SUMMED INFLATION DEVICE KIT

K133795 · Sunny Medical Device (Shenzhen) Co., Ltd. · MAV · Dec 12, 2014 · Cardiovascular

Device Facts

Record IDK133795
Device NameSUMMED INFLATION DEVICE KIT
ApplicantSunny Medical Device (Shenzhen) Co., Ltd.
Product CodeMAV · Cardiovascular
Decision DateDec 12, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1650
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sunmed™ Inflation Device is a set of accessorial device. It is composed of an inflation device, and may be composed of a stopcock and a hemostatic valve. The inflation device is intended for use in PTCA or PTA procedures to create and monitor pressure (by connecting the rotating male luer of inflation device with the balloon catheter's female luer) in the balloon and to deflate the balloon. The hemostatic valve kit is recommended for use during PTCA or PTA procedures in conjunction and/or diagnostic device (e.g. balloon dilatation catheters). Stopcock is used to release and purge any trapped air (if used).

Device Story

Sunmed Inflation Device is a manual, disposable 30ml syringe-based system for PTCA/PTA procedures. Operator (clinician) manipulates a large handle to drive a piston, creating up to 40ATM of pressure within a balloon catheter. Pressure is monitored via an integrated gauge. System may include a stopcock for air purging and a hemostatic valve (Y-connector) to provide an access port for interventional devices, including an insertion tool for guidewires and a torquer for guidewire manipulation. Device is used in clinical settings to facilitate balloon dilatation; output is mechanical pressure control and access port management. Benefits include precise balloon inflation/deflation and controlled interventional access.

Clinical Evidence

No clinical data. Evidence consists of non-clinical bench testing, including biocompatibility (pyrogen, endotoxin, acute systemic toxicity, intracutaneous reactivity, cytotoxicity, skin sensitization, hemolysis, complement activity, thrombogenicity), package integrity, asepsis, aging, and EtO/ECH residue testing.

Technological Characteristics

Disposable 30ml manual inflation device; max pressure 40ATM; piston-driven mechanism with pressure gauge. Includes optional stopcock and hemostatic valve (Y-connector) with insertion tool and torquer. Materials, packaging, and sterilization methods are equivalent to predicate devices. No software or electronic components.

Indications for Use

Indicated for patients undergoing PTCA or PTA procedures requiring balloon inflation, pressure monitoring, and deflation. Used in conjunction with balloon dilatation catheters.

