← Product Code [MHX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX) · K043439

# INFINITY MODULAR MONITORS WITH VF5 MODIFICATIONS (DELTA/KAPPA/DELTA XL/VISTA XL/GAMMA X XL AND SC 7000/8000/9000XL (K043439)

_Draeger Medical Systems, Inc. · MHX · Feb 18, 2005 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX/K043439

## Device Facts

- **Applicant:** Draeger Medical Systems, Inc.
- **Product Code:** [MHX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX.md)
- **Decision Date:** Feb 18, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Infinity Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to an R50 Bedside recorder, either directly or via the Infinity Network.

## Device Story

Infinity Modular Monitors (Delta/Kappa/Delta XL/Vista XL/Gamma X XL) perform multi-parameter patient monitoring; inputs include ECG, pressure, temperature, SpO2, EEG, and gas modules (Scio). Software version VF5 enables direct connection to Scio gas modules and integration with Infinity BisX and Trident NMT pods. Device processes physiological signals to provide real-time data display, arrhythmia detection, and ST-segment analysis; generates visual/audible alarms when parameters exceed preset limits. Used in clinical settings; operated by healthcare professionals. Output facilitates clinical decision-making through continuous patient status monitoring and alarm notification. Benefits include timely detection of physiological changes and improved patient safety via integrated monitoring network.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Multi-parameter patient monitor; software-controlled; supports modular inputs (gas, NMT, BisX). Connectivity via Infinity Network and MIB Protocol converter. Standards: IEC 60601-1.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- Infinity Modular Monitors ([K033694](/device/K033694.md))
- Infinity MIB ([K033807](/device/K033807.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY

as required per 807.92(c)

Submitters Name, Address:

Draeger Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Contact person for this submission: Penelope H. Greco Regulatory Submissions Manager Date submission was prepared: December 6, 2004

Trade Name, Common Name and Classification Name:

INFINITY Modular Monitors with VF5 Modifications Trade Name: (Delta/Kappa/Delta XL/Vista XL/Gamma X XL and SC 7000/8000/9000XL)

Common Name, Classification Name, Class and Regulation Number:

| Common Name                                                              | Product Code | Class | Regulation Number |
|--------------------------------------------------------------------------|--------------|-------|-------------------|
| Monitor, Physiological, Patient (with Arrhythmia<br>Detection or Alarms) | MHX          | II    | 8*0.1025          |
| Arrhythmia Detector & Alarm                                              | 74DSI        | II    | 8*0.1025          |
| System, Network and Communication,<br>Physiological Monitors             | MSX          | II    | 8*0.2300          |

Legally Marketed Device

| Infinity Modular Monitors | K033694 |
|---------------------------|---------|
| Infinity MIB              | K033807 |

Description of Device Modifications:

Software version VF5 is the latest release of the Infinity Modular Monitors, including the Infinity Vista XL and Gamma X XL (variants of the SC 7000 and Delta monitors) that supports direct connection to the Scio gas module, display of data received from the Infinity BisX and Infinity Trident NMT pods, as well as other minor software modifications. In addition, new devices have been tested for use with the Infinity MIB Protocol converter. Testing in accordance with internal design control procedures has verified that the Infinity Modular Monitors and the Infinity MIB Protocol Converter with VF5 modifications are as safe and effective as submitted in 510(k) K033807 and K033694.

Intended Use:

The Infinity Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to an R50 Bedside recorder, either directly or via the Infinity Network.

Page 1 of 2

## COMPANY CONFIDENTIAL

## Draeger Medical Systems, Inc.

16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879

{1}------------------------------------------------

Assessment of non-clinical performance data for equivalence: Verification and validation testing performed indicates that the modifications implemented with software version VF5 are as safe and effective as previous versions and have not altered the fundamental technology of the device(s).

Assessment of clinical performance data for equivalence: Not applicable

Biocompatability: Not applicable

Sterilization: Not applicable

and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the commen

Standards: IEC 60601-1

Page 2 of 2

## COMPANY CONFIDENTIAL

## Draeger Medical Systems, Inc.

16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or a stylized human figure. The logo is black and white.

FEB 1 8 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Draeger Medical Systems, Inc. c/o Ms. Penelope H. Greco Regulatory Submissions Manager 16 Electronics Ave. Danvers, MA 01923

Re: K043439

KU43439
Trade Name: Infinity Delta / Kappa / Delta XL / Gamma X XL / Vista XL with VF5 Software Regulation Number: 21 CFR 870.1025 Regulation Name: 21 CFR 876.1025
Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: December 10, 2004 Received: December 14, 2004

Dear Ms. Greco:

We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your bection 910(s) pe device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule for legally while in the Medical Device Amendments, or to commerce prior to May 20, 1776, the encordance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). and Costlience Act (Act) that to not require the general controls provisions of the Act. The r ou may, mercere, manel the device, equirements for annual registration, listing of general controls provisions of tactice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is crassified (see aboro) the existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations and may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Coursements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -- Ms. Penelope H. Greco

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase oc advised that I Dris brounders a your device complies with other requirements of the Act that I DA has made a and regulations administered by other Federal agencies. You must of any it call the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It Far 6077, laceming (21 CFR Part 820); and if applicable, the electronic forul in the quarty bystems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (1) (1) = 1) = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = I his letter will anow you to ough finding of substantial equivalence of your device to a legally premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your as (240) 276-0120. Also, please note the regulation entitled, Contact the Ories of Company of Set notification" (21CFR Part 807.97) you may obtain. IMISolanding of reference to premation on your responsibilities under the Act may be obtained from the Oiler general information on Jour respectional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bhimmafor

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KD4343939 510(k) Number (if known):__

Device Name: Infinity Delta / Kappa / Delta XL / Gamma X XL / Vista XL

Indications for Use:

The INFINITY Modular monitors are capable of monitoring:

- . Heart rate
- Respiration rate .
- . Invasive pressure
- . Non-invasive pressure
- . Arrhythmia
- . Temperature
- Cardiac output .
- Arterial oxygen saturation .
- Pulse rate .
- Apnea .
- . ST Segment Analysis
- 12-Lead ST Segment Analysis .
- tcp02/tcpC02 .
- EEG signals .
- FiO2 .

Prescription Use_ く (Per 21 CFR 801.109)

.

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(PI.EASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bhimmer

______________________________________________________________________________________________________________________________________________________________________________

(Division/Sign-Off) Division of Cardiovascular Dev 510(k) Number

Page 1 of 2

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX/K043439](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MHX/K043439)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
