K121385 · R4 Vascular, Inc. · LIT · Jun 14, 2012 · Cardiovascular
Device Facts
Record ID
K121385
Device Name
VECTOR
Applicant
R4 Vascular, Inc.
Product Code
LIT · Cardiovascular
Decision Date
Jun 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Vector 4 ™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae. The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilatation of stents
Device Story
Vector4 PTA Balloon Dilatation Catheter is a high-pressure, non-compliant, coaxial dual-lumen over-the-wire catheter. It features a radiopaque-coated balloon working length, enabling fluoroscopic visualization of placement without requiring contrast solution. Input: guide wires up to 0.035 inches. Operation: physician advances catheter to stenosis; balloon inflated via saline or contrast through outer lumen. Output: mechanical dilation of peripheral artery or dialysis fistula lesions. Used in clinical settings by physicians. Benefits: enables lesion visualization without contrast, reducing potential contrast-related complications; provides high-pressure dilation (30 atm rated burst pressure).
Clinical Evidence
Bench testing only. Performance testing included rated burst pressure, compliance, dimensional verification, guidewire/introducer compatibility, fluoroscopic visualization, inflation/deflation times, repeat inflation, leak, tensile, kink, torque, corrosion, luer lock compatibility, coating integrity, and distribution. Biocompatibility testing performed per ISO 10993-1 for short-duration blood contact (<24 hours). Sterilization validated to SAL 10^-6. Comparative testing confirmed radiopacity and particulate performance equivalent to predicate.
Technological Characteristics
Coaxial dual-lumen over-the-wire catheter; high-pressure non-compliant composite balloon; radiopaque coating on balloon working length; 30 atm rated burst pressure; compatible with 0.035" guidewires and 6-7F introducers; biocompatible, latex-free materials; ethylene oxide sterilization (SAL 10^-6).
Indications for Use
Indicated for patients requiring Percutaneous Transluminal Angioplasty (PTA) of femoral, iliac, popliteal, and renal arteries, or treatment of obstructive lesions in native or synthetic arterio-venous dialysis fistulae. Contraindicated for coronary artery use and stent placement or post-dilatation.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K.121385
510(k)A001 Page 9
# 510(k) Summary
JUN 1 4 2012
## Administrative Information
#### Primary Contact Person
Sew-Wah Tay, PhD Regulatory Consultant, Libra Medical Inc. 8401 73td Ave North, Suite 63 Minneapolis, MN 55428 Phone: 612-801-6782 Fax: 763-477-6357 Email: swtay@libramed.com
Date Prepared: 4/3/2012
### Sponsor's Contact
Michael Hoch r4 Vascular, Inc. 7550 Meridian Circle North, Suite 150 Maple Grove, MN 55369 Phone: 763-494-8400 x.10 Fax: 763-494-8484 Email: mhoch@r4vascular.com
### Device Name
Trade Name
Vector TM
| Common Name | Percutaneous Catheter |
|---------------------|-----------------------|
| Classification Name | Percutaneous Catheter |
| Classification | II |
| Product Code | DQY, LIT |
# Predicate Device
The device is substantially equivalent to the Bard Conquest PTA catheter (K083657).
### Indications for Use/Intended Use
The Vector " Thi PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae.
The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilatation of stents
### Device Description and Technological Characteristics
The Vector" PTA Catheter is a high pressure, non-compliant balloon catheter with a radiopaque coating on the balloon. The radio-opaque coating on the working length of the balloon
{1}------------------------------------------------
minimizes or eliminates the need for using contrast solution to inflate the balloon. The radioopaque coating delineates the working length of the balloon and provides continuous fluoroscopic visibility during balloon placement. Saline may be used to inflate the balloon without affecting visualization. Contrast mixture may be used at the discretion of the physician.
The Vector4™ PTA Balloon Dilatation Catheter is a coaxial dual lumen over-the-wire balloon dilatation catheter. The central lumen accommodates guide wires up to 0.035 inches in diameter while the outer lumen is the inflation lumen for the balloon. The catheter includes an atraumatic tip to ease advancement of the catheter to and through the stenosis. The distal tip is visible under fluoroscopy.
