← Product Code [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA) · K212719

# Micro Catheter and Guidewire System (K212719)

_Suzhou Hengrui Hongyuan Medical Co., Ltd. · KRA · Apr 11, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K212719

## Device Facts

- **Applicant:** Suzhou Hengrui Hongyuan Medical Co., Ltd.
- **Product Code:** [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA.md)
- **Decision Date:** Apr 11, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Micro Catheter and Guidewire System is intended for the infusion of contrast media into the peripheral vessels. The Micro Catheter and Guidewire system is also infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis. The system should not be used in cerebral vessels.

## Device Story

System comprises catheter, guidewire, and accessories (flushing device, shaping mandrel, insertion tool, torque device, peel-able tube). Used by clinicians in peripheral vascular procedures. Catheter features PTFE inner layer, stainless steel wire reinforcement, platinum-iridium radiopaque marker, polyamide outer layer, and hydrophilic coating. Guidewire features nitinol core, polymer jacket with PVP hydrophilic coating, and distal spring coil. Device facilitates targeted delivery of contrast, therapeutic drugs, or embolic agents to peripheral vasculature. Clinician navigates system to target site; output is delivery of therapeutic/diagnostic agents. Benefits include precise vessel access and agent delivery.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified via biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, PTT, complement activation, thrombogenicity) and mechanical bench testing (tensile force, leakage, torque, kink resistance, radiopacity, burst pressure, simulated use, tip flexibility).

## Technological Characteristics

Catheter: PTFE inner, stainless steel wire reinforcement, platinum-iridium radiopaque marker, polyamide outer layer, hydrophilic coating. Guidewire: Nitinol core, polymer jacket, PVP hydrophilic coating, distal spring coil. Sterile, single-use. Mechanical sensing/delivery principle.

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- Micro Catheter and Guidewire System ([K171665](/device/K171665.md))

## Reference Devices

- Progreat™ Angiographic Catheter ([K033913](/device/K033913.md))
- Maestro Micro catheter ([K172081](/device/K172081.md))
- Traxcess 0.014" Hydrophilic Guidewire ([K080863](/device/K080863.md))
- Rebar™ Micro Catheter ([K993672](/device/K993672.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 11, 2022

Suzhou Hengrui Hongyuan Medical Co., Ltd Wang Peipei Regulatory Affairs Manager Building B9 Unit 201, No. 218 Xinghu Road, SIP Suzhou, Jiangsu 215126 China

Re: K212719

Trade/Device Name: Micro Catheter and Guidewire System Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: March 3, 2022 Received: March 14, 2022

Dear Wang Peipei:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212719

Device Name Micro Catheter and Guidewire System

Indications for Use (Describe)

The Micro Catheter and Guidewire System is intended for the infusion of contrast media into the peripheral vessels. The Micro Catheter and Guidewire system is also infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis. The system should not be used in cerebral vessels.

Type of Use (Select one or both, as applicable)

| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary           |                                                                                                                                                                                                                                                                            |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                          | Micro Catheter and Guidewire System                                                                                                                                                                                                                                        |
|                          | (per 21 CFR 807.92)                                                                                                                                                                                                                                                        |
| Submitter:               | Suzhou Hengrui Hongyuan Medical Co., Ltd<br>Building B9 Unit 201, No. 218 Xinghu Road, SIP, Suzhou,<br>Jiangsu, China                                                                                                                                                      |
| Contact Person:          | Wang peipei<br>Regulatory Affairs Manager<br>Phone: +86-512-6835-6665<br>Email: wangpeipei@hrmedical.com.cn                                                                                                                                                                |
| Date Prepared:           | August 18th, 2021                                                                                                                                                                                                                                                          |
| Trade Name:              | Micro Catheter and Guidewire System                                                                                                                                                                                                                                        |
| Common or Usual<br>Name: | Continuous flush catheter                                                                                                                                                                                                                                                  |
| Classification:          | Class II, 21 CFR Part 870.1210                                                                                                                                                                                                                                             |
| Product Code:            | KRA                                                                                                                                                                                                                                                                        |
| Predicate Device:        | K171665-Micro Catheter and Guidewire System (Suzhou<br>Hengrui Disheng Medical Co., Ltd)                                                                                                                                                                                   |
| Reference Devices:       | K033913- Progreat™ Angiographic Catheter (Terumo Medical<br>Corporation )<br>K172081- Maestro Micro catheter (Merit Medical Systems)<br>K080863- Traxcess 0.014" Hydrophilic Guidewire (Micro<br>Vention, Inc)<br>K993672-Rebar™ Micro Catheter (Micro Therapeutics, Inc.) |

