← Product Code [LIT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT) · K093286

# POWERCROSS .018 OTW PTA DILATION CATHETER (K093286)

_Ev3, Inc. · LIT · Nov 13, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT/K093286

## Device Facts

- **Applicant:** Ev3, Inc.
- **Product Code:** [LIT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT.md)
- **Decision Date:** Nov 13, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The PowerCrossTM .018" OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also intended for stent post-dilatation.

## Device Story

PowerCross .018" OTW PTA Dilatation Catheter is an over-the-wire coaxial catheter featuring a distally mounted semi-compliant inflatable balloon and atraumatic tapered tip. Device utilizes a central lumen for guidewire tracking (max 0.018") and a separate inflation lumen for contrast/saline mixture. Radiopaque markers facilitate precise positioning relative to stenosis. Used by physicians in clinical settings for percutaneous transluminal angioplasty to dilate arterial stenoses, treat dialysis fistula lesions, and perform stent post-dilatation. Output is mechanical dilation of vascular lesions, visualized via fluoroscopy using radiopaque markers. Benefits include restoration of vessel patency and improved blood flow.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Coaxial OTW catheter design; semi-compliant inflatable balloon; hydrophilic coating on distal portion; radiopaque marker bands for positioning; 0.018" guidewire compatibility. Materials consistent with predicate NanoCross. Sterilization via standard processes.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- NanoCross™ .014" OTW PTA Dilatation Catheter ([K090849](/device/K090849.md))
- EverCross™ .035" OTW PTA Dilatation Catheter ([K082579](/device/K082579.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K093286

ev3®
heter

## 510(k) Summary

## געעים 3 ייעוץ

## PowerCross™ .018" OTW PTA Dilatation Catheter

| 510(k) Summary                        | This summary of 510(k) safety and effectiveness information is<br>being submitted in accordance with the requirements of 21 CFR<br>§ 807.92.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant                             | ev3 Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Submitter                             | ev3 Inc.<br>3033 Campus Drive<br>Plymouth, MN 55441<br>Tel: 763-398-7000<br>Fax: 763-591-3248                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Contact Person                        | Sara Bakker<br>Regulatory Affairs Specialist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Date Prepared                         | October 16, 2009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Trade Name                     | PowerCrossTM .018" OTW PTA Dilatation Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Common Name                    | PTA Dilatation Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name                   | Catheter, Percutaneous (21 CFR 870.1250, Product Code DQY)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Panel                  | Cardiovascular                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate Device                      | NanoCrossTM .014" OTW PTA Dilatation Catheter (K090849)<br>EverCrossTM .035" OTW PTA Dilatation Catheter (K082579)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Intended use                          | The PowerCrossTM .018" OTW PTA Dilatation Catheter is<br>intended to dilate stenoses in the iliac, femoral, ilio-femoral,<br>popliteal, infra-popliteal, and renal arteries, and for the<br>treatment of obstructive lesions of native or synthetic<br>arteriovenous dialysis fistulae. This device is also intended for<br>stent post-dilatation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Description                    | The PowerCrossTM .018" OTW PTA Dilatation Catheter is an<br>over the wire (OTW) 0.018" coaxial catheter with a distally<br>mounted semi-compliant inflatable balloon and an atraumatic<br>tapered tip. The distal portion of the catheter has a hydrophilic<br>coating. The manifold includes a lumen marked "THRU". This<br>is the central lumen of the catheter, which terminates at the<br>distal tip. This lumen is used to pass the catheter over a<br>guidewire with a maximum diameter of 0.018". The lumen<br>marked "BALLOON" is the balloon inflation lumen, which is<br>used to inflate and deflate the dilatation balloon with a mixture<br>of contrast medium and saline solution. The balloon has two<br>radiopaque markers for positioning the balloon relative to the<br>stenosis. The radiopaque marker bands indicate the dilating or<br>working section of the balloon. On the 150 and 200 mm devices,<br>two additional marker bands denote the middle of the balloon<br>body. The PowerCross .018" OTW PTA Dilatation Catheter is<br>available in multiple balloon sizes. Nominal balloon diameter<br>and length are printed on the strain relief. |
| Performance data                      | Bench testing was performed to support a determination of substantial equivalence. Results from this testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Summary of Substantial<br>Equivalence | The PowerCross .018" OTW PTA Dilatation Catheter has the following similarities to the legally marketed NanoCross .014" and EverCross .035" OTW PTA Dilatation Catheters:<br>Same indicated use as EverCross (K082579), Same operating principle, Same device materials as NanoCross (K090849), Same fundamental scientific technology, Same packaging and sterilization processes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Conclusion                            | Based on the intended use, technological characteristics, safety and performance testing, the PowerCross .018" OTW PTA Dilatation Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the legally marketed NanoCross .014" OTW PTA Dilatation Catheter (K090849) and EverCross .035" OTW PTA Dilatation Catheter (K082854).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

ev3 Corporate World Headquarters / Peripheral Vascular ● 9609 5-1n Avenue Morth ● Plynouth, MN 55442 ● Page Vof 298 7000 ev3 Europe SAS • • 106/108 rue La Boette • • 75008 Paris, France • • PH •33 156 88 59 10

{1}------------------------------------------------

.

:

.

:

・

:

. . . .

く

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with outstretched wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

NOV 1 3 2009

ev3 Inc. c/o Ms. Sara Bakker 9600 54" Avenue North Plymouth, MN 55442

Re: K093286

PowerCross 0.018" OTW PTA Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: October 16, 2009 Received: October 20, 2009

Dear Ms. Bakker:

We have reviewed your Section 510(k) premarket notification of intent to market the device wt have reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the onerosale) to regard) the enactment date of the Medical Device Amendments, or to Conninered price to this 20, 1978, and exacted with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). and Costicule Act (71ct) that ao novice, subject to the general controls provisions of the Act. The 1 ou may, mercere, market the act include requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (boo aco re) existing major regulations affecting your device can be may be subject to additional confroit. "Entraling may areas on to 898. In addition, FDA may found in the Oods of reactions oncerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Ms. Sara Bakker

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical or re-rated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the qualisms by oversions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rr you aboutFDA/CentersOffices/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the I ou may of amall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

er R. bihner

Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1_of _1

## Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: PowerCross™ .018" OTW PTA Dilatation Catheter Indications for Use:

The PowerCross .018" OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

unger valmin

- Sign-Off) Connof Cardiovascular Devices

Number K093286

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT/K093286](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/LIT/K093286)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
