K133222 · Hanin Medical Services Co., Ltd. · MHX · Aug 6, 2014 · Cardiovascular
Device Facts
Record ID
K133222
Device Name
SAAD PATIENT MONITOR (SPM)
Applicant
Hanin Medical Services Co., Ltd.
Product Code
MHX · Cardiovascular
Decision Date
Aug 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
SPM is indicated for monitoring heart rate, invasive blood pressure, blood oxygen saturation, and respiration on adult, pediatric and neonatal populations. SPM is indicated for monitoring and detecting arrhythmia, pacer, and QRS on adult and pediatic populations only (not intended for use on neonatal populations). SPM will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits at which point timed or alarm recordings will be produced. SPM connects to an R50 Bedside recorder, or via the Saad Patient Monitoring network (SPMS). SPM does NOT perform ST Segment analysis and it is not intended to be used in transport or ambulatory environments. MR Environment Compatibility Statement: The Saad Patient Monitors are not compatible for use in an MR Environment (MR Unsafe).
Device Story
Saad Patient Monitor (SPM) is a bedside physiological monitor for clinical settings; inputs include heart rate, invasive blood pressure, blood oxygen saturation, and respiration signals. Device processes these inputs to detect arrhythmia, pacer, and QRS complexes (adult/pediatric only); generates visual/audible alarms when parameters exceed preset limits; produces timed or alarm recordings. SPM integrates with R50 Bedside recorder or Saad Patient Monitoring network (SPMS). Clinicians use output to assess patient status and make clinical decisions. Device is not intended for transport, ambulatory use, or MR environments.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Bedside patient monitor; connectivity via R50 Bedside recorder or Saad Patient Monitoring network (SPMS); MR Unsafe; non-transportable; non-ambulatory; performs physiological parameter monitoring and alarm generation.
Indications for Use
Indicated for monitoring heart rate, invasive blood pressure, blood oxygen saturation, and respiration in adult, pediatric, and neonatal patients. Arrhythmia, pacer, and QRS detection indicated for adult and pediatric patients only. Not for use in transport, ambulatory environments, or MR environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Related Devices
K962404 — SIEMENS SC 6000 & SC 6000 P PORTABLE BEDSIDE MONITORING SYSTEM · Siemens Medical Solutions USA, Inc. · Jan 17, 1997
K150691 — Datalys Multi-Parameter Patient Monitor · Lutech Industries, Inc. · May 15, 2015
K970545 — DASH 1000 PATIENT MONITOR · Marquette Medical Systems, Inc. · May 8, 1997
K122253 — PROCARE MONITOR B20 · Ge Medical Systems China Co., Ltd. · Mar 22, 2013
K971910 — MULTI-PARAMETER MODULE (M3000A)/DISPLAY UNIT (M3046A) · Hewlett-Packard GmbH · Mar 23, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or fabric.
Auqust 6, 2014
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002
Hanin Medical Services Co. Ltd Raymond Kelly 57 Lazy Brook Road Monroe, CT 06468 US
Re: K133222
> Trade/Device Name: Saad Patient Monitor (SPM) Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Monitor Regulatory Class: Class II Product Code: MHX Dated: July 23, 2014 Received: July 24, 2014
Dear Mr. Kelly,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/1/Picture/6 description: The image shows a logo with the text "Köl" in a stylized font. Below the logo, the text "Ken Skodacek for" is present. The background of the image is plain white, and the logo and text are in black. The logo appears to be a signature or brand mark.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K133222
Device Name
Saad Patient Monitor (SPM)
#### Indications for Use (Describe)
SPM is indicated for monitoring heart rate, invasive blood pressure, blood oxygen saturation, and respiration on adult, pediatric and neonatal populations.
SPM is indicated for monitoring and detecting arrhythmia, pacer, and QRS on adult and pediatic populations only (not intended for use on neonatal populations).
SPM will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits at which point timed or alarm recordings will be produced. SPM connects to an R50 Bedside recorder, or via the Saad Patient Monitoring network (SPMS).
SPM does NOT perform ST Segment analysis and it is not intended to be used in transport or ambulatory environments.
MR Environment Compatibility Statement:
The Saad Patient Monitors are not compatible for use in an MR Environment (MR Unsafe).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
| Concurrence of Center for Devices and Radiological Health (CDRH) ( <i>Signature</i> ) |
|---------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------|

Ken Skodacek for
Bram Zuckerman
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.