K201941 · Ge Medical Systems Information Technologies, Inc. · MHX · Nov 4, 2020 · Cardiovascular
Device Facts
Record ID
K201941
Device Name
Monitor B125, Monitor B105
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Nov 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intrahospital transport. The monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner. The monitor B125/B105 is not intended for use during MRI. The monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network. The monitor B125/B105 monitors and displays: ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring during conditions of clinical patient motion or low perfusion). temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/ Core/Surface temperature. impedance respiration, respiration rate, airway Gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate), Cardiac output (C.O.) and Entropy. The monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole. Ventricular tachycardia. VT>2, Ventricular Bradycardia, Accelerated Ventricular Rhythm, Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Irregular, Multifocal PVCs. Missing Beat, SV Tachy, Premature Ventricular Contraction (PVC), Supra Ventricular Contraction (SVC) and Ventricular fibrillation.
Device Story
Portable multi-parameter patient monitor; inputs include ECG, invasive/non-invasive blood pressure, SpO2, temperature, impedance respiration, airway gases, cardiac output, and entropy via modular sensors. Transforms physiological signals into real-time waveforms, numeric trends, and alarms; utilizes EK-Pro V14 algorithm for arrhythmia detection. Used in hospital settings (bedside/transport) by clinicians. Outputs displayed on 10-inch (B105) or 12-inch (B125) LCD; supports remote printing and central station networking. Provides clinical decision support via National Early Warning Score (NEWS) and full disclosure waveform review (up to 36 hours). Benefits include continuous patient surveillance, early detection of physiological deterioration, and standardized alarm management.
Clinical Evidence
No clinical data required; substantial equivalence supported by bench testing, software verification/validation, and usability testing.
Indicated for monitoring and alarm generation for physiological parameters in adult, pediatric, and neonatal patients in hospital environments and during intrahospital transport. Not for use during MRI. Requires direct supervision by a licensed health care practitioner.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 04, 2020
GE Medical Systems Information Technologies, Inc. Joel Kent Senior Regulatory Affairs Manager 8200 West Tower Avenue Milwaukee, Wisconsin 53223
Re: K201941
Trade/Device Name: Monitor B125/B105 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX Dated: October 2, 2020 Received: October 5, 2020
Dear Joel Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201941
Device Name Monitor B125/B105
#### Indications for Use (Describe)
The monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intrahospital transport.
The monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner.
The monitor B125/B105 is not intended for use during MRI.
The monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network.
The monitor B125/B105 monitors and displays: ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/ Bladder/Axillary/Skin/Airway/Room/Myocardial/ Core/Surface temperature, impedance respiration rate, airway Gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate), Cardiac output (C.O.) and Entropy.
The monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole, Ventricular tachycardia, VT>2, Ventricular Bradycardia, Accelerated Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Irregular, Multifocal PVCs, Missing Beat, SV Tachy, Premature Ventricular Contraction (PVC), Supra Ventricular Contraction (SVC) and Ventricular fibrillation.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|------------------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K201941
Section 5: 510(k) Summary
# Monitor B125/B105
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR 807.92(c) the following summary of information is provided:
