XO Cath Microcatheter (E20-090-S, E20-110-S, E20-130-S, E20-150-S, E20-175-S, E20-220-S, E20-090-B, E20-110-B, E20-130-B, E20-150-B, E20-175-B, E20-220-B, E26-090-S, E26-110-S, E26-130-S, E26-150-S, E26-175-S, E26-220-S, E26-090-B, E26-110-B, E26-130-B, E26-150-B, E26-175-B, E26-220-B,)

K231821 · Transit Scientific, LLC · KRA · Nov 29, 2023 · Cardiovascular

Device Facts

Record IDK231821
Device NameXO Cath Microcatheter (E20-090-S, E20-110-S, E20-130-S, E20-150-S, E20-175-S, E20-220-S, E20-090-B, E20-110-B, E20-130-B, E20-150-B, E20-175-B, E20-220-B, E26-090-S, E26-110-S, E26-130-S, E26-150-S, E26-175-S, E26-220-S, E26-090-B, E26-110-B, E26-130-B, E26-150-B, E26-175-B, E26-220-B,)
ApplicantTransit Scientific, LLC
Product CodeKRA · Cardiovascular
Decision DateNov 29, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The XO Cath Microcatheter is intended for peripheral vascular use. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic or therapeutic materials into the vessel.

Device Story

XO Cath Microcatheter is a single-lumen, metal alloy shaft device with micro-cuts for flexibility; designed to support guidewire during peripheral vascular access and provide conduit for delivery of embolic/diagnostic/therapeutic materials. Available in 2.0Fr and 2.6Fr lumen sizes with straight or bern tip shapes. Features hydrophilic coating on distal outer surface and radiopaque markers at distal tip for fluoroscopic visualization. Proximal end includes standard luer for syringe connection. Used by physicians in clinical settings (e.g., interventional radiology/cardiology) to navigate peripheral vasculature. Device facilitates targeted delivery of embolics (spheres, particles, coils, plugs, liquid embolics) to treat vascular conditions. Benefits include improved access to challenging anatomy and controlled material delivery.

Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-1), dimensional verification, functional/simulated use, static burst pressure, maximum infusion pressure, flow rate, kink radius, tensile/torque testing, coating lubricity/durability, fluid/air leak testing, embolic compatibility, and radiopacity. All testing met pre-determined acceptance criteria.

Technological Characteristics

Single-lumen metal alloy shaft with micro-cuts; 2.0Fr and 2.6Fr sizes; hydrophilic coating; radiopaque marker bands; standard luer hub. Sterilization method: not specified. Connectivity: none. Energy source: none (mechanical).

Indications for Use

Indicated for peripheral vascular use in patients requiring controlled and selective infusion of diagnostic, embolic, or therapeutic materials into the vessel.

