K952317 · Localmed, Inc. · QEY · Jun 28, 1995 · Cardiovascular
Device Facts
Record ID
K952317
Device Name
Kaplan-Simpson InfusaSleeve II Catheter
Applicant
Localmed, Inc.
Product Code
QEY · Cardiovascular
Decision Date
Jun 28, 1995
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Kaplan-Simpson InfusaSleeve II is intended to infuse fluids such as heparinized saline, diagnostic agents such as contrast media, and thrombolytic agents such as urokinase into the coronary vasculature. The catheter is designed for use with a commercially available dilatation catheter to facilitate selective and subselective access and regional infusion.
Device Story
Multi-lumen infusion catheter; tracks coaxially over standard dilatation catheters; features proximal infusion port, proximal shaft, and main shaft with distal infusion region containing multiple sideholes. Input: infusion solutions (heparinized saline, contrast media, thrombolytic agents) via syringe/manometer. Operation: low-pressure delivery (max 50 psig) through peripheral lumens into coronary vasculature. Used in clinical settings by physicians for selective/subselective regional drug delivery. Benefits: enables targeted, controlled infusion of therapeutic or diagnostic agents directly to coronary sites.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including biocompatibility, tensile strength (port-to-shaft, tip-to-shaft), infusion pathway fatigue, burst pressure, infusate flow, balloon fatigue, and animal testing.
Indicated for patients requiring selective or subselective regional infusion of fluids, including heparinized saline, diagnostic contrast media, or thrombolytic agents, into the coronary vasculature via a dilatation catheter.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
February 3, 2022
Localmed, Inc.
Eleanor Chiu
Director, Regulatory Affairs
1820 Embarcadero Rd.
Palo Alto, California 94303
Re: K952317
Trade/Device Name: Kaplan-Simpson InfusaSleeve II Catheter
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy catheter
Regulatory Class: Class II
Product Code: QEY, KRA
Dear Eleanor Chiu:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 28, 1995. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell@FDA.HHS.gov.
Sincerely,
O'connell -S
Digitally signed by
Gregory W. O'connell -
S
Date: 2022.02.03
14:34:29 -05'00'
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary
and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
JUN 28 1995
Food and Drug Administration
1390 Piccard Drive
Rockville MD 20850
Ms. Eleanor V. Chiu
Director, Regulatory Affairs
LocalMed, Inc.
1820 Embarcadero Road
Palo Alto, California 94303
Dear Ms. Chiu:
Re: K952317
Kaplan-Simpson InfusaSleeve™
II Catheter
Regulatory Class: II (two)
Product Code: 74 KRA
Dated: May 11, 1995
Received: May 17, 1995
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter immediately will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), promotion, or advertising please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-300) at (301) 594-4639. Other general information on your responsibilities under the Act
0001
{3}
may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D.
Acting Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
0002
{4}
LOCALMED
K953317
# 510(k) SUMMARY AS REQUIRED BY 807.92(c)
Company Name and Address:
LocalMed, Inc.
1820 Embarcadero Road
Palo Alto, CA 94303
Phone: 415-843-6770
Fax: 415-843-6771
JUN 28 1995
Name of Contact Person: Eleanor V. Chiu
Date Prepared: May 11, 1995
Trade Name: Kaplan-Simpson InfusaSleeve™ II Catheter
Common Name: Infusion sleeve
Classification Name: Catheter, infusion (per 21 CFR 870.1250)
Substantially Equivalent To: Kaplan-Simpson InfusaSleeve Catheter.
Description of the Device: The Kaplan-Simpson InfusaSleeve™ II is a multi-lumen catheter consisting of a proximal infusion port, a proximal shaft and a main catheter shaft with a distal infusion region having multiple sideholes. The catheter has a balloon entry port and separate outer infusion lumens. The infusion sideholes are located at the catheter's infusion region near the distal tip. Radiopaque markers are located inside each infusion lumen within the infusion region. Infusion solution is delivered through the proximal infusion port, through the outer infusion lumens and exiting the sideholes into the coronary vasculature. A standard luer adapter allows for attachment of a syringe with a manometer or equivalent. The Kaplan-Simpson InfusaSleeve II is designed to track over standard dilatation catheters.