| Owner/Contact/Date (807.92(a)(1): | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | July 10th, 2020 |
| Submitter: | GE Medical Systems Information Technologies, Inc.<br>8200 West Tower Avenue<br>Milwaukee, Wisconsin 53223<br>USA |
| Primary Contact Person: | Joel Kent<br>Senior Regulatory Affairs Manager<br>GE Healthcare<br>Phone: 617-851-0943<br>E-mail: joel.kent@ge.com |
| Secondary Contact Person: | Monica Morrison<br>Sr. Regulatory Affairs Director<br>GE Healthcare<br>E-mail: monica.morrison@ge.com<br><br>Sun Yanli<br>Regulatory Affairs Program Manager<br>GE Healthcare<br>E-mail: yanli.sun@ge.com |
| Device names (807.92(a)(2): | |
| Trade Name: | Monitor B125/B105 |
| Common/Usual Name: | Multiparameter patient monitor (monitor, physiological, patient<br>(with arrhythmia detection or alarms) |
{5}------------------------------------------------
| Classification Names: | 21 C.F.R. §868.1400 Carbon dioxide gas analyzer.<br>21 C.F.R. §868.1500 Enflurane gas analyzer.<br>21 C.F.R. §868.1620 Halothane gas analyzer.<br>21 C.F.R. §868.1700 Nitrous oxide gas analyzer.<br>21 C.F.R. §868.1720 Oxygen gas analyzer.<br>21 C.F.R. §868.2375 Breathing frequency monitor.<br>21 C.F.R. §870.1025 Arrhythmia detector and alarm (including<br>ST-segment measurement and alarm).<br>21 C.F.R. §870.1110 Blood pressure computer.<br>21 C.F.R. §870.1130 Noninvasive blood pressure measurement<br>system.<br>21 C.F.R. §870.1425 Programmable diagnostic computer.<br>21 C.F.R. §870.1915 Thermodilution probe.<br>21 C.F.R. §870.2300 Cardiac monitor (including<br>cardiotachometer and rate alarm).<br>21 C.F.R. §870.2700 Oximeter<br>21 C.F.R. §870.2710 Ear oximeter.<br>21 C.F.R. §880.2910 Clinical electronic thermometer.<br>21 C.F.R. §882.1400 Electroencephalograph<br>21 C.F.R. §870.1100 alarm, blood-pressure |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code: | MHX |
| Subsequent Product<br>Code: | BZQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSJ,<br>DSK, GWQ, FLL, NHO, NHP, NHQ, DPZ, DQK, DSI, DSJ,<br>KRB, MLD, OLT, OLW, OMC, ORT. |
| Predicate Device(s)<br>(807.92(a)(3)): | Primary Predicate: K171580 Monitor B125/B105<br><br>Additional Predicate Devices:<br>K151063 Monitor B40<br>K191323 CARESCAPE B850<br>K143676 B40i Patient Monitor<br>K162012 CARESCAPE Central Station V2<br>K122223 IntelliVue MP5SC with IntelliVue Guardian Software |
| Device Description<br>(807.92(a)(4)): | The proposed Monitor B125/B105 is a multi-parameter patient<br>monitor that was developed based on predicate Monitor<br>B125/B105 (K171580) to provide additional monitored<br>parameters: Airway gases, Entropy and Cardiac output, by<br>supporting additional optional modules previously cleared by<br>FDA: CARESCAPE Respiratory module (K171028), Airway Ga<br>Option N-CAiO(K151063), E-Entropy Module (K150298) (E-<br>ENTROPY-01) and E-COP module (K052976) with extension<br>interface rack and/or second frame(B1X5-F2). |
In addition to the added parameters, the proposed Monitor
{6}------------------------------------------------
B125/B105 offers several enhanced software features:
- . Enabled the Impedance Respiration measurement for Neonates (in addition to Ped/Adult);
- Option called "Full Disclosure" to allow the display of . parameter waveforms for up to 36 hours;
- . National Early Warning Score (NEWS) calculation provided ;
- . Additional printing capabilities to a remote recorder/Laser printer connected to a central station;
- Adoption of EK-Pro V14 ECG algorithm (previously ● cleared K191323) to support enhanced arrhythmia detection performance;
- Added Irregular, SV Tachy and Supra Ventricular . Contraction (SVC), three more arrhythmia alarms;
- Enhanced parameter alarm priority adjustment/ . configuration options,
- . Cybersecurity enhancements.
The proposed monitor B125 and B105 is based on the previous design, and therefore shares a common software and hardware platform with its primary predicate, the Monitor B125/B105 (K171580). The primary function and operation of the monitors remain unchanged from the predicate. The difference between the two models (B125 and B105) is the LCD screen size. B125 has a 12-inch display; B105 has a 10-inch display. There is no change from the predicate in the display size.