{4}------------------------------------------------

| Device Description:                  | Micro Catheter and Guidewire System consists of a catheter, a<br>guidewire, and accessories. The accessories include a flushing<br>device, a shaping mandrel, an insertion tool, a torque device and<br>a peel-able tube.<br><br>The catheter is consist of a hub, a stress relief tube and a catheter<br>shaft. The catheter shaft has three layers. The inner layer is a<br>PTFE tube, the middle layer is consist of stainless steel wire<br>reinforce and platinum-iridium alloy radiopaque distal marker.<br>And the outer layer is polyamide. There is also a hydrophilic<br>coating on the catheter surface.<br><br>The guidewire is consist of a nitinol core, a polymer jacket with<br>PVP hydrophilic coating over its entire surface, and a spring coil<br>at distal. It has a white marker at the proximal to indicate the<br>length inserted into human body and its relative position with the<br>catheter.                        |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use:                 | The Micro Catheter and Guidewire System is intended for the<br>infusion of contrast media into the peripheral vessels. The Micro<br>Catheter and Guidewire system is also intended for drug<br>infusion in intra-arterial therapy and the infusion of embolic<br>materials for hemostasis. The system should not be used in<br>cerebral vessels.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Comparison with<br>Predicate Device: | The Micro Catheter and Guidewire System is similar to the<br>Micro Catheter and Guidewire System in following ways:<br>Each of the devices is intended to be used for the infusion<br>of contrast media into all peripheral vessels, drug infusion<br>in intra-arterial therapy and the infusion of embolic<br>materials for hemostasis. Each of the devices is provided with catheter, guidewire<br>and accessories. Each of the devices is provided sterile. Each of the devices is intended to be single use. Each of the devices has a hydrophilic coating. Each of the catheter has a platinum-iridium alloy<br>radiopaque marker. Each of the catheter has the same raw materials. Each of the guidewire has the same structure. Each of the guidewire has the same raw materials.          The following technological differences exist between the<br>subject and predicate devices:  Catheter outer diameter Catheter effective length |

{5}------------------------------------------------

- Guidewire effective length ●
- . Accessories

#### Biocompatibility Testing Performance Data:

Biocompatibility evaluation for the Micro Catheter and Guidewire System was conducted in accordance with current standards and the following tests were included:

- Cytotoxicity ●
- Sensitization ●
- Irritation/Intracutaneous Reactivity ●
- Acute systemic Toxicity ●
- Pyrogenicity ●
- Hemolysis study
- Partial Thromboplastin time ●
- Complement Activiation ●
- . Thrombogenicity

## Bench Testing

The tests included the following:

- Catheter Sizes
- Catheter Surface
- Catheter Hub .
- Peak tensile force of catheter
- . Coating test of catheter
- Freedom from leakage ●
- . Distal tip of catheter
- Torque Strength of catheter ●
- Kink resistant of catheter .
- Radiopacity of catheter .
- Hydration judgment of catheter
- Burst pressure under static conditions ●
- Guidewire Sizes .
- Guidewire surface ●
- Coating test of guidewire ●
- Peak tensile force of guidewire ●
- . Torque strength testing of guidewire
- Torqueability test of guidewire ●
- Kink resistance of guidewire ●
- Simulated Use ●
- Tip flexibility of guidewire .

{6}------------------------------------------------

- Radio-detectability of guidewire .
- Sterile
- Bacterial endotoxin ●
- . Particulate test
- The data provided of the Micro Catheter and Guidewire System Conclusion: and the mechanical testing results demonstrate that the device should perform as intended in the specified use conditions. Nonclinical tests demonstrate that the Micro Catheter and Guidewire System is substantially equivalent to the predicate device which is currently marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K212719](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K212719)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
