← Product Code [MHX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX) · K100939

# INTELLIVUE PATIENT MONITOR MODEL: MP2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP75, MP80 AND MP90 (K100939)

_Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · MHX · Apr 30, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX/K100939

## Device Facts

- **Applicant:** Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
- **Product Code:** [MHX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX.md)
- **Decision Date:** Apr 30, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Philips MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP75, MP80, and MP90 IntelliVue patient monitors are intended for monitoring and recording of and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP2, X2, MP5, MP5T, MP20, MP30, MP40, and MP50 are additionally intended for use in transport situations within hospital environments. The MP2, X2 and MP5 are also intended for use during patient transport outside of a hospital environment. The monitors are not intended for home use. They are intended for use by health care professionals.

## Device Story

IntelliVue patient monitors (MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP75, MP80, MP90) collect physiological signals (ECG, blood pressure, SpO2, CO2, respiration, temperature, EEG) via sensors/probes. Device processes signals to display real-time waveforms, numerical values, and generate alarms for abnormal physiological states. Used in hospital environments (ICU, OR, general wards) and during patient transport by healthcare professionals. Output provides clinicians with continuous patient status updates to facilitate timely clinical decision-making and intervention. Software revision H.00 introduces updated functionality across the monitor family. Benefits include continuous patient surveillance and early detection of physiological deterioration.

## Clinical Evidence

Bench testing only. Verification and validation activities included system-level testing, regression testing, and hazard analysis. Pass/Fail criteria were based on specifications cleared for predicate devices. Results demonstrated that the modified monitors meet all reliability requirements and performance claims.

## Technological Characteristics

Multi-parameter physiological monitors. Inputs include ECG, blood pressure, SpO2, CO2, respiration, temperature, and EEG. Connectivity includes network and communication systems. Software revision H.00. Class II devices. Sterilization not applicable (external monitoring equipment).

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- [K013199](/device/K013199.md)
- [K031481](/device/K031481.md)
- [K032858](/device/K032858.md)
- [K040304](/device/K040304.md)
- [K050762](/device/K050762.md)
- [K053522](/device/K053522.md)
- [K062392](/device/K062392.md)
- K71426
- [K081793](/device/K081793.md)
- [K082633](/device/K082633.md)
- [K083517](/device/K083517.md)
- [K091927](/device/K091927.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510 (k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

1. The submitter of this premarket notification is:

APR 80 2010

100939 1/3

Andreas Suchi Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1291 Fax: ++49 7031 463-2442 e-mail: andreas.suchi@philips.com

This summary was prepared on April 1, 2010.

- 2. The names of the devices are the Philips MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP75, MP80, and MP90 IntelliVue patient monitors Classification names are as follows:

