K101062 · Trireme Medical, Inc. · LIT · May 12, 2010 · Cardiovascular
Device Facts
Record ID
K101062
Device Name
GLIDER PTA BALLOON CATHETER
Applicant
Trireme Medical, Inc.
Product Code
LIT · Cardiovascular
Decision Date
May 12, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The GliderXtreme™ PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries. NOT for use in the coronary or cerebral vasculature.
Device Story
Standard PTA balloon dilatation catheter; braided shaft; atraumatic tapered tip; semi-compliant balloon; radiopaque markers for positioning; hydrophilic coating. Available in Rapid Exchange or Over-The-Wire configurations. Used in clinical settings by physicians for peripheral artery angioplasty. Device inserted over guidewire (0.014" or 0.018") via introducer sheath; balloon inflated via luer connector to dilate stenotic lesions. Benefits include restoration of blood flow in peripheral arteries. Single-use, sterile device.
Clinical Evidence
Bench testing only. In vitro tests included balloon rated burst pressure, inflation/deflation, fatigue, bond strength, torsional strength, catheter diameter, balloon profile, tip configuration, balloon compliance, trackability, pushability, kink resistance, and device interface compatibility.
Technological Characteristics
Braided shaft; semi-compliant balloon; hydrophilic coating; radiopaque markers; hypotube proximal end with plastic hub/strain relief. Compatible with 0.014"/0.018" guidewires and 5F/6F introducers. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for balloon dilatation of peripheral vascular lesions in iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries. Contraindicated for coronary or cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K120677 — CHOCOLATE PTA BALLON CATHETER (OVER THE WIRE) GUIDE WIRE (0.014) · Trireme Medical, Inc. · Mar 15, 2012
K083919 — AMPHIRION DEEP 0.014 OTW PTA BALLOON CATHETER · Invatec Innovative Technologies, S.R.L. · Mar 13, 2009
K093965 — ULTRAVERSE 014 AND 018 BALLOON DILATATION CATHETERS · C.R. Bard, Inc. · Mar 17, 2010
K131199 — ULTRAVERSE RX PTA BALLOON DILATATION CATHETER · Bard Peripheral Vascular, Inc. · May 30, 2013
K112513 — EBONY PTA 0.014 PERIPHERAL DILATATION CATHETER · Natec Medical , Ltd. · Dec 9, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K101062
#### 510(k) Summary [As required by 21 CFR 807.92(c)]
- 1. Submitter's Name / Contact Person
| Submitter: | TriReme Medical, Inc.<br>7060 Koll Center Parkway, Suite 300<br>Pleasanton, CA 94566 |
|-----------------|--------------------------------------------------------------------------------------|
| Contact Person: | Shiva Ardakani<br>VP of RA/QA<br>Phone: 925-931-1300 Ext 209<br>Fax: 925-931-1361 |
| Date Prepared: | May 4, 2010 |
General Information
| Trade Name: | GliderXtreme™ PTA Balloon Catheter |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Angioplasty Catheter |
| Classification Name: | Percutaneous Catheter (21 CFR 870.1250) |
| Product Code: | DQY / LIT |
| 510(k) number: | |
| Predicate Devices: | NanoCross PTA Dilatation Catheter (K090849)<br>Sterling ES MR and OTW PTA Balloon Dilatation Catheters |
MAY 12 2010Balloon Dilatation Catheters (K080982)
EverCross .035" OTW PTA Dilatation Catheter (K082579)
#### 3. Intended Use
2.
The GliderXtreme™ PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries. NOT for use in the coronary or cerebral vasculature.
- 4. Device Description
The GliderXtreme™ PTA Balloon Catheter is a standard balloon dilatation catheter with a braided shaft and an atraumatic tapered tip. The device is compatible with commonly used accessories, including standard 0.014" and 0.018" guidewires and 5F introducer sheaths (or 6F guide catheters). Overall catheter length is approximately 135 cm.
The distal end of the catheter has a semi-compliant balloon that expands to known diameters at specific pressures. The balloon is available in multiple sizes and contains one or two radiopaque markers to assist with positioning. The shaft is braid reinforced and has a lubricious hydrophilic coating. The proximal end of the device is a common PTA catheter design of a hypotube connected to a plastic hub and strain relief. The hub is used to inflate the balloon; the luer connector is compatible with standard inflation
{1}------------------------------------------------
devices. A second lumen intended for guidewire use is contained within the catheter. The balloon catheters are available in Rapid Exchange or Over-The-Wire configuration.
The GliderXtreme™ PTA Balloon Catheters are supplied sterile and intended for single use only.
- 5. Performance Data
Bench testing was performed to support a determination of substantial equivalence. Results from this testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. The following in vitro tests were performed:
- Balloon Rated Burst Pressure .
- . Balloon Inflation and Deflation
- . Balloon Fatigue
- Catheter Body Strength (Bond Strength) .
- . Torsional Strength
- Catheter Diameter, Balloon Profile and Tip Configuration
- Balloon Compliance ●
- . Trackability, Pushability
- Kink Resistance .
- Device Interface Compatibility ◆
#### ર્ભ . Substantial Equivalence Comparison and Conclusion
All GliderXtreme™ PTA Balloon Catheters are substantially equivalent to the predicate devices in design, materials, packaging, fundamental scientific technology, manufacturing, sterilization and intended use. Performance testing demonstrated that the GliderXtreme™ PTA Balloon Catheter reliably achieved the desired effect and is safe for its intended use. No new questions of safety or effectiveness were identified during device testing. Therefore, All GliderXtreme™ PTA Balloon Catheters are considered substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized image of a bird-like figure, possibly representing a symbol of health or service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
TriReme Medical, Inc. c/o Chris Lavanchy Primary Reviewer CITECH 5200 Butler Pike Plymouth Meeting, PA 19462-1298
MAY 1 2 2010
Re: K101062
Trade/Device Name: Glider PTA Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: May 10, 2010 Received: May 11, 2010
Dear Mr. Lavanchy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 -- Mr. Chris Lavanchy
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
er R. lahmen
Bram D. Zuckerman, M.D.
Director. Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and · Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (K094049): K i U (06 2
Device Name: Glider™ PTA Balloon Catheter
Indications for Use:
The Glider™ PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, NOT for use in the coronary or cerebral vasculature.
The indication for use is same for all Glider PTA Balloon Catheter Family.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
unner R. Vc. Anner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.