Regulatory Classification

Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure in profile, with three faces overlapping to represent the department's focus on people and health. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 12, 2014 Sunny Medical Device (Shenzhen) Co., Ltd. c/o James Qi Zhang 56 Lehigh Aisle Irvine, CA 92612 Re: K133795 Trade/Device Name: Sunmed™ Inflation Device Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic injector and syringe Regulatory Class: Class II Product Code: MAV Dated: September 17, 2014 Received: November 10, 2014 Dear Mr. Zhang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {1}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Melissa A. Torres -S For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## 2. Indications for Use Statement ## Indications for Use 510(k) Number (if known): K133795 Device Name: Sunmed™ Inflation Device Indications for Use: The Sunmed The Inflation Device is a set of accessorial device. It is composed of an inflation device, and may be composed of a stopcock and a hemostatic valve. The inflation device is intended for use in PTCA or PTA procedures to create and monitor pressure (by connecting the rotating male luer of inflation device with the balloon catheter's female luer) in the balloon and to deflate the balloon. The hemostatic valve is recommended for use during PTCA or PTA procedures in conjunction and/or diagnostic device (e.g. balloon dilatation catheters).Stopcock is used to release and purge any trapped air (if used). Prescription Use_ (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (Part 21 CFR 801Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1of 1 Sunny Medical Device (Shenzhen) Co., Ltd. {3}------------------------------------------------ ## 3. 510(k) Summary - Sunny Medical Device (Shenzhen) Co., Ltd. 1. Submitted by: 4/F and 5/F, Dongpingxing Creative Science Park, Registered Address: Longgang District, Shenzhen, Guangdong, P.R. China 518172 Contact Address: 56 Lehigh Aisle Irvine, CA 92612 Telephone: (949)216-8838 Fax: (949)423-0168 Contact: JamesQi Zhang, General Manager jamesqizhang@gmail.com E-mail: Date: Sep17, 2014 - 2. Proposed Device: | Trade/Proprietary Name: | Sunmed™ Inflation Device | |-------------------------|-----------------------------------| | Common/Usual Name: | Disposable Inflation Device | | Classification: | II | | Classification Name: | Angiographic injector and syringe | | Regulation Number: | 870.1650 | | Product Code: | MAV | - 3. Predicate Device: | 510(k)<br>Number | Trade or Proprietary or Model<br>Name | Manufacture | |------------------|------------------------------------------------------------|----------------------------------------------| | K102648 | ANT Inflation Device/ ANT<br>Inflation Device Compact Pack | Shenzhen ANT Hi-Tech<br>Industrial Co., Ltd. | | K032840 | Atrion Medical QL® Inflation<br>Device | Atrion Medical Products, Inc. | ## 4. Device description Sunny Medical Device (Shenzhen) Co., Ltd. {4}------------------------------------------------ The Sunmed The Inflation Device is a set of accessorial device. It is composed of an inflation device, and may be composed of a stopcock and a hemostatic valve. The inflation device is 30ml disposable device capable of producing a maximum pressure of 40ATM. Pressure can be monitored via a pressure gauge. The manually operation of the device is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Hemostatic valve is a single Y connector, and may be composed of an insertion tool, and a torquer. The hemostasis valve is designed to provide a port for interventional system. The insertion tool is used to facilitate placement of a guide wire tip through the hemostasis valve. The torque device is designed to hold the guide wire and provide a handle for manipulating. - 5. Intended Use The Sunmed™ Inflation Device is a set of accessorial device. It is composed of an inflation device, and may be composed of a stopcock and a hemostatic valve. The inflation device is intended for use in PTCA or PTA procedures to create and monitor pressure (by connecting the rotating male luer of inflation device with the balloon catheter's female luer) in the balloon and to deflate the balloon. The hemostatic valve kit is recommended for use during PTCA or PTA procedures in conjunction and/or diagnostic device (e.g. balloon dilatation catheters). Stopcock is used to release and purge any trapped air (if used). - 6. Technological Comparison to Predicate Device The proposed device will use technology that is similar to the technology already in use by Shenzhen ANT Hi-Tech Industrial Co., Ltd. and Atrion Medical Products, Inc. The maximum pressure of the proposed device is the same as The Atrion Medical OL® Inflation Device. It has the same intended use, fundamental concepts and principles of operation as the ANT Inflation Device .The proposed device uses the same materials, packaging and sterilization method like each of the predicate devices. {5}------------------------------------------------ A detailed comparison of technological characteristics with the predicate devices is included insubstantial equivalence of the submission. - 7. Summary of Non-Clinical Testing The following tests were performed on the Sunmed™ Inflation Device: - Biocompatibility Testing: Pyrogen Test Endotoxin Test Acute Systemic Toxicity Test(two kinds of solvent) Intracutaneous Reactivity Test In Vitro Cytotoxicity Test Skin sensitization Test (two kinds of solvent) In Vitro Hemolysis Study Complement Activity Test (C3a, SC5b-9) In Vivo Thrombogenicity Package Penetrate Testing Asepsis Testing Aging Testing EtO and ECH Residue Testing Bench Testing - 8. Clinical Evaluation was not applicable. - 9. Conclusions Based on the information presented in this 510(k) premarket notification, the Sunmed™ Inflation Device is considered substantially equivalent to the ANT Inflation Device and Atrion Medical QL® Inflation Device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...