← Product Code [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA) · K172081

# Maestro Microcatheter (K172081)

_Merit Medical Systems, Inc. · KRA · Aug 4, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K172081

## Device Facts

- **Applicant:** Merit Medical Systems, Inc.
- **Product Code:** [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA.md)
- **Decision Date:** Aug 4, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the microcatheter can be used for the controlled and selective infusion of diagnostic, embolic or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.

## Device Story

Maestro Microcatheter is a sterile, single-use intravascular catheter; available in 2.8F/2.1F sizes and various lengths (110cm, 130cm, 150cm). Features include hydrophilic coating on distal 80cm shaft, radiopaque distal tip marker for fluoroscopic visualization, and molded winged hub with strain relief. Distal tip configurations include straight, 45-degree, or swan neck. Used by physicians in clinical settings to access subselective regions of peripheral or coronary vasculature; facilitates controlled infusion of diagnostic, embolic, or therapeutic agents. Device provides mechanical pathway for delivery of coils, microspheres, and particles. Benefits include precise navigation and targeted delivery of therapeutic materials to specific vascular sites.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included surface, tensile force, dimensions, radio-detectability, leakage, flow rate, kink resistance, lubricity, coating coverage, delivery of microspheres/particles/coils, torque strength, particulate, tip shape retention, hub migration, guidewire/catheter compatibility, torsion, negative pressure collapse, hoop removal, tip straightener, trackability, and pushability. All tests met predetermined acceptance criteria.

## Technological Characteristics

Intravascular catheter; 2.8F proximal/2.1F distal diameter. Features hydrophilic coating, radiopaque distal marker. Materials meet ISO 10993 biocompatibility standards. Sterilized via ethylene oxide (ISO 11135, ISO 10993-7). Complies with ISO 10555-1 for general requirements. Mechanical design includes winged hub and tapered strain relief.

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- Maestro Microcatheter ([K082613](/device/K082613.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W 066-G609 Silver Spring, MD 20993-0002

August 4, 2017

Merit Medical Systems, Inc. Niloufar Samimi Regulatory Affairs Specialist II 1600 West Merit Parkway South Jordan, Utah 84095

Re: K172081

Trade/Device Name: Maestro Microcatheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: July 6, 2017 Received: July 10, 2017

Dear Niloufar Samimi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

M. G. Hillemann

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

..

510(k) Number (if known)

K172081

Device Name Maestro Microcatheter

#### Indications for Use (Describe)

The Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the microcatheter can be used for the controlled and selective infusion of diagnostic, embolic or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <div> <span> <span style="font-size: 20px;"><b>                              ✓                      </b></span>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff
   
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (101) 443-6740 EF

{3}------------------------------------------------

# 510(k) Summary

| General<br>Provisions  | Submitter Name:<br>Address:<br>Telephone Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4583<br>Niloufar Samimi<br>07/05/2017<br>1721504                 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device         | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Maestro Microcatheter<br>Microcatheter<br>Continuous Flush Catheter<br>2<br>KRA<br>870.1210<br>Cardiovascular                                                  |
| Predicate Device       | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Maestro Microcatheter<br>Continuous Flush Catheter<br>K082613<br>Merit Medical Systems, Inc.<br>This predicate has not been subject to a design-related recall |
| Reference Device       | No reference devices were used in this submission.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                |
| Device<br>Description  | The Maestro Microcatheter is available in 2.8F (proximal) / 2.1F (distal)<br>size and 110cm, 130cm and 150cm lengths. The distal tip of the<br>microcatheter is offered in straight or pre-shaped 45 degree and swan<br>neck configurations. The proximal end of the catheter consists of a<br>molded winged hub with a tapered strain relief. The outer surface of<br>the distal 80cm of the microcatheter shaft is coated with a hydrophilic<br>coating designed to facilitate the introduction of the catheter into the<br>vasculature. The microcatheter incorporates a radiopaque marker at<br>the distal tip to facilitate fluoroscopic visualization.<br>The Maestro Microcatheter is offered with two 3ml syringes. |                                                                                                                                                                |
| Indications for<br>Use | There is no change in the Indications for Use Statement from the<br>predicate to the subject device.<br>The Microcatheter is intended for general intravascular use, including<br>peripheral and coronary vasculature. Once the subselective region has<br>been accessed, the microcatheter can be used for the controlled and<br>selective infusion of diagnostic, embolic or therapeutic materials into<br>vessels. The catheter should not be used in the cerebral vessels.                                                                                                                                                                                                                                              |                                                                                                                                                                |

{4}------------------------------------------------

The design and technological characteristics of the subject 2.8F/2.1F Maestro Microcatheter are substantially equivalent to those of the predicate Maestro Microcatheter. The subject device has the same basic design as the predicate device. The main difference between the subject and the predicate devices is in the French size to expand the product line to include the 2.8F/2.1F microcatheter. The comparison between the subject and the predicate devices is based on the following:

- Same intended use ●
- . Same indications for use
- Similar material types that meet ISO 10993 biocompatibility . requirements
- Same design
- Same sterilization methods ●
- Same fundamental technology/principle of operation ●

No performance standards have been established under Section 514 of the Food. Drug and Cosmetic Act for these devices. Performance testing of the subject 2.8F/2.1F Maestro Microcatheter was conducted based on the risk analysis and based on the requirements of the following international standard:

- ISO 10555-1:2013, Intravascular Catheters Sterile and single-● use catheters - Part 1: General requirements
- ISO 11135:2014, Sterilization of health care products Ethylene ● oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
- ISO 10993-1:2009. Biological Evaluation of Medical Devices . Part 1: Evaluation and Testing within a risk management process, and FDA guidance Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices, May 1, 1995
- ISO 10993-7:2008, Biological evaluation of medical devices -● Part 7: Ethylene oxide sterilization residuals
- ISO 2233:2000, Packaging complete, filled transport ● packages and unit loads - conditioning for testing
- ASTM D4169-14:2014, Standard practice for performance ● testing of shipping containers and systems

## Comparison to Predicate Device

{5}------------------------------------------------

The following tests were performed to demonstrate there were no unacceptable risks associated with the changes made to the device:

## Performance Testing-Bench

- Surface ●
- Tensile Force ●
- Dimensions ●
- . Radio-detectability
- Freedom from Liquid Leakage ●
- Flow Rate
- Kink Resistance ●
- Lubricity ●
- Coating Coverage ●
- Delivery of Microspheres and Particles ●
- Torque Strength ●
- Particulate ●
- Tip Shape Retention
- Hub Migration from Clip ●

Safety & Performance Tests cont.

Design Validation

- Delivery of Coils ●
- Guidewire Compatibility ●
- Catheter Compatibility ●
- Torsion ●
- Negative Pressure Collapse .
- Hoop Removal ●
- Distal Tip
- Tip Straightener ●
- Soft Distal Section
- Trackability ●
- Pushability ●
- Reposition of Guidewire .

The results of the testing demonstrated that the subject 2.8F/2.1F Maestro Microcatheter met the predetermined acceptance criteria applicable to the safety and efficacy of the device. This has demonstrated the subject is substantially equivalent to the predicate device.

{6}------------------------------------------------

| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, safety and performance<br>testing, the subject 2.8F/2.1F Maestro Microcatheter meets the<br>requirements that are considered essential for its intended use and is<br>substantially equivalent to the predicate device, the Maestro<br>Microcatheter, K082613, manufactured by Merit Medical Systems, Inc |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K172081](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K172081)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