| Device Panel                                    | Classification | ProCode | Description                                                                  |
|-------------------------------------------------|----------------|---------|------------------------------------------------------------------------------|
| Cardiovascular<br>Devices                       | \$870.1025, II | DSI     | Detector and alarm, arrhythmia                                               |
|                                                 | \$870.1025, II | MLD     | Monitor, ST Segment with Alarm                                               |
|                                                 | \$870.1025, II | MHX     | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms)  |
|                                                 | \$870.1100, II | DSJ     | Alarm, Blood Pressure                                                        |
|                                                 | \$870.1110, II | DSK     | Computer, Blood Pressure                                                     |
|                                                 | \$870.1130, II | DXN     | System, Measurement, Blood-<br>Pressure, Non-Invasive                        |
|                                                 | \$870.1435, II | DXG     | Computer, Diagnostic, Pre-<br>Programmed, Single-Function                    |
|                                                 | \$870.1915, II | KRB     | Probe, Thermodilution                                                        |
|                                                 | \$870.2060, II | DRQ     | Amplifier and Signal<br>Conditioner, Transducer Signal                       |
|                                                 | \$870.2300, II | DRT     | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm)                    |
|                                                 | \$870.2340, II | DPS     | Electrocardiograph                                                           |
|                                                 | \$870.2340, II | MLC     | Monitor, ST Segment                                                          |
|                                                 | \$870.2350, II | DRW     | Electrocardiograph, Lead<br>Switching Adapter                                |
|                                                 | \$870.2370, II | KRC     | Tester, Electrode, Surface,<br>Electrocardiograph                            |
|                                                 | \$870.2600, I  | DRJ     | System, Signal Isolation                                                     |
|                                                 | \$870.2700, II | DQA     | Oximeter                                                                     |
|                                                 | \$870.2770, II | DSB     | Plethysmograph, Impedance                                                    |
|                                                 | \$870.2800, II | DSH     | Recorder, Magnetic tape,<br>Medical                                          |
|                                                 | \$870.2810, I  | DSF     | Recorder, Paper Chart                                                        |
|                                                 | \$870.2850, II | DRS     | Extravascular Blood Pressure<br>Transducer                                   |
|                                                 | \$870.2900, I  | DSA     | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector               |
| Device Panel                                    | Classification | ProCode | Description                                                                  |
|                                                 | -              | MSX     | System, Network and Communication, Physiological<br>Monitors                 |
| Anesthesiology<br>Devices                       | \$870.2910, II | DRG     | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency       |
|                                                 | \$868.1400, II | CCK     | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase                              |
|                                                 | \$868.1500, II | CBQ     | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                 | \$868.1500, II | NHO     | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                 | \$868.1500, II | NHP     | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)   |
|                                                 | \$868.1500, II | NHQ     | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                 | \$868.1620, II | CBS     | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                 | \$868.1700, II | CBR     | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
|                                                 | \$868.1720, II | CCL     | Analyzer, Gas, Oxygen, Gaseous-<br>Phase                                     |
|                                                 | \$868.1880, II | BZC     | Data calculator Pulmonary-<br>function                                       |
|                                                 | \$868.2375, II | BZQ     | Monitor, Breathing Frequency                                                 |
|                                                 | \$868.2480, II | LKD     | Monitor, Carbon Dioxide,<br>Cutaneous                                        |
|                                                 | \$868.2500, II | KLK     | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia           |
| General Hospital<br>and Personal Use<br>Devices | \$880.2910, II | FLL     | Thermometer, Electronic,<br>Clinical                                         |
| Neurological<br>Devices                         | \$882.1400, II | GWR     | Electroencephalograph                                                        |
|                                                 | \$882.1420, I  | GWS     | Analyzer, Spectrum,                                                          |

{1}------------------------------------------------

K100939
R2/3

- 3. The modified devices are substantially equivalent to previously cleared Philips IntelliVue patient monitors marketed pursuant to K013199, K031481, K032858, K040304, K050762, K053522, K062392, K71426, K081793, K082633, K083517, and K091927 .
- 4. The modification is the introduction of software revision H.00 for the entire IntelliVue patient monitors family, models MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP80, and MP90 .

Additionally a new model IntelliVue MP75 is introduced.

{2}------------------------------------------------

- 5. The modified devices have the same intended use as the legally marketed predicate devices. The Philips MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP75, MP80, and MP90 IntelliVue patient monitors are intended for monitoring and recording of and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP2, X2, MP5, MP5T, MP20, MP30, MP40, and MP50 are additionally intended for use in transport situations within hospital environments. The MP2, X2 and MP5 are also intended for use during patient transport outside of a hospital environment. The monitors are not intended for home use. They are intended for use by health care professionals.
- 6. The modified devices have the same technological characteristics as the legally marketed predicate device.
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified devices with respect to the predicate. Testing involved system level and regression tests as well as testing from the hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue patient monitors meet all reliability requirements and performance claims.

(10093

\$3/3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

APR 3 0 2010

Philips Medizin Systeme Boeblingen GmbH c/o Mr. Andreas Suchi Senior Regulatory Affairs Engineer Hewlett-Packard Str. 2 Boeblingen GERMANY 71034

## Re: K100939

Trade/Device Name: Philips MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP75, MP80, and MP90 IntelliVue patient monitors, Release H.00 Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MHX Dated: April 1, 2010 Received: April 5, 2010

Dear Mr. Suchi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Andreas Suchi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Prem D. Zuckerman, M.D.

Bram D. Zuckerman, M.D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name: Philips MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP75, MP80, and MP90 IntelliVue patient monitors, Release H.00.

Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological paramenters of patients.

Prescription Use (Part 21 CFR 801 Subpart D)

Yes AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

_No_

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE

OF NEEDED)

## Concurrence of CDRH, Office of Device Evaluation (ODE)

W. Wood

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K100939

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX/K100939](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX/K100939)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
