← Product Code [MTD](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MTD) · K210766

# AcQMap 3D Imaging and Mapping Catheter, Model 900009 (K210766)

_Acutus Medical, Inc. · MTD · Jun 21, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MTD/K210766

## Device Facts

- **Applicant:** Acutus Medical, Inc.
- **Product Code:** [MTD](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MTD.md)
- **Decision Date:** Jun 21, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The AcQMap 3D Imaging and Mapping Catheter is intended to be used in the right and left atrial chambers to collect ultrasound data for visualizing the selected chamber and recording electrical impulses in patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

## Device Story

AcQMap Catheter is a sterile, single-use, invasive device inserted via femoral vein into heart chambers. It features 48 electrodes and 48 ultrasound transducers on six distal splines. Used with AcQMap System, it collects ultrasound data and non-contact electrograms for Charge Density/Voltage mapping. The device is steerable via sheath; does not require heart wall contact. It provides 3D anatomic maps and chamber-wide electrical activation data. Used by clinicians in cardiac electrophysiology settings to identify complex arrhythmias, aiding clinical decision-making for ablation or other therapies. Benefits include improved visualization and mapping of complex cardiac electrical activity.

## Clinical Evidence

No new clinical testing performed. Clinical evidence incorporated by reference from the original AcQMap Catheter (K170819/K170948). Evidence includes the DDRAMATIC-SVT study: a prospective, non-randomized, open-label study of 84 patients at eight clinical sites, demonstrating safety and effectiveness.

## Technological Characteristics

Sterile, single-use catheter with 48 electrodes and 48 ultrasound transducers (10 MHz center frequency). M-mode ultrasound. 25mm deployed diameter. 100cm working length. Electrical source from AcQMap System. EO sterilization. Optional shaft marker band. Compatible with 0.035" guidewire and AcQGuide MAX steerable sheath.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- AcQMap 3D Imaging and Mapping Catheter ([K201341](/device/K201341.md))

## Reference Devices

- AcQMap 3D Imaging and Mapping Catheter ([K170819](/device/K170819.md))
- AcQMap 3D Imaging and Mapping Catheter ([K170948](/device/K170948.md))
- AcQGuide MAX 12 F Steerable Sheath ([K162925](/device/K162925.md))
- AcQGuide MAX 12 F Steerable Sheath ([K211100](/device/K211100.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 21, 2022

Acutus Medical, Inc. Karla Schaffner Principal Regulatory Affairs Specialist 2210 Faraday Ave, Suite 100 Carlsbad, California 92008

Re: K210766

Trade/Device Name: AcQMap® 3D Imaging and Mapping Catheter, Model 900009 Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD. ITX Dated: June 13, 2022 Received: June 14, 2022

Dear Karla Schaffner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

542 of the Act); 21 CFR 1000-1050.

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210766

Device Name

AcQMap® 3D Imaging and Mapping Catheter, Model 900009

Indications for Use (Describe)

The AcQMap 3D Imaging and Mapping Catheter is intended to be used in the right and left atrial chambers to collect ultrasound data for visualizing the selected chamber and recording electrical impulses in patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in gray, with a small orange circle inside the letter "A". Below the word "ACUTUS" is the word "MEDICAL" in orange. The logo is simple and modern, and it is likely used to represent the company's brand.

## 510(K) Notification K_210766

# General Information [807.92(a)(1)]

#### Date Prepared: 21 June 2022

| Applicant                                                                                                                         | Contact Person                                                                                                                                                                                                                       |
|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Acutus Medical, Inc.<br>2210 Faraday Ave., Suite 100<br>Carlsbad, CA 92008<br>USA<br>Phone: 1-442-232-6080<br>Fax: 1-442-232-6081 | Karla Schaffner<br>Principal Regulatory Affairs Specialist<br>Acutus Medical, Inc.<br>2210 Faraday Ave., Suite 100<br>Carlsbad, CA 92008<br>USA<br>Phone: 1-442-232-6161<br>FAX: 1-442-232-6081<br>Email: karla.schaffner@acutus.com |

