← Product Code [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA) · K961668

# MICROFERRET CATHETER (K961668)

_Cook, Inc. · KRA · Nov 4, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K961668

## Device Facts

- **Applicant:** Cook, Inc.
- **Product Code:** [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA.md)
- **Decision Date:** Nov 4, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Microferret™ catheter is used for percutaneous access to small vessel, distal anatomy.

## Device Story

Single-lumen catheter for percutaneous access to small vessel, distal anatomy; graduated stiffness shaft (stiff proximal, soft distal) constructed of polyethylene; proximal hub conforms to ANSI Luer lock standards for syringe compatibility. Used by clinicians in vascular procedures to navigate distal anatomy. Device facilitates infusion or access; performance verified via design validation studies. Marketed internationally since 1992 with established safety profile.

## Clinical Evidence

No clinical trials presented. Evidence based on worldwide marketing history since 1992, design validation studies, and established biocompatibility of materials. No adverse health effects reported in marketing history.

## Technological Characteristics

Single-lumen catheter; polyethylene construction with graduated flexibility; proximal hub compliant with ANSI Luer lock standards. Sterilization via standard methods used for existing COOK catheters.

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- Tracker™ catheters ([K853997](/device/K853997.md), [K862117](/device/K862117.md), [K874751](/device/K874751.md))
- Evolution™ Microcatheters ([K953114](/device/K953114.md))
- Venture™ Infusion Catheters ([K931335](/device/K931335.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K96/1668

510(k) Premarket Notification
Microferret™ Catheter
COOK INCORPORATED

NOV  A  1996

## J. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

**Submitted By:**
April Lavender, RAC
Vice President Regulatory Affairs
COOK INCORPORATED
925 South Curry Pike
P.O. Box 489
Bloomington, IN 47402
24 April 1996

**Device:**
Trade Name: Microferret™ Catheter
Proposed Classification Name: Class II 74 DQY

## Predicate Devices:

The Microferret™ catheter is substantially equivalent to other devices intended for percutaneous vascular use in terms of indications for use, design, construction and materials equivalence. Specifically, this device is similar to numerous pre-Amendment COOK catheters, Tracker™ catheters (K#853997, K#862117, K#874751), manufactured by Target Therapeutics®, San Jose, California, Evolution™ Microcatheters (K#953114) manufactured by Boston Scientific, Watertown, Massachusetts, Venture™ Infusion Catheters (K#931335) manufactured by Scimed Life Systems and distributed by Boston Scientific, Watertown, Massachusetts and others.

## Device Description:

The Microferret™ catheter is used for percutaneous access to small vessel, distal anatomy. Its single lumen construction provides graduated stiffness, from proximal being stiff to distal being very soft. The material used to construct this device is polyethylene with a graduated transition of flexibility along the shaft. The proximal hub of the catheter is designed to conform with Luer lock standards (ANSI) which are compatible with commercially available syringes. These materials are widely used in catheter device manufacturing and their biocompatibility has been verified. In addition, design validation studies have been performed to assure the device can be expected to perform its intended function when used according to the recommendations in the suggested instructions for use provided with the device.

## Summary of Clinical Use

This device has been marketed in Europe since 1992 by a European manufacturing firm owned by COOK Group Incorporated. There has been good clinical success with the device in its worldwide use to date. The firm has identified the primary risks of devices such as the Microferret™ catheter and have included these potential risks in the device labeling. These risks are not specific to the Microferret™ catheter, but apply to vascular catheters in general. There have been no adverse health effects associated with the use of this catheter during its marketing history.

## Substantial Equivalence

This device will be manufactured according to specified process controls and a Quality Assurance Program. This device will undergo packaging similar to the devices currently marketed and distributed by COOK INCORPORATED. This device will undergo sterilization similar to the devices currently marketed and distributed. Being similar with respect to indications for use, materials and physical construction to predicate devices, this device meets the requirements for section 510(k) substantial equivalence.

12

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K961668](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K961668)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