As with the predicate Monitor B125/B105 (K171580), the proposed Monitor B125/B105 is a multi-parameter patient monitor, utilizing an LCD display with an integrated keypad and an identical pre-configuration patient parameter measurement module (Hemo module) which provides basic parameters: ECG, RESP, NIBP, IBP, TEMP, SpO2. The proposed Monitor B125/B105 uses the identical E-MiniC module (K052582) and equivalent optional thermal recorder module as the predicate B125/B105 (K171580).
As with the predicate Monitor B125/B105 (K171580), the proposed Monitor B125/B105 interfaces to a variety of existing central station systems via a cabled or wireless network interface. The wireless interface is implemented with the identical integrated WIFI module (WLAN module type: B1x5-01. FCC ID: OU5B1X501) as used in the predicate Monitor B125/B105 (K171580).
{7}------------------------------------------------
As with the predicate Monitor B125/B105 (K171580), proposed Monitor B105/B125 includes features and subsystems that are optional or configurable. It can be mounted in a variety of ways (e.g. shelf, countertop, table, wall, pole, or head/foot board) using existing mounting accessories.
Indications for Use The monitor B125/B105 is a portable multi-parameter unit to be used for monitoring, recording, and to generate alarms for (807.92(a)(5)): multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intrahospital transport.
> The monitor B125/B105 is intended for use under the direct supervision of a licensed health care practitioner.
> The monitor B125/B105 is not intended for use during MRI.
The monitor B125/B105 can be a stand-alone monitor or interfaced to other devices via network.
The monitor B125/B105 monitors and displays: ECG (including ST segment, arrhythmia detection, ECG Diagnostic Analysis and Measurement), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring( including monitoring during conditions of clinical patient motion or low perfusion). temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/S kin/Airway/Room/Myocardial/ Core/Surface temperature. impedance respiration, respiration rate, airway Gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate), Cardiac output (C.O.) and Entropy.
The monitor B125/B105 is able to detect and generate alarms for ECG arrhythmias: Asystole. Ventricular tachycardia. VT>2. Ventricular Bradycardia, Accelerated Ventricular Rhythm, Ventricular Couplet, Bigeminy, Trigeminy, "R on T", Tachycardia, Bradycardia, Pause, Atrial Fibrillation, Irregular, Multifocal PVCs. Missing Beat, SV Tachy, Premature Ventricular Contraction (PVC), Supra Ventricular Contraction (SVC) and Ventricular fibrillation.
### Contraindications for using the monitor
{8}------------------------------------------------
The monitor B125/B105 is not intended for use during MRI
The Monitor B125/B105 is a multi-parameter patient monitor. Technology (807.92(a)(6)): The hardware functionality is unchanged compared to the predicate Monitor B125/B105 (K171580), however this 510(k) introduces monitoring of additional (previously cleared) parameters through existing cleared and marketed measurement modules, enhancements of several software features and additional cybersecurity enhancements.
> The fundamental technology of the proposed Monitor B125/B105 is the same as the predicate devices.
The proposed Monitors B125/B105 is substantially equivalent to the predicate devices.
A summary of the main changes compared to the predicate are listed below in the comparison table.
