Paladin Carotid Post-Dilation Balloon System with Integrated Embolic Protection
K221339 · Contego Medical, Inc. · LIT · Jun 9, 2022 · Cardiovascular
Device Facts
Record ID
K221339
Device Name
Paladin Carotid Post-Dilation Balloon System with Integrated Embolic Protection
Applicant
Contego Medical, Inc.
Product Code
LIT · Cardiovascular
Decision Date
Jun 9, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Paladin Carotid PTA Balloon System with Integrated Embolic Protection (IEP), is indicated for Percutaneous Transluminal Angioplasty (PTA) in the carotid arteries with capture and removal of embolic material. This device is also indicated for post-dilation of self-expanding stents in the carotid arteries with capture and removal of embolic material. The diameter of the arterial site for filter deployment should be no more than 7.0 mm. The Paladin Carotid PTA Balloon System with Integrated Embolic Protection (IEP) should always be used in conjunction with an available embolic protection device when used for post-dilation of self-expanding stents.
Device Story
Sterile, single-use, rapid exchange (RX) PTA catheter; features nitinol-based filter between PTA balloon and distal tip. Operates via hydraulic pressurization for balloon inflation; filter deployed/collapsed via handle-mounted activation knob. Used in carotid arteries under fluoroscopic guidance; physician-operated. Captures embolic material during angioplasty or stent post-dilation. Filter consists of nitinol braid chassis with polyurethane membrane (40-micron pores). Radiopaque markers on balloon and filter aid positioning. Device introduced via 6F vascular sheath; compatible with 0.014" guidewires. Benefits patient by reducing risk of distal embolization during carotid interventions.
Clinical Evidence
No clinical data provided. Substantial equivalence established via risk analysis and comparison of technological characteristics to the predicate device.
Technological Characteristics
Nitinol-based filter frame; polyurethane membrane (40-micron pores); nylon semi-compliant PTA balloon. Rapid exchange (RX) catheter platform. Radiopaque markers on balloon and filter. 6F sheath compatibility; 0.014" guidewire compatibility. EO gas sterilization. 12-month shelf life. Mechanical activation via handle knob. No software or electronic components.
Indications for Use
Indicated for patients undergoing Percutaneous Transluminal Angioplasty (PTA) or post-dilation of self-expanding stents in carotid arteries requiring embolic material capture. Contraindicated for coronary artery use, patients unable to tolerate anticoagulant/anti-platelet therapy, patients with nitinol hypersensitivity, or patients with unresolved bleeding disorders.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Carotid Post-Dilation Balloon System with Integrated Embolic Protection (K181128)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 9, 2022
Contego Medical, LLC. James Clossick Vice President of Regulatory Affairs 3801 Lake Boone Trail, Suite 100 Raleigh, North Carolina 27607
Re: K221339
Trade/Device Name: Paladin Carotid Post-Dilation Balloon System with Integrated Embolic Protection Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: NTE, LIT Dated: May 10, 2022 Received: May 11, 2022
Dear James Clossick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221339
#### Device Name
Paladin Carotid Post-Dilation Balloon System with Integrated Embolic Protection
Indications for Use (Describe)
The Paladin Carotid PTA Balloon System with Integrated Embolic Protection (IEP), is indicated for Percutaneous Transluminal Angioplasty (PTA) in the carotid arteries with capture and removal of embolic material. This device is also indicated for post-dilation of self-expanding stents in the carotid arteries with capture and removal of embolic material. The diameter of the arterial site for filter deployment should be no more than 7.0 mm. The Palaon System with Integrated Embolic Protection (IEP) should always be used in conjunction with an available embolic protection device when used for post-dilation of self-expanding stents.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) - K221339
| Date Prepared | May 26, 2022 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Contego Medical, Inc.<br>3801 Lake Boone Trail, Suite 100<br>Raleigh, NC 27607 |
| FDA RegistrationNumber | 3011471056 |
