The JADE PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilation of balloon expandable and self-expanding stents in the peripheral vasculature.
Device Story
The Jade PTA Balloon Dilatation Catheter is an over-the-wire catheter used by physicians in clinical settings for peripheral vascular interventions. It consists of a balloon mounted on a shaft with a guidewire lumen (compatible with 0.014-inch guidewires). The device is inserted into the vasculature to reach a target lesion; the balloon is inflated to dilate the vessel or post-dilate a stent. Radiopaque marker bands assist in positioning under fluoroscopy. The device features a hydrophilic coating on the distal section to facilitate tracking and a silicone-coated inner lumen. By restoring vessel patency or optimizing stent apposition, the device aims to improve blood flow and treat obstructive lesions. It does not include distal dye injection or pressure measurement capabilities.
Clinical Evidence
Bench testing only. Testing included visual inspection, marker band radiopacity, dimensional verification, coating integrity, particulate evaluation, balloon preparation/deployment/retraction, balloon burst (including in-stent), balloon compliance, balloon fatigue (including in-stent), inflation/deflation time, flexibility/kink, torque strength, tip pull strength, and catheter bond strength. Biocompatibility testing included cytotoxicity, irritation, sensitization, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility (hemolysis, complement activation, in vivo thromboresistance), and genotoxicity. All results met acceptance criteria.
Indicated for patients requiring percutaneous transluminal angioplasty in peripheral vasculature (iliac, femoral, ilio-femoral, popliteal, infra-popliteal, renal arteries) or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae; also indicated for post-dilation of peripheral stents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 28, 2020
OrbusNeich Medical (Shenzhen) Co., Ltd. Mr. Daniel Zhang Regulatory Director No.1st Jinkui Road Futian Free Trade Zone Shenzhen, Guangdong 518038 China
# Re: K201794
Trade/Device Name: Jade PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: June 23, 2020 Received: June 30, 2020
# Dear Mr. Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
### Device Name
Jade PTA Balloon Dilatation Catheter
### Indications for Use (Describe)
The Jade PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, poplited, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilation of balloon expandable and self-expanding stents in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR 807.92(c).
| Submitter: | OrbusNeich Medical (Shenzhen) Co., Ltd.<br>1 Jinkui Road, Futian Free Trade Zone<br>Shenzhen, 518038, China |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Daniel Zhang |
| Date Prepared: | June 19, 2020 |
| Trade Name: | Jade PTA Balloon Dilatation Catheter |
| Common Name: | Percutaneous Transluminal Angioplasty (PTA) Catheter |
| Classification Name: | Catheter, Angioplasty, Peripheral, Transluminal (21 CFR 870.1250) |
| Product Code: | LIT |
| Predicate Device: | OrbusNeich Jade (K173894; cleared February 9, 2018) |
| Reference Device: | Bard Ultraverse (K192318; cleared October 3, 2019)<br>Boston Scientific Sterling (K141112; cleared July 16, 2014 & K132430;<br>cleared October 17, 2013) |
| Device Description: | The JADE PTA Balloon Dilatation Catheters is now also available as an over-the-<br>wire balloon catheter for peripheral indications. The balloon diameters range from<br>1.5mm to 6.0mm and balloon lengths range from 20mm to 240mm, with 90cm,<br>150cm and 200cm catheter lengths. The balloon material is made of a minimally<br>compliant material with a rated burst pressure of 18 atmospheres for 5.0-6.0mm and<br>20atm for 1.5-4.0mm. Hydrophilic lubricious coatings are applied to the distal<br>section of the catheter.<br>The shaft of the catheter is composed of a proximal shaft and a distal shaft. The<br>distal shaft is composed of a distal outer tube and tri-extrusion inner tube with a<br>balloon welded to both tubes at the distal tip to aid in tracking through vasculature.<br>The proximal shaft is composed of a proximal outer tube and tri-extrusion inner tube<br>that are bonded to a female luer connector, with the proximal outer tube allowing<br>for proximal pushability with a smooth transition to the distal shaft. The inner lumen<br>of the catheter accepts a maximum 0.014 inch (0.36mm) guidewire. The guidewire<br>enters the catheter tip and advances coaxially out the hub guidewire lumen. Two<br>radiopaque marker bands are positioned within the balloon shoulder, and for<br>balloons of working length 180-240mm, two more marker bands (four marker bands<br>in total) are positioned in the middle of the balloon. One marked section, for the<br>90cm catheter working length, or two marked sections, for the 150cm and 200cm<br>catheter working lengths, are located on the proximal shaft to indicate catheter<br>position relative to the tip of the guiding catheter or introducer sheath. The Y-type<br>hub is bonded on the proximal end of the catheter at the entrance to the inflation<br>lumen and the guidewire lumen. The design of this dilatation catheter does not<br>incorporate a lumen for distal dye injections or distal pressure measurements. |
| Intended Use: | The JADE PTA Balloon Dilatation Catheter is indicated for Percutaneous<br>Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral,<br>ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of<br>obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device<br>is also indicated for post-dilation of balloon expandable and self-expanding stents<br>in the peripheral vasculature. |
| Technological Characteristics: | The subject and predicate devices are the same based on the following<br>technological elements:<br>■ the same indications for use<br>■ minimally-compliant balloon |
{4}------------------------------------------------
- nominal pressure of 12 atm
- rated burst pressure of 18/20 atm
- distal shaft
- hydrophilic coating on the distal section of the catheter
- silicone coating in the inner lumen
- . 0.014" guidewire compatibility
- . EO sterilization
The technological differences of the subject comparing with the predicate device are as follows:
- over-the-wire catheter shaft design (for catheter working lengths of 90, 150 and 200cm)
- . additional balloon lengths of 150, 180, and 240mm
- over-the-wire hub design
- specific materials selected for proximal shaft
- dimensions of proximal shaft components and catheter
{5}------------------------------------------------
Performance Data:
The following testing was performed to support the over-the-wire shaft design of the Jade PTA balloon dilatation catheter:
- . Performance Testing
- Visual Inspection ●
- Marker Band Radiopacity .
- Dimensional Verification .
- . Coating Integrity
- Particulate Evaluation ●
- . Balloon Preparation, Deployment, and Retraction
- . Balloon Burst.
- . Balloon Burst (in-stent)
- . Balloon Compliance
- Balloon Fatigue
- . Balloon Fatigue (in-stent)
- . Balloon Inflation and Deflation Time
- Flexibility and Kink .
- Torque Strength ●
- . Tip Pull Strength
- Catheter Bond Strength
- Biocompatibility ■
- Cytotoxicity ●
- ISO Intracutaneous Irritation .
- Sensitization ISO Guinea Pig .
- Acute Systemic Toxicity
- Material-mediated Pyrogenicity ●
- Hemocompatibility ●
- Hemolysis
- . Complement Activation
- . In vivo Thromboresistance
- . Genotoxicity
- Bacterial Mutagenicity ●
- In Vitro Mouse Lymphoma Assay .
- Sterilization
The test results met all acceptance criteria, which are the same or similar to the predicate device, and ensure that the Jade PTA balloon dilation catheter design and construction are suitable for its intended use.
Conclusion:
This information supports a determination of substantial equivalence between the Jade PTA balloon dilatation catheter and the predicate device described above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.