| Specification | Predicate monitor<br>B125/B105 (K171580) | Proposed monitor<br>B125/B105 | Differences |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient type | Adult, pediatric & neonatal | Adult, pediatric & neonatal | Identical |
| Use environments | The Patient Monitor is<br>intended for use in multiple<br>areas within a professional<br>healthcare facility | The Patient Monitor is<br>intended for use in multiple<br>areas within a professional<br>healthcare facility | Identical |
| Intrahospital<br>transport within a<br>professional<br>healthcare facility | Yes | Yes | Identical |
| Size (H x W x D) | Physical size without<br>extension rack and modules:<br>B105: 270 mm x 290 mm x<br>150 mm<br>B125: 280 mm x 317 mm x<br>150 mm<br>Horizontal space projection<br>size without extension rack<br>and modules:<br>B105: 265 mm x 290 mm x<br>165 mm<br>B125: 280 mm x 320 mm x<br>165 mm | Physical size without<br>extension rack and modules:<br>B105: 270 mm x 290 mm x<br>150 mm<br>B125: 280 mm x 317 mm x<br>150 mm<br>Horizontal space projection<br>size without extension rack<br>and modules:<br>B105: 265 mm x 290 mm x<br>165 mm<br>B125: 280 mm x 320 mm x<br>165 mm | Identical |
| Weight | Weight with battery, without<br>extension rack and modules:<br>B105: $\[le]$ 3.9 kg<br>B125: $\[le]$ 4.3 kg | Weight with battery, without<br>extension rack and modules:<br>B105: $\[le]$ 3.9 kg<br>B125: $\[le]$ 4.3 kg | Identical |
| Module Housing | Optional rack which provides<br>one slot for E-Modules | Optional rack which provides<br>one slot for E-Modules.<br>Optional Second Frame<br>B1x5-F2 which provides two<br>slots for E-Modules. | Equivalent to predicate.<br>The Proposed monitor supports the<br>optional Second Frame to provide<br>more slots for E-Modules to improve<br>parameter acquisition capability.<br>One monitor can connect to one rack<br>and one Second Frame at a time. |
| Display/screen | TFT LCD with Touch screen<br>B105: 10.1-inch<br>B125: 12.1-inch | TFT LCD with Touch screen<br>B105: 10.1-inch<br>B125: 12.1-inch | Equivalent to predicate<br>12.1inch LCD qualified a new<br>supplier.<br>The display from new supplier meets<br>the same specifications and<br>functionality. |
| Battery Type | Rechargeable Lithium-Ion<br>batteries | Rechargeable Lithium-Ion<br>batteries | Identical |
| Waveforms and<br>parameter windows | Up to 6 waveforms<br>Up to 10 parameter windows | Up to 6 waveforms<br>Up to 10 parameter windows | Identical |
| Operating System | Linux<br>(Rev 3.18) | Linux<br>(Rev 4.4) | Equivalent to predicate.<br>The two versions of the Operation<br>System use the same fundamental<br>technology.<br>The newer version of Linux includes<br>equivalent clinical user functionality<br>with improved operating system<br>security. |
| Modules | E-MiniC | E-MiniC<br>E-sCO, E-sCAiO<br>N-CAiO<br>E-ENTROPY-01<br>E-COP | Equivalent to predicate.<br>Support added for additional<br>parameter measurement modules<br>including: CARESCAPE Respiration<br>Module (K123195) (E-sCO, E-sCAiO)<br>Airway Gas Module N-CAiO (Cleared<br>with B40(K151063))<br>E-Entropy Module E-ENTROPY-<br>01(K150298)<br>And COP module (K052976).<br>The measurement modules are cleared<br>and available on the market. This<br>change solely adds the capability to<br>display these additional parameters on |
| | | | |
| Available parameters | ECG, ST segment, arrhythmia<br>detection, ECG diagnostic<br>analysis and measurement,<br>invasive pressure, non-<br>invasive blood pressure, pulse<br>oximetry, temperature,<br>impedance respiration, airway<br>gases (CO2) | ECG, ST segment, arrhythmia<br>detection, ECG diagnostic<br>analysis and measurement,<br>invasive pressure, non-<br>invasive blood pressure, pulse<br>oximetry, temperature,<br>impedance respiration, airway<br>gases (CO2, O2, N2O, and<br>anesthetic agents), Entropy,<br>cardiac output. | Equivalent to predicate.<br>Additional airway gases parameters<br>(O2, N2O, Anesthetic Agents) are<br>added and can be displayed on the<br>screen, through the<br>compatibility/support for<br>CARESCAPE Respiration module<br>(K123195) and Airway Gas Module<br>N-CAIO (cleared with B40(K151063)<br>Additional Entropy monitoring<br>capability is added through the<br>compatibility/support for additional<br>supporting for E-Entropy Module<br>E-ENTROPY-01(K150298)<br>Additional Cardiac Output monitoring<br>is added through the<br>compatibility/support for additional<br>supporting for E-COP module<br>(K052976)<br>This change solely adds the capability<br>to display these additional parameters<br>on the B105/B125 display, similar to<br>the other displayed parameters. |
