The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring; 12 Lead ECG recording with interpretation; real-time arritythmia detection / alarming: OT measurement / alarming and ST measurements / alarming: impedance pneumography; noninvasiveblood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate: pulse oximetry (SpO2); contact temperature; and invasive pressure and extended pulse oximetry capability including; carboxy haemoglobin (SpCO), methaemoglobin (SpMet), total haemoglobin (SpHb) and total oxygen content (SpOC) measurements. The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere). The device is indicated for adults, paediatrics and neonates. The monitor can be used to display images from Interson 3.5 MHz General Purpose (GP) and 7.5 MHz Small Parts / Vascular (SR) USB ultrasound probes or a Karl Storz C-MAC S USB video laryngoscope. These optional accessories are to be used in accordance with their Indications For Use.
Device Story
Tempus Pro is a portable multi-parameter vital signs monitor for pre-hospital and clinical use by trained professionals. Inputs include ECG leads, pulse oximetry sensors, NIBP cuffs, capnometry, temperature probes, invasive pressure transducers, and optional USB ultrasound probes or video laryngoscopes. Device processes physiological signals to provide real-time monitoring, arrhythmia detection, ST/QT analysis, and co-oximetry. Integrated GPS, camera, and voice communication support telemedicine workflows, transmitting data via wired/wireless links to remote response centers. Output includes visual display of waveforms/vitals and optional thermal printing of medical records. Device functions as a stand-alone monitor or telemedicine hub. Benefits include rapid patient assessment and remote clinical decision support in austere or pre-hospital environments.
Clinical Evidence
No clinical data; substantial equivalence demonstrated via non-clinical bench testing and design control processes.
Technological Characteristics
Portable vital signs monitor; includes optional thermal printer and secure 1-to-1 data link (TDL). Connectivity via wired/wireless for telemedicine. Supports USB peripherals (ultrasound, video laryngoscope). Monitoring includes ECG, SpO2, NIBP, ETCO2, temperature, invasive pressure. Software-based arrhythmia and ST/QT analysis.
Indications for Use
Indicated for adults, paediatrics, and neonates requiring monitoring of vital signs including ECG, arrhythmia, ST-segment, QT interval, impedance pneumography, NIBP, ETCO2, respiration rate, SpO2, contact temperature, invasive pressure, and co-oximetry (SpCO, SpMet, SpHb, SpOC) in clinical or pre-hospital settings.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Remote Diagnostic Technologies Ltd. James Hamlyn Senior Regulatory Affairs Specialist Pavilion C2, Ashwood Park, Ashwood Way Basingstoke, Hampshire RG23 8BG UNITED KINGDOM
Re: K173768
Trade/Device Name: Tempus Pro Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, MWI, CCK, DPS, DSB, DXN, DRT, IYO, ITX, MNR, DQA, FFL, DSK, DRG
Dear James Hamlyn:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 10, 2018. Specifically, FDA is updating this SE Letter to remove the secondary product code NSX as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter, please contact Aneesh Deoras, OHT2: Office of Cardiovascular Devices, 240-402-4363, Aneesh.Deoras@fda.hhs.gov.
Sincerely,
Aneesh S. Deoras -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
September 1, 2023
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
January 10, 2018
Remote Diagnostic Technologies Ltd. James Hamlyn Senior Regulatory Affairs Specialist Pavilion C2. Ashwood Park. Ashwood Wav Basingstoke, RG23 8BG UK
# Re: K173768
Trade/Device Name: Tempus Pro Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, MWI, CCK, DPS, DSB, DXN, DRT, IYO, ITX, MNR, DQA, FFL, DSK, DRG, NSX Dated: December 4, 2017 Received: December 11, 2017
Dear James Hamlyn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
# Page 2 - James Hamlyn
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Hillebrand
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K173768
Device Name Tempus Pro Patient Monitor
#### Indications for Use (Describe)
The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring; 12 Lead ECG recording with interpretation; real-time arritythmia detection / alarming: OT measurement / alarming and ST measurements / alarming: impedance pneumography; noninvasiveblood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate: pulse oximetry (SpO2); contact temperature; and invasive pressure and extended pulse oximetry capability including; carboxy haemoglobin (SpCO), methaemoglobin (SpMet), total haemoglobin (SpHb) and total oxygen content (SpOC) measurements.
The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).
The device is indicated for adults, paediatrics and neonates.