| Contact Person: | Mr. Jim Clossick<br>Contego Medical, Inc.<br>3801 Lake Boone Trail, Suite 100<br>Raleigh, NC 27607<br>Phone: + 1 305 607 1708<br>Email: Jclossick@contegomedical.com |
| Proprietary Name: | Paladin® System |
| Common Name: | Paladin Carotid PTA Balloon System with Integrated<br>Embolic Protection |
| Device Classification: | Class II per 21 CFR §870.1250 |
| Classification Name: | Catheter, Angioplasty, Peripheral, Transluminal<br>Temporary Carotid Catheter for Embolic Capture |
| Product Code: | LIT, NTE |
| Predicate Device: | Carotid Post-Dilation Balloon System with<br>Integrated Embolic Protection (K181128) |
{5}------------------------------------------------
## Device Description
The Paladin System is a sterile, single use, rapid exchange (RX) PTA catheter with a nitinol-based filter between the PTA balloon and the distal tip of the catheter. The balloon has radiopaque markers to aid in accurate positioning under fluoroscopic guidance. The dilation balloon is designed to inflate to a known diameter and length at a specific inflation pressure. Paladin has balloon diameters Ø of 5.0 mm and balloon lengths of 20 mm and 30 mm. The embolic protection filter is composed of a nitinol braid chassis with an overlying polyurethane membrane with radiopaque markers on either end and radiopaque markers on the filter. The catheter has a guidewire lumen, an inflation/deflation lumen, and a filter activation wire lumen. The quidewire lumen permits the passage of a guidewire to facilitate advancement of the catheter to and through the stenosis to be dilated. The proximal end of the catheter has a handle that controls activation. The filter is identical across all sizes of balloons and can be expanded to a maximum diameter of 7 mm. The Paladin System is designed to be introduced through a 6 French vascular sheath.
## Indication for Use
"The Paladin Carotid PTA Balloon System with Integrated Embolic Protection (IEP), is indicated for Percutaneous Transluminal Angioplasty (PTA) in the carotid arteries with capture and removal of embolic material. This device is also indicated for post-dilation of self-expanding stents in the carotid arteries with capture and removal of embolic material. The diameter of the arterial site for filter deployment should be no more than 7.0 mm. The Paladin Carotid PTA Balloon System with Integrated Embolic Protection (IEP) should always be used in conjunction with an available embolic protection device when used for post-dilation of self-expanding stents."
| Parameters | Contego Medical Paladin<br>System<br>(modified device) | Contego Medical Paladin<br>System<br>(predicate device K181128) |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | The Paladin Carotid PTA | The Paladin System with |
| | Balloon System with | Integrated Embolic Protection |
| | Integrated Embolic Protection | (IEP) is indicated for |
| | (IEP), is indicated for | Percutaneous Transluminal |
| | Percutaneous Transluminal | Angioplasty (PTA) in the carotid |
| | Angioplasty (PTA) in the | arteries with capture and removal |
| | carotid arteries with capture | of embolic material. This device |
| | and removal of embolic | is also indicated for post-dilation |
| | material. This device is also | of self-expanding stents in the |
| | indicated for post-dilation of | carotid arteries with capture and |
| | self-expanding stents in the | removal of embolic material. The |
| | carotid arteries with capture | diameter of the arterial site for |
| Parameters | Contego Medical Paladin<br>System<br>(modified device) | Contego Medical Paladin<br>System<br>(predicate device K181128) |
| | and removal of embolic<br>material. The diameter of the<br>arterial site for filter<br>deployment should be no<br>more than 7.0 mm. The<br>Paladin Carotid PTA Balloon<br>System with Integrated<br>Embolic Protection (IEP)<br>should always be used in<br>conjunction with an available<br>embolic protection device<br>when used for post-dilation of<br>self-expanding stents. | filter deployment should be no<br>more than 7.0 mm. The Paladin<br>System with IEP should always<br>be used in conjunction with an<br>available embolic protection<br>device. |
| Common name | Carotid PTA Balloon System<br>with Integrated Embolic<br>Protection | Carotid Post-Dilation Balloon<br>System with Integrated Embolic<br>Protection |
| Classification | SAME as 510(k) cleared<br>Paladin System | Class II (21 CFR 870.1250) |