| EK-Pro arrhythmia<br>detection algorithm | EK-Pro V12 | EK-Pro V14 | Equivalent to predicate.<br>The Ek-Pro V14 algorithm was<br>cleared with CARESCAPE B850 in<br>K191323.<br>This cleared algorithm provides<br>equivalent functionality and was<br>updated to further enhance the<br>analysis. |
| Printing | Local thermal recorder for<br>printings waveforms, alarms<br>waveforms, numeric trends | Local thermal recorder for<br>printing waveforms, alarms<br>waveforms, numeric trends<br>Remote printer via central<br>station for printing waveforms,<br>alarms waveforms, numeric<br>trends | Equivalent to predicate. Equivalent<br>functionality with an additional<br>supported printer option.<br>Proposed device supports additional<br>remote printing via central station to<br>address customer requests. |
| Thermal Recorder<br>Module | B1X5-REC thermal recorder<br>module with "off the shelf"<br>printer core XE-50 | B1X5-REC thermal recorder<br>module with "off the shelf"<br>printer core GPR212-M0 | Equivalent to predicate.<br>Both XE-50 and GPR212-M0 are<br>Industry's standard 50mm thermal<br>printer.<br>XE-50 and GPR212-M0 have the<br>same specifications except GPR212-<br>M0 (APS) has lower power<br>consumption. |
| Alarm Classification<br>(IEC) | Four levels — High, Medium,<br>Low and Informational | Four levels — High, Medium,<br>Low and Informational | Identical |
| Alarm Notification | Audible and visual | Audible and visual | Identical |
| Alarm Priority<br>adjustment | Support to adjust ECG (only<br>Arrhythmia) alarm priority | Support to adjust ECG, IBP,<br>SpO2, NIBP, Respiration,<br>CO2, Gas and Temperature<br>technical alarm and<br>physiological alarms priority | Equivalent to predicate.<br>The predicate already included the<br>option to adjust and configure some<br>alarm settings and priorities. In the<br>proposed monitor, more flexibility was<br>added for the user to adjust alarm<br>priorities for additional parameter<br>alarms to meet the user requirements<br>of different hospitals and departments<br>within each hospital.<br>This alarm priority configuration<br>option and mechanism has already<br>been implemented in the identical way<br>in the predicate CARESCAPE B850<br>(K191323).<br>The default settings are according to<br>IEC 60601-1-8 |
| Full Disclosure | Not supported | Acquires ECG and mixed<br>parameters waveforms data<br>from monitor and displays<br>for review and storage for up<br>to 36 hours. | Equivalent to predicate.<br>Full disclosure is an optional license-<br>based feature to display parameter<br>waveforms up to 36 hours. It provides<br>the clinician the flexibility to view<br>historical waveforms on bedside<br>monitors, rather than solely relying on<br>an additional central station. It is<br>equivalent to the full disclosure<br>functionality within the predicate<br>CARESCAPE Central Station V2<br>(K162012). |
| Early Warning Score | Not supported | NEWS (National Early Warning Score)<br><br>NEWS is an optional license-based feature which allows the display of the NEWS Early Warning Score on the monitor. The monitor calculates the score based on the measured parameters following the exact formula for NEWS published from the Royal College of Physicians (a published/well-established clinical score). | Equivalent to Predicate.<br>The proposed Monitor B125/B105 V1.5 supports display/calculation of the National Early Warning Score (NEWS). NEWS is an optional license-based feature which allows for the display of a calculated/ well-established "Early Warning Score" that clinicians often use as one part of patient assessments/ clinical decision support looking for early signs of deterioration. The Early Warning Score concept is used in many monitors and even EMRs to provide standardized calculations. This is equivalent to the early warning score implemented in the legally marketed predicate Monitor IntelliVue MP5SC with IntelliVue Guardian Software (K122223). |
