← Product Code [MXD](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD) · K171514

# BioMonitor 2-AF, BioMonitor 2-S, FIT - Fast Insert Tool 1 + 2, Set (K171514)

_Biotronik, Inc. · MXD · Jun 29, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD/K171514

## Device Facts

- **Applicant:** Biotronik, Inc.
- **Product Code:** [MXD](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD.md)
- **Decision Date:** Jun 29, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The BioMonitor 2 is indicated to detect the following cardiac arrhythmias: - Atrial fibrillation, - Bradycardia, - Sudden rate drop. - High ventricular rate (HVR), - Asystole. The BioMonitor 2 is indicated for: - Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias - Patients who experience transient symptoms that may suggest a cardiac arrhythmia - The device has not been tested for and it is not intended for pediatric use

## Device Story

Insertable cardiac monitor; senses subcutaneous electrocardiograms (SECG) via two integrated electrodes. Detects arrhythmias (atrial fibrillation, bradycardia, sudden rate drop, HVR, asystole); triggers automatic SECG recording. Patients use Remote Assistant for manual recording. Device stores SECG episodes; transmits data via BIOTRONIK Home Monitoring to Service Center for physician review. Used in clinical/home settings; physician-implanted. Output aids diagnosis of otherwise undetected arrhythmias; informs clinical decision-making regarding cardiac management.

## Clinical Evidence

No clinical data. Bench testing only, including mechanical system verification, usability testing, sterilization validation, biocompatibility, and packaging validation.

## Technological Characteristics

Insertable cardiac monitor; subcutaneous ECG sensing via two integrated electrodes. Includes FIT1/FIT2 insertion accessories. Data transmission via BIOTRONIK Home Monitoring. Sterilization validation performed. No changes to core sensing/detection technology compared to predicate.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- BIOTRONIK BioMonitor 2 ([K152995](/device/K152995.md))

## Reference Devices

- BIOTRONIK Home Monitoring Service Center ([P950037](/device/P950037.md)/S66)
- BIOTRONIK Home Monitoring Service Center ([P950037](/device/P950037.md)/S69)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 29, 2017

BIOTRONIK, Inc. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, Oregon 97035

Re: K171514

Trade/Device Name: BioMonitor 2 Insertable Cardiac Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: May 22, 2017 Received: May 24, 2017

Dear Jon Brumbaugh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

M.A. Hillebrand

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K171514

#### Device Name

BioMonitor 2 Insertable Cardiac Monitor

Indications for Use (Describe)

The BioMonitor 2 is indicated to detect the following cardiac arrhythmias:

- · Atrial fibrillation,
- · Bradycardia,
- · Sudden rate drop.
- · High ventricular rate (HVR),
- · Asystole.

The BioMonitor 2 is indicated for:

- · Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- Patients who experience transient symptoms that may suggest a cardiac arrhythmia
- · The device has not been tested for and it is not intended for pediatric use

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 1. Submitter

BIOTRONIK 6024 SW Jean Road Lake Oswego, OR 97035 Phone: (888) 345-0374 Fax: 503-451-8519

Contact Person: Jon Brumbaugh

Date Prepared: June 29, 2017

#### 2. Device

| Name of Device       | BioMonitor 2 Insertable Cardiac Monitor                                       |
|----------------------|-------------------------------------------------------------------------------|
| Common or Usual Name | Insertable Cardiac Monitor                                                    |
| Classification Name  | Arrhythmia detector and alarm (including ST-segment measurement<br>and alarm) |
| Classification       | Class II (21 CFR 870.1025)                                                    |
| Product Code         | MXD                                                                           |

# 3. Predicate Devices

BIOTRONIK BioMonitor 2 (K152995, cleared April 11, 2016)

# 4. Device Description

The BioMonitor 2 senses subcutaneous electrocardiograms (SECG) using two integrated electrodes and has the capability of detecting a number of arrhythmias. The BioMonitor 2 uses BIOTRONIK's Home Monitoring to send recorded SECG and statistics to the Home Monitoring Service Center (P950037/S66, dated November 21, 2008; and P950037/S69, dated March 23, 2009) for further analysis. BioMonitor 2 is intended to aid in the diagnosis of cardiac arrhythmias that might otherwise go undetected.

BioMonitor 2 is designed to detect and trigger automatic recording of certain cardiac arrhythmias; these arrhythmias can be classified as follows:

- atrial fibrillation
- bradycardia ●
- sudden rate drop
- high ventricular rate (HVR)
- asystole. .

Patients can also manually trigger recording of the cardiac information by use of the Remote Assistant. The memory capacity of BioMonitor 2 is such that the device can record at least 66 minutes of subcutaneous electrocardiograms (SECG). The device automatically stores a maximum of 55 separately recorded SECG-episodes of 40 seconds each (60 seconds maximum), and a maximum of 4 patient triggered SECG-episodes of 7.5 minutes. The BioMonitor 2 reports the recorded episodes and related statistical data through the physician's programmer and BIOTRONIK's Home Monitoring®

{4}------------------------------------------------

# 5. Indications for Use

The BioMonitor 2 Insertable Cardiac Monitor is indicated to detect the following cardiac arrhythmias:

- Atrial fibrillation, .
- Bradycardia, ●
- Sudden rate drop,
- High ventricular rate (HVR), ●
- . Asystole.

The BioMonitor 2 Insertable Cardiac Monitoris indicated for:

- . Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- . Patients who experience transient symptoms that may suggest a cardiac arrhythmia
- The device has not been tested for and it is not intended for pediatric use.

#### 6. Comparison of Technological Characteristics with the Predicate Device

The technological principles of the subject and predicate device are the same. The differences represent minor modifications to the currently marketed BioMonitor 2 as follows:

- . Dimensional changes to the FIT accessories:
	- FIT1 assists to create the device pocket for insertion. Dimensions at the tip and handle of ● this accessory were modified for improved handling.
	- . FIT2 supports the device during insertion. Dimensions of this accessory were modified for improved support during insertion.
- Packaging changes to the FIT accessories:
	- FIT1 and FIT2 were provided separately in double sterile pouches, and may now be ● provided together in a sterile blister package.
- A visual inspection was added to the receiving inspection of batteries. Subsequent testing . and inspections, including acceptance criteria for device performance, are unchanged.

# 7. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

To demonstrate that the modified BioMonitor 2 meets the same performance criteria, the following tests were conducted using the same test methods and acceptance criteria as for the predicate device.

- Mechanical system verification of FIT1 and FIT2
- . Visual inspection and handling of FIT1 and FIT2
- Usability testing: system verification and validation
- Visual inspection of labeling ●
- Compatibility testing of BioMonitor 2 with FIT 2 ●
- . Procedure of implantation: Product Validation
- Bioburden
- Pvrogenicitv
- Sterilization validation ●
- . Biocompatibility
- Packaging validation, including seal integrity, environmental conditioning, sealing process, and label integrity

{5}------------------------------------------------

No clinical testing was deemed necessary or completed in the premarket notification submission for a determination of substantial equivalence.

# 8. Conclusions

The subject device results from minor modifications to the predicate device. The performance testing demonstrates that the subject device meet the same functional acceptance criteria for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD/K171514](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MXD/K171514)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
