USCI IDEAL PTCA KIT
Device Facts
| Record ID | K953522 |
|---|---|
| Device Name | USCI IDEAL PTCA KIT |
| Applicant | Ryder Intl. Corp. |
| Product Code | MAV · Cardiovascular |
| Decision Date | Apr 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1650 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Ryder device is intended for single use inflating of a balloon catheter, monitoring the pressure within the catheter and deflating the balloon catheter during angioplasty procedures.
Device Story
Ryder Balloon Catheter Inflation Syringe; manual screw-type plunger mechanism with locking lever and rotating palm grip; 20 cc fluid capacity; operating pressure range vacuum to 30 bars; includes manometer for pressure monitoring and high-pressure connecting tube; used by clinicians during angioplasty; provides controlled inflation/deflation of balloon catheters; enables precise pressure management; single-use device.
Clinical Evidence
Bench testing only; USP material biocompatibility testing performed on patient-contacting components.
Technological Characteristics
Plastic syringe construction; screw-type plunger; locking lever; rotating palm grip; integrated manometer; high-pressure connecting tube; 20 cc capacity; 0-30 bar pressure range; single-use.
Indications for Use
Indicated for patients undergoing angioplasty procedures requiring balloon catheter inflation, pressure monitoring, and deflation.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
- Merit Medical Systems Monarch™ Balloon Catheter Inflation Syringe
Related Devices
- K962611 — BARD EAGLE INFLATION DEVICE · Ryder Intl. Corp. · Oct 21, 1996
- K130566 — BASIXTOUCH · Merit Medical Systems, Inc. · Jun 19, 2013
- K122321 — BASIXCONPAK ANALOG INFLATION SYRINGE · Merit Medical Ireland, Ltd. · Dec 19, 2012
- K040138 — VICEROY INFLATION DEVICE · Merit Medical Systems, Inc. · Feb 13, 2004
- K143522 — Presto Inflation Device · C.R. Bard, Inc. · Jan 12, 2015