# Device Information [807.92(a)(2)]

#### Trade Name:

AcQMap® 3D Imaging and Mapping Catheter, Model 900009

#### Generic/Common Name:

Electrode Recording Catheter or Electrode Recording Probe/ Transducer Ultrasonic

# Classification:

Class II / 21 CFR § 870.1220 and Class II / 21 CFR § 892.1570

### Product Code(s):

MTD/ITX

# Predicate Device [807.92(a)(3)]

AcQMap 3D Imaging and Mapping Catheter (K201341)

#### Reference Device

AcQMap 3D Imaging and Mapping Catheter (K170819)

# Device Description [807.92(a)(4)]

{4}------------------------------------------------

The AcQMap 3D Imaging and Mapping Catheter ("AcQMap Catheter") is provided sterile and is a singleuse, non-pyrogenic, invasive device that is inserted into the femoral vein and advanced through the venous circulatory system to the inferior vena cava and into the right and/or left atrium of the heart.

The AcQMap Catheter consists of a shaft with a lumen, an integral handle with a deployment mechanism and a flush port, and a connector. The AcQMap Catheter is intended to be used with the AcQMap High Resolution Imaging and Mapping System ("AcQMap System"). The AcQMap Catheter will provide ultrasound imaging and non-contact electrograms for Charge Density/Voltage heart chamber mapping. The AcQMap Catheter has six (6) splines at the distal end that support a configuration of 48 electrodes and 48 ultrasound transducers. The AcQMap Catheter is placed within the desired heart chamber and the distal end is deployed. There is no requirement for the electrodes or transducers to be in contact with the heart wall. The AcQMap Catheter is capable of over-the-wire delivery and contains a flexible distal segment that allows it to be directed via a steerable sheath to various locations of interest within the heart. Refer to for images of the device.

The AcQMap Catheter, Model 900009 is sterilized by EO method, has a reusable interface cable, is compatible with a 0.035" guidewire, and has minor design changes to improve manufacturability. The shaft of the catheter has a shaft marker band, as an optional feature, to allow the user to visualize whether the distal array is inside or outside the AcQGuide MAX Steerable Sheath (900200).

## INDICATIONS FOR USE [807.92(a)(5)]

The AcQMap 3D Imaging and Mapping Catheter is intended to be used in the right and left atrial chambers to collect ultrasound data for visualizing the selected chamber and recording electrical impulses in patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