Intended Use: The Kaplan-Simpson InfusaSleeve II is intended to infuse fluids such as heparinized saline, diagnostic agents such as contrast media, and thrombolytic agents such as urokinase into the coronary vasculature. The catheter is designed for use with a commercially available dilatation catheter to facilitate selective and subselective access and regional infusion.
# Summary of Technological Characteristics of InfusaSleeve II Compared to InfusaSleeve:
InfusaSleeve II has similar technological characteristics as the InfusaSleeve that has 510(k) clearance to market. The following is a summary of the characteristics comparison chart with the predicate device (InfusaSleeve):
| | LOCALMED™ InfusaSleeve™ II Catheter | LOCALMED™ InfusaSleeve™ (K950615) Catheter |
| --- | --- | --- |
| Intended Use General Indications: | General intravascular use, delivery of therapeutic solutions. Selective and sub-selective, controlled, infusion of solutions. | General intravascular use, delivery of therapeutic solutions. Selective and sub-selective, controlled, infusion of solutions. |
| Frequency of Use: Technological Characteristics Materials: | Single use, disposable. | Single use, disposable. |
| Infusion Port: | Polycarbonate | ABS, Polyurethane, Polycarbonate |
20
EMBARCADERO ROAD, PALO ALTO, CALIFORNIA 94303 TELEPHONE: 415-843-6770 FACSIMILE: 415-843-6771
{5}
| Proximal Shaft: | Stainless Steel, Nylon | Stainless Steel, Nylon |
| --- | --- | --- |
| Main Shaft & Tip: | Nylon | Nylon |
| Lubricious Coating: | Silicone | Silicone |
| Radiopaque Marker: | Gold/Stainless Steel | Gold/Stainless Steel |
| Support Elements: | Stainless Steel | None |
| Design and Construction | | |
| Catheter Configuration: | Introduced coaxially over dilatation device. Infusion through peripheral lumens | Introduced coaxially over dilatation device. Infusion through peripheral lumens |
| Shaft Configuration: | Multi-lumen. Central lumen accommodates dilatation device. Peripheral lumens for solution infusion. | Multi-lumen. Central lumen accommodates dilatation device. Peripheral lumens for solution infusion. |
| Distal Infusion Region: | Multiple sideholes (9 per peripheral lumen). Each hole has a diameter of 0.04 mm. | Multiple sideholes (9 per peripheral lumen). Each hole has a diameter of 0.04 mm. |
| Solution Delivery Mode: | Low pressure delivery directly into the vasculature. | Low pressure delivery directly into the vascular lumen. |
| Maximum Infusion Pressure: | 50 psig (3 ATM) | 50 psig (3 ATM) |
| Mode of Operation: | | |
| Vascular Access | Percutaneous | Percutaneous |
| Tracking Mechanism | Over-the-Balloon | Over-the-Balloon |
| General and Dimensional Characteristics | | |
| Total Length: | 1.4 m | 1.3 m |
| Working Length: | 1.3 m | 1.2 m |
| Outer Diameter: | 1.5 - 2.1 mm | 1.8 mm |
| Inner Diameter: | 0.9 - 1.5 mm | 1.2 mm |
| Sterilization Method: | Ethylene Oxide | Ethylene Oxide |
| Compatibility | | |
| Balloon Catheter Shaft Size: | 3.5 French | 3.5 French |
| Guiding Catheter: | Standard | Standard |
| Infusion Port: | Luer fitting | Luer fitting |
| Radiopacity: | Markers at infusion region. | Markers at infusion region. |
| Length of Infusion Region: | 2 cm | 2 cm |
Nonclinical Tests Performed: Testing and evaluation included biocompatibility, proximal infusion port to shaft tensile strength, tip to shaft tensile strength, infusion pathway fatigue, burst pressure, infusate flow, balloon fatigue, proximal shaft to main shaft tensile, and animal test. The test results verified that the performance of the InfusaSleeve II is equivalent to, or exceeds, other commercially marketed infusion catheters.
Conclusions Drawn from Nonclinical Tests: The InfusaSleeve II is substantially equivalent to the InfusaSleeve.
2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.