Reveal LINQ Insertable Cardiac Monitor, LINQ II Insertable Cardiac Monitor, AccuRhythm AI ECG Classification System
Applicant
Medtronic, Inc.
Product Code
MXD · Cardiovascular
Decision Date
Apr 5, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Acceptance
Observed
Dev DS
Test DS
Atrial Fibrillation (AF) Detection
Met or exceeded the criteria in the Validation Plan.
All results met or exceeded the criteria in the Validation Plan.
Not specified.
Performance validation testing and analysis conducted to ensure the algorithms were able to reduce false alerts from ICM detected AF episodes while retaining true alerts.
Pause Detection
Met or exceeded the criteria in the Validation Plan.
All results met or exceeded the criteria in the Validation Plan.
Not specified.
Performance validation testing and analysis conducted to ensure the algorithms were able to reduce false alerts from ICM detected Pause episodes while retaining true alerts.
Indications for Use
The Reveal LINQ ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: patients with clinical syndromes or situations at increased risk of cardiac arrhythmias; patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia. The device has not been tested specifically for pediatric use.
Device Story
Reveal LINQ ICM is a small, leadless, subcutaneous insertable cardiac monitor; implanted in chest; continuously monitors subcutaneous ECG via two electrodes. Device automatically detects and records arrhythmias (AT/AF, bradyarrhythmia, pause, ventricular tachyarrhythmia) and supports patient-activated event recording. AccuRhythm AI ECG Classification System processes ECG data to classify rhythms and reduce false alerts. Data is transmitted to clinical monitoring systems (e.g., CareLink Network) for physician review. Output assists clinicians in diagnosing cardiac arrhythmias and managing patient care. Benefits include improved detection of abnormal heart rhythms in patients who may not trigger manual activation.
Clinical Evidence
Bench testing and design validation only. Software verification confirmed design output met specifications. Performance validation testing demonstrated the AccuRhythm AI algorithm reduces false alerts from ICM-detected AF and Pause episodes while retaining true alerts, meeting all pre-defined acceptance criteria.
Technological Characteristics
Leadless, subcutaneous implantable monitor. Sensing: two electrodes for subcutaneous ECG. Connectivity: wireless data transmission to external monitors/network. Software: AccuRhythm AI ECG Classification System (Models ZA400, ZA410, ZA420). Standards: ISO 14971:2019, ISO 15223-1:2016, IEC 82304-1:2017, IEC 62304:2006/AMD 1:2015, IEC 62304:2006/AC:2008, ANSI/AAMI EC57:2012.
Indications for Use
Indicated for patients at increased risk of cardiac arrhythmias or experiencing transient symptoms (dizziness, palpitation, syncope, chest pain) suggestive of cardiac arrhythmia. Not tested for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
LINQ II ICM, Model LNQ22, with AccuRhythm AI ECG Classification System (K210484)
K210484 — LINQ II Insertable Cardiac Monitor, Zelda AI ECG Classification System · Medtronic, Inc. · Jun 11, 2021
K162855 — Reveal LINQ · Medtronic, Inc. · Feb 23, 2017
K221962 — LINQ II Insertable Cardiac Monitor · Medtronic, Inc. · Aug 25, 2022
K163460 — Reveal LINQ Insertable Cardiac Monitor · Medtronic, Inc. · Feb 14, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 5, 2023
Medtronic, Inc. Kerry Luyster Senior Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, Minnesota 55112
Re: K223630
Trade/Device Name: Reveal LINQ Insertable Cardiac Monitor (ICM) with AccuRhythm AI ECG Classification System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: March 3, 2023 Received: March 6, 2023
Dear Kerry Luyster:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Sara M.
Royce -S
Hetal Odobasic
Assistant Director
DHT2A: Division of Cardiac
Electrophysiology, Diagnostics
and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Digitally signed by
Sara M. Royce -S
Date: 2023.04.05
14:20:21 -04'00'
1 D
1
Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K223630
Device Name
Reveal LINQ Insertable Cardiac Monitor (Model LNQ11)
Indications for Use (Describe)
The Reveal LINQ ICM is an insertable automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:
• patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
· patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia
The device has not been tested specifically for pediatric use.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
# 510(k) Summary
| Date Prepared: | April 5, 2023 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>Cardiovascular Diagnostics and Services<br>8200 Coral Sea Street NE<br>Mounds View, MN 55112<br>Establishment Registration Number: 2182208 |
| Contact Person: | Kerry C Luyster<br>Senior Regulatory Affairs Specialist<br>Cardiovascular Diagnostics and Services<br>Phone: (763) 505-2124<br>Email: kerry.c.luyster@medtronic.com |
| Alternate Contact: | Ryan Calabrese<br>Sr Regulatory Affairs Director<br>Cardiovascular Diagnostics and Services<br>Phone: (763) 526-3515<br>Email: ryan.s.calabrese@medtronic.com |
## General Information
| Trade Name: | Reveal® LINQ™ Insertable Cardiac Monitor (ICM) with AccuRhythm AI<br>ECG Classification System |
|-----------------------|------------------------------------------------------------------------------------------------|
| Common Name: | Insertable Cardiac Monitor |
| Regulation Number: | CFR 870.1025 |
| Product Code: | MXD |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
| Special Controls: | Class II Special Controls Guidance Document: Arrhythmia Detector and<br>Alarm |
| Predicate Device: | LINQ II ICM, Model LNQ22, with AccuRhythm AI ECG Classification<br>System (K210484) |
{4}------------------------------------------------
## Device Description
The Reveal LINQ Model LNQ11 Insertable Cardiac Monitors (ICM) is designed to automatically record the occurrence of arrhythmias in a patient. Arrhythmia may be classified as atrial tachyarrhythmia/atrial fibrillation (AT/AF), bradyarrhythmia, pause, or (fast) ventricular tachyarrhythmia. The Reveal LINQ ICM provides storage of ECG and Marker Channel during patient-activated and automatically-detected (auto-activated) events. Auto activation may help to detect abnormal heart rhythms in patients who may not activate/trigger the ICM. The Reveal LINQ ICM Model LNQ11 is a small, leadless device that is typically implanted under the skin, in the chest. Two electrodes on the body of the device continuously monitor the patient's subcutaneous ECG.