# Comparison of Technological Characteristics with The Predicate Devices [807.92(A)(6)]

| Characteristics                                                                                             | SUBJECT<br>Modified AcQMap 3D Imaging<br>and Mapping Catheter, Model<br>900009                                                                                                                                                                                                                                                                        | PREDICATE<br>AcQMap 3D Imaging and<br>Mapping Catheter, Model<br>900009                                                                                                                                                                                                                                                                               | Rationale for<br>Substantial<br>Equivalence                                                                                                                                                                           |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory                                                                                                  |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                       |
| 510(k) Number                                                                                               | K210766                                                                                                                                                                                                                                                                                                                                               | K201341                                                                                                                                                                                                                                                                                                                                               | --                                                                                                                                                                                                                    |
| Classification/<br>Regulation Number/<br>Regulation Name/<br>Product Code                                   | Class II / 21 CFR §870.1220 /<br>Electrode recording catheter or<br>electrode recording probe / MTD<br>Class II / 21 CFR §892.1570 /<br>Transducer Ultrasonic/ ITX                                                                                                                                                                                    | Class II / 21 CFR §870.1220 /<br>Electrode recording catheter or<br>electrode recording probe / MTD<br>Class II / 21 CFR §892.1570 /<br>Transducer Ultrasonic/ ITX                                                                                                                                                                                    | Identical                                                                                                                                                                                                             |
| Characteristics                                                                                             | SUBJECT<br>Modified AcQMap 3D Imaging<br>and Mapping Catheter, Model<br>900009                                                                                                                                                                                                                                                                        | PREDICATE<br>AcQMap 3D Imaging and<br>Mapping Catheter, Model<br>900009                                                                                                                                                                                                                                                                               | Rationale for<br>Substantial<br>Equivalence                                                                                                                                                                           |
| Indications for Use                                                                                         | The AcQMap 3D Imaging and<br>Mapping Catheter is intended to be<br>used in the right and left atrial<br>chambers to collect ultrasound<br>data for visualizing the selected<br>chamber and recording electrical<br>impulses in patients with complex<br>arrhythmias that may be difficult to<br>identify using conventional mapping<br>systems alone. | The AcQMap 3D Imaging and<br>Mapping Catheter is intended to be<br>used in the right and left atrial<br>chambers to collect ultrasound<br>data for visualizing the selected<br>chamber and recording electrical<br>impulses in patients with complex<br>arrhythmias that may be difficult to<br>identify using conventional mapping<br>systems alone. | Identical                                                                                                                                                                                                             |
| Physical Characteristics                                                                                    |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                       |
| Configuration                                                                                               | Electrode/Transducer mounted on<br>splines                                                                                                                                                                                                                                                                                                            | Electrode/Transducer mounted on<br>splines                                                                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                             |
| Deployed Diameter                                                                                           | 25mm                                                                                                                                                                                                                                                                                                                                                  | 25mm                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                             |
| Working Length                                                                                              | 100 cm                                                                                                                                                                                                                                                                                                                                                | 100 cm                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                                                                                                             |
| Number of Spines                                                                                            | 6                                                                                                                                                                                                                                                                                                                                                     | 6                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                             |
| Heart Wall Contacting                                                                                       | No                                                                                                                                                                                                                                                                                                                                                    | No                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                             |
| Deflection Y/N                                                                                              | No                                                                                                                                                                                                                                                                                                                                                    | No                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                             |
| Flush Port                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                             |
| Marker Band                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                                                                                                                    | Shaft marking<br>provides visual<br>feedback for distal<br>tip location<br>relative to the<br>AcQGuide MAX<br>sheath tip.<br>Marker is optional<br>for use and does<br>not improve<br>safety.                         |
| Technical Specifications - Electrical                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                       |
| Voltage                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                             |
| Charge Density                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                             |
| Number of Electrodes                                                                                        | 48                                                                                                                                                                                                                                                                                                                                                    | 48                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                             |
| Data Points/heartbeat                                                                                       | 2500                                                                                                                                                                                                                                                                                                                                                  | 2500                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                             |
| Power source                                                                                                | Electrical source from AcQMap<br>System                                                                                                                                                                                                                                                                                                               | Electrical source from AcQMap<br>System                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                             |
| Technical Specifications - Ultrasound                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                       |
| Characteristics                                                                                             | SUBJECT<br>Modified AcQMap 3D Imaging<br>and Mapping Catheter, Model<br>900009                                                                                                                                                                                                                                                                        | PREDICATE<br>AcQMap 3D Imaging and<br>Mapping Catheter, Model<br>900009                                                                                                                                                                                                                                                                               | Rationale for<br>Substantial<br>Equivalence                                                                                                                                                                           |
| Mode                                                                                                        | M-mode                                                                                                                                                                                                                                                                                                                                                | M-mode                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                                                                                                             |
| Number of<br>transducers                                                                                    | 48                                                                                                                                                                                                                                                                                                                                                    | 48                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                       |
| Phased Array Y/N                                                                                            | N                                                                                                                                                                                                                                                                                                                                                     | N                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                       |
| Center Frequency                                                                                            | 10 MHz                                                                                                                                                                                                                                                                                                                                                | 10 MHz                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                       |
| Acoustic Output                                                                                             |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                       |
| MI (Mechanical Index)                                                                                       | 0.06                                                                                                                                                                                                                                                                                                                                                  | 0.06                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                             |
| ISTPA.3 (Derated<br>Spatial-Peak<br>Temporal-Average<br>Intensity (milliwatts<br>per square<br>centimeter)) | 0.08 (mW/cm²)                                                                                                                                                                                                                                                                                                                                         | 0.08 (mW/cm²)                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                       |
| ISPPA.3 (Derated<br>Spatial-Peak Pulse<br>Average Intensity<br>(watts per square<br>centimeter))            | 1.03                                                                                                                                                                                                                                                                                                                                                  | 1.03                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                       |
| Accessories                                                                                                 |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                       |
| Dimensionally<br>Compatible Sheath                                                                          | For Use with Acutus<br>Medical<br>AcQGuide MAX 12 F Steerable<br>Sheath (K162925 and K211100)                                                                                                                                                                                                                                                         | For Use with<br>Acutus   Medical<br>AcQGuide 12 F Steerable Sheath<br>(K162925)                                                                                                                                                                                                                                                                       | The compatibility<br>with the recently<br>cleared<br>2nd<br>generation<br>AcQGuide<br>MAX<br>Steerable<br>Sheath<br>does not create a<br>new risk as the<br>two sheaths<br>are<br>representative<br>of<br>each other. |
| Compatible Guidewire                                                                                        | 0.035" (0.89 mm) diameter J-tip<br>guidewire                                                                                                                                                                                                                                                                                                          | 0.035" (0.89 mm) diameter J-tip<br>guidewire                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                             |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in gray, block letters, with a small orange circle in the lower left corner of the "A". Below the word "ACUTUS" is the word "MEDICAL" in smaller, orange letters. The logo is simple and modern, and the use of color helps to make it stand out.

