← Product Code [MXD](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD) · K231328

# LUX-Dx II (M302); LUX-Dx II+ (M312) (K231328)

_Boston Scientific Corp · MXD · Aug 19, 2023 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD/K231328

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** [MXD](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD.md)
- **Decision Date:** Aug 19, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx has not been tested specifically for pediatric use.

## Device Story

Subcutaneously implanted cardiac monitor (ICM) records S-ECG signals; detects cardiac arrhythmias via onboard algorithms. System includes ICM device, mobile applications (myLUX Patient app, LUX-Dx Clinic Assistant app) on OTS mobile devices, and LATITUDE Clarity server. ICM communicates with mobile app via Bluetooth Low Energy (BLE); app transmits data to server via cellular/Wi-Fi. Clinicians access server to review event data, manage patient parameters, and monitor workflows. Device aids clinical diagnosis of arrhythmias; provides remote monitoring capability. Enhancements include PVC detection, improved Pause/AF algorithms, and BYOD (bring-your-own-device) support.

## Clinical Evidence

No clinical trial data presented. Substantial equivalence supported by bench testing, system/sub-system design verification, software validation, cybersecurity testing, human factors/usability testing, and algorithm validation.

## Technological Characteristics

Subcutaneous ICM; LiMnO2 battery; 1.2 cc volume. Electrodes at header and battery base. Connectivity: BLE to mobile device, cellular/Wi-Fi to server. MR Conditional (1.5T/3T). Software includes detection algorithms for AF, Tachy, Brady, Pause, AT, and PVC. Sterile, single-use.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- LUX-Dx Insertable Cardiac Monitor ([K193473](/device/K193473.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 19, 2023

Boston Scientific Corp Emily Jallen Senior Regulatory Affairs Specialist 4100 Hamline Ave North St. Paul, Minnesota 55112

Re: K231328

Trade/Device Name: LUX-Dx II (M302); LUX-Dx II+ (M312) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: July 17, 2023 Received: July 18, 2023

Dear Emily Jallen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows a digital signature. The signature includes the name "Sara M. Royce -S" on the left side of the image. On the right side of the image, it says "Digitally signed by Sara M. Royce -S" and "Date: 2023.08.19 20:33:18-04'00'".

Sara Royce Acting Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

Submission Number (if known)

#### K231328

Device Name

LUX-Dx II (M302): LUX-Dx II+ (M312)

Indications for Use (Describe)

The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx has not been tested specifically for pediatric use.

Type of Use (Select one or both, as applicable)

> | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

4100 Hamline Avenue North St Paul, MN 55112-5798 651.582.4000 Tel 651.582.5134 Fax www.bostonscientific.com

# 510(k) Summary

## 1. Submitter

Boston Scientific Corporation 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798

- Contact: Emily Jallen Senior Regulatory Specialist Phone: 651.581.1565 Email: emily.jallen@bsci.com
Date Prepared: 12 July 2023

#### 2. Device

Trade Names: LUX-Dx II™ Insertable Cardiac Monitor and LUX-Dx II+™ Insertable Cardiac Monitor Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025

## 3. Predicate Device

Trade Name: LUX-Dx™ Insertable Cardiac Monitor Manufacturer: Boston Scientific Corp Clearance Number: K193473, 26 June 2020 Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025

{4}------------------------------------------------

# 4. Indication for Use

The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath.

The LUX-Dx has not been tested specifically for pediatric use.

## 5. Device Description

The LUX-Dx II (M302) and LUX-Dx II+ (M312) ICM devices evaluate S-ECG waveform data for indications of cardiac arrhythmias and "marks" the S-ECG signal for clinical presentation and evaluation when the algorithm criteria are met. The ICM device is inserted into the subcutaneous layer of the fourth intercostal space of the left chest wall. The ICM device is powered by an integrated battery. The electrodes used for detecting the S-ECG signal are located on each end of the ICM device, in the header and at the base of the battery. The LUX-Dx system includes the following main components:

