RADIUS PTCA GUIDEWIRE

K970466 · Radius Medical Technologies, Inc. · LIT · May 20, 1997 · Cardiovascular

Device Facts

Record IDK970466
Device NameRADIUS PTCA GUIDEWIRE
ApplicantRadius Medical Technologies, Inc.
Product CodeLIT · Cardiovascular
Decision DateMay 20, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Radius PTCA Guidewire is indicated to facilitate the placement of balloon dilatation catheters during PTCA and/or PTA.

Device Story

The Radius PTCA Guidewire is a cardiovascular guide wire used by clinicians during percutaneous transluminal coronary angioplasty (PTCA) or percutaneous transluminal angioplasty (PTA). The device facilitates the navigation and placement of balloon dilatation catheters within the vasculature. It consists of a solid stainless steel core with a tapered distal section and an attached coil. The proximal section is PTFE-coated, while the distal coil is silicone-coated. The device is provided sterile via ETO gas. It is intended for prescription use in a clinical setting.

Clinical Evidence

Bench testing only. In vitro performance testing included tensile strength, torque strength, torqueability, tip flexibility, and coating adherence/integrity. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous toxicity, acute systemic toxicity, material-mediated pyrogenicity, and ASTM hemolysis. All results were acceptable and comparable to the predicate device.

Technological Characteristics

Solid stainless steel core; PTFE-coated proximal section; silicone-coated distal coil. Outer diameter: .014". Lengths: 180cm or 300cm. Sterilization: ETO gas. Packaging: Tyvek/mylar pouch.

Indications for Use

Indicated for patients undergoing PTCA or PTA procedures to facilitate the placement of balloon dilatation catheters. Compatible with approved PTCA balloon catheters labeled for use with .014 guidewires.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K970466 MAY 20 1997 # 510(K) SUMMARY Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Radius Medical Technologies, Inc. is required to submit with this Premarket Notification either an "...adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Radius Medical Technologies, Inc. chooses to submit a summary of information respecting safety and effectiveness. ## A. GENERAL INFORMATION Submitter's Name: Radius Medical Technologies, Inc. Address: 63 Great Road Maynard, MA 01754 Contact Person: Maureen A. Finlayson Date of Preparation: February 5, 1997 Device Generic Name: PTCA Guidewire Device Trade Name: Radius PTCA Guidewire Classification Name: Wire, Guide, Cardiovascular (74DQX) ## B. INDICATIONS The Radius PTCA Guidewire is intended to facilitate the placement of balloon dilatation catheters during PTCA and/or PTA. Radius PTCA Guidewires are compatible with all currently approved and marketed PTCA balloon catheters which are labeled for use with an .014 guidewire. ## C. DESCRIPTIVE CHARACTERISTICS The Radius PTCA Guidewire is constructed from a solid stainless steel core to which a coil is attached to the tapered distal section. Table 1 lists the key dimensional characteristics for the proposed device. The proximal section of the wire is coated with PTFE, and the distal coil portion of the wire is silicone coated. The device is packaged in a protective hoop sealed into a Tyvek/mylar pouch, and is sterilized using ETO gas. Table 1 | | Specification | | --- | --- | | G/W Outer Diameter | .014" | | G/W Overall Length | 180cm or 300cm | | Overall Coil Length | 15cm to 30cm | | Radiopaque Length | 3cm to 30cm | | Floppy Tip Length | 2.5cm to 3cm | Radius Medical Technologies, Inc. 510(k) premarket notification PTCA Guidewire Feb. 5, 1997 Pg. 7 {1} D. COMPARATIVE INFORMATION The Radius PTCA Guidewires are substantially equivalent to the currently marketed ACS High Torque Guidewire family. ACS High Torque Guidewires, marketed by: Advanced Cardiovascular Systems, Inc. 26531 Ynez Road Temecula, California 92390-1856 E. PERFORMANCE TESTING The following in vitro performance tests were performed on the Radius PTCA Guidewire: 1. Tensile Strength 2. Torque Strength 3. Torqueability 4. Tip Flexibility 5. Coating Adherence/Integrity In all cases, the performance of the Radius PTCA Guidewires was comparable to that of the predicate ACS guidewires. F. BIOCOMPATIBILITY Biocompatibility testing was conducted on the proposed line of Radius PTCA Guidewires according to General Program Memorandum - #G95-1 issued May 1, 1995 from the ODE. Following is a summary of the tests performed: 1. Cytotoxicity 2. Sensitization 3. Intracutaneous Toxicity 4. Acute systemic toxicity 5. Material Mediated Pyrogenicity 6. ASTM Hemolysis All test results were acceptable. CONCLUSION: Based on the indications for use, technological characteristics, and safety and performance testing, the proposed Radius PTCA Guidewire meets the minimum requirements that are considered adequate for its intended use. Radius Medical Technologies, Inc. 510(k) premarket notification PTCA Guidewire Feb. 5, 1997 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 20 1997 Ms. Maureen A. Finlayson President Radius Medical Technologies, Inc. 63 Great Road Maynard, Massachusetts 01754 Re: K970466 Radius PTCA Guidewire Regulatory Class: II (two) Product Code: LIT Dated: April 23, 1997 Received: April 24, 1997 Dear Ms. Finlayson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in {3} regulatory action. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597. Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} Page 1 of 1 510(k) Number (if known): K970466 Device Name: Radius PTCA Guidewire Indications for Use: The Radius PTCA Guidewire is indicated to facilitate the placement of balloon dilatation catheters during PTCA and/or PTA. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K970466 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-the-counter Use ☐
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...