Regulatory Classification

Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 29, 2023 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Transit Scientific, LLC % Spencer Walker Director of Regulatory Affairs University of Utah, Center for Medical Innovation 10 North 1900 East, EHSL Rm. 22 Salt Lake City, Utah 84112 ## Re: K231821 Trade/Device Name: XO Cath Microcatheter (E20-090-S, E20-110-S, E20-130-S, E20-175-S, E20-220-S, E20-090-B, E20-110-B, E20-130-B, E20-150-B, E20-175-B, E20-220-B, E26-090-S, E26-110-S, E26-130-S, E26-150-S, E26-175-S, E26-220-S, E26-090-B, E26-110-B, E26-130-B, E26-150-B, E26-175-B, E26-220-B,) Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: October 27, 2023 Received: October 30, 2023 Dear Spencer Walker: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Samuel G. Raben -S for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices {2}------------------------------------------------ OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K231821 #### Device Name XO Cath Microcatheter (E20-090-S, E20-130-S, E20-175-S, E20-175-S, E20-220-S, E20-090-B, E20-110-B, E20-130-B, E20-150-B, E20-175-B, E20-220-B, E26-110-S, E26-130-S, E26-175-S, E26-175-S, E26-220-S, E26-110-B, E26-130-B, E26-150-B, E26-175-B, E26-220-B,) Indications for Use (Describe) The XO Cath Microcatheter is intended for peripheral vascular use. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic or therapeutic materials into the vessel. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> | |----------------------------------------------|-------------------------------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | □ | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Transit Scientific. The logo is in black and white and features the words "TRANSIT" stacked on top of "SCIENTIFIC." The word "TRANSIT" is in a bold, sans-serif font, while the word "SCIENTIFIC" is in a thinner, sans-serif font. There is a trademark symbol to the right of the word "SCIENTIFIC." ## 510(K) SUMMARY #### (21 CFR 807.92) ### GENERAL INFORMATION | Submitter: | Transit Scientific, LLC | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Spencer Walker, MSc – Director of Regulatory Affairs<br>University of Utah/ Center for Medical Innovation<br>10 North 1900 East, Rm. 22B<br>Salt Lake City, UT 84112<br>(801) 708-2238<br>Spencer.walker@hsc.utah.edu | | Date Prepared: | June 20, 2023 | | Trade Name: | XO Cath Microcatheter | | Classification Name: | Continuous Flush Catheter<br>21 CFR §870.1210, Product Code KRA | | Device Class: | Class II | | Predicate Device: | 510(k) No.: K163701<br>Model: Direxion and Direxion Hi-Flo Torqueable Microcathete<br>Manufacture: Boston Scientific Corporation<br>Classification: KRA | #### Device Description: The subject XO Cath Microcatheter is a single-lumen, metal alloy shaft with micro-cuts for flexibility, designed to support a guidewire during access of the peripheral vasculature, and to provide a conduit for the delivery of embolic applications. The XO Cath Microcatheter is available in two lumen sizes, 2.0Fr and 2.6Fr. The XO Cath 2.0Fr and 2.6Fr Microcatheters are available with a straight or bern tip shape to aid with accessing challenging anatomy. The distal outer surface of the microcatheter is coated with a hydrophilic coating. Radiopaque markers are located at the distal tip to facilitate fluoroscopic visualization. The proximal end incorporates a standard luer for ease of use and to connect with a syringe. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Transit Scientific. The logo is in black and white and features the word "TRANSIT" in large, bold letters on the top line. Below that, the word "SCIENTIFIC" is written in smaller letters. The logo has a clean and modern design. | Table 1: XO Cath Microcatheter Model Numbers | | |----------------------------------------------|-------------------------------------------------------------------------| | Model No. | Description | | E20-090-S | XO Cath Microcatheter – 2.0 Fr diameter, and 90cm length, Straight Tip | | E20-110-S | XO Cath Microcatheter - 2.0 Fr diameter, and 110cm length, Straight Tip | | E20-130-S | XO Cath Microcatheter – 2.0 Fr diameter, and 130cm length, Straight Tip | | E20-150-S | XO Cath Microcatheter – 2.0 Fr diameter, and 150cm length, Straight Tip | | E20-175-S | XO Cath Microcatheter - 2.0 Fr diameter, and 175cm length, Straight Tip | | E20-220-S | XO Cath Microcatheter – 2.0 Fr diameter, and 220cm length, Straight