| Classification<br>name | SAME as 510(k) cleared<br>Paladin System | Catheter, Angioplasty,<br>Peripheral, Transluminal<br>Temporary Carotid Catheter for<br>Embolic Capture |
| Product code | SAME as 510(k) cleared<br>Paladin System | LIT, NTE |
| Contraindications | SAME as 510(k) cleared<br>Paladin System | Contraindicated for use in<br>coronary arteries.<br>Contraindicated for use in<br>patients who cannot tolerate<br>anticoagulant or anti-platelet<br>therapy<br>Contraindicated for use in<br>patients with known<br>hypersensitivity to nitinol.<br>Contraindicated for use in<br>patients with unresolved bleeding<br>disorders. |
| Principle of<br>operations | SAME as 510(k) cleared<br>Paladin System | Device operates on the principle<br>of hydraulic pressurization<br>applied through an inflatable<br>balloon attached to the distal<br>end.<br>0.014" guidewire compatible<br>catheter based platform with |
| Parameters | Contego Medical Paladin<br>System<br>(modified device) | Contego Medical Paladin<br>System<br>(predicate device K181128) |
| | | adjustable nitinol filter frame and<br>urethane filter used to capture<br>and remove embolic material<br>during the carotid angioplasty /<br>post-dilation procedure. Platform<br>is advanced to the carotid lesion,<br>filter is expanded via turning the<br>filter deployment knob on the<br>handle. PTA, or post-dilation of<br>stent, in the carotid is performed,<br>where after the filter is collapsed<br>by turning the knob and the<br>device is removed. |
| Sterilization | SAME as 510(k) cleared<br>Paladin System | EO gas |
| Shelf life | SAME as 510(k) cleared<br>Paladin System | 12 months |
| Packaging<br>system | SAME as 510(k) cleared<br>Paladin System | Balloon and filter have a<br>protective sheath. Catheter is<br>protected by a HDPE spiral<br>dispenser, placed in a PET/G<br>snap-top tray, sealed in a sterile<br>pouch. Product is packed in an<br>outer cardboard carton. |
| Radiopaque<br>markers | SAME as 510(k) cleared<br>Paladin System | Two balloon markers - distal and<br>proximal to balloon.<br>Two marker bands distal and<br>proximal to the filter, and 6<br>circumferential marker bands on<br>the proximal portion of the filter. |
| Angioplasty<br>Balloon<br>Characteristics | SAME as 510(k) cleared<br>Paladin System | Nylon semi-compliant PTA<br>balloon |
| Embolic Capture<br>Mechanism | SAME as 510(k) cleared<br>Paladin System | Self-expanding nitinol filter frame<br>with urethane membrane (40<br>micron pores) for capture and<br>removal of embolic material from<br>the body |
| Filter diameter | SAME as 510(k) cleared<br>Paladin System | Adjustable filter fits carotid artery<br>diameters up to 7mm diameter |
| Usable length<br>[cm] | SAME as 510(k) cleared<br>Paladin System | 140 |
| Introducer | SAME as 510(k) cleared<br>Paladin System | 6 F |
| Parameters | Contego Medical Paladin<br>System<br>(modified device) | Contego Medical Paladin<br>System<br>(predicate device K181128) |
| sheath<br>compatibility | Paladin System | |
| Guide wire<br>compatibility | SAME as 510(k) cleared<br>Paladin System | 0.014" |
| Balloon diameter<br>[mm] | SAME as 510(k) cleared<br>Paladin System | 5.0 mm and 5.5 mm |
| Balloon length<br>[mm] | SAME as 510(k) cleared<br>Paladin System | 20 mm and 30 mm |
| Balloon RBP<br>[atm] | SAME as 510(k) cleared<br>Paladin System | 14 atm (5.0 mm)<br>11 atm (5.5 mm) |
# Table 4-1. Substantial Equivalency Table
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Performance Data:
The risk analysis activities for the device changes summarized in this Special 510(k) identified that no additional verification or validation activities are required to establish substantial equivalence with predicate devices.
The changes highlighted for the Paladin System do not create any new risks, harms nor were any existing risks altered. No additional risk control measures or mitigations are required.
### Conclusion:
The information provided is sufficient to establish the substantial equivalency of the Paladin System (modified device) to the predicate device (Contego Medical, Paladin System 510(k) K181128).
There is no change to device design, its principles of operation or expected performance. The information provided in this submission demonstrates that modified Paladin System is equivalent to the currently cleared Paladin system.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.