The balloon is a high pressure composite balloon with a radiopaque coating on the working length to help visualize placement within the lesion during both placement and inflation without the use of contrast. The balloon's rated burst pressure is 30 atm.
The catheter is supplied sterile and non-pyrogenic. Different balloon sizes (5 mm to 10 mm) and catheter lengths (50 to 135 cm) are available. The proximal end of the catheter contains a hub with two female luer locks.
The device is compatible with 6-7F introducers. The introducer compatibility for each device size is listed on the device label itself. The materials used in the construction are biocompatible and latex free.
### Performance Data
The Vector 4™ PTA Balloon Dilatation Catheter is tested and meets all its physical and performance specifications including:
- Balloon rated burst pressure .
- Balloon compliance ●
- Critical dimension verifications .
- Guidewire and introducer compatibility .
- Fluoroscopic visualization .
- . Inflation/deflation times
- . Repeat inflation
- Leak
- Tensile
- . Kink
- . Torque
- Corrosion ●
- . Luer lock compatibility
- Coating integrity .
- Distribution ●
In addition, the device was tested for biocompatibility ISO 10993-1 for short duration contact with blood (<24 hour).
{2}------------------------------------------------
The device is sterilized by ethylene oxide to an SAL 10° level. These performances are similar to that described by the predicate device as these are standard tests for PTA catheters. In addition, direct comparative testing for radiopacity and particulates was conducted for both devices and results found to be similar.
After accelerated aging on devices, a subset of the original tests that may be affected by aging were repeated.
The preclinical testing showed that the device meets specifications before and after aging indicating that the device is as safe and effective as the predicate device.
#### Substantial Equivalence
The Vector" is substantially equivalent to the Conquest PTA catheter (K083657). They have similar intended use, and treat the same target population. Both devices are intended to treat peripheral arteries. The Vector4" is indicated for an additional artery - popliteal artery which is also a peripheral artery with similar sizes and access as the other peripheral arteries. The manner in accessing and treating lesions in all these arteries is similar. Therefore, this minor change in indication for use does not change the intended use of the device and does not raise new safety and effectiveness question.
Both devices employ a composite balloon construction to attain the high rated burst pressures. Both devices are over the wire catheters with coaxial lumens. Both devices are compatible with 0.035" wires and have similar construction and principles of operation. Both devices are used by the physician in a similar manner typical of all PTA balloon catheters. Both devices are sterilized with ethylene oxide with a sterility assurance of 10°.
Fluoroscopic visualization for the Conquest balloon is achieved by injecting contrast into the balloon of the Conquest catheter while the radiopaque coating on the Vector balloon allows visualization without contrast. This difference does not raise new safety and effectiveness questions as the visualization under fluoroscopy has been shown to be equivalent for both devices.
#### Conclusion
The non-clinical testing of the Vector" PTA Balloon Dilatation Catheter met all the product specifications, and all the required standards for sterility and biocompatibility. The Vector" and the predicate Conquest PTA catheters have similar intended use and technological characteristics and are therefore substantially equivalent. These tests showed that the Vector"™ PTA Balloon Dilatation Catheter is as safe and effective as the predicate device.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle or bird-like symbol with flowing lines, representing the department's mission related to health and human well-being.
#### Public Health Service
JUN 1 4 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
r4 Vascular, Inc. c/o Mark Job Responsible Third Party Official Regulatory Technical Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K121385
Trade/Device Name: Vector4 PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (Two) Product Code: LIT, DQY Dated: May 7, 2012 Received: May 8, 2012
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MIT Alleler
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 510(k) A001
# Indications For Use Statement
### Indications for Use
510(k) Number (if known):
Device Name: Vector4TM PTA Catheter
#### Indications for Use:
The Vector 4 ™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae.
The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilatation of stents
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices
1421795 510(k) Number_
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.