Tip | | E20-090-B | XO Cath Microcatheter - 2.0 Fr diameter, and 90cm length, Bern Tip | | E20-110-B | XO Cath Microcatheter – 2.0 Fr diameter, and 110cm length, Bern Tip | | E20-130-B | XO Cath Microcatheter – 2.0 Fr diameter, and 130cm length, Bern Tip | | E20-150-B | XO Cath Microcatheter – 2.0 Fr diameter, and 150cm length, Bern Tip | | E20-175-B | XO Cath Microcatheter – 2.0 Fr diameter, and 175cm length, Bern Tip | | E20-220-B | XO Cath Microcatheter – 2.0 Fr diameter, and 220cm length, Bern Tip | | E26-090-S | XO Cath Microcatheter – 2.6 Fr diameter, and 90cm length, Straight Tip | | E26-110-S | XO Cath Microcatheter – 2.6 Fr diameter, and 110cm length, Straight Tip | | E26-130-S | XO Cath Microcatheter – 2.6 Fr diameter, and 130cm length, Straight Tip | | E26-150-S | XO Cath Microcatheter – 2.6 Fr diameter, and 150cm length, Straight Tip | | E26-175-S | XO Cath Microcatheter - 2.6 Fr diameter, and 175cm length, Straight Tip | | E26-220-S | XO Cath Microcatheter – 2.6 Fr diameter, and 220cm length, Straight Tip | | E26-090-B | XO Cath Microcatheter - 2.6 Fr diameter, and 90cm length, Bern Tip | | E26-110-B | XO Cath Microcatheter - 2.6 Fr diameter, and 110cm length, Bern Tip | | E26-130-B | XO Cath Microcatheter – 2.6 Fr diameter, and 130cm length, Bern Tip | | E26-150-B | XO Cath Microcatheter – 2.6 Fr diameter, and 150cm length, Bern Tip | | E26-175-B | XO Cath Microcatheter - 2.6 Fr diameter, and 175cm length, Bern Tip | | E26-220-B | XO Cath Microcatheter - 2.6 Fr diameter, and 220cm length, Bern Tip | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Transit Scientific. The word "TRANSIT" is in all caps and bolded on the top line. The word "SCIENTIFIC" is in all caps on the second line, but it is not bolded. ### Indications for Use: The XO Cath Microcatheter is intended for peripheral vascular use. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic or therapeutic materials into the vessel. #### Comparative Analysis: It has been demonstrated that the XO Cath Microcatheter is comparable to the predicate device in intended use, fundamental scientific technology, design, principles of operation and functional performance evaluations. The XO Cath Microcatheter has been fully assessed within the Transit Scientific Risk Management and Design Controls systems. All necessary verification steps met pre-determined acceptance criteria to determine performance as compared to the predicate. It has been demonstrated that the XO Cath Microcatheter is comparable to the predicate device in the following manner: - Same intended use ● - Same indications for use - Same fundamental scientific technology ● - Same or similar material properties ● - Same or similar operating principle - . Similar performance specifications - Similar patient-user interface | | Subject Device –<br>(XO Cath 2.0 Fr) | Predicate –<br>K163701 (Direxion<br>Microcatheter) | Subject Device –<br>(XO Cath 2.6 Fr) | Predicate –<br>K163701 (Direxion<br>Hi-Flo<br>Microcatheter) | | |---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Ind. for Use | The XO Cath<br>Microcatheter is intended<br>for peripheral vascular<br>use. The microcatheter<br>can be used for<br>controlled and selective<br>infusion of diagnostic,<br>embolic, or therapeutic<br>materials into the vessel | The Direxion<br>Microcatheters is<br>intended for peripheral<br>vascular use. The<br>microcatheter can be<br>used for controlled and<br>selective infusion of<br>diagnostic, embolic, or<br>therapeutic materials into<br>the vessel. | The XO Cath<br>Microcatheter is<br>intended for peripheral<br>vascular use. The<br>microcatheter can be<br>used for controlled and<br>selective infusion of<br>diagnostic, embolic, or<br>therapeutic materials<br>into the vessel. | The Direxion HI-FLO<br>Torqueable<br>Microcatheters is<br>intended for peripheral<br>vascular use. The<br>microcatheter can be<br>used for controlled and<br>selective infusion of<br>diagnostic, embolic, or<br>therapeutic materials<br>into the vessel. | | | Classification<br>Name | Cardiovascular Device - Continuous Flush Catheter<br>21 CFR §870.1210<br>Product Code: KRA<br>Class II | | | | | | Single Use | Yes | | | | | | Prescription (Rx<br>Only) | Yes | | | | | | Table 2: Substantial Equivalence Comparison Chart | | | | | | | | Subject Device -<br>(XO Cath 2.0 Fr) | Predicate -<br>K163701 (Direxion<br>Microcatheter) | Subject Device -<br>(XO Cath 2.6 Fr) | Predicate -<br>K163701 (Direxion<br>Hi-Flo<br>Microcatheter) | | | Anatomical<br>Access | Peripheral vasculature | | | | | | Effective<br>Length (cm) | 90cm, 110cm,<br>130cm, 150cm,<br>175cm, and 220cm | 105cm, 130cm,<br>155cm | 90cm, 110cm,<br>130cm, 150cm,<br>175cm, and 220cm | 105cm, 130cm,<br>155cm | | | Shaft Outer<br>Diameter | Distal: 2.0 Fr<br>Proximal: 2.6 Fr | Distal: 2.4 Fr<br>Proximal: 2.7 Fr | Distal: 2.6 Fr<br>Proximal: 2.75Fr | Distal: 2.8 Fr<br>Proximal: 3.0 Fr | | | Inner Lumen | 0.021" | 0.021" | 0.027" | 0.027" | | | Max Pressure<br>(PSI) | 1200 | | | | | | Max Flow Rate<br>(mL/s) | 2.0Fr x 130cm:<br>3.0 mL/s | 2.4Fr x 130cm:<br>3.1 mL/s | 2.6Fr x 130cm:<br>5.8 mL/s | 2.8Fr x 130cm:<br>5.6 mL/s | | | Dead Space<br>Volume (mL) | 2.0Fr x 130cm:<br>0.47 mL | 2.4Fr x 130cm:<br>0.46 mL | 2.6Fr x 130cm:<br>0.66 mL | 2.8Fr x 130cm:<br>0.64 mL | | | Hydrophilic<br>Coating | Yes | | | | | | Visibility/<br>Radiopacity | Radiopaque Marker Bands on catheter body | | | | | | Guidewire<br>Compatibility | Max 0.018" | Max 0.018" | Max 0.021" | Max 0.021" | | | Guide<br>Catheter<br>Compatibility | Min. 0.038 in | Min. 0.035 in | Min. 0.038 in | Min. 0.038 in | | | Embolic<br>Compatibility | ≤ 700 Micron spheres<br>≤ 500 Micron particles<br>Coils & Plugs (max<br>0.021" recommended<br>microcatheter<br>compatible)<br>Liquid Embolics<br>(DMSO.<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives) | ≤ 700 Micron spheres<br>≤ 500 Micron particles<br>Coils (max 0.018"<br>recommended<br>microcatheter<br>compatible)<br>Liquid Embolics<br>(DMSO.<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives, Ethanol) | ≤ 900 Micron spheres<br>≤ 700 Micron particles<br>Coils & Plugs (max<br>0.027" recommended<br>microcatheter<br>compatible)<br>Liquid Embolics<br>(DMSO,<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives) | ≤ 900 Micron spheres<br>≤ 710 Micron particles<br>Liquid Embolics<br>(DMSO,<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives, Ethanol) | | | Tip Shape | Straight, Bern | Straight, Bern, J,<br>Swan | Straight, Bern | Straight, Bern, J,<br>Swan | | | Shapable | No | Some SKUs (via<br>steam) | No | Some SKUs (via<br>steam) | | {7}------------------------------------------------ # Functional Testing: The following testing was conducted to validate and verify that the subject device was substantially equivalent to the predicate devices. All data met pre-determined acceptance criteria. {8}------------------------------------------------ TRANSIT SCIENTIFIC™ - Biocompatibility Biocompatibility of the complete and finished XO Cath Microcatheter ● has been verified according to the requirements and testing prescribed in ISO 10993-1 and in accordance with FDA guidance document "Use of International Standard ISO 10993-1" for an external communicating device with limited exposure (<24hrs) to circulating blood. - Cytotoxicity O - Sensitization O - Irritation/ Intracutaneous Toxicity O - Acute Systemic Toxicity O - Material Mediated Pyrogenicity O - Hemolysis Assay O - Complement Activation Assay O - Partial Thromboplastin Time (PTT) O - Blood Platelet and Leukocyte Count Testing O - LAL Pyrogenicity O - Design Verification Performance bench testing was conducted to ensure that the XO ● Cath Microcatheter met the applicable design and performance requirements throughout its shelf life, verify conformity to applicable standards, and demonstrate substantial equivalence to the predicate system. The following performance testing was performed or fulfilled with the XO Cath Microcatheter. - o Material Verification - Dimensional Verification o - Visual Verification O - Functional/ Simulated Use Testing O - Static Burst Pressure o - Maximum Infusion Pressure O - O Maximum Flow Rate - Kink Radius O - Tensile Testing O - Torque Testing o - Coating Lubricity O - Coating Durability O - Fluid Leak Testing O - Hub Assembly Air Leak O - Physical Embolic Testing O - Liquid Embolic Exposure Testing O - Radiopacity o - Packaging O - Sterilization O {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for Transit Scientific. The word "TRANSIT" is in large, bold, sans-serif font on the top line. Below that, in a smaller, thinner font, is the word "SCIENTIFIC" with a trademark symbol to the right. The logo is simple and modern. ## Conclusion: Based on the similarities in design between the subject and predicate devices, and the performance testing performed, the subject XO Cath Microcatheter is substantially equivalent to the cited predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...