Reveal LINQ ICM includes the following accessories: LINQ Tool Kit Model LNQTOOL, Patient Assistant PA96000, Reveal LINQ™ Mobile Manager Model MSW002 used with Patient Connector Model 24967, CareLink Programmer Model 2090, Encore Programmer Model 29901, Reveal LINO Application Software Model SW026, MyCareLink Patient Monitor Models 24950 and 24955, CareLink Express Monitor Model 2020B, Device Data Management Application Model 2491, Device Command Library Model 2692 and Instrument Command Library Model 2691, CareLink Express Mobile Application Models 31302, and CareLink Network. New to the Reveal LINO ICM system is the AccuRhythm AI ECG Classification System Models ZA400, ZA410, ZA420, included in this submission.
#### Indications for Use
There are no changes to the Indications for Use for the Reveal LINQ ICM device as a result of this submission. The subject device Reveal LINQ ICM when used with the AccuRhythm AI ECG Classification System has the same indications for use as the predicate device (K210484). The Indications for Use are provided below:
The Reveal LINO ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:
- patients with clinical syndromes or situations at increased risk of cardiac arrhythmias ●
- . patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia
The device has not been tested specifically for pediatric use.
#### Technological Characteristics
Substantial equivalence between the Reveal LINO II ICM and LINO II ICM was established in K200795 (03JUL2020). There are no attribute changes to the device as part of this 510(k) Notification, other than the addition of AccuRhythm AI to the Reveal LINQ ICM ecosystem.
There is no change in the design and technology of the Reveal LINO ICM or the LINO II ICM device, other than the addition of AccuRhythm AI ECG Classification System to the Reveal LINQ ICM ecosystem. The subject device Reveal LINQ ICM when used with the AccuRhythm AI ECG Classification System has the same operating principle, device technology and functionality, indications for use and biological safety as the predicate device LINQ II with AccuRhythm AI ECG Classification System (K210484).
{5}------------------------------------------------
#### Substantial Equivalence
Differences between the subject and predicate devices have been evaluated through bench testing to provide evidence of substantial equivalence. The Reveal LINQ ICM when used with the AccuRhythm AI ECG Classification System is substantially equivalent to the predicate LINQ II ICM with AccuRhythm AI ECG Classification System (K210484) based on comparisons of indications for use, operating principle, device technology and functionality, and safety.
#### Summary of Testing
Design verification and design validation were performed to demonstrate that the Reveal LINQ ICM when used with the AccuRhythm AI ECG Classification System met both design requirements and established performance criteria to support substantial equivalence to the predicate LINQ II ICM with AccuRhythm AI ECG Classification System (K210484).
- . Design Verification: Software design verification was completed to ensure the design output meets specifications outlined in the design inputs. The AccuRhythm ECG Classification System meets the functionality per the requirements and all test executions resulted in a status of Passed.
- Design Validation: Performance validation testing and analysis were completed to . ensure the algorithms were able to reduce false alerts from ICM detected AF and Pause episodes while retaining true alerts. All results met or exceeded the criteria in the Validation Plan.
Since there were no changes to the Reveal LINQ ICM itself, there was no development or testing specific to the ICM; therefore, no standards are referenced for the Reveal LINQ ICM.
The following standards were used for development and testing of the AccuRhythm AI ECG Classification System.
| Standard Number | Standard<br>Organization | Recognition<br>Number | Standard Title |
|---------------------------|--------------------------|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| 14971:2019 | ISO | 5-125 | Medical Devices - Application of<br>Risk Management to Medical<br>Devices |
| 15223-1:2016 | ISO | 5-117 | Medical devices - Symbols to be<br>used with medical device labels,<br>labelling, and information to be<br>supplied - Part 1: General<br>requirements |
| 82304-1:2017 | IEC | 13-97 | Health software - Part 1: General<br>requirements for product safety |
| 62304:2006/ AMD<br>1:2015 | IEC | 13-79 | Medical device software - Software<br>life cycle processes |
| 62304:2006/AC:2008 | IEC | 13-79 | Medical device software - Software<br>life cycle processes |
{6}------------------------------------------------
| Standard Number | Standard<br>Organization | Recognition<br>Number | Standard Title |
|-----------------|--------------------------|-----------------------|---------------------------------------------------------------------------------------------------------|
| EC57: 2012 | ANSI/AAMI | 3-118 | Testing and Reporting Performance<br>Results of Cardiac Rhythm and ST<br>Segment Measurement Algorithms |
## Conclusion
The results of the testing met the design requirements and specified acceptance criteria and did not raise new safety or performance issues. Therefore, the Reveal LINQ ICM Model LNQ11 when used with the AccuRhythm AI ECG Classification System Models ZA400, ZA410, ZA420 described in this submission results in a device that is substantially equivalent to the predicate LINQ II ICM, Model LNQ22, with AccuRhythm AI ECG Classification System (K210484).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.