, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLG/K190785

## Device Facts

- **Applicant:** Innovative Health, LLC
- **Product Code:** [NLG](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLG.md)
- **Decision Date:** Jun 25, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed PentaRay Nav eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e. recording or stimulation only. The catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The catheter provides location information when used with the compatible Carto 3 EP Navigation Systems. (This catheter is not compatible with Carto 3 EP Navigation Systems prior to Version 3.x.)

## Device Story

Reprocessed diagnostic electrophysiology mapping catheter; used for cardiac mapping, recording, and stimulation. Input: cardiac electrical signals via platinum electrodes; magnetic location data via embedded sensor. Operation: deflectable distal tip with multiple spines; inserted via 8F guiding sheath; thumb knob controls tip deflection. Output: electrograms and location data transmitted to Carto 3 EP Navigation System. Used in clinical electrophysiology labs by physicians. Benefits: enables high-density mapping of heart chambers to guide clinical decision-making for arrhythmias. Reprocessed by Innovative Health; includes cleaning, decontamination, and functional testing.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, cleaning and sterilization validation, visual inspection, dimensional verification, electrical continuity and resistance, simulated use, leak/occlusion testing, mechanical characteristics, and electrical safety (dielectric and current leakage).

## Technological Characteristics

Deflectable catheter with multiple 3F spines; platinum electrodes; magnetic location sensor. Irrigation lumen for anticoagulant fluid. 7F French size; 115cm usable length. Compatible with Carto 3 EP Navigation System (v3.x+). Reprocessed device; limited to one reprocessing cycle. Sterilization and cleaning validation performed.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- PentaRay NAV eco High-Density Mapping Catheter ([K123837](/device/K123837.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 25, 2019

Innovative Health, LLC. Amanda Babcock Regulatory Affairs Manager 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257

Re: K190785

Trade/Device Name: Reprocessed PentaRay Nav eco High-Density Mapping Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLG Dated: March 26, 2019 Received: March 27, 2019

Dear Amanda Babcock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark Fellman Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Page | 3 |
|------|---|
|------|---|

| Item<br>Number | Usable<br>Length (cm) | French Size<br>(Fr) | Curve | Electrode<br>Spacing | Number of<br>Electrodes |
|----------------|-----------------------|---------------------|-------|----------------------|-------------------------|
| D128207        | 115                   | 7                   | F     | 4-4-4                | 20                      |
| D128208        | 115                   | 7                   | F     | 2-6-2                | 20                      |
| D128210        | 115                   | 7                   | D     | 4-4-4                | 20                      |
| D128211        | 115                   | 7                   | D     | 2-6-2                | 20                      |

The following device models are included in the scope of this 510(k) submission:

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190785

Device Name

Reprocessed PentaRay Nav eco High-Density Mapping Catheter

Indications for Use (Describe)

The Reprocessed PentaRay Nav eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e. recording or stimulation only. The catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

The catheter provides location information when used with the compatible Carto 3 EP Navigation Systems. (This catheter is not compatible with Carto 3 EP Navigation Systems prior to Version 3.x.)

Type of Use (Select one or both, as applicable)

| <div> <span></span> <span></span>Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|-----------------------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C)    </div>                |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### SECTION 5: 510(k) SUMMARY

As required by 21 CFR 807.92(c)

#### Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

### Contact Name and Information:

Amanda Babcock Regulatory Affairs Manager Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7705 (fax) ababcock@innovative-health.com

#### Date prepared:

March 26, 2019

#### Device Information:

| Trade/Proprietary Name: | Reprocessed PentaRay Nav eco High-Density Mapping Catheter |
|-------------------------|------------------------------------------------------------|
| Common or Usual Name:   | Diagnostic Electrophysiology Mapping Catheter              |
| Classification Name:    | Electrode Recording Catheter or Electrode Recording Probe  |
| Classification Number:  | Class II, 21 CFR 870.1220                                  |
| Product Code:           | NLG                                                        |

#### Predicate Device:

| 510(k) Number | 510(k) Title                                      | Manufacturer           |
|---------------|---------------------------------------------------|------------------------|
| K123837       | PentaRay NAV eco High-Density Mapping<br>Catheter | Biosense Webster, Inc. |

#### Device Description:

The Reprocessed PentaRay Nav eco High-Density Mapping Catheter is designed to facilitate electrophysiological mapping of the heart with the Carto 3 EP Navigation System. It is designed for deployment in a heart chamber through an 8 F guiding sheath. This deflectable catheter consists of multiple 3F spines on its distal tip, each spine having multiple platinum electrodes that are used for stimulation and recording. A magnetic location sensor embedded in the deflectable tip transmits location information to the Carto 3 EP Navigation System. The catheter has two electrodes on the deflectable tip to provide for visualization of the tip when used with the Carto 3 EP Navigation System. Pushing forward on the catheter thumb knob deflects the tip; pulling back on the thumb knob straightens the tip. This device includes an irrigation lumen for connection to a source of continuous anticoagulant fluid.

This catheter interfaces with standard recording equipment and the Carto 3 EP Navigation System via interface cables with the appropriate connectors. Please consult the manufacturer for the appropriate interface cables.

{5}------------------------------------------------

For use in mapping procedures, refer to the instructions for the Carto 3 EP Navigation System.

| Item Number | Description                                       | Usable<br>Length<br>(cm) | French<br>Size | Curve | Spacing<br>(mm) | Electrodes |
|-------------|---------------------------------------------------|--------------------------|----------------|-------|-----------------|------------|
| D128207     | PentaRay Nav eco High-Density<br>Mapping Catheter | 115                      | 7F             | F     | 4-4-4           | 20         |
| D128208     | PentaRay Nav eco High-Density<br>Mapping Catheter | 115                      | 7F             | F     | 2-6-2           | 20         |
| D128210     | PentaRay Nav eco High-Density<br>Mapping Catheter | 115                      | 7F             | D     | 4-4-4           | 20         |
| D128211     | PentaRay Nav eco High-Density<br>Mapping Catheter | 115                      | 7F             | D     | 2-6-2           | 20         |

### The item numbers in scope of this submission are as follows:

Table 5.1: Device Scope

## Indications for Use:

The Reprocessed PentaRay NAV eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e. recording or stimulation only. The catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

The catheter provides location information when used with the compatible Carto 3 EP Navigation Systems. (This catheter is not compatible with Carto 3 EP Navigation Systems prior to Version 3.x).

### Technoloqical Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed PentaRay Nav eco High-Density Mapping Catheter are identical to the predicate devices. There are no changes to the claims, clinical applications, patient populations, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of the Catheter includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

# Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed PentaRav Nav eco High-Density Mapping Catheter. This included the following:

- Biocompatibilitv ●
- Cleaning Validation .
- Sterilization Validation ●
- Functional testing
	- . Visual Inspection
	- . Dimensional Verification
	- . Electrical Continuity and Resistance
	- . Simulated Use
	- . Leak/Occlusion
	- . Inner lumen occlusion
	- I Mechanical Characteristics
- . Electrical Safety Testing
	- Dielectric and Current Leakage .
- Packaging Validation

{6}------------------------------------------------

The Reprocessed Reprocessed PentaRay Nav eco High-Density Mapping Catheter are reprocessed no more than one (1) time. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

### Conclusion:

Innovative Health concludes that the Reprocessed PentaRay Nav eco High-Density Mapping Catheter is as safe and effective as the predicate devices described herein.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLG/K190785](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/NLG/K190785)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
