LUX-Dx Insertable Cardiac Monitor, myLUX Software Mobile Application, LUX-Dx Clinic Assistant Software, Application Server Device Services (DS) - US
K193473 · Boston Scientific Corporation · MXD · Jun 26, 2020 · Cardiovascular
Device Facts
Record ID
K193473
Device Name
LUX-Dx Insertable Cardiac Monitor, myLUX Software Mobile Application, LUX-Dx Clinic Assistant Software, Application Server Device Services (DS) - US
Applicant
Boston Scientific Corporation
Product Code
MXD · Cardiovascular
Decision Date
Jun 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous ECG (S-ECG) for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition, are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a heart condition such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx has not been tested specifically for pediatric use.
Device Story
LUX-Dx is a subcutaneously-implanted cardiac monitor (ICM) inserted in the fourth intercostal space of the left chest wall. It continuously records subcutaneous ECG (S-ECG) signals via electrodes on the device header and battery. An onboard algorithm evaluates S-ECG waveform data for cardiac arrhythmias and marks events when criteria are met. The device transmits data via Bluetooth Low Energy (BLE) to a mobile monitor (myLUX Patient app or LUX-Dx Clinic Assistant app on an OTS mobile device). The mobile app forwards data to the LATITUDE Clarity server via cellular/Wi-Fi. Clinicians access the server to perform remote monitoring, manage patient data, and review marked events. This system enables long-term cardiac rhythm monitoring, assisting physicians in diagnosing arrhythmias and managing patients with cardiac symptoms, potentially improving clinical decision-making and patient outcomes.
Clinical Evidence
Bench testing only. Evidence includes safety risk management, design verification, design validation, and usability testing. Specific testing categories: sterilization, biocompatibility, electromagnetic compatibility, electrical safety, MRI compatibility, mechanical verification, implant tool verification, packaging verification, software verification, algorithm validation, and human factors/usability testing. Pre-clinical evidence includes GLP and non-GLP studies.
Technological Characteristics
Subcutaneously implanted sensor (Model M301). Dimensions: 1.2 cc. Energy source: LiMnO2 battery. Connectivity: BLE to mobile device, cellular/Wi-Fi to cloud server. Sterilization: Supplied sterile. Software: Algorithm-based arrhythmia detection. MR Conditional (1.5T and 3T).
Indications for Use
Indicated for patients with known heart conditions, risk of abnormal heart rhythm, or symptoms suggesting cardiac arrhythmia (dizziness, palpitations, syncope, chest pain, shortness of breath). Not tested for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
K231328 — LUX-Dx II (M302); LUX-Dx II+ (M312) · Boston Scientific Corp · Aug 19, 2023
K210608 — LUX-Dx Insertable Cardiac Monitor · Boston Scientific Corporation · May 25, 2021
K150614 — Reveal LINQ · Medtronic, Inc. · Aug 6, 2015
K200795 — LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device Command Library, Instrument Command Library, LINQ Tool Kit · Medtronic, Inc. · Jul 3, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
June 26, 2020
Boston Scientific Corporation Melissa Klamerus Principal Regulatory Affairs Specialist 4100 Hamline Ave North St. Paul, Minnesota 55112
#### Re: K193473
Trade/Device Name: LUX-DxTM Insertable Cardiac Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: June 10, 2020 Received: June 12, 2020
Dear Melissa Klamerus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193473
Device Name LUX-Dx™ Insertable Cardiac Monitor
#### Indications for Use (Describe)
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous ECG (S-ECG) for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition, are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a heart condition such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath.
The LUX-Dx has not been tested specifically for pediatric use.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 10px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
### 1. Submitter
Boston Scientific Corporation 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798
Contact: Melissa Klamerus Principal Regulatory Specialist Phone: (651) 582-6771 Email: melissa.klamerus@bsci.com
Date Prepared: 11 December 2019
### 2. Device
Trade Name: LUX-Dx™ Insertable Cardiac Monitor Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025
# 3. Predicate Device
Trade Name: Reveal LINQ Insertable Cardiac Monitor Manufacturer: Medtronic Inc Clearance Number: K132649, 14 February 2014 Common Name: Arrythmia detector and alarm Product Code/Panel: DSI, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025
# 4. Indication for Use
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous ECG (S-ECG) for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition, are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a heart condition such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath.
