← Product Code [LIT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT) · K103354

# EBONY PTA .035 PERIPHERAL DILATATION CATHETER (K103354)

_Natec Medical , Ltd. · LIT · Feb 23, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT/K103354

## Device Facts

- **Applicant:** Natec Medical , Ltd.
- **Product Code:** [LIT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT.md)
- **Decision Date:** Feb 23, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Ebony® .035 PTA Catheter is intended for the dilation of lesions in the femoral, iliac, popliteal, infrapopliteal and renal arteries. The Ebony® .035 PTA Catheter is not for use in coronary arteries or neuro-vasculature.

## Device Story

Over-the-wire (OTW) peripheral transluminal angioplasty (PTA) catheter; features semi-compliant inflatable balloon at distal end; atraumatic tapered tip for crossing stenoses; double lumen design (one for guidewire/inflation, one for distal access). Radiopaque markers on balloon provide visual positioning reference. Hydrophilic coating on distal shaft improves trackability/pushability. Luer lock fitting connects to external inflation device. Used by clinicians in clinical settings to dilate arterial lesions; balloon diameter/length selected based on vessel anatomy. Inflation device expands balloon to known diameter at specific pressures per compliance chart. Benefits patient by restoring blood flow in peripheral arteries.

## Clinical Evidence

Bench testing only. No clinical data. Testing included balloon compliance, burst pressure, fatigue, shaft resistance, bond strength, dimensions, deflation time, and compatibility with guidewires/introducers. Results met all predetermined acceptance criteria.

## Technological Characteristics

Double lumen OTW catheter; semi-compliant balloon; hydrophilic coating; radiopaque markers. Materials biocompatible per ISO 10993 (Parts 1, 2, 4, 5, 10, 11, 12) and ASTM F756-00. Available in diameters 5-12mm, lengths 20-80mm, catheter lengths 80-150cm. Sterile, single-use.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- ADMIRAL XTREME™ PTA Catheter ([K062809](/device/K062809.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103354

# 510(k) Summary 807.92(c)

# FEB 2 3 2011

### SPONSOR

#### 807.92(a)(1)

Company Name: Company Address

Natec Medical Ltd. Maeva Centre Building - Ebene Business Park Reduit, Mauritius

Telephone: Fax:

+230 466 30 54 +230 466 67 70

Contact Person:

Alain Valorge Quality manager info@natec-medical.com

Summary Preparation Date: October 28, 2010

| DEVICE NAME          | 807.92(a)(2)                                   |  |
|----------------------|------------------------------------------------|--|
| Trade Name:          | Ebony® PTA .035 Peripheral Dilatation Catheter |  |
| Common/Usual Name:   | PTA Catheter                                   |  |
| Classification Name: | Percutaneous Catheter                          |  |
| Regulation Number:   | 21 CFR 870.1250                                |  |
| Product Code:        | LIT                                            |  |
| Device Class:        | Class II                                       |  |

#### PREDICATE DEVICE

#### 807.92(a)(3)

Legally Marketed Equivalent Device

| K Number | Trade Name                      | Manufacturer                       |
|----------|---------------------------------|------------------------------------|
| K062809  | ADMIRAL XTREME™ PTA<br>Catheter | Invatec Innovative<br>Technologies |

#### DEVICE DESCRIPTION

#### 807.92(a)(4)

The Ebony® PTA .035 Peripheral Dilatation Catheter is a standard Over the Wire (OTW) PTA Catheter with a semi-compliant inflatable balloon at the distal part and an atraumatic, tapered tip to aid in crossing tight stenoses.

It is a double lumen catheter, one lumen for the guide wire used for inflation/deflation of the balloon and one lumen for the guide wire used to access to the distal tip of the catheter (max/0.035″). A luer lock fitting allows the connection with an inflation device (Y hub). The balloon has two radiopaque markers to provide visual reference points for balloon positioning relative to the stenosis within the vessel.

{1}------------------------------------------------

The distal catheter, proximal to the balloon, is covered with a hydrophilic coating to improve the trackability and pushability characteristics.

The balloon material expands to a known diameter at specific pressure defined in a compliance chart supplied with the catheter. The device is available in balloon. diameters of 5, 6, 7, 8, 9, 10 or 12 mm, balloon lengths of 20, 40, 60, or 80 mm and catheter lengths of 80, 130, or 150 cm.

lt will be supplied sterile and is intended for one time use.

## DEVICE INTENDED USE

#### 807.92(a)(5)

The Ebony® .035 PTA Catheter is intended for the dilation of lesions in the femoral, iliac, popliteal, infrapopliteal and renal arteries.

The Ebony® .035 PTA Catheter is not for use in coronary arteries or neuro-vasculature.

## COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a)(6)

The Ebony® .035 PTA Catheter and the predicate device, Admiral Xtreme™ PTA Catheter have the same intended use for the dilatation of lesion in the femoral, iliac, infra popliteal and renal arteries. The technological characteristics such as material, biocompatibility, mode of operation, performance properties, sterilization and packaging are substantially equivalent to the predicate device.

## NONCLINICAL TESTS

807.92(b)

## SAFETY and EFFECTIVENESS

#### BIOCOMPATIBILITY

All materials used in the Ebony® .035 PTA Catheter are biocompatible based on biocompatibility testing results. The device has been tested according to ISO 10993 Part 1, 2, 4, 5, 10, 11, 12, ASTM F756-00 and 21 CFR 58 (GLP regulations)

## PERFORMANCE DATA

The safety and effectiveness of the Ebony® .035 PTA Catheter has been demonstrated through data collected from in vitro bench tests and analyses. Testing results demonstrated equivalent performance of the Ebony® .035 PTA Catheter with the predicate device.

The testing included balloon compliance, balloon burst pressure, balloon fatigue, shaft resistance, bond strength, catheter dimensions, deflation time and guide wire and introducer compatibility.

### CONCLUSION

## 807.92(b)(3)

The subject device, the Ebony® .035 PTA Catheter, met all the predetermined acceptance criteria of design verification and validation as specified by applicable

{2}------------------------------------------------

:

:

:

. . . . .

. .

.

1. 1. 1. 1.

standards, guidance, and test protocols. It is the conclusion of Natec Medical, Ltd that the Ebony® .035 PTA Catheter is as safe and effective as the predicate device.

and the state of the state of the states of the states

:

:

:

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes wrapped around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Natec Medical Ltd c/o Mr. E.J. Smith Smith Associates 1468 Harwell Ave. Crofton, MD 21114

FEB 2 3 2011

ﻟﻤﻤﻠﻜﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ

Re: K103354

Trade/Device Name: Ebony .035 PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Regulatory Class: Class II (two) Product Code: LIT Dated: February 8, 2011 Received: February 9, 2011

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

## Page 2 - Mr. E.J. Smith

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

er D. V. hner

Image /page/4/Picture/8 description: The image shows a black and white drawing of a symbol. The symbol appears to be a stylized letter or character, with curved lines and a loop at the bottom. The lines are thick and bold, and the overall impression is abstract and calligraphic. The symbol is isolated against a white background.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use Form

Indications for Use

510(k) Number (if known): K103354

Device Name: Ebony .035 PTA Catheter

Indications for Use:

The Ebony® .035 PTA Catheter is intended for the dilation of lesions in the femoral, illac, popliteal, infrapopliteal and renal arteries.

The Ebony® .035 PTA Catheter is not for use in coronary arteries or neuro-vasculature.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

R. Vihnes

(Division Sign-Off) Division of Cardiovascular Devices

Page _1_of ।

510(k) Number_KIO 3354

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT/K103354](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT/K103354)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
