← Product Code [KRB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRB) · K020587

# PULSION PULSIOCATH PICCO MONITORING KIT, INJECTATE SENSOR TEMPERATURE HOUSING PV 4046 (K020587)

_Pulsion Medical Systems AG · KRB · May 23, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRB/K020587

## Device Facts

- **Applicant:** Pulsion Medical Systems AG
- **Product Code:** [KRB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRB.md)
- **Decision Date:** May 23, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1915
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The PULSION Pulsiocath Thermodilution Catheter and Pressure Monitoring Kit with Injectate In-Line Sensor are intended for use with the PULSION PiCCO Cardiac Output System for the measurement of cardiac output by the thermodilution method, measurement of arterial blood pressure, and for cardiac output determination by arterial pulse contour analysis. The PULSION PiCCO Cardiac Output System with Thermodilution Catheter and Pressure Monitoring Kit with Injectate In-Line Sensor is indicated in patients where cardiovascular and circulatory volume status monitoring is necessary. Such as patients in surgical, medical, cardiac, and burn specialty units as well as other specialty units where cardiovascular monitoring is desired and patients undergoing surgical interventions of such magnitude that cardiovascular monitoring is necessary.

## Device Story

The PV4046 Injectate In-Line Sensor is a component of the Pulsion PiCCO Cardiac Output System. It functions as a thermodilution probe to measure injectate temperature. The system uses this input, alongside arterial blood pressure data, to calculate cardiac output via the thermodilution method and arterial pulse contour analysis. Used in clinical settings (surgical, medical, cardiac, burn units) by healthcare professionals to monitor hemodynamic status. The output provides real-time cardiovascular data to assist clinicians in managing patient volume status and cardiac function, potentially improving outcomes for critically ill or surgical patients.

## Clinical Evidence

Bench testing only. Verification and validation activities included laboratory studies for performance, mechanical properties, and biocompatibility to establish safety and reliability. No clinical data was required.

## Technological Characteristics

Injectate in-line sensor and housing (PV4046). Thermodilution probe sensing principle. Designed as a component of the Pulsion PiCCO Cardiac Output System. Modifications involve sensor design, materials, and dimensions.

## Regulatory Identification

A thermodilution probe is a device that monitors cardiac output by use of thermodilution techniques; this device is commonly attached to a catheter that may have one or more probes.

## Predicate Devices

- [K991886](/device/K991886.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 2 3 2002

## 510(k) Summary KO20587

This summary of 510(k) safety and,effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

| Submitter of this<br>premarket notification | Pulsion Medical Systems, AG<br>C/O James M. Delaney<br>EXPERTech Associates, Inc.<br>100 Main Street, Ste 120<br>Concord, MA 01742<br>Phone 978.371.0066<br>Fax 978.371.1676<br>e-mail: jdelaney@fdahelp.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                       |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Date of summary                             | January 29, 2002                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                       |
| Device name                                 | Pulsion Pulsiocath PiCCO Monitoring Kit, Injectate Sensor Temperature<br>Housing PV 4046                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                       |
| Common name                                 | Injectate in-line sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                       |
| Classification names                        | Regulation Number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Classification Name   |
|                                             | §870.1915                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Probe, Thermodilution |
| Predicate Devices                           | The modified device is substantially equivalent to the device previously<br>cleared under K991886.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                       |
| Modifications                               | The primary modifications are changes to the injectate in-line sensor design,<br>materials, and dimensions. Corresponding dimensional changes have been<br>made to the housing to accommodate additional components and their<br>positioning within the housing resulting from the change.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                       |
| Intended Use                                | The modified device has the same intended use as previously cleared in<br>K991886. The PV4046 injectate in-line sensor is a component of a<br>prescription use accessory package to the Pulsion PiCCO Cardiac Output<br>System. The accessory package was cleared with the following Indications<br>for Use:<br>The PULSION Pulsiocath Thermodilution Catheter and Pressure<br>Monitoring Kit with Injectate In-Line Sensor are intended for use with the<br>PULSION PiCCO Cardiac Output System for the measurement of cardiac<br>output by the thermodilution method, measurement of arterial blood<br>pressure, and for cardiac output determination by arterial pulse contour<br>analysis. The PULSION PiCCO Cardiac Output System with<br>Thermodilution Catheter and Pressure Monitoring Kit with Injectate In-Line<br>Sensor is indicated in patients where cardiovascular and circulatory volume<br>status monitoring is necessary. Such as patients in surgical, medical,<br>cardiac, and burn specialty units as well as other specialty units where<br>cardiovascular monitoring is desired and patients undergoing surgical |                       |

{1}------------------------------------------------

KOZO587

p.2/2

interventions of such magnitude that cardiovascular monitoring is necessary.

Technological characteristics

Testing

The modified device has the same technological characteristics as the legally marketed predicate devices.

Verification and validation testing activities were conducted to establish the performance and reliability characteristics of the modified device. Testing involved safety testing from the risk analysis, including laboratory studies for performance, mechanical properties, and biocompatibility. Acceptance criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure.

Public Health Service

**MAY 23 2002**

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Pulsion Medical Systems, Inc. c/o Mr. James M. Delaney EXPERTECH Associates, Inc. 100 Main Street. Suite 120 Concord, MA 01742

Re: K020587

Pulsion Pulsiocath PiCCO Regulation Number: 870.1915 Regulation Name: Thermodilution Probe Regulatory Class: Class II (two) Product Code: 74 KRB Dated: February 19, 2002 Received: February 22, 2002

#### Dear Mr. Delaney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 - Mr. James M. Delaney

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Elias Mallis

Image /page/3/Picture/6 description: The image contains a handwritten word that appears to be "for". The writing is cursive and the letters are connected. The "f" is a loop that extends above the other letters, and the "r" has a small curve at the end.

Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### 3.1 ODE Indications Statement

## Indications for. Use Statement

| 510(k) Number<br>(if known) | K 02 0587                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                 | The Pulsion PV4046 Injectate Sensor component of the<br>PULSION Pulsiocath Thermodilution Catheter and Pressure<br>Monitoring Kit with Injectate In-Line Sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for Use         | The PULSION Pulsiocath Thermodilution Catheter and Pressure<br>Monitoring Kit with Injectate In-Line Sensor are intended for use<br>with the PULSION PiCCO Cardiac Output System for the<br>measurement of cardiac output by the thermodilution method,<br>measurement of arterial blood pressure, and for cardiac output<br>determination by arterial pulse contour analysis. The PULSION<br>PiCCO Cardiac Output System with Thermodilution Catheter and<br>Pressure Monitoring Kit with Injectate In-Line Sensor is indicated<br>in patients where cardiovascular and circulatory volume status<br>monitoring is necessary. Such as patients in surgical, medical,<br>cardiac, and burn specialty units as well as other specialty units<br>where cardiovascular monitoring is desired and patients<br>undergoing surgical interventions of such magnitude that<br>cardiovascular monitoring is necessary. |

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elias Mallis

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number_KO20597

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

8

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRB/K020587](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRB/K020587)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