| ECG Alarms | ECG arrhythmias alarms:<br>Asystole, Ventricular tachycardia, VT>2,<br>Ventricular Bradycardia,<br>Accelerated Ventricular Rhythm, Ventricular Couplet,<br>Bigeminy, Trigeminy, “R on T”, Tachycardia, Bradycardia,<br>Pause, Atrial Fibrillation,<br>Multifocal PVCs, Missing Beat, Premature Ventricular Contraction (PVC), and Ventricular fibrillation. | Three additional alarm types added (highlighted in <b>bold</b> ).<br><br>ECG arrhythmias alarms:<br>Asystole, Ventricular tachycardia, VT>2,<br>Ventricular Bradycardia,<br>Accelerated Ventricular Rhythm, Ventricular Couplet,<br>Bigeminy, Trigeminy, “R on T”, Tachycardia, Bradycardia,<br>Pause, Atrial Fibrillation, <b>Irregular</b> , Multifocal PVCs, Missing Beat, <b>SV Tachy</b> , Premature Ventricular Contraction (PVC), <b>Supra Ventricular Contraction (SVC)</b> and Ventricular fibrillation. | Equivalent to predicate.<br><br>Added additional ECG arrhythmia alarms types based on the implemented algorithm. The additions are: Irregular, SV Tachy and Supra Ventricular Contraction (SVC). Provides additional options within the same monitored parameter as the predicate. |
| Neonatal respiration | Adult, Pediatric | Adult, Pediatric, Neonatal | Equivalent to predicate<br><br>The proposed device uses the identical Impedance Respiration design and algorithm as both the predicate Monitor B125/B105 V1(K171580) and B40i (K143676). It is now being implemented for neonates in the equivalent way to the Monitor B40i (K143676). |
| Networking Interface | LAN<br>WLAN (Optional) | LAN<br>WLAN (Optional) | Identical |
| Networking Protocol | CARESCAPE Network HL7 | CARESCAPE Network HL7 | Identical |
Product Comparison versus Predicate Main Features:
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
Determination of Substantial Equivalence (807.92(b)(1):
### Summary of Non-Clinical Tests:
Bench testing related to software, hardware and performance including applicable consensus standards was conducted on the proposed Monitor B125/B105, demonstrating the design meets the specifications.
The hardware bench testing included electromagnetic compatibility, electrical safety, environmental, WLAN, and usability.
The proposed Monitor B125/B105 has been found to be equivalent to the predicate device(s) for the intended users, uses and use environments. Usability evaluation has been completed for the proposed monitor and the predicate devices including critical task identification through use-based hazard analysis, multiple rounds of formative usability testing and summative testing, among other activities.
Software testing included software design, development, verification, validation and traceability. Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered "Major" level of concern, the same as the predicate.
Patient safety, security, and privacy risks have been addressed in the design and development of the proposed B125/B105 including a Security Risk Assessment and Threat model. This includes system integrity controls, access controls, audit controls and network controls which map to the General Principles and Security Capabilities outlined in the FDA Guidance for Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on October 2, 2014.
Testing was completed to show the device can withstand network storm, i.e. continue to monitor patients without rebooting when connected by wire to GE Unity networks which suffer broadcast storm traffic.
{14}------------------------------------------------
## Clinical (807.92(b)(2)): Summary of Clinical Tests:
The subject of this premarket submission, the proposed Monitor B125/B105 did not require clinical studies to support substantial equivalence.
Conclusion (807.92(b)(3)): GE Healthcare considers the proposed Monitor B125/B105 to be substantially equivalent to the predicate device(s).