K143036 · Natec Medical , Ltd. · LIT · Mar 30, 2015 · Cardiovascular
Device Facts
Record ID
K143036
Device Name
Ebony PTA 0.035 Peripheral Dilatation Catheter
Applicant
Natec Medical , Ltd.
Product Code
LIT · Cardiovascular
Decision Date
Mar 30, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ebony® PTA .035 Peripheral Dilatation Catheter is intended for dilatation of lesions in the femoral, iliac, popliteal, infrapopliteal & renal arteries. The Ebony® PTA . 035'' OTW Peripheral Dilatation Catheter is not for use in the coronary arteries and neuro-vasculature.
Device Story
The Ebony PTA 0.035 Peripheral Dilatation Catheter is an Over-The-Wire (OTW) catheter featuring a semi-compliant inflatable balloon and an atraumatic tapered tip. Two radiopaque markers assist in positioning the balloon across arterial stenoses; a hydrophilic coating aids navigation through the vasculature. Used by physicians in clinical settings for percutaneous transluminal angioplasty (PTA) to dilate arterial lesions. The device is advanced through the vasculature to the target site, where the balloon is inflated to restore vessel patency. This procedure benefits patients by improving blood flow in peripheral arteries. The current device is a line extension of a previously cleared catheter, adding specific balloon diameters (3.0 mm and 4.0 mm) and increased balloon lengths.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including balloon compliance, burst pressure, fatigue, shaft torque, bond strength, catheter dimensions, coating integrity, deflation time, guide wire/introducer compatibility, and biocompatibility testing.
Technological Characteristics
Over-the-wire (OTW) PTA catheter; semi-compliant inflatable balloon; radiopaque markers; hydrophilic coating. Materials and design are consistent with predicate K103354. Bench testing performed for balloon performance (compliance, burst, fatigue), mechanical integrity (bond strength, torque), and biocompatibility.
Indications for Use
Indicated for dilatation of lesions in femoral, iliac, popliteal, infrapopliteal, and renal arteries. Contraindicated for use in coronary arteries and neuro-vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 30, 2015
Natec Medical, Ltd. % Judith Danielson Senior Regulatory Consultant CardioMed Device Consultants, LLC 5523 Research Park Drive, Suite 205 Baltimore, MD 21228
Re: K143036
Trade/Device Name: Ebony PTA 0.035 Peripheral Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: March 23, 2015 Received: March 25, 2015
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K143036
Device Name
Ebony® PTA .035 Peripheral Dilatation Catheter
Indications for Use (Describe)
The Ebony@ PTA. 035 Peripheral Dilatation Catheter is intended for dilatation of lesions in the femoral, iliac, popliteal, infrapopliteal & renal arteries.
The Ebony® PTA . 035'' OTW Peripheral Dilatation Catheter is not for use in the coronary arteries and neuro-vasculature.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FOR FDA USE ONLY CONTRACTOR
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
{3}------------------------------------------------
### Ebony® PTA 0.035'' Peripheral Dilatation Catheter 510(k) SUMMARY 21 CFR 807.92
### Applicant:
| Company Name: | Natec Medical Ltd. |
|---------------------------|------------------------------------------------------------------------------------------|
| Company Address: | Maeva Centre Building - Ebene Business Park<br>Reduit, Mauritius |
| Telephone: | +230 466 30 54 |
| Fax: | +230 466 67 70 |
| Contact Person: | Xavier De Buchere<br>Regulatory Affairs & Quality Manager<br>xdbuchere@natec-medical.com |
| Summary Preparation Date: | October 14th. 2014 |
### Device Name:
| Trade Name: | Ebony® PTA 0.035 Peripheral Dilatation Catheter |
|----------------------|-------------------------------------------------|
| Common/Usual Name: | PTA Catheter |
| Classification Name: | Percutaneous Catheter |
| Regulation Number: | 21 CFR 870.1250 |
| Product Code: | LIT |
| Device Class: | Class II |
### Predicate Devices:
- Ebony® PTA 0.035 Peripheral Dilatation Catheter (K103354) ●
### Device Description:
The Ebony® PTA 0.035 Peripheral Dilatation Catheter is a standard Over The Wire (OTW) PTA catheter with a semi-compliant inflatable balloon at the distal part with an atraumatic, tapered tip at its distal end. Two radiopaque markers facilitate proper balloon positioning across the stenosis, and a hydrophilic coating facilitates advancement of the catheter through the vasculature.
The modified Ebony® PTA 0.035 Peripheral Dilatation Catheter line extension adds 3.0 mm and 4.0 mm diameter balloon and additional balloon lengths to the currently available sizes.
{4}------------------------------------------------
## Indication for Use:
Ebony® PTA 0.035 Peripheral Dilatation Catheter is intended for dilatation of lesions in the femoral, iliac, popliteal, infra popliteal and renal arteries.
The Ebony® PTA .035'' OTW Catheter is not for use in the coronary arteries and neurovasculature.
# Technological Characteristics:
The modified Ebony® PTA 0.035 Peripheral Dilatation Catheter has the same indication for use, and is manufactured with the same over the wire design and materials as the Ebony® PTA 0.035 Peripheral Dilatation Catheter cleared under K103354. The modification involved adding balloon diameter 3.0mm and 4.0mm and longer balloon lengths (100, 120, 150 and 200 mm) to the previously cleared balloon size matrix and a change in the hydrophilic coating material.
# Non-Clinical Testing Summary:
The following bench testing was performed according to the same testing method used with the original Traditional 510(k) cleared under K103354.
- I Balloon compliance
- 트 Balloon burst pressure
- 트 Balloon fatigue
- 트 Biocompatibility testing (A summary of biocompatibility results is provided in file 002).
- I Shaft resistance (Torque Test)
- I Bond strength
- 트 Catheter dimensions
- I Coating integrity (Visual inspection)
- 트 Deflation time
- I Guide wire and introducer compatibility.
# Clinical Testing Summary:
No clinical study was performed.
# Conclusion
The proposed device and the predicate Ebony® PTA 0.035 Peripheral Dilatation Catheter (K103354) have the same indication, similar materials and a similar design. The new balloon sizes met all of the verification/validation acceptance criteria and the biocompatibility results demonstrated that the modified Ebony® PTA 0.035 Peripheral Dilatation Catheter is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.