The monitor can be used to display images from Interson 3.5 MHz General Purpose (GP) and 7.5 MHz Small Parts / Vascular (SR) USB ultrasound probes or a Karl Storz C-MAC S USB video laryngoscope. These optional accessories are to be used in accordance with their Indications For Use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
## K173768
## 510(k) SUMMARY
#### Submitter's Information
| Name: | Remote Diagnostic Technologies Limited |
|------------------------|----------------------------------------------------------------------------------|
| Address: | Pavilion C2, Ashwood Park, Ashwood Way, Basingstoke, RG23<br>8BG, United Kingdom |
| Phone No: | +44 (0) 1256 362 400 |
| Fax No: | +44 (0) 1256 362 415 |
| Contact Person: | James Hamlyn, Senior Regulatory Affairs Specialist |
| Date Summary Prepared: | December 04, 2017 |
#### Device Identification
| Device Name: | Tempus Pro Patient Monitor |
|--------------------------|------------------------------------------------------------------------|
| Common Name: | Patient Monitoring System |
| Classification: | II |
| Product Code: | MHX |
| Secondary Product Codes: | MWI, CCK, DPS, DSB, DXN, DRT, IYO, ITX, MNR, DQA, FLL<br>DSK, DRG, NSX |
| Regulation Number: | 21 CFR 870.1025 |
## Predicate Device(s)
Tempus Pro Patient Monitor, K170567
#### Device Description
The Tempus Pro is a multi-parameter vital signs monitoring system designed for use in pre-hospital care and remote clinical locations by trained healthcare professionals. It provides 3 & 5 lead ECG recording, real-time arrhythmia detection & alarming, QT interval measurement & alarming, ST segment & alarming, impedance respiration, pulse oximetry (including Masimo Rainbow® co-oximetry measurements i.e. SpOC, SpHb, SpMet, SpCO, PVI and Pl), non-invasive blood pressure, sidestream capnometry, contact temperature, invasive pressure, and user configurable alarms. Third-party video laryngoscopes and ultrasound probes may also be connected to the device. In addition, it provides the ability to transmit all vital signs data via wired or wireless connections to a telemedicine software system expected to be based in a facility far from the user e.g. a response centre facility. In addition to sending all vital signs, the system can also send pictures or video via an integrated camera, geographic position by an integrated GPS receiver, and voice via a wired or wireless headset.
This submission is to notify the FDA of our intent to market the Tempus Pro modified by the addition of an optional thermal printer, which enables the user of a Tempus Pro to print medical records on-demand. Additionally, compatible external devices may also print medical records on the Tempus Pro's optional thermal printer when paired to the Tempus Pro via a secure 1-to-1 data link called Tempus Data Link (TDL).
> Page 1 of 2 K173768
{5}------------------------------------------------
### Indications for Use
No changes to the Indications for Use are proposed by this submission. The Indications for Use are as follows: The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring: 12 Lead ECG recording with interpretation: real-time arrhythmia detection / alarming: QT measurement / alarming and ST measurements / alarming; impedance pneumography; non-invasive blood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate; pulse oximetry (SpO2); contact temperature; and invasive pressure and extended pulse oximetry capability including; carboxy haemoglobin (SpCO), methaemoglobin (SpMet), total haemoglobin (SpHb) and total oxygen content (SpOC) The monitor is intended to be used as a stand-alone monitor or as a telemedicine system measurements. (transmitting patient data to other medical professionals located elsewhere). The device is indicated for adults, paediatrics and neonates. The monitor can be used to display images from Interson 3.5 MHz General Purpose (GP) and 7.5 MHz Small Parts / Vascular (SR) USB ultrasound probes or a Karl Storz C-MAC S USB video laryngoscope. These optional accessories are to be used in accordance with their Indications for Use.
## Comparison of Technological Characteristics
The unmodified Tempus Pro currently enables a user to print medical records on an external printer connected via USB cable, and to also export patient records to ePCR, email, and USB flash drives. The modified Tempus Pro will continue to provide these features but with the addition of an optional printing function that will enable a user to print medical data directly from the device's thermal printer. Additionally, compatible external devices may also print records on the Tempus Pro's optional thermal printer when paired with the Tempus Pro via a secure 1-to-1 data link. All other functionality of the Tempus Pro monitor remains unchanged. No changes have been made to the device's indications for use, intended use, or fundamental scientific technology as a result of the modification that has prompted this submission.
## Substantial Equivalence – Non-Clinical Evidence
The modifications to the device have been designed and assessed under design control processes compliant with FDA 21 CFR 820. Non-clinical testing has been performed to demonstrate that the safety and efficacy of the device remains substantially equivalent to its predicate.
## Substantial Equivalence - Clinical Evidence
Clinical evidence was not necessary to demonstrate substantial equivalence.
# Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, Remote Diagnostic Technologies Limited conclude that the Tempus Pro Patient Monitoring System with the changes described in this pre-market notification is as safe, as effective, and substantially equivalent to the predicate device.
> Page 2 of 2 K173768
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.