- ICM Device - a subcutaneously-implanted cardiac monitor device for cardiac arrhythmia event data collection and transmission. In addition, symptom events are collected and transmitted from the device.
- . Mobile Monitor (MM) - mobile applications (myLUX™ Patient app and LUX-Dx™ Clinic Assistant app) running on an OTS mobile device that communicates with the ICM device (using Bluetooth Low Energy (BLE)) and the LATITUDE Clarity™ server (using cellular/Wi-Fi) for collection and transmission of event, patient, and device data.
- . LATITUDE Clarity™ server - a server that communicates with the Mobile Monitor for bidirectional data transmission and provides web access for clinicians to perform remote monitoring activities and manage general patient and system parameters and workflow activities.
- . System Accessories- for insertion of the ICM device, an insertion tool and incision tool are provided. In addition, a magnet is provided to initiate ICM/MM app communication.

# 6. Technological Characteristics

The intended use, principles of operation, and fundamental technology remain unchanged from the predicate LUX-Dx M301 ICM device. In addition, there are no changes to the ICM device hardware or physical characteristics as part of this 510(k) Notification. The changes incorporated into the LUX-Dx ICM device models (M302, M312) include new and enhanced detection algorithms, bring-your-own-device (BYOD) capability, and additional changes for sustaining/maintenance and continuous improvement.

The LUX-Dx ICM device models (M302, M312) include a PVC detection algorithm as well as an enhancement to the Pause algorithm to reduce the potential for false positive detections. The M312 ICM device model also includes an enhancement to the AF algorithm to reduce the potential for false positive detections. In

{5}------------------------------------------------

addition, both ICM device models (M302, M312) allow for the use of a personal mobile device (referred to as BYOD) as part of the LUX-Dx system.

# 7. Substantial Equivalence

In support of a substantial equivalence determination, BSC has evaluated the differences between the LUX-Dx II M302 and LUX-Dx II+ M312 ICM devices and the predicate LUX-Dx ICM M301, K193473. Based on the indications for use, fundamental technological characteristics, and performance testing, the LUX-Dx ICM device models (M302, M312) have been shown to be appropriate for the intended use and are substantially equivalent to the predicate device.

| Characteristic         | Predicate: Model M301 LUX-Dx ICM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Model M302 LUX-Dx II ICM<br>Model M312 LUX-Dx II+ ICM |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Indications for<br>use | The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to<br>monitor and record subcutaneous electrocardiogram (S-<br>ECG). The recorded S-ECG is used for the clinical evaluation<br>and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated<br>for use in patients that have a known heart condition and are<br>at risk of developing an abnormal heart rhythm, or have<br>symptoms that may suggest a cardiac arrhythmia such as<br>dizziness, palpitations, syncope, chest pain, and/or shortness<br>of breath.<br>The LUX-Dx has not been tested specifically for pediatric use. | Identical                                             |
| Usage                  | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                             |
| Sterility              | Supplied sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                             |
| Volume<br>Dimension    | 1.2 cc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                             |
| MR<br>Conditional      | 1.5 T and 3T                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                             |
| Energy Source          | Battery powered - LiMnO2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                             |
| Longevity              | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                             |
| Arrythmias<br>Detected | AF Tachy Brady Pause AT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | AF Tachy Brady Pause AT PVC                           |
| Remote<br>Monitoring   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                             |
| Packaging              | Sterile kit including ICM device and custom implant tools<br>Accessories: External, non-sterile devices are packaged<br>separately                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                             |

{6}------------------------------------------------

## 8. Summary of Performance Testing

Boston Scientific performed safety risk management activities, design verification, design validation, and usability testing to demonstrate that the LUX-Dx system meets the design intent, conforms to user needs and intended use, and supports substantial equivalence to the predicate device. Testing to support the changes for the LUX-Dx ICM device models (M302, M312) includes system and sub-system design verification testing, various software validations, cybersecurity testing, human factors/usability, and design validation testing including algorithm validation.

## 9. Conclusion

Based on the intended use, fundamental technological characteristics, and performance testing, LUX-Dx II M302 and LUX-Dx II+ M312 have been shown to be appropriate for the intended use and are substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD/K231328](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD/K231328)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
