The Maxi PTA catheter is intended to dilate lesions in large vessels.
Device Story
Over-the-wire balloon catheter; distal balloon; proximal hub; two radiopaque marker bands. Used by physicians in clinical settings for percutaneous transluminal angioplasty (PTA) to dilate stenotic lesions in large vessels. Device facilitates vessel patency by mechanical inflation of balloon at lesion site.
Clinical Evidence
Bench testing only. In vitro tests performed to ensure safety and effectiveness; all results met or exceeded specifications.
K092286 — FOX SV PTA CATHETER · Abbott Vascular · Aug 27, 2009
K093965 — ULTRAVERSE 014 AND 018 BALLOON DILATATION CATHETERS · C.R. Bard, Inc. · Mar 17, 2010
K955886 — MEGA PTA BALLOON CATHETER · Cordis Corp. · Mar 11, 1996
K123358 — PACIFIC PLUS · Medtronic Vascular · Jan 29, 2013
Submission Summary (Full Text)
{0}
07-00032
K963000
# Section 513(j) of the Federal Food, Drug and Cosmetic Act
## Summary of Safety and Effectiveness
July 26, 1996
FEB 20 1997
## I. General Provisions
Common or Usual name: PTA Balloon Catheter
Proprietary name: Cordis Maxi PTA Balloon Catheter
Name and Address of Applicant: Cordis Corporation
Miami Lakes Operation Center
14201 NW 60 Avenue
Miami Lakes, FL 33014
## II. Name of Predicate Devices
Cordis Opta 5 PTA Balloon Catheter
Cordis Powerflex 5F PTA Balloon Catheter
Cook Omega N.V. Balloon Dilatation Catheter
## III. Classification
PTA catheters are class II devices according to 21 CFR 870.1340.
## IV. Performance Standards
Performance standards have not been established by the FDA under Section 514 of the Food, Drug and Cosmetic Act.
## V. Intended Use and Device Description
The Maxi PTA catheter is intended to dilate lesions in large vessels. The device is an over-the-wire balloon catheter, with a distal balloon and a proximal hub. The balloon is indicated by two radiopaque marker bands.
## VI. Biocompatibility
All appropriate biocompatibility testing was performed, and successfully passed, on the materials used for the Maxi PTA Catheter.
## VII. In vitro Testing
{1}
00-00033
A series of in vitro tests were performed to assure that the introduction of the Maxi PTA catheters does not raise new issues of safety and effectiveness. All test results met or exceeded established specifications.
## VIII. Summary of Substantial Equivalence
The Maxi PTA catheter is designed for dilatation of stenotic lesions in large vessels. The Maxi PTA catheters have the same intended use, and several of the same materials, design characteristics and dimensions as the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.