The LUX-Dx has not been tested specifically for pediatric use.
{4}------------------------------------------------
# 5. Device Description
The LUX-Dx ICM sensor device (Model M301) evaluates S-ECG waveform data for indications of cardiac arrhythmias and "marks" the S-EG signal for clinical presentation when the algorithm criteria are met. The device is inserted into the subcutaneous layer of the fourth intercostal space of the left chest wall. The sensor device is powered by an integrated battery. The electrodes used for detecting the S-ECG signal are located on each end of the sensor device, in the header and at the battery. The LUX-Dx system includes the following main components:
- LUX-Dx™ sensor device a subcutaneously-implanted sensor device for cardiac arrhythmia event ● data collection and transmission. In addition, symptom events are collected and transmitted from the sensor device.
- Mobile Monitor (MM) mobile applications (myLUX™ Patient app and LUX-Dx™ Clinic Assistant . app) running on an OTS mobile device that communicates with the LUX-Dx sensor device (using Bluetooth Low Energy (BLE)) and the LATITUDE server (using cellular/Wi-Fi) for collection and transmission of event, patient, and device data.
- . LATITUDE Clarity™ server - a server that communicates with the Mobile Monitor for bidirectional data transmission and provides web access for clinicians to perform remote monitoring activities and manage general patient and system parameters and workflow activities.
- . System Accessories- for insertion of the sensor device, an insertion tool and incision tool are provided. In addition, a magnet is provided to initiate sensor/MM app communication.
| Characteristic | BSC Model M301 LUX-Dx ICM | Medtronic Reveal LINQ LNQII |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The LUX-Dx™ Insertable Cardiac Monitor (ICM)<br>is intended to monitor and record<br>subcutaneous ECG (S-ECG) for the clinical<br>evaluation and diagnosis of cardiac<br>arrhythmias. The LUX-Dx is indicated for use in<br>patients that have a known heart condition,<br>are at risk of developing an abnormal heart<br>rhythm, or have symptoms that may suggest a<br>heart condition such as dizziness, palpitations,<br>syncope, chest pain, and/or shortness of<br>breath. | The Reveal LINQ ICM is an insertable<br>automatically-activated and patient-activated<br>monitoring system that records subcutaneous<br>ECG and is indicated in the following cases:<br>● patients with clinical syndromes or<br>situations at increased risk of cardiac<br>arrhythmias<br>● patients who experience transient<br>symptoms such as dizziness, palpitation,<br>syncope, and chest pain that may suggest a<br>cardiac arrhythmia |
| Usage | Single use | Single use |
| Sterility | Supplied sterile | Supplied sterile |
| Volume Dimension | 1.2 cc | 1.2 cc |
| MR Conditional | 1.5 T and 3T | 1.5 T and 3T |
| Energy Source | Battery powered - LiMnO₂ | Battery powered - LiCFx |
# 6. Substantial Equivalence
{5}------------------------------------------------
| Characteristic | BSC Model M301 LUX-Dx ICM | Medtronic Reveal LINQ LNQII |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Longevity | 3 years | 3 years |
| Remote Monitoring | Yes | Yes |
| Packaging | Sterile kit including ICM device and custom implant tools.<br>Accessories: External, non-sterile devices are packaged separately. | Sterile kit including ICM device and custom implant tools.<br>Accessories: External, non-sterile devices are packaged separately. |
# 7. Summary of Performance Testing
Boston Scientific performed safety risk management activities, design verification, design validation, and usability testing to demonstrate that the LUX-Dx system is substantially equivalent to the predicate device. The system conforms to user needs and intended use.
Performance Testing - Bench:
- . Sterilization
- Biocompatibility
- Electromagnetic compatibility ●
- Electrical safety ●
- MRI compatibility
- Mechanical verification ●
- Implant tool verification
- Packaging Verification
- Software Verification
- Algorithm Validation
- Human Factors/Usability Testing ●
Performance Testing – Pre-clinical:
- . LUX-Dx GLP and non-GLP studies
# 8. Conclusion
Based on the intended use, fundamental technological characteristics, and performance testing, the proposed LUX-Dx system has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.