← Product Code [MAV](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MAV) · K122152

# OXTI INFLATION DEVICE (K122152)

_Oxti Corporation · MAV · Dec 14, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MAV/K122152

## Device Facts

- **Applicant:** Oxti Corporation
- **Product Code:** [MAV](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MAV.md)
- **Decision Date:** Dec 14, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1650
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The OXTI Inflation Device is indicated for use during cardiovascular procedures to create, maintain, and monitor pressure in a balloon dilation catheter.

## Device Story

Manual balloon inflation syringe used during cardiovascular procedures; consists of 30ml polycarbonate barrel, inflation mechanism, and reinforced polyurethane tubing with 6% conical fitting. Operator (clinician) manually pushes/pulls plunger to inflate/deflate balloon catheter; locking latch secures plunger to maintain constant pressure. Integrated dual-scale pressure gauge (up to 30atm) allows monitoring of inflation pressure. Device connects to PTCA catheters via 3-way stopcock. Benefits include controlled balloon dilation and pressure maintenance during interventional procedures.

## Clinical Evidence

Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility) and functional performance (ISO 594.1 & 2 conical fitting efficacy, water leakage resistance, tensile pull resistance, plunger force, and pressure gauge accuracy up to 30atm).

## Technological Characteristics

30ml polycarbonate barrel; high-impact polypropylene pressurization system/plunger; reinforced polyurethane tubing; 6% conical locking connector. Manual operation. Pressure gauge accuracy tested at 2atm intervals up to 30atm. Conforms to ISO 594.1 & 2 for fittings.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Dolphin Inflation Device ([K042449](/device/K042449.md))
- Indeflator Plus 30 Priority Pack (Kae2495)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K122152 e 16f2

## DEC 1 4 2012

## Pre-market Notification for OXTI Inflation Devices

## 510(k) Summary of Safety and Effectiveness

1. Submitter

OXTI Corporation.

8th Floor, Unit 146, Jian 1st Road, Jhonghe District '

New Taipei City, TAIWAN 23585

## Contact:

Mr. John Sung

Phone: +886 2 2225 4998 Ext. 20

## 2. Name and Classification of Device

| Trade Name:                   | OXTI Inflation Device             |
|-------------------------------|-----------------------------------|
| Common/Usual Name:            | Balloon inflation syringe         |
| Classification Name:          | Angiographic injector and syringe |
| Regulation Medical Specialty: | Cardiovascular                    |
| Review Panel:                 | Cardiovascular                    |
| Classification Number:        | 21CFR 870.1650                    |
| Product Code:                 | MAV                               |

## 3. Predicate Device

| Trade Name                       | 510(k) Number | Decision Date |
|----------------------------------|---------------|---------------|
| Dolphin Inflation Device         | K042449       | 5/20/2005     |
| Indeflator Plus 30 Priority Pack | Kae2495       | 9/12/1996     |

{1}------------------------------------------------

#### 4. Device Description

The device consists of three major components; a) 30ml barrel, b) inflation mechanism and c) 6% standard conical fitting and stopcocks system for connecting to the PTCA catheter. The barrel has graduation markings at 0.5ml interval to the full 30ml capacity A dual scale pressure gauge mounted at distal end of the barrel is capable of indicating pressure up to 30atm at 1atm or 10psi intervals. Reinforced polyurethane tubing extending from the svringe exit port connects to a 6% male/female conical fitting for attaching to the 3-way stopcock and to the angiographic catheter.

#### 5. Indications for Use

The OXTI Inflation Device is indicated for use during cardiovascular procedures to create, maintain, and monitor pressure in a balloon dilation catheter.

#### 6. Technological Characteristics

The 30 ml barrel is made of polycarbonate for clarity and added strength against high system pressure. The pressurization system and the plunger are made of high impact resistant polypropylene. Controlled inflation and deflation is achieved by manually pushing or pulling the plunger. A unique locking latch secures the plunger in place to maintain a constant pressure even if the user lets go of the device. High-pressure polyurethane tubing provides links from the inflation barrel to the angioplastic catheter using a 6% conical locking connector.

#### 7. Performance Summary

- The whole device has been tested for material biocompatibility as listed below:
	- Cytotoxicity:
	- Sensitization;
	- . Intracutaneous reactivity;
	- Acute systematic toxicity; and
	- 1 Hemocompatibility.
- Functional aspects and performance characteristics of the inflation device are verified through a series of bench tests which include:
	- 6% conical fitting efficacy, per ISO 594.1 & 2, 0
	- Whole device resistance to water leakage; D
	- Tubing joints resistance to tensile pull; 1
	- Barrel and plunger resistance to water leakage; and
	- Forces to initiate and continue pushing of plunger; 1
- Pressure gauge accuracy was tested against a standard, calibrated gauge at 2atm ● pressure intervals for up to 30atms.

Results of various tests support the claim that the OXTI Inflation Device meets requirements specified by various international voluntary standards and is substantially equivalent to legally marketed predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract eagle or bird in flight, composed of three curved, parallel lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

DEC 1 4 2012

Oxti Corporation c/o Joseph Chang Consultant 7128 Staffordshire Street Houston, Texas 77030

Re: K122152

Trade/Device Name: OXTI Inflation Device Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic injector and syringe Regulatory Class: Class II Product Code: MAV Dated: October 31, 2012 Received: November 5, 2012

Dear Mr. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Chang

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

# Matthew G. Hillebrenner

for Bram D. Zuckerman, M.D.

Director

Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indication for Use

510(k) Number (if known):

Device Name: OXTI Inflation Device

#### Indications for Use:

The OXTI Inflation Device is indicated for use during cardiovascular procedures to create, maintain, and monitor pressure in the balloon dilation catheter.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.A. Hille

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number K/27/52 42

Page 1 of

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MAV/K122152](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MAV/K122152)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