ACQMAP® 3D IMAGING AND MAPPING CATHETER

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Acutus Medical. The logo has the word "ACUTUS" in gray, block letters. Below that, the word "MEDICAL" is spelled out in smaller, orange letters. There is a registered trademark symbol to the right of the letter "S" in "ACUTUS".

ACQMAP® 3D IMAGING AND MAPPING CATHETER

## Substantial Equivalence

The indications for use of the subject device are identical to those of the predicate device. The difference between the subject and predicate AcQMap Catheter model 900009 is the addition of the marker band on the shaft of the catheter. This is an optional feature, to allow the user to visualize whether the distal

{7}------------------------------------------------

array is inside or outside the AcQGuide MAX Steerable Sheath (900200). The marker band does not improve safety, nor is it detrimental if used with the AcQGuide MAX Steerable Sheath (K162925). This difference between the devices does not raise any different questions of safety or effectiveness. Thus, the modified AcQMap 3D Imaging and Mapping Catheter, is substantially equivalent to the predicate AcQMap 3D Imaging and Mapping Catheter device.

## PERFORMANCE DATA [807.92(b)]

All necessary bench testing was conducted on the modified AcQMap 3D Imaging and Mapping Catheter to support a determination of substantial equivalence to the predicate device. The necessary clinical testing was completed for the original AcQMap Catheter (K170819) and is incorporated by reference. No further clinical testing is required to support the subject device.

# Nonclinical Testing Summary [807.92(b)(1)]

The nonclinical, bench testing is repeated or included in this submission by reference. A full list of the nonclinical testing is provided below:

- . Design Verification
	- Dimensional Inspection O
	- O Visual Inspection
	- 0 Functional and Compatibility Testing
	- Mechanical Testing O
	- 0 Corrosion Testing
	- Electrode Coating Particulate Testing O
	- Acoustic Output Testing O
	- 0 Accuracy Testing
- Design Validation
	- Usability Testing 0
	- 0 Animal Testing
- Biocompatibility Testing

In addition, Acutus performed sterilization, shelf life and packaging validations. The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the AcQMap 3D Imaging and Mapping Catheter meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the AcQMap 3D Imaging and Mapping Catheter does not raise different questions of safety or effectiveness for collecting data that enables the creation of 3D anatomic maps that display chamber-wide electrical activation when compared to the predicate device.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Acutus Medical. The logo is composed of the word "ACUTUS" in gray, with the word "MEDICAL" in orange underneath. The "A" in "ACUTUS" has an orange dot in the middle. There is a registered trademark symbol next to the "S" in "ACUTUS".

## Clinical Testing Summary [807.92(b)(2)]

As discussed above, no further clinical testing is required to support the modified AcQMap 3D Imaging and Mapping Catheter. The necessary clinical testing was completed for the original AcQMap 3D Imaging and Mapping Catheter (K170948) and is incorporated by reference. That study, entitled, "Dipole Density Right (and left) Atrial Mapping and Assessment of Therapy In Comblex Supraventricular Tachycardia, (DDRAMATIC-SVT)" was a prospective, non-randomized, open-label study conducted at eight clinical sites outside the U.S. The results for 84 patients demonstrated that the AcQMap 3D Imaging and Mapping Catheter is safe and effective for its intended use.

### Conclusions

Extensive nonclinical performance testing, either repeated for the modified device or incorporated by reference to the original AcQMap 3D Imaging and Mapping Catheter 510(k), was conducted on the modified AcQMap 3D Imaging and Mapping Catheter to evaluate the overall performance of the device. The clinical validation of the original AcQMap 3D Imaging and Mapping Catheter (K170819) is applicable to the modified device. The subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MTD/K210766](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MTD/K210766)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
